← Product Code [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY) · K223423

# 32HQ713D (K223423)

_Lg Electronics.Inc · PGY · Mar 1, 2023 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K223423

## Device Facts

- **Applicant:** Lg Electronics.Inc
- **Product Code:** [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY.md)
- **Decision Date:** Mar 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, and diagnosis by trained medical practitioners.

## Device Story

Medical monitor for high-resolution color and grayscale radiological imaging; used in PACS and radiology environments. Input: digital image data from PACS/radiology systems. Operation: 31.5-inch TFT LCD module with LED backlight; 3840x2160 resolution; 10-bit color drivers. Output: visual display of medical images for clinician review and diagnosis. Operated by trained medical practitioners. Benefits: provides high-resolution visualization of diagnostic images, including mammography and tomosynthesis, to support clinical decision-making.

## Clinical Evidence

No clinical studies were performed. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and display performance metrics (spatial resolution, luminance, pixel defects, artifacts, temporal response, and grayscale conformance) per FDA guidance.

## Technological Characteristics

31.5" TFT LCD module (LB315WR6-SSZ1) with LED backlight; 3840x2160 resolution; 10-bit color drivers; 30-pin eDP interface; DVI and DP inputs. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6 standards.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- 31HN713D ([K201777](/device/K201777.md))
- RX560 ([K172738](/device/K172738.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

LG Electronics Inc. % Jonghyun Kim Chief Consultant GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 REPUBLIC OF KOREA

# Re: K223423

March 1, 2023

Trade/Device Name: 32H0713D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: February 1, 2023 Received: February 3, 2023

Dear Jonghyun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K223423

Device Name 32HQ713D

Indications for Use (Describe)

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, and diagnosis by trained medical practitioners.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

[As Required by 21 CFR 807.92]

# 1. Date Prepared [21 CFR 807.92(a)(a)]

Nov 11, 2022

# 2. Submitter's Information [21 CFR 807.92(a)(1)]

| • Name of Sponsor:      | LG Electronics Inc.                                                                           |
|-------------------------|-----------------------------------------------------------------------------------------------|
| - Address:              | 222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea |
| • Contact Name:         | Jinhwan Jun / Chief Research Engineer                                                         |
| - Telephone No.:        | +82-31-8066-5641                                                                              |
| - Email Address:        | jinhwan.jun@lge.com                                                                           |
| • Name of Manufacturer: | LG Electronics Inc.                                                                           |

- 168, Suchul-daero, Gumi-si, Gyeongsangbuk-do, 39368, Address: -Republic of Korea
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

- Trade Name: ● 32HQ713D
- . Common Name:

Medical Monitor

- Classification: ●

| Classification Name   | Picture archiving and communications system |
|-----------------------|---------------------------------------------|
| Classification Number | 21 CFR 892.2050                             |
| Product Code          | PGY                                         |
| Device Class          | II                                          |
| Review Panel          | Radiology                                   |

{4}------------------------------------------------

# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow;

#### Primary Predicate Device

- 510(k) Number:
- Applicant: ●

Classification Name: ●

- Trade Name: ●
# Predicate Device

●

.

●

●

- K201777 510(k) Number: Applicant: LG Electronics Classification Name: Picture archiving and communications system Trade Name: 31HN713D
EIZO Corporation

Picture archiving and communications system

K172738

RX560

# 5. Description of the Device [21 CFR 807.92(a)(4)]

The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images

# 6. Indications for use [21 CFR 807.92(a)(5)]

This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

{5}------------------------------------------------

# 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

The table below presents comparisons between the subject device (32HQ713D) and the legally marketed predicate devices (K172738 and K201777):

