← Product Code [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY) · K222717

# CL24N (K222717)

_Wide Corporation · PGY · Oct 31, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K222717

## Device Facts

- **Applicant:** Wide Corporation
- **Product Code:** [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY.md)
- **Decision Date:** Oct 31, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

CL24N LCD Monitor is intended to be used to display and view digital medical images for review by trained medical practitioners. It does not support the display of mammography images for diagnosis.

## Device Story

CL24N is a 2.1MP color LCD monitor for displaying digital medical images in PACS and radiology environments. Operated by trained medical practitioners; device displays images for clinical review. System includes EzCal software to calibrate display output to DICOM Part 14 GSDF standards. Users periodically check monitor performance using EzCal; if performance falls outside specifications, device requires professional recalibration. Device does not contact patients or control life-sustaining equipment; human intervention remains essential for image interpretation. Benefits include high-resolution visualization of medical imaging data to support clinical decision-making.

## Clinical Evidence

No clinical data. Effectiveness validated through bench testing per FDA guidance 'Display Devices for Diagnostic Radiology'. Tests included pixel defects, artifacts, luminance, reflection, uniformity, veiling glare, color/gray tracking, and MTF. Safety validated via ANSI/AAMI ES60601-1 and IEC 60601-1-2.

## Technological Characteristics

AAS TFT color LCD panel; LED backlighting; 2.1MP resolution (1920x1080); 16:9 aspect ratio; 400 cd/m² max luminance. Connectivity via Mini HDMI and Mini DisplayPort. Includes IQ, ambient light, and human sensors. Software-based calibration (EzCal) for DICOM Part 14 GSDF compliance. Complies with ANSI/AAMI ES60601-1 and IEC 60601-1-2.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RadiForce MX242W ([K143261](/device/K143261.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

WIDE Corporation % Josh Baker Consultant OT Consulting Inc. 33781 Bayside Lane DANA POINT CA 92629

### Re: K222717

October 31, 2022

Trade/Device Name: CL24N Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: September 8, 2022 Received: September 8, 2022

Dear Josh Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K222717

Device Name CL24N

#### Indications for Use (Describe)

The CL24N LCD Monitor System is intended to be used in displaying digital medical images for review by trained medical practitioners. It does not support the display of mammography images for diagnosis.

Type of Use (Select one or both, as applicable):

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

[As required by 21 CFR 807.92]

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92

### 1. Date Prepared [21 CFR 807.92(a) (1)]

06/13/2022

### 2. Submitter's Information [21 CFR 807.92(a) (1)]

| Name of Sponsor:      | WIDE Corporation.                                                                                      |
|-----------------------|--------------------------------------------------------------------------------------------------------|
| Address:              | 15F, The First Tower III, 602, Dongtangiheung-Ro,<br>Hwaseong-Si, Gyeonggi-Do 18469, Republic of Korea |
| Contact Name:         | YeoJin Yun                                                                                             |
|                       | Telephone #: +82-31-218-1675                                                                           |
|                       | Fax #: +82-31-376-9600                                                                                 |
|                       | Email: yyjin@widecorp.com                                                                              |
| Registration Number:  | 3004082357                                                                                             |
| Name of Manufacturer: | Same as Sponsor                                                                                        |
| 510(k) Contact Person | Josh Baker – Consultant, OT Consulting Inc.                                                            |
|                       | 33781 Bayside Lane, Dana Point, California 92629 USA                                                   |
|                       | Tel: 714-788-8152                                                                                      |
|                       | Email: josh@otconsulting.tech                                                                          |

### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

| Model Name:          | CL24N                                          |
|----------------------|------------------------------------------------|
| Common Name:         | 2.1MP Color LCD Monitor                        |
| Classification Name: | Medical Image management and Processing System |
| Regulation Number:   | 21 CFR 892.2050                                |
| Product Code:        | PGY                                            |
| Device Class:        | 2                                              |
| Review Panel:        | Radiology                                      |

### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

| 510(k) Number:       | K143261                       |
|----------------------|-------------------------------|
| Applicant:           | EIZO CORPORATION              |
| Model Name:          | RadiForce MX242W              |
| Common Name:         | 2.3MP Color LCD Monitor       |
| Classification Name: | Display, Diagnostic Radiology |

510(k) Summary 1/4

{4}------------------------------------------------

Regulation Number: 21 CFR 892.2050 Product Code: PGY Device Class: 2

### 5. Description of the Device [21 CFR 807.92(a) (4)]

CL24N is intended to display high resolution color and gravscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying and reviewing of medical images.

