← Product Code [QIH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QIH) · K230150

# OptimMRI (K230150)

_Rebrain, Sas · QIH · Jul 21, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QIH/K230150

## Device Facts

- **Applicant:** Rebrain, Sas
- **Product Code:** [QIH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QIH.md)
- **Decision Date:** Jul 21, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

OptimMRI is a software application intended to aid qualified medical professionals in processing, visualizing, and interpreting anatomical structures from medical images. The software can be used to process pre-operative DICOM compatible MR images to generate 3D annotated models of the brain that aid the user in neurosurgical functional planning. The annotated MR images can further be used in conjunction with other clinical methods as an aid in localization of the Subthalamic Nuclei (STN) and Ventral Intermediate Nucleus (VIM) regions of interest.

## Device Story

OptimMRI is a software application for neurosurgical functional planning. It takes pre-operative DICOM MR images as input. Using semi-automatic machine learning models, it processes these images to generate 3D annotated markers for STN and VIM brain regions. The software is used by qualified medical professionals in a clinical setting. Output consists of 3D annotated models and images, which are reviewed and interpreted by the clinician to assist in surgical planning and localization of target regions. The device aids in clinical decision-making by providing visual anatomical guidance for neurosurgical procedures.

## Clinical Evidence

Performance evaluation based on retrospective analysis of 44 cerebral MRIs (88 hemispheres) for STN and 31 cerebral MRIs (62 hemispheres) for VIM. Accuracy compared against reference devices Guide XT (K213930) and SureTune4 (DEN210003). Results showed ≥90% of surface distances for STN and VIM were ≤ 2.0mm. STN performance was identical to the predicate SIS Software.

## Technological Characteristics

Software-based medical image processing system. Utilizes machine learning models for semi-automatic anatomical annotation. Compatible with DICOM standard. Complies with ISO 14971, IEC 62304, IEC 82304-1, ISO 15223-1, IEC 62366-1, ANSI UL 2900-1, and IEC 80001-1 standards. Cybersecurity controls implemented per AAMI TIR57.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- SIS Software ([K162830](/device/K162830.md))

## Reference Devices

- Guide XT ([K213930](/device/K213930.md))
- SureTune4 ([DEN210003](/device/DEN210003.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 21, 2023

RebrAIn, SAS % Ram Bedi Consultant Plateforme Technologique d'Innovation Biomédicale (PTIB) Hôpital Xavier Arnozan Avenue du Haut Lévêque 33600 Pessac. France

Re: K230150

Trade/Device Name: OptimMRI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH, LLZ Dated: June 21, 2023 Received: June 21, 2023

Dear Ram Bedi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230150

Device Name OptimMRI

#### Indications for Use (Describe)

OptimMRI is a software application intended to aid qualified medical professionals in processing, visualizing, and interpreting anatomical structures from medical images. The software can be used to process pre-operative DICOM compatible MR images to generate 3D annotated models of the brain that aid the user in neurosurgical functional planning. The annotated MR images can further be used in conjunction with other clinical methods as an aid in localization of the Subthalamic Nuclei (STN) and Ventral Intermediate Nucleus (VIM) regions of interest.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for RebrAIn. The word "RebrAIn" is written in a sans-serif font, with the "Rebr" portion in teal and the "AIn" portion in dark purple. Above the "AIn" portion of the logo is a stylized flower-like design with teal leaves and dark purple berries.

# Chapter 6. 510(k) Summary

| Submitter's Name:     | RebrAln, SAS                                                                                                                                                                                                                                                                                                                                                                                                      |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:              | Plateforme Technologique d'Innovation Biomédicale (PTIB) Hôpital Xavier Arnozan Avenue du Haut Lévêque 33600 Pessac, France                                                                                                                                                                                                                                                                                       |
| Contact: Telephone:   | Ram Bedi, PhD, MBA (425) 985 86780                                                                                                                                                                                                                                                                                                                                                                                |
| Date:                 | January 17, 2023                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device/Trade Name:    | OptimMRI                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:          | Medical Image Processing Software                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name:  | System, Image Processing, Radiological Picture Archiving and Communications System                                                                                                                                                                                                                                                                                                                                |
| Classification Panel: | Radiology                                                                                                                                                                                                                                                                                                                                                                                                         |
| CFR Code:             | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification:       | II                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code:         | Primary: QIH                                                                                                                                                                                                                                                                                                                                                                                                      |
|                       | Secondary: LLZ                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device:     | K162830 – SIS Software, Surgical Information Sciences, Inc.                                                                                                                                                                                                                                                                                                                                                       |
| Device Description:   | OptimMRI is a software application for processing medical images of the brain that enables 3D visualization and analysis of anatomical structures. Specifically, the software can be used to read DICOM compatible pre-operative MR images acquired by commercially available imaging devices. These images can be processed to generate 3D markers in specific regions of the anatomy to allow qualified medical |