|                       | Proposed Device                                                                                                                                                                                                                                                                                                                                  | Primary Predicate Device                                                                                                                                                                                                                                                                             |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number              | Not known                                                                                                                                                                                                                                                                                                                                        | K172738                                                                                                                                                                                                                                                                                              |
| Manufacturer          | LG Electronics Inc.                                                                                                                                                                                                                                                                                                                              | EIZO Corporation                                                                                                                                                                                                                                                                                     |
| Model Name            | 32HQ713D                                                                                                                                                                                                                                                                                                                                         | RX560                                                                                                                                                                                                                                                                                                |
| Classification Name   | Picture archiving and communications system                                                                                                                                                                                                                                                                                                      | Picture archiving and communications system                                                                                                                                                                                                                                                          |
| Classification Number | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                  | 21 CFR 892.2050                                                                                                                                                                                                                                                                                      |
| Indications for Use   | This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.                                                                                                                    | This product is intended to be used in displaying and viewing digital images, including standard and multiframe digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.                            |
| Display Technology    | The 32HQ713D has a LB315WR6-SSZ1 module for displaying. The a LB315WR6-SSZ1 is a 31.5" TFT Liquid Crystal Display module with LED Backlight unit and one 30pin eDP port and 1ch-eDP interface. This module supports 3840x2160 Pixel mode and can display color driven by 10bit drivers. The LCD module includes built-in converter for Backlight | This is a color LCD monitor for viewing digital images, including standard and multiframe digital mammography, for review, analysis and diagnosis. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 2048x2560 pixels. |
| Power Consumption     | MAX. 120W<br>Off Mode ≤ 0.3W                                                                                                                                                                                                                                                                                                                     | 87 W / Less than 1 W                                                                                                                                                                                                                                                                                 |
| Screen size           | 730.70 x 425.20 mm                                                                                                                                                                                                                                                                                                                               | 337.9 mm x 422.4 mm                                                                                                                                                                                                                                                                                  |
| LCD Screen            | TFT LCD                                                                                                                                                                                                                                                                                                                                          | TFT LCD                                                                                                                                                                                                                                                                                              |
| Pixel Pitch           | 0.18159 x 0.18159 mm                                                                                                                                                                                                                                                                                                                             | 0.165 x 0.165 mm                                                                                                                                                                                                                                                                                     |
| Resolution            | 3,840 x 2,160 pixels                                                                                                                                                                                                                                                                                                                             | 2,048 x 2,560                                                                                                                                                                                                                                                                                        |
| Horizontal Frequency  | 30 kHz to 135kHz                                                                                                                                                                                                                                                                                                                                 | 31 kHz to 135 kHz                                                                                                                                                                                                                                                                                    |
| Vertical Frequency    | 24 Hz to 61 Hz                                                                                                                                                                                                                                                                                                                                   | 59 Hz to 61 Hz                                                                                                                                                                                                                                                                                       |
| Input video signals   | DVI IN x 1,<br>DP IN x 1                                                                                                                                                                                                                                                                                                                         | DVI x 1<br>DP IN x 1                                                                                                                                                                                                                                                                                 |

[Table 1. Comparison of Proposed Device to Primary Predicate Device]

The comparison table shows that the subject device (32HQ713D) has the similar indications for use the primary predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.