EzCal is a software solution which enables the user to modify display output to meet DICOM Part 14 GSDF and other key industry standards.

EzCal is packed with the display.

While using the CL24N product, use the EzCal S/W provided as a bundle to periodically check whether the product meets the intended use.

If the product does not meet the intended use, the product must be returned to the manufacturer or an authorized service center to be calibrated to a product that can be used normally.

### 6. Intended Use [21 CFR 807.92(a) (5)]

CL24N LCD Monitor is intended to be used to display and view digital medical images for review by trained medical practitioners. It does not support the display of mammography images for diagnosis.

### 7. Technological Characteristics [21 CFR 807.92(a) (6)]

The device is an image display system which consists of computer software and components. The device does not contact the patient, nor does it control any life sustaining devices. A physician or trained medical practitioner provides ample opportunity for competent human intervention to interpret images and information being displayed.

| Attributes                                  | Predicate Device                                                                                                                                                                                                  | Subject Device                                                                      | Discussion<br>of<br>Differences                                                                                                                                                                                           |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product                                     | RadiForce MX242W                                                                                                                                                                                                  | CL24N                                                                               |                                                                                                                                                                                                                           |
| Screen<br>technology                        | IPS TFT Color LCD<br>Panel                                                                                                                                                                                        | AAS TFT Color LCD<br>Panel                                                          | Provided by the panel<br>manufacturers.                                                                                                                                                                                   |
| Viewing<br>angle (H, V)                     | H: 178°, V: 178°                                                                                                                                                                                                  | H: 178°, V: 178°                                                                    | -                                                                                                                                                                                                                         |
| Active screen<br>size                       | 518.4 mm x 324.0 mm                                                                                                                                                                                               | 527.04 mm x 296.48 mm                                                               | Provided by the panel<br>manufacturers.                                                                                                                                                                                   |
| Resolution                                  | 2.3MP (1,920 x 1,200)                                                                                                                                                                                             | 2.1MP (1920 x 1080)                                                                 | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Aspect ratio                                | 16 : 10                                                                                                                                                                                                           | 16 : 9                                                                              | Provided by the panel<br>manufacturers.                                                                                                                                                                                   |
| Pixel pitch                                 | 0.270 mm x 0.270 mm                                                                                                                                                                                               | 0.2745 mm x 0.2745<br>mm                                                            | Provided by the panel<br>manufacturers.                                                                                                                                                                                   |
| Maximum<br>luminance                        | 350 cd/m²                                                                                                                                                                                                         | 400 cd/m²                                                                           | Provided by the panel<br>manufacturers.                                                                                                                                                                                   |
| DICOM<br>calibrated<br>luminance            | 180 cd/m²                                                                                                                                                                                                         | 150 cd/m²                                                                           | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Contrast ratio                              | 1000 : 1                                                                                                                                                                                                          | 1000 : 1                                                                            | -                                                                                                                                                                                                                         |
| Backlighting                                | LED                                                                                                                                                                                                               | LED                                                                                 |                                                                                                                                                                                                                           |
| Display<br>Colors                           | From a palette of 68<br>billion colors:<br>- 10-bit input<br>(DisplayPort): 1.07<br>billion colors<br>(maximum)<br>- 8-bit input: 16.77<br>million colors                                                         | From a palette of 1.07<br>billion colors:<br>- 8-bit input: 16.77<br>million colors | Tone between the<br>predicate device and<br>our subject devices<br>are different. But It<br>pass the exams in<br>AAPM-TG18 4.3 "<br>Luminance response".<br>Therefore, they are<br>equivalent to the<br>predicate device. |
| Luminance<br>non-uniformity<br>compensation | Digital Uniformity<br>Equalizer                                                                                                                                                                                   | -                                                                                   | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Input video<br>signals                      | DVI-I x 1,<br>DisplayPort x 1                                                                                                                                                                                     | Mini HDMI x 1,<br>Mini DisplayPort x 1                                              | It is only a difference<br>connector types, but<br>the functions are<br>similar.                                                                                                                                          |
| Scanning<br>Frequency<br>(H, V)             | Digital: 31 - 76 kHz / 59<br>- 61 Hz (VGA Text: 69 -<br>71 Hz)<br>Frame synchronous<br>mode: 59 - 61 Hz<br>Analog: 26 - 76 kHz /<br>49 - 71 Hz (VGA Text:<br>69 - 71 Hz)<br>Frame synchronous<br>mode: 59 - 61 Hz | 67.5KHz, 60Hz                                                                       | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Power<br>Requirements                       | AC 100 - 240 V,<br>50 / 60 Hz                                                                                                                                                                                     | AC 100~240 V: 50 /<br>60 Hz<br>DC +12V, 3A                                          | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Power<br>Consumption<br>/ Save Mode         | 68 W / Less than 0.5 W                                                                                                                                                                                            | 30 W / Less than 5 W                                                                | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Power<br>Management                         | Digital : DVI DMPM,<br>DisplayPort 1.1a<br>Analog : VESA DPM                                                                                                                                                      | DVI DMPM,<br>DisplayPort 1.1a                                                       | Differences by<br>manufacturers.                                                                                                                                                                                          |
| QC software                                 | RadiCS                                                                                                                                                                                                            | EzCal                                                                               | It is only a difference<br>in terms of each<br>manufacturer, but the<br>functions are similar.                                                                                                                            |
| Sensors                                     | Backlight Sensor                                                                                                                                                                                                  | IQ Sensor,<br>Ambient Light Sensor,<br>Human Sensor                                 | Differences by<br>manufacturers                                                                                                                                                                                           |
| USB Ports /<br>Standard                     | 1 upstream,<br>2 downstream / Rev.<br>2.0                                                                                                                                                                         | -                                                                                   | Differences by<br>manufacturers.                                                                                                                                                                                          |
| Dimensions<br>w/o stand<br>(W x H x D)      | 575 x 398 x 71 mm                                                                                                                                                                                                 | 538.52 x 326.22 x 35.1<br>mm                                                        | Different housing<br>design.                                                                                                                                                                                              |

