← Product Code [QDQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QDQ) · K221347

# Transpara 1.7.2 (K221347)

_Screenpoint Medical B.V. · QDQ · Aug 3, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QDQ/K221347

## Device Facts

- **Applicant:** Screenpoint Medical B.V.
- **Product Code:** [QDQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QDQ.md)
- **Decision Date:** Aug 3, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2090
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

Transpara® software is intended for use as a concurrent reading aid for physicians interpreting screening full-field digital mammography exams and digital breast tomosynthesis exams from compatible FFDM and DBT systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes locations of calcifications groups and soft-tissue regions, with scores indicating the likelihood that cancer is present, and an exam score indicating the likelihood that cancer is present in the exam. Patient management decisions should not be made solely on the basis of analysis by Transpara®.

## Device Story

Transpara 1.7.2 is a software-only concurrent reading aid for physicians interpreting screening FFDM and DBT exams. It accepts DICOM-formatted mammography/DBT images as input. Deep learning algorithms analyze images to detect suspicious calcifications and soft-tissue lesions (masses, architectural distortions, asymmetries). The device outputs CAD marks highlighting suspicious regions, region-specific malignancy scores (0-100), and an exam-level malignancy score (1-10). It links corresponding regions across different breast views to assist workflow. Used in clinical facilities/hospitals, the software processes images on a server and transmits results to workstations, PACS, or RIS. It serves as an adjunct tool; physicians retain final diagnostic responsibility. By providing automated detection and risk stratification, it aims to improve interpretation accuracy and clinical workflow efficiency.

## Clinical Evidence

Bench testing only. Standalone performance evaluated on 10,690 exams (9,218 non-cancer, 1,472 cancer) from multiple US/EU centers. 2D FFDM sensitivity: 95.0% (93.5-96.4) at 0.30 FP/image; AUC 0.945 (0.935-0.954). DBT sensitivity: 93.2% (91.0-95.1) at 0.34 FP/volume; AUC 0.945 (0.936-0.954). Dataset included diverse manufacturers (Hologic, GE, Philips, Siemens, Giotto, Fujifilm).

## Technological Characteristics

Software-only device; runs on processing server. Uses deep learning algorithms for image analysis. Inputs/outputs via DICOM protocol. Compatible with various FFDM/DBT systems. Complies with IEC 62366-1, ISO 14155, ISO 14971, IEC 62304, IEC 82304-1, and ISO 15223-1 standards.

## Regulatory Identification

A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.

## Special Controls

A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.

## Predicate Devices

- Transpara® 1.7.0 ([K210404](/device/K210404.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 3, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ScreenPoint Medical B.V. % Robin Barwegen Head of Regulatory and Quality Affairs Mercator II, 7th floor, Teornooiveld 300 Nijmegen, Gelderland 6525EC NETHERLANDS

## Re: K221347

Trade/Device Name: Transpara 1.7.2 Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: May 5, 2022 Received: May 9, 2022

## Dear Robin Barwegen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Michael D. O'Hara, Ph.D. Deputy Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

## K221347

Device Name Transpara® 1.7.2

### Indications for Use (Describe)

Transpara® software is intended for use as a concurrent reading aid for physicians interpreting screening full-field digital mammography exams and digital breast tomosynthesis exams from compatible FFDM and DBT systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes locations of calcifications groups and soft-tissue regions, with scores indicating the likelihood that cancer is present, and an exam score indicating the likelihood that cancer is present in the exam. Patient management decisions should not be made solely on the basis of analysis by Transpara®.

| Type of Use (Select one or both, as applicable)                                                                                |                                                                                                              |
|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> <b>                          Prescription Use (Part 21 CFR 801 Subpart D)                     </b> </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K221347

# 510(k) Summary Transpara®

This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR § 807.92.

#### Submitter 1.

## Manufacturer:

ScreenPoint Medical B.V.

Mercator II, 7th floor

Toernooiveld 300

6525 EC Nijmegen

Netherlands

www.screenpoint-medical.com

## Contact person:

Robin Barwegen

Office: +31 24 3030045 | +31 24 2020020

Mobile: +31 6 44077104

Mercator II, 7th floor, Toernooiveld 300, 6525 EC Nijmegen, Netherlands

## Date:

August 3, 2022

{4}------------------------------------------------

#### 2. Device

| Device trade name         | Transpara® 1.7.2                                                   |
|---------------------------|--------------------------------------------------------------------|
| Device                    | Radiological Computer Assisted Detection and Diagnosis<br>Software |
| Classification regulation | 21 CFR 892.2090                                                    |
| Panel                     | Radiology                                                          |
| Device class              | II                                                                 |
| Product code              | QDQ                                                                |
| Submission type           | Traditional 510(k)                                                 |