{6}------------------------------------------------

|                       | Proposed Device                                                                                                                                                                                                                                                                                                                                                           | Predicate Device                                                                                                                                                                                                                             |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number              | Not known                                                                                                                                                                                                                                                                                                                                                                 | K201777                                                                                                                                                                                                                                      |
| Manufacturer          | LG Electronics Inc.                                                                                                                                                                                                                                                                                                                                                       | LG Electronics Inc.                                                                                                                                                                                                                          |
| Model Name            | 32HQ713D                                                                                                                                                                                                                                                                                                                                                                  | 31HN713D                                                                                                                                                                                                                                     |
| Classification Name   | Picture archiving and communications system                                                                                                                                                                                                                                                                                                                               | Picture archiving and communications system                                                                                                                                                                                                  |
| Classification Number | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                           | 21 CFR 892.2050                                                                                                                                                                                                                              |
| Indications for Use   | This Medical Monitor is indicated for use<br>in displaying radiological images<br>(including full-field digital mammography<br>and digital breast tomosynthesis) for<br>review, analysis, and diagnosis by trained<br>medical practitioners.                                                                                                                              | This Medical Monitor is indicated for use<br>in displaying radiological images<br>(including full-field digital mammography<br>and digital breast tomosynthesis) for<br>review, analysis, and diagnosis by trained<br>medical practitioners. |
| Display Technology    | The 32HQ713D has a LB315WR6-SSZ1<br>module for displaying. The a LB315WR6-<br>SSZ1 is a 31.5" TFT Liquid Crystal Display<br>module with LED Backlight unit and one<br>30pin eDP port and 1ch-eDP interface.<br>This module supports 3840x2160 Pixel<br>mode and can display color driven by<br>10bit drivers. The LCD module includes<br>built-in converter for Backlight | The 31HN713D has a LB310FTM module<br>for displaying. The LB310FTM is a 31"<br>TFT Liquid Crystal Display module with<br>LED Backlight unit. This module supports<br>4,200 x 2,800 pixels and can display<br>color driven by 10bit drivers.  |
| Power Consumption     | MAX. 120W<br>Off Mode ≤ 0.3W                                                                                                                                                                                                                                                                                                                                              | MAX. 150W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W                                                                                                                                                                                            |
| Screen size           | 730.70 x 425.20 mm                                                                                                                                                                                                                                                                                                                                                        | 676.9 x 459.7 mm                                                                                                                                                                                                                             |
| LCD Screen            | TFT LCD                                                                                                                                                                                                                                                                                                                                                                   | TFT LCD                                                                                                                                                                                                                                      |
| Pixel Pitch           | 0.18159 x 0.18159 mm                                                                                                                                                                                                                                                                                                                                                      | 0.1554 x 0.1554 mm                                                                                                                                                                                                                           |
| Resolution            | 3,840 x 2,160 pixels                                                                                                                                                                                                                                                                                                                                                      | 4,200 x 2,800 pixels                                                                                                                                                                                                                         |
| Horizontal Frequency  | 30 kHz to 135kHz                                                                                                                                                                                                                                                                                                                                                          | 30 kHz to 135kHz                                                                                                                                                                                                                             |
| Vertical Frequency    | 24 Hz to 61 Hz                                                                                                                                                                                                                                                                                                                                                            | 56 Hz to 61 Hz                                                                                                                                                                                                                               |
| Input video signals   | DVI IN x 1,<br>DP IN x 1<br>DP OUT x 1                                                                                                                                                                                                                                                                                                                                    | DisplayPort x 2<br>HDMI x 1                                                                                                                                                                                                                  |

| [Table 2. Comparison of Proposed Device to Predicate Device] |  |  |
|--------------------------------------------------------------|--|--|
|                                                              |  |  |

The comparison table shows that the subject device (32HQ713D) has the similar indications for use the predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.

{7}------------------------------------------------

#### 8. Non-Clinical Test summary

- 1) Electrical Safety and Electromagnetic Compatibility
The test results demonstrated that the proposed device complies with the following standards:

- · Electrical Basic Safety and Essential Performance requirements in accordance with IEC 60601-1:2015+COPR. 1:2006+COPR. 2:2007 + A1:2012 +A2:2020
- · Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014
- · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010+A1:2013+A2:2020
- 2) Software Validation

