K231149 · Sep 22, 2023
qXR-CTR
Qure.Ai Technologies
Retrospectively collected chest X-ray scans
Retrospective clinical chest X-ray scans were used to validate the performance of the qXR-CTR software in measuring cardiac and thoracic diameters compared to radiologist-established ground truth.
Adults (implied by device indications); Sample Size: 435 scans
Not applicable for this study
Accuracy of cardiac diameter and thoracic diameter measurements (RMSE and Mean Absolute Error)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiothoracic Ratio (CTR)
Deep-learning based software
—
RMSE: 7.55 mm (Cardiac Diameter), 5.43 mm (Thoracic Diameter); Mean Absolute Error: 5.66 mm (Cardiac Diameter), 4.04 mm (Thoracic Diameter); CTR: 0.02
—
—
Clinical study: 435 scans with PA view from various parts of the US.
3 (ABR thoracic radiologists)
Indications for Use
qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans. qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm. The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities. The system is suitable for adults ≥ 22 years of age. The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.
Device Story
qXR-CTR is a non-invasive software medical device for automated cardiothoracic ratio (CTR) assessment on PA chest X-rays. Input: DICOM PA chest X-ray images. Processing: AI algorithm measures maximal transverse heart diameter (CD) and maximal inner thoracic diameter (TD) to calculate CTR. Output: Structured report with CTR value, annotation lines, and preview image. Usage: Installed on hospital/clinic computers; operated by physicians or licensed practitioners. Workflow: Integrates into PACS or web-based environments; provides quantitative data to aid clinical assessment. Benefit: Assists clinicians in evaluating cardiac size; does not replace clinical judgment.
Clinical Evidence
Retrospective clinical study of 435 PA chest X-rays from US sites. Ground truth established by 3 ABR-certified thoracic radiologists (>10 years experience). Primary endpoints: RMSE and Mean Absolute Error (MAE) for cardiac and thoracic diameter measurements. Results: Cardiac diameter RMSE 7.55 mm (95% CI: 6.95, 8.34); Thoracic diameter RMSE 5.43 mm (95% CI: 4.95, 6.11). MAE (SD) for cardiac diameter: 5.66 (5.0) mm; for thoracic diameter: 4.04 (3.63) mm. Performance exceeded predicate device metrics.
Technological Characteristics
Software-only medical device; runs on off-the-shelf hardware (Ubuntu 20.04+). Uses deep-learning AI algorithm for image analysis. Inputs: DICOM PA chest X-rays. Outputs: Structured reports (PDF/JSON/DICOM SR). Connectivity: PACS integration or web-based download. No physical materials or energy sources.
Indications for Use
Indicated for adults ≥ 22 years of age for automated assessment of cardiothoracic ratio (CTR) on posterior-anterior (PA) chest X-rays in clinical settings. Not for stand-alone clinical decision-making.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System (K212624)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
Qure.ai Technologies % Srinidhi Ragunathan Regulatory Clinical Affairs Manager Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon(E) Mumbai, Maharashtra 400063 INDIA
September 22, 2023
Re: K231149
Trade/Device Name: qXR-CTR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: August 24, 2023 Received: August 24, 2023
Dear Srinidhi Ragunathan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231149
Device Name qXR-CTR
## Indications for Use (Describe)
qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.
qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.
The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.
The system is suitable for adults > 22 years of age.
The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Qure.ai's qXR-CTR
#### 1 SUBMITTER
Qure.ai Technologies Level 7, Commerz II, International Business Park Oberoi Garden City, Goregaon (E), Mumbai 400 063 Phone: +91-9768123013 Primary Contact Person: Srinidhi Ragunathan Secondary contact person: Bunty Kundnani
Date Prepared: August 23, 2023
#### 2 DEVICE
| Name of Device: | qXR-CTR |
|-----------------------|--------------------------------------------------|
| Common or Usual Name: | Automated radiological image processing software |
| Classification Name: | Medical image management and processing system |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | QIH |
#### 3 PREDICATE DEVICE
| Name of Device: | EFAI Intelligent Cardiothoracic Ratio (iCTR) Assessment System |
|-----------------|----------------------------------------------------------------|
| Manufacturer: | Ever Fortune.Al Co., Ltd. |
| 510(k) Number: | K212624 |
{4}------------------------------------------------
#### 4 INTENDED USE / INDICATIONS FOR USE:
qXR-CTR is a deep-learning based software for use by hospitals and clinics for automated assessment of the CTR on chest X-ray (CXRs) scans.
