← Product Code [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY) · K231026

# 12MP Color LCD Monitor C1216W, C12*** (* = 0 to 9, A to Z or blank, and the difference among models means the product is named according to different appearance colors and customer models) (K231026)

_Shenzhen Beacon Display Technology Co., Ltd. · PGY · Aug 18, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K231026

## Device Facts

- **Applicant:** Shenzhen Beacon Display Technology Co., Ltd.
- **Product Code:** [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY.md)
- **Decision Date:** Aug 18, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

These products are indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

## Device Story

The 12MP Color LCD Monitor (C1216W, C12***) is a high-resolution medical display designed for diagnostic radiology. It accepts digital video inputs (HDMI, DisplayPort) from medical imaging systems and renders them on a 31-inch TFT color LCD panel with 4200 x 2800 resolution. The device features an LED backlight, integrated front sensor, and ambient light sensor to maintain brightness stability and DICOM GSDF calibration. It is intended for use in clinical environments by trained medical practitioners to review and diagnose radiological images, including mammography and tomosynthesis. The monitor includes an anti-glare screen to improve image clarity under varying lighting conditions. Healthcare providers use the high-definition output to perform diagnostic tasks; the device's stable luminance and color accuracy are intended to ensure consistent image quality, supporting accurate clinical decision-making.

## Clinical Evidence

No clinical or animal testing was performed. Substantial equivalence was demonstrated through bench testing in accordance with AAPM TG18 guidelines, including measurements of DICOM GSDF conformance, luminance non-uniformity, chromaticity, spatial resolution (MTF), veiling glare, and pixel defect analysis.

## Technological Characteristics

31-inch TFT color LCD (IPS) panel; 4200 x 2800 resolution; LED backlight; 1200 cd/m² typical brightness; 1500:1 contrast ratio; 60 Hz refresh rate; 14ms response time. Connectivity: HDMI, DisplayPort, USB. Features integrated optical sensor and ambient light sensor for calibration. Powered by AC 100-240V. Housing: Plastic/Metal. IP20 protection. Complies with IEC 60601-1, EN 60601-1, ANSI/AAMI ES60601-1.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- EIZO Corporation, 12MP Color LCD Monitor, RadiForce RX1270, RX1270-AR ([K200485](/device/K200485.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date August 18, 2023. The text is in a simple, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The text is centered and takes up most of the image.

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health and Human Services seal. To the right of the seal is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shenzhen Beacon Display Technology Co., Ltd. % Li Yafei Official Correspondent 15F, Building 6, Hengda Shishang Huigu (East), Fulong Road, Dalang Subdistrict. Longhua Shenzhen, Guangdong 518109 CHINA

Re: K231026

Trade/Device Name: 12MP Color LCD Monitor C1216W, C12*** ("*" = 0 to 9, A to Z or blank, and the difference among models means the product is named according to different appearance colors and customer models) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: July 19, 2023 Received: July 19, 2023

Dear Li Yafei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Zhkke

, for

Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K231026

#### Device Name

12MP Color LCD Monitor (C1216W, C12** (1*) = 0 to 9, A to Z or blank, and the difference among models means the product is named according to different appearance colors and customer models))

#### Indications for Use (Describe)

These products are indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

[As required by 21 CFR 807.92]

## 1. Date Prepared [21 CFR807.92 (a) (1)]

August 17, 2023

# 2. Submitter's Information [21 CFR807.92 (a) (1)]

| Name of Sponsor: | Shenzhen Beacon Display Technology Co., Ltd.                      |
|------------------|-------------------------------------------------------------------|
| Address:         | 15F, Building 6, Hengda Shishang Huigu(East), Fulong Road, Dalang |
|                  | Subdistrict, Longhua, Shenzhen, 518109 China                      |
| Contact Name:    | Li Yafei                                                          |
| Telephone No.:   | +86-024-88087610                                                  |
| Fax No.:         | +86-024-88087629                                                  |
| Email Address:   | liyf@beacon-display.cn                                            |

