TraumaCad Neo (1.1)

K243810 · Brainlab , Ltd. · QIH · Jun 4, 2025 · Radiology

Device Facts

Record IDK243810
Device NameTraumaCad Neo (1.1)
ApplicantBrainlab , Ltd.
Product CodeQIH · Radiology
Decision DateJun 4, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAcceptanceObservedDev DSTest DS
Implant Presence DetectionNot specified99% accuracyTrained with supervision on X-ray image data from multiple clinical sites, including a wide variety of scanner models, implants, and patient characteristics.349 original X-ray images (augmented to over 1000 images) from 186 patients, including an independent test dataset comprising approximately 28% of the test images from an independent clinical site and X-ray manufacturer.
2D Landmark DetectionAt least 80% of the analyzed femur and implant stem shafts within 4 mm distance to ground-truth landmark annotation.92% of landmarks successfully detected within 4 mm distance from ground-truth.Trained with supervision on X-ray image data from multiple clinical sites, including a wide variety of scanner models, implants, and patient characteristics.184 images from 184 patients (augmented to over 1000 images), including an independent test dataset comprising approximately 28% of the test images from an independent clinical site and X-ray manufacturer.

Indications for Use

TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones.

Device Story

TraumaCad Neo 1.1 is a software application for orthopedic surgical planning and post-operative assessment. It accepts digital X-ray images from PACS, EMR, or the Quentry cloud platform. Surgeons use the software to perform measurements, select and position digital prosthesis templates from a manufacturer library, and conduct pre-operative planning. Post-operatively, the device provides automated outcome analysis for total hip arthroplasty using an AI/ML algorithm that detects implants and anatomical landmarks. The software operates on PCs, Macs, and tablets via web or local installation. Healthcare professionals review the overlaid visual information to guide surgical decisions and workflow. The device facilitates a film-less practice and assists in pre-operative preparation and post-operative review, potentially improving surgical accuracy and efficiency.

Clinical Evidence

Bench testing only. Performance evaluated using 349 original standing pelvic X-ray images (augmented to >1000 images) from 186 patients. Dataset included diverse manufacturers, implants, and patient demographics (57% female, 43% male, ages 30-90). Implant detection accuracy was 99%. Landmark detection accuracy showed 92% of landmarks were within 4mm of ground-truth annotations, exceeding the 80% acceptance criterion. Testing included an independent dataset (28% of images) to validate generalizability.

Technological Characteristics

Software-based medical image management and processing system. Supports DICOM. Connectivity via PACS, EMR, and Quentry cloud. AI/ML-based algorithm for automatic landmark detection (non-adaptive). Operates on Windows, macOS, iOS, and Android platforms. Web and local deployment. No patient-contacting components. No life-sustaining control functions.