### 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

510(k) Summary 2/4

{5}------------------------------------------------

{6}------------------------------------------------

CL24N Device is substantially equivalent to the currently marketed and predicate devices in terms of design features, indications for use, and safety and effectiveness.

#### 9. Summary of Non-Clinical Data

CL24N comply with the following international and FDA-recognized consensus standards: ANSI/AAMI ES60601-1: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests

The bench tests below were performed on the CL24N following the instructions in 'Display Devices for Diagnostic Radiology - Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017.

- · Test Item
Pixel Defects, Artifacts, Luminance, Reflection, Luminance Uniformity, Veiling Glare, Color Uniformity, Luminance Response, Luminance at 30° and 45° in horizontal, and vertical directions, Luminance Stability test, Color Tracking, Gray Tracking, MTF

The tests showed that the device has similar characteristics compared to the predicated device.

### 10. Summary of Clinical Data

No clinical studies were considered necessary and performed.

### 11. Conclusion [21 CFR 807.92(b) (3)|

Subject Device is substantially equivalent to the currently marketed and predicate devices in terms of design features, indications for use, and safety and effectiveness.

Additionally, the safety of the subject device was validated through tests including ANSI/AAMI ES 60601-1 and IEC 60601-1-2. The effectiveness of the device was validated through bench tests.

The results of these tests demonstrate that CL24N meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing demonstrates that the device is as safe and effective as the predicate device and performs as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K222717](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K222717)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