### 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]

| • Trade Name/Model: | 12MP Color LCD Monitor (C1216W, C12*** ("*" = 0 to 9, A to Z or blank, and the difference among models means the product is named according to different appearance colors and customer models)) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Common Name:      | 12MP Color LCD Monitor                                                                                                                                                                           |

- 0 Classification Name: Medical Image Management and Processing System
- Regulation Number: 21 CFR 892.2050
- PGY 0 Product code:
- O Classification Panel: Radiology
- Device Class: ll O

### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicates within this submission are as follows:

{4}------------------------------------------------

EIZO Corporation, 12MP Color LCD Monitor, RadiForce RX1270, RX1270-AR has been cleared by FDA through 510(k) No. K200485 (Decision Date - March 23, 2020).

### 5. Description of the Device [21 CFR 807.92(a) (4)]

The C1216W and C12*** are 12MP color LCD monitors, which are specifically designed to provide the high definition image output for general radiography. They are using the latest generation of LED back light panel, with high resolution 4200 x 2800, built-in brightness stabilization circuit, front sensor and ambient light sensor, stable brightness and persistent calibration can be quaranteed throughout its life. The anti-qlare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

#### 6. Intended Use [21 CFR 807.92(a)(5)]

These products are indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

#### 7. Indications for Use

These products are indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

| Panel characteristics:            |                                                                                        |
|-----------------------------------|----------------------------------------------------------------------------------------|
| Screen technology                 | 31", TFT, color, LCD screen, AG type                                                   |
| Active area (H x V)               | 652.68 x 435.12 mm                                                                     |
| Pixel pitch                       | 0.1554 mm (H) x 0.1554 mm (V)                                                          |
| Resolution                        | 4200 x 2800 Pixels                                                                     |
| Contrast ratio                    | 1050:1 (minimum), 1500:1 (typical)                                                     |
| Viewing angle (CR > 10)           | Horizontal: 178° (typical)                                                             |
|                                   |                                                                                        |
|                                   | Vertical: 178° (typical)                                                               |
| Screen brightness                 | 1200cd/m² (typical)                                                                    |
| Refresh rate                      | 60 Hz                                                                                  |
| Backlighting                      | LED                                                                                    |
| Lifetime of backlight             | 50000 hours (minimum)                                                                  |
| Response time (Gray to Gray)      | 14ms (typical )                                                                        |
| Color support                     | 1.07G                                                                                  |
| Power supply:                     |                                                                                        |
| Power connection                  | Power cord connector with protective conductor,<br>IEC60320-C14                        |
| Input voltage                     | AC 100-240V 50/60Hz                                                                    |
| Current consumption               | 2.4-1.0 A                                                                              |
| Power consumption                 | ≤200W                                                                                  |
| Standby                           | < 0.5 W                                                                                |
| Control and connection:           |                                                                                        |
| Front                             | 1 operation LED, 6 functional keys                                                     |
| Back                              | Power switch * 1 AC socket * 1 HDMI * 2 Display Port * 3 Service Connector * 1 USB * 7 |
| Mechanical characteristics:       |                                                                                        |
| Housing components                | Plastic, Metal                                                                         |
| Ventilation openings              | Natural heat dissipation                                                               |
| Degree of protection              | IP20                                                                                   |
| Climatic conditions:              |                                                                                        |
| Operational temperature           | 0°C ~40°C                                                                              |
| Operational humidity              | 15%~ 85% Relative humidity, no condensation                                            |
| Transport and storage temperature | -20°C ~ +60°C                                                                          |
| Transport and storage humidity    | 10% ~ 90% Relative humidity, no condensation                                           |
| Operational pressure              | 700 hPa~1060 hPa                                                                       |
| Safety regulations:               |                                                                                        |
| Safety standards                  | IEC 60601-1<br>EN 60601-1<br>ANSI/AAMI ES60601-1<br>CAN/CSA-C22.2 NO. 60601-1:14       |
| Conformity                        | CCC,CE,FCC,TUV,CB,CECP                                                                 |
| Dimension:                        |                                                                                        |
| With packing (W x H x D)          | 840 x 643 x321 mm                                                                      |
| Weight:                           |                                                                                        |
| Net weight                        | 18.3± 1 kg                                                                             |
| Gross weight                      | 22.8 ± 1 kg                                                                            |

