UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program
K250328 · Canon Medical Systems Corporation · QIH · Apr 30, 2025 · Radiology
Device Facts
Record ID
K250328
Device Name
UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program
Applicant
Canon Medical Systems Corporation
Product Code
QIH · Radiology
Decision Date
Apr 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Acceptance
Observed
Dev DS
Test DS
2D Wall Motion Tracking (Left Ventricle)
Substantial equivalence to performance in reference device K223017 using same acceptance criteria
Performs as intended and with substantial equivalence to the reference device
Not specified in document
Representative subset of clinical data acquired for the original performance testing of these features in reference device K223017
Left Ventricular Ejection Fraction (Auto EF)
Substantial equivalence to performance in reference device K223017 using same acceptance criteria
Performs as intended and with substantial equivalence to the reference device
Not specified in document
Representative subset of clinical data acquired for the original performance testing of these features in reference device K223017
Indications for Use
UltraExtend NX CUW-U001S Ultrasound Image Analysis Program is designed to allow the user to observe images and perform analysis based on examination data acquired using the following diagnostic ultrasound systems; TUS-AI900, TUS-AI800, and TUS-AI700. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
Device Story
UltraExtend NX V2.0 is an ultrasound image analysis software workstation; processes RAW or Image+RAW data acquired from Canon TUS-AI900, TUS-AI800, and TUS-AI700 systems. Used in hospital/clinical settings by trained physicians/qualified personnel. Software enables post-processing, including standard/application/obstetrics measurements, Cardiac4D, 2D/3D Wall Motion Tracking, Aortic Valve Analysis, Smart Fetal Heart, Flex-M, Vascularity Index, and AI/ML-enabled Full-assist functions for left ventricle (LV) analysis and Auto EF. Output allows clinicians to review images and perform quantitative analysis to support clinical decision-making and patient assessment.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed device meets specifications. Performance of AI/ML-based features (2D Wall Motion Tracking with Full-assist, Auto EF with Full-assist) was validated by re-running original clinical datasets used in reference device (K223017) clearance, applying identical acceptance criteria and ground truth comparisons. Results demonstrate performance is consistent with reference device.
Technological Characteristics
Software-based medical image management and processing system. Features AI/ML-enabled algorithms for cardiac analysis (LV Full-assist, Auto EF). Compliant with IEC 62304 (software lifecycle) and IEC 62366-1 (usability).
Indications for Use
Indicated for use by physicians or qualified persons in hospital and clinical settings to observe and analyze ultrasound examination data acquired from TUS-AI900, TUS-AI800, and TUS-AI700 systems.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K251342 — EchoPAC Software Only / EchoPAC Plug-in · GE Medical Systems Ultrasound and Primary Care Diagnostics · Jul 16, 2025
K200852 — EchoPAC Software Only, EchoPAC Plug-In · GE Medical Systems Ultrasound and Primary Care Diagnostics · Sep 18, 2020
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 30, 2025
Canon Medical Systems Corporation
% Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems USA, Inc.
2441 Michelle Drive
Tustin, California 92780
Re: K250328
Trade/Device Name: UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, IYN, IYO
Dated: February 5, 2025
Received: February 5, 2025
Dear Yoshiaki Cook:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250328 - Yoshiaki Cook
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250328 - Yoshiaki Cook
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250328
Device Name
UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program
Indications for Use (Describe)
UltraExtend NX CUW-U001S Ultrasound Image Analysis Program is designed to allow the user to observe images and perform analysis based on examination data acquired using the following diagnostic ultrasound systems; TUS-AI900, TUS-AI800, and TUS-AI700.
This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
Canon
CANON MEDICAL SYSTEMS USA, INC.
K250328
Made For life
# 510(k) SUMMARY
1. SUBMITTER'S NAME
Fumiaki Teshima
Sr. Manager, Quality Assurance Dept.
Quality, Safety and Regulation Center
Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-shi, Tochigi-ken, Japan 324-8550
2. ESTABLISHMENT REGISTRATION
9614698
3. OFFICIAL CORRESPONDENT/CONTACT PERSON
Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems USA, Inc.
2441 Michelle Drive
Tustin, CA 92780
ycook@us.medical.canon
+1 (657) 270-5595
4. DATE PREPARED
February 05, 2025
5. DEVICE NAME/TRADE NAME
UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program
6. COMMON NAME
Radiological Image Processing System
7. DEVICE CLASSIFICATION
Class II
Medical Image Management and Processing System – Product Code: QIH [per 21 CFR892.2050]
Subsequent Product Codes:
Ultrasonic Pulsed Doppler Imaging System – Product Code: IYN [per 21 CFR 892.1550]
Ultrasonic Pulsed Echo Imaging System – Product Code: IYO [per 21 CFR 892.1560]
Page 1 of 4
{5}
Page 2 of 4
# 8. PREDICATE DEVICE
| Product | Marketed by | Regulation Number & Classification Product Code | 510(k) Number | Clearance Date |
| --- | --- | --- | --- | --- |
| UltraExtend FX, Ultrasound Workstation Package, V2.02
(Primary predicate device) | Canon Medical Systems USA, Inc. | 892.2050
LLZ | K121076 | 10/09/2012 |
| Aplio i900/i800/i700 Diagnostic Ultrasound System, Software Version 7.0
(Reference device) | Canon Medical Systems USA, Inc. | 892.1550
IYN | K223017 | 03/31/2023 |
# 9. REASON FOR SUBMISSION
Introduction of a new device.