Indications for Use

Indicated for assisting healthcare professionals in analyzing orthopedic conditions and planning orthopedic procedures for adult patients (≥18 years old) using radiological images. Not for primary image interpretation or appraisal. Not for use on mobile phones.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 4, 2025 Brainlab Ltd. Veronika Kravtsov RA Manager 35 Efal Street Petach-Tikva, 4951132 Israel Re: K243810 Trade/Device Name: TraumaCad Neo (1.1) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: May 13, 2025 Received: May 13, 2025 Dear Veronika Kravtsov: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243810 - Veronika Kravtsov Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243810 - Veronika Kravtsov Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243810 Device Name TraumaCad Neo (1.1) Indications for Use (Describe) TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} BRAINLAB 510(k) Summary # K243810 510(k) Summary Pursuant to CFR 807.92, the following 510(k) Summary is provided: | 1. (a) | Submitter Address: | Brainlab Ltd. 35 Efal Street Petach-Tikva, Israel 4951132 | | --- | --- | --- | | 1. (b) | Manufacturer Address: | Brainlab Ltd 35 Efal Street Petach-Tikva, Israel 4951132 | | | Mfg. Phone: | Tel.: +972-3-929-0929 | | | Contact Person: | Veronika Kravtsov, RA Manager | | | Date: | May 8, 2025 | | 2. | Device & Classification: Name: | Radiological Image Processing System – classified as Class 2 QIH and LLZ Regulation Number 21 CFR 892.2050 TraumaCad Neo (1.1) | | 3. | Predicate Devices: | TraumaCad Neo (K231498) PeekMed Web (V1) (K222767) | | 4. | Description: | TraumaCad Neo 1.1 allows surgeons to evaluate digital images while performing various pre-operative surgical planning and evaluation of images. This software application enables surgeons to plan operations on screen, execute measurements, and facilitates a film-less orthopedic practice. TraumaCad Neo 1.1 also allows post-operative review and assessment of X-ray images obtained after the surgical procedure, with a feature for automatic surgery outcome analysis of postoperative total hip arthroplasty images. The program features an extensive regularly updated library of digital templates from leading manufacturers. TraumaCad Neo supports DICOM and is communicating with Quentry®, a proprietary web-based cloud service from Brainlab and with other healthcare data platforms, such as PACS solutions. It is through these healthcare data platforms, where the medical staff can upload images to plan their expected results prior to the procedure to create a smooth surgical workflow from start to finish. | | 5. | Indications for Use: | TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are | 510k Notification: TraumaCad Neo 1.1 Page 1 of 7 {5} BRAINLAB 510(k) Summary | | | required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones. | | --- | --- | --- | | 6. | Comparison of Technological Characteristics: | With respect to technology and intended use, TraumaCad Neo 1.1 is substantially equivalent to its predicate devices. Based upon the outcomes from the Risk Analysis and Performance Testing Evaluation, Brainlab Ltd. believes that the modification of TraumaCad Neo 1.0 (predicate device) which allows it to become TraumaCad Neo 1.1 does not raise additional safety or efficacy concerns. The following comparison tables depict the changes. | | Features/ Characteristics | Submitted Device: TraumaCad Neo 1.1 | Predicate Device: TraumaCad Neo 1.0 | | --- | --- | --- | | Product Code | QIH, LLZ | LLZ | | Indication for Use | TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones. | TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones. | 510k Notification: TraumaCad Neo 1.1 Page 2 of 7 {6} BRAINLAB 510(k) Summary | Features/ Characteristics | Submitted Device: TraumaCad Neo 1.1 | Predicate Device: TraumaCad Neo 1.0 | | --- | --- | --- | | Operating System | Microsoft Windows 10 and above iOS 17.x and above MAC Sequoia and above Android 14 and above | Microsoft Windows 10 and above iOS 16.x and above MAC OS 11 and above Android 11 and above | | Devices Supported | PC MAC iPads Android tablets | PC MAC iPads Android tablets | | Browsers Supported | Microsoft Edge Firefox Chrome Safari (MAC/iOS) | Microsoft Edge Firefox Chrome Safari (MAC/iOS) | | Image Input | Can receive digital images from PACS solutions, EMR systems, or from Quentry® | Can receive digital images from Quentry® | | Number of Images that can simultaneously viewed on the screen | Up to 3 | Up to 3 | | Hip Module | Yes | Yes | | Knee Module | Yes | Yes | | Foot and Ankle Module | Yes | Yes | | Upper Limb Module | Yes | Yes | | Digital Prosthetic Templates | Yes | Yes | | Interactive template positioning | Yes | Yes | | Automatic Scaling | Yes | Yes | 510k Notification: TraumaCad Neo 1.1 Page 3 of 7 {7} BRAINLAB 510(k) Summary | Features/ Characteristics | Submitted Device: TraumaCad Neo 1.1 | Predicate Device: TraumaCad Neo 1.0 | | --- | --- | --- | | Template support from manufacturers | Yes | Yes | | Permits Template Rotation | Yes | Yes | | Pre-Operative Planning | Yes | Yes | | Post Operative Review | Yes | Yes | | Patient Contacting | No | No | | Control of Life Sustaining Devices | No | No | | Healthcare professional intervention for interpretation of images | Yes | Yes | | Software Delivery | Web and locally | Web | | Post Operative Landmark Placement | Automatically (AI) for Hip images, manually for other anatomies modules | Manually for all anatomies modules | | 510(k) # | Pending | K231498 | 510k Notification: TraumaCad Neo 1.1 Page 4 of 7 {8} BRAINLAB 510(k) Summary | Features/ Characteristics | Submitted