______________________________________________________________________________________________________________________________________________________________________________

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Rebrain. The word "Rebrain" is written in a sans-serif font, with the "Rebr" portion in teal and the "AIn" portion in dark purple. Above the "AIn" portion of the logo is a teal flower-like design with dark purple accents.

professionals to display, review, analyze, annotate, interpret, export, and plan neurosurgical functional procedures. OptimMRI is used as an aid to localize regions of the brain such as Subthalamic Nuclei (STN) and Ventral Intermediate Nucleus (VIM).

OptimMRI complies with the following standards:

ISO, 14971 Third Edition 2019-12, Medical devices -Application of Risk Management to medical devices

IEC, 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, Medical device software - Software life cycle processes

IEC, 82304-1 Edition 1.0 2016-10, Health software - Part 1 : General requirements for product safety

ISO 15223-1:2021, Medical devices - Symbols to be used with information to be supplied by the manufacturer — Part 1

IEC, /TR 80002-1 Edition 1.0 2009-09, Medical device software – Part 1 : Guidance of the application of ISO 14971 to medical device

ISO 20417 First edition 2021-04 Corrected version 2021-12 Medical devices - Information to be supplied by the manufacturer

IEC 62366-1:2015+AMD1:2020 Medical devices - Part 1: Application of usability engineering to medical devices + Amend. 1

ANSI UL 2900-1 Ed.1 2017 Standard for Software Cybersecurity Network-Connectable Products, Part I: General Requirements.

IEC 80001-1 Edition 1.0 2010-10 Application of risk management for IT-networks incorporating medical devices -Part 1: Roles, responsibilities and activities.

AAMI TIR57:2016 Principles for medical device security -Risk management.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for RebrAIn. The word "Rebr" is in a teal color, and the "AIn" is in a dark red color. Above the "AIn" is a teal and dark red design that looks like leaves.

Indications for Use:

OptimMRI is a software application intended to aid qualified medical professionals in processing, visualizing, and interpreting anatomical structures from medical images. The software can be used to process pre-operative DICOM compatible MR images to generate 3D annotated models of the brain that aid the user in neurosurgical functional planning. The annotated MR images can further be used in conjunction with other clinical methods as an aid in localization of the Subthalamic Nuclei (STN) and Ventral Intermediate Nucleus (VIM) regions of interest.

### Comparison to Predicate Device

The OptimMRI software is substantially equivalent to the predicate that has already been cleared for USA distribution with 510(k) premarket notification number K162830. Briefly, the subject and predicate devices are based on the following same technological elements:

- . Software tools for the visualization of DICOM compliant medical images
- . Permit specialized annotation of anatomic structures using machine learning models
- . Output data compatible with multiple surgical planning systems
- Support for 2D and 3D visualization tools ●
- Export of images and reports in DICOM format ●

The following technological differences exist between the subject and predicate device:

- . Subject device is indicated for only pre-operative usaqe
- Segmentation process is semi-automatic for subject . device compared to fully automatic for the predicate device
- . Fusion imaging capability is not available in the subject device

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for RebrAIn. The first part of the word, "Rebr" is in teal, while the "A" is in maroon. The "In" is in teal again. Above the "A" is a teal flower-like design with a maroon center.

- Region of interest is annotated with 3D crosses on the ● subject device compared to 3D segmented anatomical volumes on the predicate
- Localization of VIM region of interest is not available on . the predicate device

Accuracy of segmentations for OptimMRI was compared to previously cleared commercially available comparable software tools. A total of 44 cerebral MRIs (88 hemispheres) were retrospectively annotated for the STN performance study, and 31 cerebral MRIs (62 hemispheres) were used for the VIM study. Qualified and experienced medical professionals performed the segmentation, and all validation criteria were met. The performance evaluation studies against reference devices Guide XT (K213930) and SureTune4 (DEN210003) demonstrated that at least 90% of surface distances of STN or VIM were not greater than 2.0mm when using segmentation tools for OptimMRI. The STN result is identical to the predicate SIS Software that used highresolution 7T MRIs of the brain.

No new safety or efficacy issues were introduced by OptimMRI compared to the predicate device. There are no differences in indications for use between OptimMRI and the Software (K162830). predicate ટાટ Furthermore, performance data demonstrate that the functionality, output and clinical usage of OptimMRI is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QIH/K230150](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QIH/K230150)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