#### 3. Legally marketed predicate device

| Device trade name         | Transpara® 1.7.0                                                   |
|---------------------------|--------------------------------------------------------------------|
| Legal Manufacturer        | ScreenPoint Medical B.V.                                           |
| Device                    | Radiological Computer Assisted Detection and Diagnosis<br>Software |
| Classification regulation | 21 CFR 892.2090                                                    |
| Panel                     | Radiology                                                          |
| Device class              | II                                                                 |
| Product code              | QDQ                                                                |
| Clearance number          | K210404                                                            |

#### Device description 4.

Transpara® is a software only application designed to be used by physicians to improve interpretation of digital mammography and digital breast tomosynthesis. The system is intended to be used as a concurrent reading aid to help readers with detection and characterization of potential abnormalities suspicious for breast cancer and to improve workflow. 'Deep learning' algorithms are applied to FFDM images and DBT slices for recognition of suspicious calcifications and soft tissue lesions (including densities, masses, architectural distortions, and asymmetries). Algorithms are trained with a large database of biopsy-proven examples of breast cancer, benign abnormalities, and examples of normal tissue.

Transpara® offers the following functions which may be used at any time during reading (concurrent use):

{5}------------------------------------------------

- a) Computer aided detection (CAD) marks to highlight locations where the device detected suspicious calcifications or soft tissue lesions.
- b) Decision support is provided by region scores on a scale ranging from 0-100, with higher scores indicating a higher level of suspicion.
- c) Links between corresponding regions in different views of the breast, which may be utilized to enhance user interfaces and workflow.
- d) An exam score which categorizes exams on a scale of 1-10 with increasing likelihood of cancer. The score is calibrated in such a way that approximately 10 percent of mammograms in a population of mammograms without cancer falls in each category.

Results of Transpara® are computed in processing server which accepts mammograms or DBT exams in DICOM format as input, processes them, and sends the processing output to a destination using the DICOM protocol. Common destinations are medical workstations, PACS and RIS. The system can be configured using a service interface. lmplementation of a user interface for end users in a medical workstation is to be provided by third parties.

#### 5. Indications for use

Transpara® is a software medical device for use in a healthcare facility or hospital with the following indications for use:

Transpara® software is intended for use as a concurrent reading aid for physicians interpreting screening full-field digital mammography exams and digital breast tomosynthesis exams from compatible FFDM and DBT systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes locations of calcifications groups and soft-tissue regions, with scores indicating the likelihood that cancer is present, and an exam score indicating the likelihood that cancer is present in the exam. Patient management decisions should not be made solely on the basis of analysis by Transpara®.

## Intended user population

Intended users of Transpara® are physicians qualified to read screening mammography exams and digital breast tomosynthesis exams.

## Intended patient population

The device is intended to be used in the population of women undergoing screening mammography and digital breast tomosynthesis.

{6}------------------------------------------------

## Warnings and precautions

Transpara® is an adjunct tool and not intended to replace a physicians' own review of a mammogram. Decisions should not be made solely based on analysis by Transpara®.

#### 6. Predicate device comparison

The indication for use of Transpara® 1.7.2 is similar to that of the predicate device. Both devices are intended for concurrent use by physicians interpreting breast images to help them with localizing and characterizing abnormalities. The devices are not intended as a replacement for the review of a physician or their clinical judgement.

The overall design of Transpara® 1.7.2 is the same as that of the predicate device. Both versions detect and characterize findings in radiological breast images and provide information about the presence, location, and characteristics of the findings to the user in a similar way. There are differences in the algorithmic components, which have changed to improve detection accuracy for DBT. Support for Giotto FFDM and General Electric (DBT) has been added.

Changes do not raise different questions of safety and effectiveness of the device when used as labeled.

#### Summary of non-clinical performance data 7.

In the design and development of Transpara® 1.7.2, ScreenPoint applied the following voluntary FDA recognized standards and guidelines:

| Standard ID                              | Standard Title                                                                                                           | FDA Recognition # |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------|
| IEC 62366-1 Edition<br>1.0 2015-02       | Medical devices - Part 1: Application of usability<br>engineering to medical devices [Including<br>CORRIGENDUM 1 (2016)] | 5-114             |
| ISO, 14155 Second<br>edition 2011-02-01, | Clinical investigation of medical devices for<br>human subjects - Good clinical practice                                 | 2-205             |
| ISO 14971:2019                           | Medical Devices - Application Of Risk<br>Management To Medical Devices                                                   | 5-125             |
| IEC 62304:2015                           | Medical Device Software - Software Life Cycle<br>Processes                                                               | 13-79             |
| IEC 82304-1: 2016                        | Health software - Part 1: General requirements<br>for product safety                                                     | 13-97             |

