K221100 · Viz. Ai, Inc. · QIH · Aug 29, 2022 · Radiology
Device Facts
Record ID
K221100
Device Name
Viz RV/LV
Applicant
Viz. Ai, Inc.
Product Code
QIH · Radiology
Decision Date
Aug 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Right and left ventricular diameter measurement
Artificial intelligence algorithm
MAE < 7.2 mm
MAE < 7.2 mm
13 sites used as part of the training data set.
—
Pivotal study using 4 clinical sites.
>1 (trained radiologists)
Indications for Use
The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.
Device Story
Viz RV/LV is a software-only device analyzing volumetric CT pulmonary angiogram (CTPA) scans. It uses a locked AI/ML algorithm to automatically segment the heart ventricles, identify their maximal diameters, and calculate the RV/LV ratio. The device operates on Viz.ai's backend server; input scans are forwarded from CT scanners. Output includes an annotated DICOM image series with RGB overlays (marking diameters and interventricular septum) and a summary report. These are sent to a PACS destination for review by radiologists, pulmonologists, or cardiologists. The output assists clinicians in assessing right ventricle enlargement; it is not for stand-alone clinical decision-making. The device provides objective measurements to support clinical assessment, potentially improving diagnostic efficiency and consistency in identifying ventricular enlargement.
Clinical Evidence
Clinical performance evaluated via comparison of device output against ground truth established by trained radiologists. Study included 4 clinical sites (subset of 13 training sites) with no data overlap between training and pivotal study sets. Results showed high agreement between algorithm measurements and manual measurements. Mean absolute error (MAE) was < 7.2 mm, meeting performance goals. Performance was stratified by clinical site, gender, age, slice thickness, and scanner manufacturer/model.
Technological Characteristics
Software-only device; DICOM-compliant; hosted on Viz.ai backend server. Uses locked AI/ML algorithm for automated segmentation and diameter measurement of right and left ventricles from non-gated CTPA. Outputs include annotated DICOM image series and summary reports. No specific hardware materials or sterilization required as it is a software-only medical image management and processing system.
Indications for Use
Indicated for patients undergoing volumetric CT pulmonary angiography (CTPA) to assist physicians in assessing right ventricle enlargement by measuring maximal right and left ventricular diameters and their ratio.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Viz.ai, Inc. % Vi Ma Regulatory Affair Specialist 201 Mission Street, 12th Floor SAN FRANCISCO CA 94105
Re: K221100
August 29, 2022
Trade/Device Name: Viz RV/LV Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: August 4, 2022 Received: August 5, 2022
Dear Vi Ma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below
510(k) Number (if known)
### K221100
Device Name
Viz RV/LV
Indications for Use (Describe)
The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
□ Over-The-Counter Use (21 CFR 801
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY-K221100
### Viz.ai, Inc.'s Viz RV/LV
| Applicant Name: | Viz.ai, Inc. |
|-----------------|--------------------------------|
| | 201 Mission Street, 12th Floor |
| | San Francisco, CA 94105 |
- Contact Person: Vi Ma Regulatory Affairs Specialist 201 Mission Street, 12th Floor San Francisco, CA 94105 Tel. (415) 663-6130 vi.ma@viz.ai
- Date Prepared: August 24, 2022
Device Name and Classification
| Name of Device: | Viz RV/LV |
|-----------------------|--------------------------------------------------|
| Common or Usual Name: | Automated Radiological Image Processing Software |
| Classification Panel: | Radiology |
| Regulation No: | 21 C.F.R. § 892.2050 |
| Regulatory Class: | Class II |
| Product Code: | QIH |
### Predicate Device
| Manufacturer | Device Name | Application No. |
|--------------|----------------------|-----------------|
| Imbio | Imbio RV/LV Software | K203256 |
### Device Description
The Viz RV/LV is a software-only device that uses a locked artificial intelligence machine learning (AI/ML) algorithm to measure the maximal diameters of the right and left ventricles of the heart from a computed tomography pulmonary angiogram (CTPA) and report the ratio of those measurements.
Viz RV/LV produces an Annotated Image Series (Figure 1) and an RV/LV Summary Report (Figure 2) in DICOM format. The Annotated Image Series shows an RGB overlay on each slice of the input scan:
- . The red and blue solid lines indicate the maximum ventricular diameter for each ventricle.
- . The dashed line indicates a diameter measured on a slice that is within 10 slices of the global maximum ventricular diameter.
{4}------------------------------------------------
- . The interventricular septum is marked in solid green on all images where diameters are marked.
- . The maximal diameter is presented along with solid lines on slices with global maximum diameter.
Image /page/4/Picture/2 description: The image shows a CT scan of a human chest. The scan shows the heart, lungs, and other organs in the chest. The scan also shows the measurements of the right ventricle (RV) and left ventricle (LV) of the heart. The RV measures 58.3mm, and the LV measures 36.5mm.
Figure 1: Annotated Image Series
Image /page/4/Picture/4 description: The image shows a Viz RV/LV Analysis Summary. The RV/LV Diameter ratio is 1.6, the LV is 36.5mm (slice 218), and the RV is 58.3mm (slice 212). There are two CT scan images, one showing the Maximal RV Diameter Slice and the other showing the Maximal LV Diameter Slice. The image also includes a legend indicating the Largest RV diameter, Detected Interventricular Septum, and Largest LV diameter.