# 10. DEVICE DESCRIPTION
The UltraExtend NX, V2.0 is designed to allow the user to observe images and perform analysis based on examination data acquired using the Aplio i900/i800/i700 diagnostic ultrasound systems. RAW only or data saved in Image + RAW should be used for UltraExtend NX.
# 11. INDICATIONS FOR USE
UltraExtend NX CUW-U001S Ultrasound Image Analysis Program is designed to allow the user to observe images and perform analysis based on examination data acquired using the following diagnostic ultrasound systems; TUS-AI900, TUS-AI800, and TUS-AI700.
This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
# 12. SUBSTANTIAL EQUIVALENCE
The UltraExtend NX, V2.0 is substantially equivalent to the UltraExtend FX, V2.02 (K121076). The subject device employs the same fundamental scientific technology as the primary predicate device and functions in a manner similar to, and is intended for the same use as the primary predicate device. The subject device differs from the primary predicate device primarily by the latter's compatibility with image data acquired by the Aplio 300, 400, and 500 Ultrasound system models, and several software features and improvements to software functionality cleared in the reference device (K223017), which are not implemented in the primary predicate device. These differences do not raise any new questions about the safety and effectiveness of the subject device, and evidence is provided in this submission to demonstrate the substantial equivalence of the subject and predicate devices.
- The subject device has the same clinical intended use as the predicate device, and both are application software products enabling post processing of image data acquired by compatible Ultrasound system models
{6}
- The software features and functionality supported in the subject and primary predicate devices are identical except for the following features and improvements to software functionality which are available with the reference device, but not implemented in the primary predicate device
- Availability of certain measurement packages (Standard measurements, Application measurements, Obstetrics measurements) which are not available, or limited in the primary predicate device
- Availability of capabilities with the Cardiac4D feature which are limited in the primary predicate device
- 2D Wall Motion Tracking, which is expanded to include left atrial analysis and the AI/ML-enabled Full-assist function for left ventricle, both of which are not available with the primary predicate device
- Auto EF LV, including with the AI/ML-enabled Full-assist function for the left ventricle, which are not available with the primary predicate device
- 3D Wall Motion Tracking, Aortic Valve Analysis, Smart Fetal Heart, Flex-M, and Vascularity Index features which are not available with the primary predicate device
## 13. SAFETY
The subject device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC 62304 Edition 1.1 2015-06 and IEC 62366-1 Edition 1.1 2020-06 standards.
## 14. TESTING
Risk Analysis and verification and validation activities demonstrate that the established specifications for these devices have been met. Additional performance testing was conducted in order to demonstrate that the requirements for the features and software functionality were met. The results of this testing demonstrate that the subject device meets established specifications and performs as intended and in accordance with labeling.
Software documentation appropriate for the Basic Documentation Level, per the FDA guidance document, "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, was included in this submission.
Additionally, cybersecurity documentation, per the FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", issued on September 27, 2023, was included in this submission.
## PERFORMANCE TESTING OF AI/ML-BASED FUNCTIONALITY MIGRATED FROM REFERENCE DEVICE (K223017)
In this submission, 2D Wall Motion Tracking with Full-assist function for left ventricle (LV) and Auto EF with Full-assist function for LV, as cleared with the reference device, were both integrated without modification into UltraExtend NX, V2.0.
Page 3 of 4
{7}
In order to demonstrate that when integrated into the subject device, both features perform with substantial equivalence as with the reference device, these studies utilized a representative subset of the clinical data acquired for the original performance testing of these features; additionally these studies applied the same acceptance criteria to evaluate the performance of these features compared to the same ground truth as utilized in the original performance evaluation of these features with the reference device.
The results of this testing demonstrate that both features perform as intended when integrated into the subject device, and with substantial equivalence as with the reference device. Appropriate instructions for use and performance information related to the AI/ML-enabled functionality are included in user labeling as included in this submission.
## 15. CONCLUSION
UltraExtend NX CUW-U001S V2.0 Ultrasound Image Analysis Program functions in a manner similar to, and is intended for the same use as the UltraExtend FX, Ultrasound Workstation Package, as described in labeling. Based upon the evidence presented in this submission, it is concluded that the subject device has demonstrated substantial equivalence to the predicate devices, and is as safe and effective for its intended use.
Page 4 of 4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.