AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K232613 · Feb 28, 2024
CT Cardiomegaly
Innolitics, LLC
University of Alabama at Birmingham imaging database (retrospective clinical CT studies)
Retrospective clinical CT imaging data was used to train and test the non-adaptive machine learning algorithm for the automated calculation of cardiothoracic ratio (CTR).
Retrospective analysis of clinical imaging database
Adults aged 18-96 years (median 62) with various clinical conditions and imaging protocols (abdominal CT, chest CT, PET/CT, high-resolution chest CT).; Sample Size: 1,875 total cases (1,600 training, 275 testing); Number of Sites: Over 300 unique imaging sites (training); 12 U.S. states (testing)
Manual segmentation by three board-certified radiologists (reference standard)
Accuracy of linear and area-based CTR measurements; Key heart slice detection; Segmentation Dice scores.
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiothoracic Ratio
Non-adaptive machine learning algorithm using the MONAI framework
Mean difference < 0.0100
Mean difference 0.0012 [95% CI: -0.0008, 0.0033]
1600 unique patient cases from 267 unique sites across 20 U.S. states.
>1 (board-certified radiologists)
275 unique patient cases from sites across 12 U.S. states.
3 (board-certified radiologists)
Heart to Chest Area Ratio
Non-adaptive machine learning algorithm using the MONAI framework
Mean difference < 0.0100
Mean difference 0.0036 [95% CI: 0.0017, 0.0055]
1600 unique patient cases from 267 unique sites across 20 U.S. states.
>1 (board-certified radiologists)
275 unique patient cases from sites across 12 U.S. states.
3 (board-certified radiologists)
Indications for Use
CT Cardiomegaly is a command line software application intended to be run on its own or as part of another medical device to automatically calculate linear and area based cardiothoracic ratio (CTR) from a CT image containing the heart. CT Cardiomegaly is designed to measure the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity and calculate the CTR from an axial CT slice containing the heart using a non-adaptive machine learning algorithm. Intended users of the software are aimed to the physicians or other licensed practitioners in the healthcare institutions, such as clinics, hospitals, healthcare facilities, residential care facilities and long-term care services. The system is suitable for adults and transitional adolescents (18 to 21 years old but treated as an adult). Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease.
Device Story
CT Cardiomegaly is a command-line SaMD that processes chest CT images to calculate cardiothoracic ratio (CTR). It automatically identifies the heart and thoracic cavity, segments these structures, and performs linear and area-based measurements on an axial CT slice. The software is intended for use by physicians and licensed practitioners in clinical settings. Output is provided as a summary report (PDF/JSON) containing CTR values and patient information. The device does not provide a diagnosis; clinicians use the output to inform further diagnostic work-up. The device benefits patients by providing automated, standardized measurements to assist in the detection of cardiomegaly.
Clinical Evidence
Performance assessment using 275 patient cases (142 male, 133 female, ages 18-96) from 12 U.S. states. Primary endpoints: mean difference between device and human reader measurements. Results: Mean difference for CTR was 0.0012 (95% CI: -0.0008, 0.0033) and for Heart to Chest Area Ratio was 0.0036 (95% CI: 0.0017, 0.0055), both meeting targets (<0.01). Secondary endpoints: Key heart slice detection mean difference 0.46 mm; segmentation Dice scores 0.95 (heart) and 0.98 (inner chest).
Technological Characteristics
SaMD; non-adaptive machine learning algorithm using MONAI framework. Inputs: CT image studies. Outputs: PDF/JSON reports. Software runs on Ubuntu 20.04.5 LTS via Docker container. Automated segmentation and measurement of heart and thoracic cavity.