Figure 2: RV/LV Summary Report
The RV/LV Summary Report summarizes the results of the ventricle analysis and shows the slices with the maximum right and left ventricular diameters. The lines measuring the maximum RV and LV
{5}------------------------------------------------
diameters are displayed over the original CTPA slice image, along with the lengths of the largest RV and LV diameters, and the RV/LV ratio.
Viz RV/LV is hosted on Viz.ai's Backend Server and analyzes applicable CTPA scans that are acquired on CT scanners and are forwarded to Viz.ai's Backend Server. The results of the analysis are exported in DICOM format are sent to a PACS destination for review by thoracic radiologists, general radiologists, pulmonologists, cardiologists, or other similar physicians to assist in the assessment of right ventricle enlargement. The basic principle of operations is outlined in Figure 3 below.
Image /page/5/Figure/2 description: The image shows a diagram of a medical imaging workflow. It starts with a CT scanner, which sends data to a Viz.ai backend server. The server uses an RV/LV measurement algorithm to analyze the data and generate a report, which is then sent to a PACS server. Finally, the report is displayed on a monitor.
Figure 3. Data and Event Workflow
# Intended Use / Indications for Use
Viz.ai RV/LV software and the predicate device, Imbio RV/LV software (K203256), are Medical Image Management and Processing System software intended to calculate of the ratio of right ventricular diameter to left ventricular diameter from contrast enhanced CT images of the chest acquired using a standard CT pulmonary angiogram acquisition.
Viz RV/LV has the following indications for use:
The Viz RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. Viz RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The Viz RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.
The predicate device, Imbio RV/LV Software (K203256), has the following indications use:
The Imbio RV/LV Software device is designed to measure the maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the ventricles. The RV/LV
{6}------------------------------------------------
software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking or otherwise preclude clinical assessment of CTPA cases.
In comparing the intended use and indications for use of Viz RV/LV and the predicate device, Viz RV/LV has the same intended use as its predicate device.
A substantial equivalence chart (Table 1) comparing the subject device, Viz RV/LV, and the predicate, the Imbio RV/LV software, is provided below:
| Table 1 Technological Characteristic Comparison | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | Subject Device |
| | Imbio RV/LV Software | Viz RV/LV |
| Device Class | 2 | 2 |
| Product Code | QIH | QIH |
| Regulation<br>Number: | 21 C.F.R. § 892.2050 | 21 C.F.R. § 892.2050 |
| Indications for<br>Use: | The Imbio RV/LV Software device is<br>designed to measure the maximal<br>diameters of the right and left ventricles of<br>the heart from a volumetric CTPA<br>acquisition and report the ratio of those<br>measurements. RV/LV analyzes cases<br>using an artificial intelligence algorithm to<br>identify the location and measurements of<br>the ventricles. The RV/LV software<br>provides the user with annotated images<br>showing ventricular measurements. Its<br>results are not intended to be used on a<br>stand-alone basis for clinical decision-<br>making or otherwise preclude clinical<br>assessment of CTPA cases. | The Viz RV/LV Software device is<br>designed to measure the maximal<br>diameters of the right and left ventricles of<br>the heart from a volumetric CTPA<br>acquisition and report the ratio of those<br>measurements. Viz RV/LV analyzes cases<br>using an artificial intelligence algorithm to<br>identify the location and measurements of<br>the ventricles. The Viz RV/LV software<br>provides the user with annotated images<br>showing ventricular measurements. Its<br>results are not intended to be used on a<br>stand-alone basis for clinical decision-<br>making or otherwise preclude clinical<br>assessment of CTPA cases. |
| Input Data<br>Requirements: | Non-gated, CT Pulmonary Angiography<br>(CTPA) | Same |
| DICOM<br>Compliant | Yes | Same |
| Left Ventricle<br>Segmentation | Yes | Same |
| Right Ventricle<br>Segmentation | Yes | Same |
| Diameter<br>Measurements | Yes - Automated | Same |
| Fully automated<br>segmentation | Yes | Same |
| Outputs | Reports, DICOM Secondary Capture<br>Series | Same |
# Table 1 Technological Characteristic Comparison
{7}------------------------------------------------
### Performance Data
Clinical testing was performed as a study comparing the Viz RV/LV's output to the ground truth as established by trained radiologists. There was a high degree of agreement between the different trained radiologist as demonstrated by statical analysis. The study demonstrated that the MAE (Mean absolute error) was less than 7.2 mm between the established ground truth, which was aligned with the performance goal.
Additional statistical analysis was performed to describe the degree of agreement between the ventricle diameters measured by the RV/LV algorithm in comparison to the measurements that were obtained manually. The algorithm's ventricle diameter measurements were aligned when compared against the measurements that were obtained manually.
Stratification of device performance was divided by clinical site, gender, age, slice thickness, scanner manufacture and model.
The 4 clinical sites used in the pivotal study were a subset of a larger 13 sites used as part of the training data set. There was no overlapping data between the training sets and the pivotal study in terms of time and patient images.
# Conclusions
A comparison of intended use and technological characteristics, along with clinical performance data, demonstrated that the device is as safe and effective as the previously cleared Imbio RV/LV software (K203256). Thus, Viz RV/LV is substantially equivalent to Imbio RV/LV.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.