The 32HQ713D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

- 3) Guidance
Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017

| Measurements          | Description                                                                                                                                                                                                                                                                                                                                                                                                      | SE Note                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                  |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| a. Spatial resolution | Measurements of the transfer<br>of information from the<br>image data to the luminance<br>fields at different spatial<br>frequencies of interest<br>typically done by reporting<br>the modulation transfer<br>function. Non-isotropic<br>resolution properties should<br>be characterized properly by<br>providing two-dimensional<br>measurements or<br>measurements along at least<br>two representative axes. | As demonstrated through the<br>function of the resolution for<br>the TG18 test pattern and<br>the comparison of MTF<br>measurements, the<br>specification of the proposed<br>device is equivalent to the<br>predicate device.                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                  |
| b. Pixel defects      | Measurements (count, types<br>(e.g., sub-pixel or entire<br>pixel, always-on, always-off),<br>and locations (map) of pixel<br>defects. This is typically<br>provided as a tolerance limit.<br>Pixel defects can interfere<br>with the visibility of small<br>details in medical images.                                                                                                                          | As demonstrated through the<br>measurements of the pixel<br>defects, the specification of<br>the proposed device is an<br>equivalent predicate device.                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                  |
| c. Artifacts          | Evaluate for image artifacts<br>such as ghosting and/or<br>image sticking from<br>displaying a fixed test<br>pattern for a period of time.                                                                                                                                                                                                                                                                       | Demonstrated through the<br>measurements of the Using<br>5x5 mosaic pattern, 64Gray /<br>127 Gray judgment, the<br>specification of the proposed                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                  |
| Measurements          |                                                                                                                                                                                                                                                                                                                                                                                                                  | Description                                                                                                                                                                                                                                                                                                                                                                                                                  | SE Note                                                                                                                                                                                                                                                                                          |
| d.                    | Temporal response                                                                                                                                                                                                                                                                                                                                                                                                | Measurements of the<br>temporal behavior of the<br>display in responding to<br>changes in image values<br>from frame to frame. Since<br>these transitions are typically<br>not symmetric, rise and fall<br>time constants are needed to<br>characterize the system.<br>Slow displays can alter<br>details and contrast of the<br>image when large image<br>stacks are browsed or in<br>video, panning, and zooming<br>modes. | device is equivalence to that<br>of the predicate device.<br>Demonstrated through the<br>measurements of the rise<br>and fall time constants for 5-<br>95% and 40-60% luminance<br>transitions, the specification<br>of the proposed device is<br>equivalent to that of the<br>predicate device. |
| e.                    | Luminance                                                                                                                                                                                                                                                                                                                                                                                                        | Measurements of the<br>maximum and minimum<br>luminance that the device<br>outputs as used in the<br>application under<br>recommended conditions and<br>the achievable values if the<br>device is set to expand the<br>range to the limit.                                                                                                                                                                                   | Demonstrated through the<br>function of the maximum<br>measured, minimum,<br>achievable, and<br>recommended luminance<br>values analysis, the<br>specification of the proposed<br>device is equivalence to that<br>of the predicate device.                                                      |
| f.                    | Conformance to a<br>grayscale-to-luminance<br>function                                                                                                                                                                                                                                                                                                                                                           | Measurements of the<br>mapping between image<br>values and the luminance<br>output following a target<br>model response for 256 or<br>more levels.                                                                                                                                                                                                                                                                           | Demonstrated through the<br>function of the mapping<br>between image values and<br>the luminance output<br>following a target model<br>response for 256 or more<br>levels of analysis, the<br>specification of the proposed<br>device is equivalent to that of<br>the predicate device.          |
| g.                    | Luminance at 30° and 45°<br>in diagonal, horizontal,<br>and vertical directions at<br>center and four corners                                                                                                                                                                                                                                                                                                    | Measurements of the<br>luminance response at off-<br>normal viewing related to the<br>target model for the<br>luminance response.                                                                                                                                                                                                                                                                                            | Demonstrated through the<br>function of luminance<br>response for luminance<br>response, the specification of<br>the proposed device is<br>equivalency to that of the<br>predicate device.                                                                                                       |
| h.                    | Luminance uniformity or<br>Mura test                                                                                                                                                                                                                                                                                                                                                                             | Measurements of the<br>uniformity of the luminance<br>across the display screen.                                                                                                                                                                                                                                                                                                                                             | As demonstrated through the<br>measurements of the<br>uniformity of the luminance<br>across the display screen,<br>the specification of the<br>proposed device is an<br>equivalence predicate device.                                                                                            |
