Browse hierarchy: [Radiology (RA)](/submissions/RA) → [Subpart B — Diagnostic Devices](/submissions/RA/subpart-b%E2%80%94diagnostic-devices) → [21 CFR 892.2100](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/892.2100) → QJU — Image Acquisition And/Or Optimization Guided By Artificial Intelligence

# QJU · Image Acquisition And/Or Optimization Guided By Artificial Intelligence

_Radiology · 21 CFR 892.2100 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU

## Overview

- **Product Code:** QJU
- **Device Name:** Image Acquisition And/Or Optimization Guided By Artificial Intelligence
- **Regulation:** [21 CFR 892.2100](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/892.2100)
- **Device Class:** 2
- **Review Panel:** [Radiology](/submissions/RA)

## Identification

A radiological acquisition and/or optimization guidance system is a device that is intended to aid in the acquisition and/or optimization of images and/or diagnostic signals. The device interfaces with the acquisition system, analyzes its output, and provides guidance and/or feedback to the operator for improving image and/or signal quality.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed, technical device description, including a detailed description of the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, and the associated inputs and outputs.
(ii) A detailed, technical report on the non-clinical performance testing of the subject device in the intended use environments, using relevant consensus standards when applicable.
(iii) A detailed report on the clinical performance testing, obtained from either clinical testing, accepted virtual/physical systems designed to capture clinical variability, comparison to a closely-related device with established clinical performance, or other sources that are justified appropriately. The choice of the method must be justified given the risk of the device and the general acceptance of the test methods. The report must include the following:
(A) A thorough description of the testing protocol(s).
(B) A thorough, quantitative evaluation of the diagnostic utility and quality of images/data acquired, or optimized, using the device.
(C) A thorough, quantitative evaluation of the performance in a representative user population and patient population, under anticipated conditions and environments of use.
(D) A thorough discussion on the generalizability of the clinical performance testing results.
(E) A thorough discussion on use-related risk analysis/human factors data.
(iv) A detailed protocol that describes, in the event of a future change, the level of change in the device technical specifications or indications for use at which the change or changes could significantly affect the safety or effectiveness of the device and the risks posed by these changes. The assessment metrics, acceptance criteria, and analytical methods used for the performance testing of changes that are within the scope of the protocol must be included.
(v) Documentation of an appropriate training program, including instructions on how to acquire and process quality images and video clips, and a report on usability testing demonstrating the effectiveness of that training program on user performance, including acquiring and processing quality images.
(2) The labeling required under § 801.109(c) of this chapter must include:
(i) A detailed description of the device, including information on all required and/or compatible parts.
(ii) A detailed description of the patient population for which the device is indicated for use.
(iii) A detailed description of the intended user population, and the recommended user training.
(iv) Detailed instructions for use, including the information provided in the training program used to meet the requirements of paragraph (b)(1)(iv) of this section.
(v) A warning that the images and data acquired using the device are to be interpreted only by qualified medical professionals.
(vi) A detailed summary of the reports required under paragraphs (b)(1)(ii) and (iii) of this section.
(vii) A statement on upholding the As Low As Reasonably Achievable (ALARA) principle with a discussion on the associated device controls/options.

## Recent Cleared Devices (7 of 7)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K251416](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K251416.md) | UltraSight Guidance | Ultrasight , Ltd. | Dec 17, 2025 | SESE |
| [K242807](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K242807.md) | HeartFocus (V.1.1.1) | Deski | Apr 4, 2025 | SESE |
| [K243065](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K243065.md) | Cardiac Guidance | Caption Health, Inc. | Jan 15, 2025 | SESE |
| [K223347](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K223347.md) | UltraSight AI Guidance | Ultrasight, Inc. | Jul 24, 2023 | SESE |
| [K201992](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K201992.md) | Caption Guidance | Caption Health | Sep 18, 2020 | SESE |
| [K200755](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/K200755.md) | Caption Guidance | Caption Health | Apr 16, 2020 | SESE |
| [DEN190040](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU/DEN190040.md) | Caption Guidance | Bay Labs, Inc. | Feb 7, 2020 | DENG |

## Top Applicants

- Caption Health — 2 clearances
- Ultrasight, Inc. — 1 clearance
- Ultrasight , Ltd. — 1 clearance
- Deski — 1 clearance
- Caption Health, Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QJU)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
