← Product Code [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY) · K170949

# RadiForce MX315W (K170949)

_Eizo Corporation · PGY · Apr 27, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K170949

## Device Facts

- **Applicant:** Eizo Corporation
- **Product Code:** [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY.md)
- **Decision Date:** Apr 27, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This product is intended to be used in displaying digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.

## Device Story

RadiForce MX315W is a 31.1-inch color TFT IPS LCD monitor; 8MP resolution (4,096 x 2,160); LED backlighting. Used in clinical settings by trained medical practitioners for viewing/analyzing medical images (excluding mammography). Device receives digital video signals via DVI-D or DisplayPort; internal lookup tables apply factory-calibrated tone curves (DICOM GSDF compliant) to ensure accurate luminance and color reproduction. Digital Uniformity Equalizer compensates for luminance non-uniformity. Integrated sensors (backlight, front, presence, ambient light) support quality control. RadiCS application software installed on workstations performs quantitative and visual QC tests. Output is a high-resolution visual display of medical images, aiding clinicians in diagnostic decision-making.

## Clinical Evidence

Bench testing only. Conformance to DICOM GSDF, luminance/chromaticity non-uniformity, MTF spatial resolution, and pixel defect testing performed per AAPM TG18 guidelines. No clinical or animal testing performed.

## Technological Characteristics

TFT Color LCD Panel (IPS); 8MP (4,096 x 2,160) resolution; LED backlight; 1300:1 contrast ratio; 450 cd/m2 max luminance. Connectivity: DVI-D, DisplayPort. Integrated sensors: backlight, front, presence, ambient light. Software: RadiCS for QC. Digital Uniformity Equalizer. Power: 100-240V AC.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RadiForce RX840 ([K113226](/device/K113226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three overlapping profiles of human faces, oriented to the right. The profiles are depicted in a minimalist style, with flowing lines suggesting hair and facial features.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 27, 2017

EIZO Corporation Hiroaki Hashimoto Manager Medical System Standards 153 Shimokashiwano. Hakusan Ishikawa 924-8566 JAPAN

Re: K170949

Trade/Device Name: RadiForce MX315W Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: PGY Dated: March 27, 2017 Received: March 31, 2017

Dear Hiroaki Hashimoto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Robert Oals

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name RadiForce MX315W

Indications for Use (Describe)

This product is intended to be used in displaying digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for EIZO. The logo consists of a stylized, geometric shape on the left and the word "EIZO" in bold, sans-serif font on the right. The geometric shape is a square divided into smaller squares, some of which are shaded to create a checkered pattern. The "EIZO" text is in black, and there is a small "®" symbol next to the "O".

EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan

U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

Name Department

Hiroaki Hashimoto Medical System Standards

Telephone Fax E-Mail

+81 (76) 274-2468 +81 (76) 274-2484 hiroaki.hashimoto@eizo.com

# 510(k) Summary (in accordance with 21 CFR 807.92)

### 1. Company

EIZO Corporation 153 Shimokashiwano, Hakusan Ishikawa 924-8566 Japan Tel: +81 (76) 274-2468 Fax: +81 (76) 274-2484

# 2. Contact Person

Hiroaki Hashimoto

- 3. Date of Summary March 27th, 2017
## 4. Device Information

- RadiForce MX315W Device Name: .
- Trade Name/Model: RadiForce MX315W
- Common Name: 79.0 cm (31.1 inch) class Color LCD Monitor ●
- Classification Name: ● Display, Diagnostic Radiology
- Regulation Number: ● 21 CFR 892.2050, Product Code PGY

### 5. Predicate Device

8MP Color LCD Monitor, RadiForce RX840 (K113226)

{4}------------------------------------------------

# 6. Device Description

RadiForce MX315W is a color LCD monitor for viewing medical images other than those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 4,096 x 2,160 pixels (8MP) with a pixel pitch of 0.1704 mm.

Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce MX315W based on the QC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce MX315W.

## 7. Indications for use

This product is intended to be used in displaying and viewing digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.

{5}------------------------------------------------

# 8. Comparison of Technological Characteristics

The comparison table below enumerates information derived from the product brochure of the each device and different technological characteristics are discussed in it:

| Attributes                                      | RadiForce MX315W                                                                                                                           | RadiForce RX840                                                                                                                            | Explanation of<br>Differences                                                                                                                                                               |  |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Display Performance/Specifications              |                                                                                                                                            |                                                                                                                                            |                                                                                                                                                                                             |  |
| Screen<br>technology                            | TFT Color<br>LCD Panel (IPS)                                                                                                               | TFT Color<br>LCD Panel (IPS)                                                                                                               |                                                                                                                                                                                             |  |
| Viewing angle<br>(H, V)                         | H: 178°, V: 178°                                                                                                                           | H: 176°, V: 176°                                                                                                                           | -                                                                                                                                                                                           |  |
| Resolution                                      | 8MP (4,096 x 2,160)                                                                                                                        | 8MP (4,096 x 2,160)                                                                                                                        |                                                                                                                                                                                             |  |
| Aspect ratio                                    | 17 : 9                                                                                                                                     | 17 : 9                                                                                                                                     |                                                                                                                                                                                             |  |
| Active screen<br>size                           | 697.9 mm x 368.0 mm                                                                                                                        | 817.1 mm x 430.9 mm                                                                                                                        |                                                                                                                                                                                             |  |
| Pixel pitch                                     | 0.1704 mm x 0.1704 mm                                                                                                                      | 0.1995 mm x 0.1995 mm                                                                                                                      |                                                                                                                                                                                             |  |
| Maximum<br>luminance                            | 450 cd/m²                                                                                                                                  | 700cd/m²                                                                                                                                   | Both devices'<br>Maximum and                                                                                                                                                                |  |
| DICOM<br>calibrated<br>luminance                | 270 cd/m²                                                                                                                                  | 400 cd/m2                                                                                                                                  | DICOM calibrated<br>luminance are greater<br>than 171cd/m2 which<br>is required for<br>primary displays in<br>AAPM On-line<br>Report No. 03:2005<br>(Sec. 4.3.4.2.1) with<br>enough margin. |  |
| Contrast ratio                                  | 1300 : 1                                                                                                                                   | 1000: 1                                                                                                                                    |                                                                                                                                                                                             |  |
| Response<br>Time (typical)                      | 20 ms (On/Off)                                                                                                                             | 25 ms (On/Off)                                                                                                                             |                                                                                                                                                                                             |  |
| Backlighting                                    | LED                                                                                                                                        | LED                                                                                                                                        |                                                                                                                                                                                             |  |
| Display Colors                                  | 10-bit colors<br>(DisplayPort) : 1.07 billion<br>(maximum) colors<br>8-bit colors: 16.77 million<br>from a palette of 68 billion<br>colors | 10-bit colors<br>(DisplayPort) : 1.07 billion<br>(maximum) colors<br>8-bit colors: 16.77 million<br>from a palette of 68 billion<br>colors |                                                                                                                                                                                             |  |
| Luminance<br>non-<br>uniformity<br>compensation | Digital Uniformity<br>Equalizer                                                                                                            | Digital Uniformity<br>Equalizer                                                                                                            |                                                                                                                                                                                             |  |
| Video Signals                                   |                                                                                                                                            |                                                                                                                                            |                                                                                                                                                                                             |  |
| Input video<br>signals                          | DVI-D (dual link) x 1,<br>DisplayPort x 2                                                                                                  | DVI-D (dual link) x 2,<br>DisplayPort x 2                                                                                                  |                                                                                                                                                                                             |  |

{6}------------------------------------------------

| Output video signals                  | DisplayPort x 1 (daisy chain)                                                                                 | -                                                                                                                                                        |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Scanning Frequency (H, V)             | 31 - 134 kHz / 14 - 61 Hz,<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous mode:<br>29.5 - 30.5 Hz, 59 - 61 Hz | 31 - 140 kHz / 29.5 - 30.5 Hz (2048 x 2160, 1920 x 2160),<br>59 - 61 Hz, (VGA Text: 69 - 71 Hz)<br>Frame synchronous mode:<br>29.5 - 30.5 Hz, 59 - 61 Hz |
| Power Related Specifications          |                                                                                                               |                                                                                                                                                          |
| Power Requirements                    | AC 100 - 240 V:<br>50 / 60 Hz                                                                                 | AC 100 - 120 V,<br>200 - 240 V: 50 / 60 Hz                                                                                                               |
| Power Consumption / Save Mode         | 125 W / Less than 1.6 W                                                                                       | 350 W / Less than 6 W                                                                                                                                    |
| Power Management                      | DVI DMPM,<br>DisplayPort 1.2a                                                                                 | DVI DMPM,<br>DisplayPort 1.1a                                                                                                                            |
| Miscellaneous Features/Specifications |                                                                                                               |                                                                                                                                                          |
| QC software                           | RadiCS                                                                                                        | RadiCS                                                                                                                                                   |
| Sensors                               | Backlight Sensor,<br>Integrated Front Sensor,<br>Presence Sensor,<br>Ambient Light Sensor                     | Backlight Sensor,<br>Integrated Front Sensor,<br>Presence Sensor,<br>Ambient Light Sensor                                                                |
| USB Ports / Standard                  | 2 upstream,<br>3 downstream / Rev. 2.0                                                                        | 1 upstream,<br>2 downstream / Rev. 2.0                                                                                                                   |
| Dimensions w/o stand<br>(W x H x D)   | 733 x 408 x 64.5 mm                                                                                           | 896 x 527 x 157 mm                                                                                                                                       |

It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the MX315W.

{7}------------------------------------------------

## 9. Performance Testing

The following bench tests were performed on the RadiForce MX315W.

- Verification of the conformance to DICOM GSDF as specified in Assessment of . Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- Measurement of the luminance non-uniformity characteristics of the display screen . as specified in TG18 guideline
- . Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline
- . Measurement of the chromaticity at the center of the display screen at 5%, 50% and 95% of the maximum luminance
- Visual check of presence or absence of miscellaneous artifacts on the display screen . as specified in TG18 guideline
- . Measurement of spatial resolution expressed as modulation transfer function (MTF)
- The maximum number allowed for each type of pixel defects/faults ●

The test results showed that the RadiForce MX315W has display characteristics equivalent to those of the predicate device, RadiForce RX840.

Besides, the display characteristics of the RadiForce MX315W meet the pre-defined criteria when criteria are set.

No animal or clinical testing was performed on the RadiForce MX315W.

### 10. Conclusion

The RadiForce MX315W was determined to be substantially equivalent to the predicate device due to the following reasons:

- . The stated intended use is substantially the same as that of the predicate device.
- . It was confirmed that the technological characteristics differences from those of the predicate device do not affect the safety or the effectiveness.
- . The bench tests demonstrated that the display characteristics are equivalent to those of the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K170949](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K170949)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