| i.                    | Stability of luminance and<br>chromaticity response<br>with temperature and time                                                                                                                                                                                                                                                                                                                                 | Measurements of the change<br>in luminance and<br>chromaticity response with<br>temperature and use time.                                                                                                                                                                                                                                                                                                                    | As demonstrated through the<br>measurements, there is no<br>display off, and the user can<br>see the display without                                                                                                                                                                             |
|                       | Measurements                                                                                                                                                                                                                                                                                                                                                                                                     | Description                                                                                                                                                                                                                                                                                                                                                                                                                  | SE Note                                                                                                                                                                                                                                                                                          |
|                       |                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                              | temperature on use time; the<br>specification of the proposed<br>device is an equivalence<br>predicate device.                                                                                                                                                                                   |
| j.                    | Spatial noise                                                                                                                                                                                                                                                                                                                                                                                                    | Measurements of the spatial<br>noise level as represented by<br>the noise power spectrum<br>using an appropriate ratio of<br>camera and display pixels.<br>Spatial noise and resolution<br>affect the way images are<br>presented to the viewer and<br>can alter features that are<br>relevant to the interpretation<br>process of the physician or<br>radiologist.                                                          | Demonstrated through the<br>function of the noise levels<br>for the TG18 test pattern and<br>recommended luminance<br>values analysis, the<br>specification of the proposed<br>device is equivalence to that<br>of the predicate device.                                                         |
| k.                    | Reflection coefficient                                                                                                                                                                                                                                                                                                                                                                                           | Measurements of the<br>reflection coefficients of the<br>display device. Specular and<br>diffuse reflection coefficients<br>can be used as surrogates<br>for the full bidirectional<br>reflection distribution<br>function.                                                                                                                                                                                                  | As demonstrated through the<br>measurements of the<br>reflection Coefficient (Rd),<br>the specification of the<br>proposed device is equivalent<br>predicate device.                                                                                                                             |
| l.                    | Veiling glare or small-spot<br>contrast                                                                                                                                                                                                                                                                                                                                                                          | Measurements of the<br>contrast obtained for small<br>targets.                                                                                                                                                                                                                                                                                                                                                               | Demonstrated through the<br>function of the contrast<br>obtained for the TG18 test<br>pattern and recommended<br>luminance values analysis,<br>the specification of the<br>proposed device is<br>equivalence to that of the<br>predicate device.                                                 |
| m.                    | Color tracking                                                                                                                                                                                                                                                                                                                                                                                                   | Chromaticity at different<br>luminance levels of primary<br>colors as indicated by the<br>color coordinates in an<br>appropriate units system<br>(e.g., CIE u'v') and the color<br>gamut enveloped by the<br>primary colors.                                                                                                                                                                                                 | As demonstrated through the<br>measurements of the sRGB<br>xy coverage, the specification<br>of the proposed device is<br>equivalent to that of the<br>predicate device.                                                                                                                         |
| n.                    | Gray tracking                                                                                                                                                                                                                                                                                                                                                                                                    | Chromaticity at different<br>luminance levels of gray<br>shades, including the white<br>point, as indicated by the<br>color coordinates in an<br>appropriate units system                                                                                                                                                                                                                                                    | As demonstrated through the<br>measurements of the sRGB<br>UV coverage, the<br>specification of the proposed<br>device is equivalent to that of<br>the predicate device.                                                                                                                         |

According to the quidance, we test following performance items:

{8}------------------------------------------------

{9}------------------------------------------------

According to the above test results, there are no significant performance differences between 32HQ713D and the predicate devices that would adversely affect the use of the product. It has substantially equivalent performance compared to the predicate devices.

{10}------------------------------------------------

# Clinical Test Summary:

No clinical studies were considered necessary and performed.

# 9. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 32HQ713D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K223423](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K223423)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
