Customize for shoulder arthroplasty is intended to be used as a software interface to assist in: - Visualization, modification, validation of the planning of shoulder arthroplasty - Communication of treatment options - Segmentation of CT-scan data - 3D CAD models generation - Managing timeline and cases Customize is not intended to be used for - Spine surgeries - Implant and instrument design Experience in usage and clinical assessment are necessary for a proper use of the software. It is to be used for adult patients only and should not be used for diagnostic purposes.
Device Story
Customize is a software interface for shoulder arthroplasty planning. Input: multi-slice DICOM CT-scan data. Processing: software performs image segmentation and generates 3D CAD models; provides an implant library for 3D digital representation; allows physician to review, modify position/orientation of 3D objects, and validate treatment plans. Output: 3D surgical plan saved on server for intraoperative reference. Used in clinical settings by physicians. Benefits: assists in visualization, communication of treatment options, and precise implant positioning/sizing. Does not perform diagnostic functions.
Clinical Evidence
No clinical data. Bench testing only. Validation included software verification and validation per FDA guidance, segmentation validation, repeatability and reproducibility studies on shoulder anatomy segmentation, and accuracy studies on 3D model generation for humerus and scapula.
Technological Characteristics
Software-based medical image management and processing system. Features include DICOM image segmentation, 3D CAD model generation, and an implant library. Includes a distance map tool for object proximity estimation. Operates on CT-scan data. Connectivity involves server-based storage of treatment plans.
Indications for Use
Indicated for adult patients undergoing shoulder arthroplasty planning. Not for spine surgeries or implant/instrument design. Not for diagnostic purposes.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
3D-Side SA % Mieke Janssen Regulatory Affairs Consultant OrthoGrow NV Davincilaan 1 Zaventem, Vlaams-Brabant 1930 BELGIUM
March 16, 2022
Re: K213779
Trade/Device Name: Customize Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: February 17, 2022 Received: February 18, 2022
Dear Mieke Janssen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb Associate Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213779
Device Name Customize
Indications for Use (Describe)
Customize for shoulder arthroplasty is intended to be used as a software interface to assist in:
- Visualization, modification, validation of the planning of shoulder arthroplasty
- Communication of treatment options
- Segmentation of CT-scan data
- 3D CAD models generation
- Managing timeline and cases
Customize is not intended to be used for
- Spine surgeries
- Implant and instrument design
Experience in usage and clinical assessment are necessary for a proper use of the software. It is to be used for adult patients only and should not be used for diagnostic purposes.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY (21CFR807.92)
## SUBMITTER
Company Name: Establishment registration number: Address:
| Phone number: | |
|-----------------------------------|----------------|
| Principal contact person: | |
| Principal contact e-mail address | |
| Additional contact person: | |
| Additional contact e-mail address | |
| Summary date: | March 14, 2022 |
DEVICE
Name & trade name: Classification name:
3D-Side 3013561205 Rue André Dumont 5 1435 Mont-Saint-Guibert, BE +32 (0) 10 81 35 48 Mieke Janssen mieke@ortho-grow.com Laurent Paul lp@3dside.eu
Customize Automated Radiological Image Processing Software QIH
Classification product code
# PREDICATE AND REFERENCE DEVICES
The predicate device to which substantial equivalence is claimed:
| Trade or proprietary or model name | Preview Shoulder |
|------------------------------------|---------------------------------|
| 510(k) number | K210556 |
| Decision date | 21 April 2021 |
| Classification product code | QIH |
| Manufacturer | Genesis Software<br>Innovations |
# DESCRIPTION AND FUNCTIONING OF THE DEVICE
Customize is intended to be used during the preparation of shoulder arthroplasties. It visualizes surgical treatment options that were previously created based on 3D CAD files generated from multi-slice DICOM data from a CT scanner. It consists of a single software user interface where the physician can review the CAD files in 3D and modify the position and orientation of the different 3D objects. Customize includes an implant library with 3D digital representations of various implant models so that the right implant positioning and sizing can be achieved based on the physician's input. After approval by the physician, the treatment plan is saved on the server and can be used as a reference during surgery.
Customize is prescription use only.
{4}------------------------------------------------
## INTENDED USE
Customize for shoulder arthroplasty is intended to be used as a software interface to assist in:
- Visualization, modification, validation of the planning of shoulder arthroplasty -
- Communication of treatment options -
- Segmentation of CT-scan data -
- 3D CAD models generation -
- -Managing timeline and cases
Customize is not intended to be used for
- Spine surgeries -
- Implant and instrument design -
Experience in usage and a clinical assessment are necessary for a proper use of the software. It is to be used for adult patients only and should not be used for diagnostic purposes.
## TECHNOLOGICAL CHARACTERISTICS
Comparison of technological characteristics with the predicate device (Preview Shoulder, K210556):
- . Following similarities exist between the subject device Customize and the predicate device:
- Intended use/indications for use: both the subject and the predicate device share a O similar intended use
- Patient population: both devices are applicable for adult population only о
- Device functionality: both devices enable visualization of medical images о (DICOM), include an implant library, and allow implant visualization in both the original scan images as well as the 3D model reconstruction. Both devices utilize image segmentation and allow 3D view manipulation.
- The following technological difference exists between the subject device and the predicate device:
- The subject device includes a distance map providing an approximate estimate of O the distance between two objects. This functionality is absent in the predicate device.
This difference in technological characteristics does not raise new questions of safety and effectiveness.
# Performance Data
Non-clinical performance data was included in the 510(k)-submission demonstrating Customize has been validated for its intended use and substantial equivalence to the predicate device.
{5}------------------------------------------------
Software verification and validation was performed, and documentation was provided following the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". This includes verification against defined requirements, and validation against user needs. In addition, performance testing included 1) segmentation validation of the Customize software, 2) Repeatability and Reproducibility study on the segmentation of shoulder anatomies and 3) accuracy study on 3D model generation for humerus and scapula.
## SUMMARY
The characteristics that determine the functionality and performance of Customize, the subject device, are substantially equivalent to the predicate device cleared under K210556. The nonclinical testing indicates that the subject device is as safe, as effective, and performs as well as the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.