Indications for Use
Indicated for adults and transitional adolescents (18-21 years) to automatically calculate linear and area-based cardiothoracic ratio (CTR) from axial CT slices containing the heart. Not for stand-alone clinical decision-making.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System (K212624)
Submission Summary (Full Text)
{0}------------------------------------------------
February 28, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Innolitics, LLC % Meritxell Martinez Project Manager and Regulatory Specialist 1101 West 34th St Austin, Texas 78705
Re: K232613
Trade/Device Name: CT Cardiomegaly Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: OIH Dated: January 30, 2024 Received: January 30, 2024
Dear Meritxell Martinez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Samul Zik for
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K232613
Device Name CT Cardiomegaly
#### Indications for Use (Describe)
CT Cardiomegaly is a command line software application intended to be run on its own or as part of another medical device to automatically calculate linear and area based cardiothoracic ratio (CTR) from a CT image containing the heart.
CT Cardiomegaly is designed to measure the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity and calculate the CTR from an axial CT slice containing the heart using a non-adaptive machine learning algorithm.
Intended users of the software are aimed to the physicians or other licensed practitioners in the healthcare institutions, such as clinics, hospitals, healthcare facilities, residential care facilities and long-term care services.
The system is suitable for adults and transitional adolescents (18 to 21 years old but treated as an adult).
Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for "Body Check". The logo consists of a green circle with a check mark that is partially navy blue. The text "Body Check" is to the right of the logo, with "Body" in green and "Check" in navy blue.
# K232613
## 1. ADMINISTRATIVE INFORMATION
| Submitter Name | Innolitics LLC |
|------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Address | 1101 West 34th St #550, Austin, TX 78705, US |
| Phone Number | +1 (512) 967-6088 |
| Fax Number | N/A |
| Company Representative | Meritxell Martinez |
| Additional Company Representatives | Yujan Shrestha, MD<br>Ethan Ulrich<br>Andrew Smith, MD, PhD<br>Steven Rothenberg, MD<br>Jim Luker, RN, MS |
| Email | fda@innolitics.com |
| Date Summary Prepared | January 30th, 2024 |
| Documentation Prepared Using | Innolitics Medtech OS Framework |
## 2. SUBJECT DEVICE INFORMATION
| Trade Name | CT Cardiomegaly |
|-------------------------|--------------------------------------------------|
| Subject Device K Number | K232613 |
| Common Name | Automated Radiological Image Processing Software |
| Product Code | QIH |
| Regulation Number | 892.2050 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
Image /page/3/Picture/7 description: The image contains the logo for Innolitics SaMD Simplified. The logo consists of a blue hexagon-like shape on the left, followed by the text "Innolitics" in a bold, blue font. Below "Innolitics" is the text "SaMD Simplified" in a smaller, lighter blue font. The logo is clean and professional, suggesting a company in the technology or software industry.
{4}------------------------------------------------
## 3. PREDICATE DEVICE INFORMATION
| Predicate Device Name | EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System |
|---------------------------|----------------------------------------------------------------|
| Predicate Device K Number | K212624 |
| Common Name | EFAI iCTR |
| Product Code | QIH |
| Regulation Number | 892.2050 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
## 4. DEVICE DESCRIPTION
The CT Cardiomegaly application is a software only (SaMD) medical device which includes automated algorithms and non-adaptive machine learning to analyze chest computed tomography (CT) images. CT Cardiomegaly is a command line software intended to be run on its own or as part of another medical device.
CT Cardiomegaly scans the existing images for the presence of specific anatomical structures (i.e., heart and chest cavity). If the target structures are present, the software segments and performs automated measurements which may be helpful in assisting with the detection of certain diseases and/or conditions (i.e., cardiomegaly). The quantitative information is provided to the clinician in the form of a summary report. The clinician (using all available information) then decides if further diagnostic work-up, analysis or measurements are indicated.
The results from CT Cardiomegaly are not intended to be used as the primary input for clinicaldecision making or diagnosis.
A reference standard was created to train the machine learning algorithms in CT Cardiomegaly. Data for training and testing were sampled from this database based on exclusion criteria. Data was also selected to ensure no site was shared by training and testing cohorts. The training dataset included 1600 unique patient cases from 267 unique sites across 20 U.S. states.