Device: TraumaCad Neo 1.1 | Predicate Device: PeekMed Web (V1) | | --- | --- | --- | | Product Code | QIH, LLZ | QIH , LLZ | | Indication for Use | TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software. The software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones. | PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers. | | Operating System | Microsoft Windows 10 and above iOS 17.x and above MAC Sequoia and above Android 14 and above | Web base | | Devices Supported | PC MAC iPads Android tablets | PC MAC | | Image Input | Can receive digital images from PACS solutions, EMR systems, or from Quentry® | Web | | Landmarking | Automatic landmarking postoperatively on Hip only. Manual for other anatomy pre and post operatively | Automatic landmarking is a feature designed to accelerate the process of placing the landmarks on each bone before planning the surgery | | Anatomic Regions for Templates | Hip, Knee, Upper Limb, Foot and Ankle | Hip, Knee and Upper Limb | | Input Images | Xray | CT and Xray | | Digital Prosthetic Templates | Yes | Yes | | Interactive template | Yes | Yes | 510k Notification: TraumaCad Neo 1.1 Page 5 of 7 {9} BRAINLAB 510(k) Summary | Features/ Characteristics | Submitted Device: TraumaCad Neo 1.1 | Predicate Device: PeekMed Web (V1) | | --- | --- | --- | | positioning | | | | Automatic Scaling | Yes | Yes | | Template support from manufacturers | Yes | Yes | | Permits Template Rotation | Yes | Yes | | Pre-Operative Planning | Yes | Yes | | Post Operative Review | Yes | Yes | | Patient Contacting | No | No | | Control of Life Sustaining Devices | No | No | | Healthcare professional intervention for interpretation of images | Yes | Yes | | 510(k) # | Pending | K222767 | 7. Performance Evaluation: The Performance Evaluation of TraumaCad Neo 1.1 was based upon well-established test methods which demonstrated conformity to the intended use. These test methods were the same which were used to demonstrate the substantial equivalence of the predicate device TraumaCad Neo (K231498). The AI/ML models incorporated into the TraumaCad Neo 1.1 were trained, tested and validated for their performance, by qualified personnel, based on predefined protocols and criteria. The automatic postoperative total hip arthroplasty analysis is achieved by the AI/ML based algorithm, by automatically identifying whether a relevant implant is present, and as a next step, if an implant is found, detecting certain landmarks on patient's anatomy and on the implant on the DICOM image. The AI/ML models were trained with supervision on X-ray image data from multiple clinical sites, including wide variety of scanner models, implants and patient characteristics. The mentioned AI/ML models are non-adaptive, i.e. do not learn from data once initially trained. The training data was totally separate from the performance testing data. Measures were taken to eliminate bias and prevent overfitting in the data. The performance testing was based on a data pool containing 349 original X-ray images, which were used in the following way: For implant detection evaluation, used all 349 images from 186 patients; For landmark detection evaluation, used 184 images from 184 patients. 510k Notification: TraumaCad Neo 1.1 Page 6 of 7 {10} BRAINLAB 510(k) Summary Each of the selected X-ray images have been augmented 3 times, in order to test the robustness of the detection algorithm to input variations, leading to a sample size of over 1000 images. The dataset included the following characteristics to ensure generalization: - All images in this dataset are standing pelvic X-rays - Pixel spacing is between 0.1 and 0.2 mm both in x and y axes - Approximately 57% of test set images are from females, while 43% are from male patients - All patients are adults (≥18 years old) typically in the age range of 30 to 90 and predominantly between the ages 50 and 80, which made up 68% of the entire test set - Consists of images from seven unique X-ray device manufacturers with 11 unique X-ray device models - Balanced distribution of implant laterality - Contains Cup and Stem Implants from multiple manufacturers in a range of sizes - An independent test dataset (comprising approximately 28% of the test images) from an independent clinical site and X-ray manufacturer was allocated to be able to quantitatively test the algorithm's generalizability to completely unseen data Accuracy of implant presence and 2D landmark detection have been tested against ground-truth annotations done by qualified and trained personnel. The pre-specified acceptance criteria required that at least 80% of the analyzed femur and implant stem shafts were within 4 mm distance to their ground-truth landmark annotation. The results showed that 99% of the time the machine learning algorithm was able to correctly determine the implant presence and 92% of the landmarks were successfully detected automatically within 4mm distance from their corresponding ground-truth landmark annotations. Therefore, tests showed that the AI/ML model yields acceptable performance for the intended use. 8. Conclusion: The intended use and the fundamental technological characteristics of TraumaCad Neo 1.1 are the same as those in the TraumaCad Neo (1.0), which is the predicate device. Any additions or differences do not affect the safety and effectiveness of the device. The performance tests have been completed and successfully confirm the performance of the device. Based upon this data, Brainlab Ltd believes that TraumaCad Neo 1.1 is substantially equivalent to the predicate devices. 510k Notification: TraumaCad Neo 1.1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...