{7}------------------------------------------------

| ISO, 15223-1 Third<br>Edition 2016-11-01, | Medical devices - Symbols to be used with<br>medical device labels labelling and information<br>to be supplied - Part 1: General requirements | 5-117 |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------|
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------|

The following guidance documents were used to support this submission:

- Guidance for Industry and FDA Staff Guidance for the Content of Premarket ● Submissions for Software Contained in Medical Devices (Issued on May 11, 2005)
- Guidance for Industry and Food and Drug Administration Staff Computer-. Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions (Issued on July 3, 2012)
- . Guidance for Industry and FDA Staff - Clinical Performance Assessment: Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data in Premarket notification [510(k)] Submissions (Issued on January 2020)
- Guidance for Industry and Food and Drug Administration Staff The 510(k) . Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (Issued on July 28 2014)
- Guidance for Industry and Food and Drug Administration Staff Unique Device . Identification: Direct Marking of Devices (Issued on November 17, 2017)

Transpara® 1.7.2 is a software-only device. The level of concern for the device is determined as Moderate Level of Concern.

## Non-clinical performance tests

Verification testing was conducted, which consisted of software unit testing, software integration testing and software system testing. The verification tests showed that the software application satisfied the software requirements.

Standalone performance tests were conducted to demonstrate substantial equivalence with the predicate device. For these tests an independent dataset was used, which was acquired from multiple centers and had not been used for development of the algorithms. This test set contained 2D and 3D mammograms acquired with devices from different manufacturers (2D: Hologic, GE, Philips, Siemens, Giotto and Fujifilm, 3D: Hologic, Siemens, General Electric and Fujifilm), representative for breast imaging practices performing screening and diagnostic assessment, collected from multiple clinical centers in seven EU countries and the US. For the inclusion of the normal exams in the test set the majority of exams had a normal follow-up of at least one year.

{8}------------------------------------------------

The test set consisted of 10,690 exams, including 9,218 non-cancer exams, and 1,472 exams with cancer. An overview is presented in table 1. In total, 56.5% of biopsy-proven cancer regions in the datasets of 2D and 3D are mass, 26.1% calcifications, 4.0% architectural distortions and asymmetries, and 13.3% are reported as a combination of morphological findings. The main histological cancer types are invasive non-specific type (47.1%), ductal carcinoma in situ (17.2%), invasive lobular carcinoma (6.7%), and other/unknown (28.9%). The median age of the women in the dataset was 56 years old (interquartile range: 49-64).

|       | Number of<br>Exams | Normal | Benign | Cancer |
|-------|--------------------|--------|--------|--------|
| FFDM  | 5,867              | 4,841  | 149    | 877    |
| DBT   | 4,823              | 3,988  | 240    | 595    |
| Total | 10,690             | 8,829  | 389    | 1,472  |

Table 1: Data used for evaluation of stand-alone performance.

Exam based sensitivity for cancer detection in the test set was computed by taking the fraction of cancers that were correctly localized in it least one view (MLO or CC). False positive rates were computed in exams without cancer, by dividing the number of regions detected per image by the number of images. For 2D. the sensitivity is 95,0% (93,5-96,4) at 0.30 FP/image. For DBT, sensitivity is 93.2% (91.0-95.1) at 0.34 FP/volume. ROC analysis was also performed. The AUC is 0.945 (0.935-0.954) for 2D and 0.945 (0.936-0.954) for DBT.

Based on standalone testing it was concluded that Transpara 1.7.2 breast cancer detection performance for 2D and 3D mammograms of compatible devices is non-inferior to the performance of the predicate device Transpara 1.7.0.

#### Conclusions 8.

The data presented in this 510(k) includes all required information to support the review by FDA. Standalone performance tests with FFDM and DBT demonstrate that Transpara® 1.7.2 achieves non-inferior detection performance compared to the predicate device.

ScreenPoint has applied a risk management process in accordance with FDA recognized standards to identify, evaluate, and mitigate all known hazards related to Transpara® 1.7.2. These hazards may occur when accuracy of diagnosis is potentially affected, causing either false-positives or false-negatives. All identified risks are effectively mitigated and it can be concluded that the residual risk is outweighed by the benefits.

{9}------------------------------------------------

Considering all data in this submission, the data provided in this 510(k) application supports the safe and effective use of Transpara® 1.7.2 for its indications for use and substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QDQ/K221347](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QDQ/K221347)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