Image /page/4/Picture/9 description: The image contains the logo for Innolitics. The logo consists of a blue hexagon-like shape on the left, followed by the word "Innolitics" in large, bold, blue font. Below "Innolitics" is the phrase "SaMD Simplified" in a smaller, lighter blue font. The logo is clean and modern, with a focus on simplicity and clarity.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Body Check. The logo consists of a green and blue checkmark inside of a green circle. To the right of the logo is the text "Body Check", with "Body" in green and "Check" in blue.
## 5. INDICATIONS FOR USE
CT Cardiomegaly is a command line software application intended to be run on its own or as part of another medical device to automatically calculate linear and area based cardiothoracic ratio (CTR) from a CT image containing the heart.
CT Cardiomegaly is designed to measure the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity and calculate the CTR from an axial CT slice containing the heart using a non-adaptive machine learning algorithm.
Intended users of the software are aimed to the physicians or other licensed practitioners in the healthcare institutions, such as clinics, hospitals, healthcare facilities, residential care facilities and long-term care services.
The system is suitable for adults and transitional adolescents (18 to 21 years old but treated as an adult).
Its results are not intended to be used on a stand-alone basis for clinical-decision making or otherwise preclude clinical assessment of any disease.
## 5.1. Indications for Use Equivalence Discussion
Both the subject (CT Cardiomegaly) and predicate device (EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System) have similar indications for use, with the exception that the subject device analyzes CT images while the predicate device is limited to X-ray images.
Another difference between the indications for use for both devices is that the predicate uses an artificial intelligence algorithm using an unknown framework while the subject device utilizes a non-adaptive machine learning algorithm using the MONAI framework.
However, these are differences in technological characteristics that do not affect safety and effectiveness of the device, and have been supported with appropriate verification and validation testing.
| Characteristic | CT Cardiomegaly | EFAI Intelligent<br>Cardiothoracic Ratio<br>(iCTR) Assessment<br>System (K212624) | Substantial Equivalence Discussion |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Innolitics LLC | Ever Fortune. AI Co.,<br>Ltd. | N/A |
| Characteristic | CT Cardiomegaly | EFAI Intelligent<br>Cardiothoracic Ratio<br>(iCTR) Assessment<br>System (K212624) | Substantial Equivalence Discussion |
| Regulation<br>Number | 892.2050 | 892.2050 | Same |
| Regulatory Class | Class II | Class II | Same |
| Product Code | QIH | QIH | Same |
| Regulation Name | Medical image processing<br>and management system | Medical image processing<br>and management system | Same |
| Device Property | SaMD | SaMD | Same |
| Intended Use/<br>Indications for Use | CT Cardiomegaly is<br>software intended to be run<br>on its own or as part of<br>another medical device to<br>automatically calculate<br>linear and area based<br>cardiothoracic ratio (CTR)<br>from a CT image<br>containing the heart.<br>CT Cardiomegaly is<br>designed to measure the<br>maximal transverse<br>diameter of heart and<br>maximal inner transverse<br>diameter of thoracic cavity<br>and calculate the CTR from<br>an axial CT slice<br>containing the heart using a<br>non-adaptive machine<br>learning algorithm.<br>Intended users of the<br>software are aimed to the<br>physicians or other<br>licensed practitioners in the<br>healthcare institutions,<br>such as clinics, hospitals,<br>healthcare facilities,<br>residential care facilities<br>and long-term care<br>services.<br>The system is suitable for<br>adults and transitional<br>adolescents (18 to 21 years<br>old but treated as an adult).<br>Its results are not intended<br>to be used on a stand-alone<br>basis for clinical-decision | EFAI Intelligent<br>Cardiothoracic Ratio<br>Assessment System (or<br>iCTR) is a software for use<br>by hospital and clinics to<br>automatically assess the<br>cardiothoracic ratio (CTR)<br>of a chest X-ray image from<br>the X-ray imager subject.