# 8. Technological Characteristics [21 CFR 807.92(a)(6)]

{5}------------------------------------------------

{6}------------------------------------------------

# 9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

#### 9.1 Intended uses:

|  |  |  | Table 1 Intended Use Comparison |
|--|--|--|---------------------------------|
|--|--|--|---------------------------------|

| ID                                          | Comparison<br>Item                                                                                          | Proposed Device                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                   |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                             |                                                                                                             | 12MP Color LCD Monitor<br>(C1216W ,C12***)                                                                                                                                                                                                    | 12MP Color LCD Monitor<br>(RX1270, RX1270-AR)                                                                                                                                                                                                                                      |
| 1                                           | Intended Use                                                                                                | These products are indicated<br>for use in displaying<br>radiological images (including<br>full-field digital mammography<br>and digital breast<br>tomosynthesis) for review,<br>analysis, and diagnosis by<br>trained medical practitioners. | The 12MP Color LCD Monitor<br>(RX1270, RX1270-AR)" is<br>intended to be used in displaying<br>radiological images (including<br>full-field digital mammography<br>and digital breast tomosynthesis)<br>for review, analysis, and<br>diagnosis by trained medical<br>practitioners. |
| Attributes                                  | Proposed Device<br>12MP Color LCD<br>Monitor<br>(C1216W ,C12***)                                            | Predicate Device<br>12MP Color LCD<br>Monitor<br>(RX1270, RX1270-AR)                                                                                                                                                                          | Explanation of<br>Difference                                                                                                                                                                                                                                                       |
| Display Technology                          | TFT Color LCD Panel<br>(IPS)                                                                                | TFT Color LCD Panel<br>(IPS)                                                                                                                                                                                                                  | -                                                                                                                                                                                                                                                                                  |
| Screen size                                 | 78.4cm / 31"                                                                                                | 78.4cm / 30.9"                                                                                                                                                                                                                                | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                                                                   |
| Resolution                                  | 4200 x 2800 (3:2 aspect ratio)                                                                              | 4200 x 2800 (3:2 aspect ratio)                                                                                                                                                                                                                | -                                                                                                                                                                                                                                                                                  |
| Pixel pitch                                 | 0.1554 x 0.1554 mm                                                                                          | 0.1554 x 0.1554 mm                                                                                                                                                                                                                            | -                                                                                                                                                                                                                                                                                  |
| Backlight type                              | LED                                                                                                         | LED                                                                                                                                                                                                                                           | -                                                                                                                                                                                                                                                                                  |
| Maximum luminance                           | 1200 cd/m²                                                                                                  | 1200 cd/m²                                                                                                                                                                                                                                    | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                                                                   |
| DICOM calibrated luminance                  |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | 450 cd/m²                                                                                                   | 500 cd/m²                                                                                                                                                                                                                                     | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                                                                   |
| Contrast Ratio (typical)                    |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | 1500:1                                                                                                      | 1500:1                                                                                                                                                                                                                                        | -                                                                                                                                                                                                                                                                                  |
| Response Time (typical)                     |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | 14 ms<br>(Gray to Gray)                                                                                     | 12 ms<br>(black-white-black)                                                                                                                                                                                                                  | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Frame rate and refresh rate                 |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
| Digital<br>Scanning<br>Frequency<br>(H / V) | HDMI: 31.5 - 133.32<br>kHz / 30 - 60 Hz<br>DisplayPort: 31.5 -<br>172.8 kHz / 60 Hz                         | HDMI: 31 - 160 kHz / 59<br>- 61 Hz<br>(VGA Text: 69 - 71 Hz)<br>DisplayPort: 31 - 175<br>kHz / 29 -61 Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous<br>mode:<br>29.5 - 30.5 Hz, 59 - 61<br>Hz                                             | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Analog<br>Scanning<br>Frequency<br>(H / V)  | -                                                                                                           | HDMI: 31 - 160 kHz / 59<br>- 61 Hz<br>(VGA Text: 69 - 71 Hz)<br>DisplayPort: 31 - 175<br>kHz / 29 -61 Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous<br>mode:<br>29.5 - 30.5 Hz, 59 - 61<br>Hz                                             | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Display Interface                           |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
| Input video<br>signals                      | HDMI x 2,<br>DisplayPort x 2                                                                                | DisplayPort x 2,<br>HDMI                                                                                                                                                                                                                      | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Output video<br>signals                     | DisplayPort x 1                                                                                             | -                                                                                                                                                                                                                                             | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Video bandwidth                             |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | HDMI: 594MHz<br>DisplayPort:<br>753.25MHz                                                                   | HDMI: 260MHz<br>DisplayPort: 765MHz                                                                                                                                                                                                           | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Ambient light sensing                       |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
| Ambient light<br>sensor                     | Yes                                                                                                         | Yes                                                                                                                                                                                                                                           | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Luminance calibration tools                 |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | Integrated optical<br>sensor<br>External optical sensor<br>Calibration<br>software:Beacon<br>Monitor Manage | Integrated optical<br>sensor<br>External optical sensor<br>Calibration software:<br>RadiCS                                                                                                                                                    | Different design<br>scheme.                                                                                                                                                                                                                                                        |
| Quality-control procedures                  |                                                                                                             |                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                    |
|                                             | Beacon Monitor<br>Manage                                                                                    | Software: RadiCS                                                                                                                                                                                                                              | Different design<br>scheme.                                                                                                                                                                                                                                                        |