qXR-CTR is designed to measure the ratio of the maximal transverse diameter of the heart (CD) and the maximal inner transverse diameter (TD) of the thoracic cavity and calculate the CTR value on posterior-anterior view chest view using an artificial intelligence algorithm.
The intended users of this device are physicians or licensed practitioners in healthcare institutions, such as clinics, hospitals, residential care facilities, long-term care services, and healthcare facilities.
The system is suitable for adults ≥ 22 years of age.
The device is used to aid the intended users and results are not intended to be used on a stand-alone basis for clinical decision making or otherwise preclude clinical assessment of CTR cases.
#### ഗ DEVICE DESCRIPTION
The qXR-CTR is a non-invasive software medical device designed to be installed on the computer with specific system requirements. It is a radiological computer-assisted software system that automatically analyzes DICOM chest X-ray images in PA view and outputs the cardiac diameter, thoracic diameter, and CTR through an artificial intelligence algorithm. The structured report includes a preview of the compressed chest X-ray image with the automatically derived CTR result and annotation line, indicating the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity.
#### ട COMPARISON OF THE PREDICATE DEVICE
qXR-CTR is technologically similar to the predicate device, EFIA iCTR Assessment in regard to intended use and technological characteristics. Both are medical image management and processing system intended to read chest X-rays to measure target structures (cardiac and thoracic diameter). The algorithms function similarly and there are no notable technological differences between the subject and predicate devices.
In terms of establishing substantial equivalence, the subject and predicate device have the same intended use, as an image processing tool that measures anatomical structures and produces caselevel output. The indications for use proposed for the subject device are similar those of the predicate device.
| | Predicate Device | Subject Device |
|------------------------|----------------------------------------------------------------------|--------------------------------------------|
| Device Name | EFAI Intelligent Cardiothoracic<br>Ratio (iCTR) Assessment<br>System | qXR-CTR |
| 510(k) Number | K212624 | NA |
| Regulation | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Regulation Description | radiological computer-<br>assisted software | radiological computer-assisted<br>software |
| Table 1 Comparison between gXR-CTR and the Predicate Device | | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | Subject Device |
| Product Code | QIH | QIH |
| Device type | Quantifying software | Quantifying software |
| Manufacturer | Ever Fortune.Al Co., Ltd. | Qure.ai Technologies |
| Intended use /<br>Indications for Use | EFAI Intelligent Cardiothoracic<br>Ratio Assessment System (or<br>iCTR) is a software for use by<br>hospitals and clinics to<br>automatically assess the<br>cardiothoracic ratio (CTR) of a<br>chest X-ray image from the X-<br>ray imager subject. The iCTR is<br>designed to measure the<br>maximal transverse diameter<br>of heart and maximal inner<br>transverse diameter of<br>thoracic cavity and calculate<br>the CTR of a chest X-ray image<br>in posterior-anterior (PA)<br>chest view using an artificial<br>intelligence algorithm.<br>Intended users of the<br>software are aimed to the<br>physicians or other licensed<br>practitioners in the healthcare<br>institutions, such as clinics,<br>hospitals, healthcare facilities,<br>residential care facilities and<br>long-term care services. The<br>system is suitable for adults<br>between 20 - 80 years of age.