<br>The iCTR is designed to<br>measure the maximal<br>transverse diameter of heart<br>and maximal inner<br>transverse diameter of<br>thoracic cavity and calculate<br>the CTR of a chest X-ray<br>image in posterior-anterior<br>(PA) chest view using an<br>artificial intelligence<br>algorithm.<br>Intended users of the<br>software are aimed to the<br>physicians or other licensed<br>practitioners in the<br>healthcare institutions, such<br>as clinics, hospitals,<br>healthcare facilities,<br>residential care facilities and<br>long-term care services.<br>The system is suitable for<br>adults between 20 - 80 years<br>of age.<br>Its results are not intended<br>to be used on a stand-alone<br>basis for clinical-decision<br>making or otherwise<br>preclude clinical assessment | Both the subject (CT Cardiomegaly)<br>and predicate device (iCTR) have<br>identical indications for use, with the<br>exception that the subject device<br>analyzes CT images while the predicate<br>device is limited to X-ray images.<br>Another difference between the<br>indications for use for both devices is<br>that the predicate uses artificial<br>intelligence algorithms using an<br>unknown framework while the subject<br>device utilizes non-adaptive machine<br>learning algorithms using the MONAI<br>framework.<br>However, these differences in<br>technological characteristics do not<br>affect safety and effectiveness of the<br>device, and have been supported with<br>verification and validation testing. |
| Characteristic | CT Cardiomegaly | EFAI Intelligent<br>Cardiothoracic Ratio<br>(iCTR) Assessment<br>System (K212624) | Substantial Equivalence Discussion |
| | making or otherwise<br>preclude clinical<br>assessment of any disease. | of cardiothoracic ratio<br>(CTR) cases. | |
| Technical<br>Characteristics | | | |
| Input | CT image study | Post-anterior (PA) view<br>chest X-ray image | The input for the predicate device is a<br>2-dimensional chest X-ray image, while<br>the subject device's input is an entire<br>CT study. The subject device will select<br>out of the CT study an axial slice<br>containing the heart, if present, and<br>perform measurements. The main<br>difference is that the predicate operates<br>on images that are post-anterior (PA)<br>view chest X-ray, while the subject<br>device's operates on images that are<br>axial CT. Nonetheless, this difference<br>in imaging modality does not raise any<br>new concerns and has been verified and<br>validated in testing. |
| Output files | PDF and JSON Reports | Reports, DICOM<br>Secondary Capture<br>series | The output format of the predicate<br>device includes reports and DICOM<br>secondary capture series, while the<br>subject device's only output format is a<br>report (in both PDF and JSON format).<br>Thus, the subject device's output<br>format is a subset of the predicate's<br>available output formats (i.e., report).<br>This lack of a feature does not raise any<br>new concerns. |
| Output<br>measurements | Linear-based CTR and<br>area-based CTR | Linear-based CTR | The predicate device outputs a linear-<br>based CTR measurement while the<br>subject device outputs both linear and<br>area-based CTR measurements.<br>Although this output measurement of<br>area-based CTR is not included in the<br>predicate device, it does not raise<br>different questions of safety or<br>effectiveness and has been properly<br>verified and validated in testing. |
| Report Structure | Report will be output in<br>the PDF and JSON<br>file format which is<br>structured with following | Report will be output in<br>the DICOM and JSON<br>file format which is<br>structured with following | Both devices produce reports that<br>output CTR and patient information.<br>One of the main differences between<br>both devices is that the predicate only |
| Characteristic | CT Cardiomegaly | EFAI Intelligent<br>Cardiothoracic Ratio<br>(iCTR) Assessment<br>System (K212624) | Substantial Equivalence Discussion |
| | information:<br>1) CTR (both linear and<br>area-based)<br>2) Patient information<br>3) Reference value(s) | information and function<br>tool:<br>1) CTR<br>2) adjustable annotation line<br>(maximal inner border<br>diameter of thoracic cavity<br>and maximal diameter of<br>heart)<br>3) the trajectory of CTR<br>(including chest X-ray)<br>4) Patient information…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.