{7}------------------------------------------------

### 9.2 Comparison table

#### Table 2 General Comparison

{8}------------------------------------------------

{9}------------------------------------------------

It is clear that the technological characteristic differences discussed above do not affect the safety and the effectiveness of the C1216W and C12***.

#### 9.3 Performance Testing

According to the instructions in" Guidance for Industry and Food and Drug Administration Staff Display Devices for Diagnostic Radiology", the bench tests were performed on C1216W and C12*** as below.

- Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline).
- Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 guideline.
- O Measure the luminance stability and chromaticity response with the temperature 0°C,

25°C and 40°C on Luminance response by AAPM-TG18.

- Measure the chromaticity non-uniformity characteristics of the display screen in accordance with TG18 guideline.

{10}------------------------------------------------

- Measure the luminance at the angles of 30º and 45º in diagonal, horizontal and vertical directions at center and four corners by AAPM-TG18.
- Measure the temporal response using the typical data provided by the panel manufacturer.
- 0 Measurement of Luminance
- 0 Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 guideline.
- 0 Measure the spatial noise by noise power spectrum.
- 0 Measure the reflection coefficient with specular reflection and diffuse reflection by TG18.
- Measure the veiling glare of small-spot contrast performing veiling glare test by TG-18.
- O Measure the spatial resolution expressed as modulation transfer function (MTF)
- Maximum number allowed for each type of pixel defects/faults
- 0 Measurement of Color tracking and Gray tracking
- Measure pixel fill factor like pixel structure and aperture ratio etc.

The test results showed that C1216W and C12*** are with display characteristics equivalent to those of the predicate device, RadiForce RX1270 except some items, each of which was determined that it would not affect observer's performance.

No animal or clinical testing is needed for C1216W and C12**.

### 10. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd. concludes that:

- The intended use of C1216W and C12*** is totally same as that of the predicate device.
- The technological characteristic differences between C1216W (C12***) and RadiForce RX1270 do not affect the safety and effectiveness, no new risk is raised.
- 0 Demonstrated by the bench tests, the display characteristics of C1216W and C12*** are equivalent to those of the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K231026](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K231026)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