<br>Its results are not intended to<br>be used on a stand-alone<br>basis for clinical decision<br>making or otherwise preclude<br>clinical assessment of<br>cardiothoracic ratio (CTR)<br>cases. | qXR-CTR is a deep-learning based<br>software for use by hospitals and<br>clinics for automated assessment of<br>the CTR on chest X-ray (CXRs) scans.<br><br>qXR-CTR is designed to measure the<br>ratio of the maximal transverse<br>diameter of the heart (CD) and the<br>maximal inner transverse diameter<br>(TD) of the thoracic cavity and<br>calculate the CTR value on posterior-<br>anterior view chest view using an<br>artificial intelligence algorithm.<br><br>The intended users of this device are<br>physicians or licensed practitioners in<br>healthcare institutions, such as clinics,<br>hospitals, residential care facilities,<br>long-term care services, and<br>healthcare facilities.<br>The system is suitable for adults ≥ 22<br>years of age.<br>The device is used to aid the intended<br>users and results are not intended to<br>be used on a stand-alone basis for<br>clinical decision making or otherwise<br>preclude clinical assessment of CTR<br>cases. |
| Algorithm for<br>measurement of target<br>structures | AI Algorithm | AI Algorithm |
| Modality | Chest X-ray in Digital<br>Radiography (DR) | Chest X-ray in Digital Radiography (DR) |
| Input | Post-anterior (PA) view chest<br>X-ray image | Post-anterior (PA) view chest X-ray<br>image |
| Input format | DICOM | DICOM |
| Output Format | Reports, DICOM Secondary<br>Capture series | Pdf report is generated, DICOM<br>Secondary capture is returned to the<br>PACS |
| Report Structure | Report will be output in the<br>DICOM and JSON file format | Report will be output in the DICOM<br>and JSON file format which is<br>structured with following information |
| | Predicate Device | Subject Device |
| | | |
| | following information and<br>function tool: 1) CTR 2)<br>adjustable annotation line<br>(maximal inner border<br>diameter of thoracic cavity<br>and maximal diameter of<br>heart) 3) the trajectory of CTR<br>(including chest X-ray) | and function tool: 1) CTR 2) annotation<br>line (maximal inner border diameter of<br>thoracic cavity and maximal diameter<br>of heart) |
| Intended User | Physicians or other licensed<br>practitioners in the healthcare<br>institutions | Physicians or other licensed<br>practitioners in the healthcare<br>institutions |
| Target Population | 20-80 years | ≥ 22 years |
| Modality | Chest X-ray in Digital<br>Radiography | Chest X-ray in Digital Radiography |
| Intended Use<br>Environment | Healthcare institutions<br>(Clinics, hospitals, healthcare<br>facilities, residential care<br>facilities and long-term care<br>services) | Healthcare institutions (Clinics,<br>hospitals, healthcare facilities,<br>residential care facilities and long-term<br>care services) |
| Software device that | Yes | Yes |
| operates on off the shelf | | |
| hardware | | |
| Software devices uses | Yes | Yes |
| software algorithms for<br>image | | |
| Diameter Measurement | Yes | Yes |
| Storage | Saved in JSON and DICOM file<br>format text for DICOM and<br>DICOM file format | The text report is generated with the CXR<br>scan on a Structured Report (SR) format for<br>PACS-based mode. For the Web-based<br>mode, the user can download a PDF<br>version of the report with the same<br>information |
| Software Requirement | Ubuntu 18.04 (Web browser:<br>chrome 88.0.4324.182 or<br>above) | Ubuntu 20.04+ |
| Performance metrics of the predicate device and qXR-CTR | | |
| Performance level | RMSE: | RMSE: |
| | Cardiac Diameter = 8.81 mm<br>Thoracic Diameter = 14.4 mm | Cardiac Diameter = 7.55 mm<br>Thoracic Diameter = 5.43 mm |
| | Mean Absolute Error:<br>NR | Mean Absolute Error:<br>5.66 (5.0) mm |
| Measurement | Root Mean Squared Error (95%<br>CI) in mm | Mean Absolute Error (SD) in mm…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.