← Product Code [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY) · K151134

# 21.3 inch (54cm) Color LCD Monitor CCL210 (CL21210) (K151134)

_Jvckenwood Corporation · PGY · May 27, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K151134

## Device Facts

- **Applicant:** Jvckenwood Corporation
- **Product Code:** [PGY](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY.md)
- **Decision Date:** May 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210) is intended to be used in displaying medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.

## Device Story

The CCL210 is a 21.3-inch color LCD monitor designed for clinical diagnostic imaging. It accepts medical image data via DVI-I or DisplayPort interfaces from a workstation. The device utilizes an LED backlight system to provide stable, high-luminance output (250 cd/m² DICOM calibrated) and displays images in accordance with the DICOM Grayscale Standard Display Function (GSDF) using internal lookup tables. It is intended for use by trained medical practitioners in clinical environments to assist in diagnostic decision-making. The monitor includes a backlight sensor for luminance stabilization and supports optional external photo sensors for calibration via software. By providing high-resolution, calibrated visual output, the device enables clinicians to view medical images with consistent quality, supporting accurate diagnosis.

## Clinical Evidence

Bench testing only. Performance was validated through system and performance testing, including luminance uniformity (AAPM-TG18 4.4), chromaticity (AAPM-TG18 4.8.4), and pixel defect analysis (ISO13406-2). No clinical data was required or provided.

## Technological Characteristics

21.3-inch color LCD with LED backlight; 1600x1200 resolution; DVI-I and DisplayPort connectivity; 1500:1 contrast ratio. Complies with ANSI/AAMI ES60601-1, IEC60601-1, IEC60601-1-2, and ISO13406-2. Features internal lookup tables for DICOM GSDF compliance and a backlight sensor for luminance stabilization.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- 20.1 inch (51 cm) Color LCD Monitor CCL208 (CDL2013A) ([K092728](/device/K092728.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 27, 2015

JVC KENWOOD Corporation % Mr. Tsukasa Tashiro Engineering Specialist 3-12 Moriya-cho, Kanagawa-ku Yokohama-shi 221-0022 JAPAN

Re: K151134

Trade/Device Name: 21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: April 28, 2015 Received: April 29, 2015

Dear Mr. Tashiro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K151134

### Device Name

21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210)

Indications for Use (Describe)

21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210) is intended to be used in displaying medical images for diagnosis by trained medical practitioners. It is not meant to be used in digital mammography.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                         |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# TOTOKU

# 510(k) SUMMARY

| Submitted Information:   | JVC KENWOOD CORPORATION<br>3-12, MORIYA-CHO, KANAGAWA-KU,<br>YOKOHAMA-SHI, KANAGAWA, 221-0022 JAPAN                                                                                                                                                 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:          | Tsukasa Tashiro, Engineering Specialist<br>Email: tashiro.tsukasa@jvckenwood.com<br>Tel: +81.258.24.6611<br>Fax: +81.258.24.6617                                                                                                                    |
| Date Prepared:           | April 28, 2015                                                                                                                                                                                                                                      |
| Device Name:             | 21.3 inch (54 cm) Color LCD Monitor CCL210 (CL21210)                                                                                                                                                                                                |
| Common Name:             | CCL210, CL21210                                                                                                                                                                                                                                     |
| Classification Name:     | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System)                                                                                                                                                |
| Predicate Device:        | 20.1 inch (51 cm) Color LCD Mo<br>nitor CCL208 (CDL2013A)<br>(K092728)                                                                                                                                                                              |
| Device Description:      | CCL210 (CL21210) is a 21.3-inch (54 cm) Color LCD monitor<br>whose display resolution is 1600 x 1200 (landscape), 1200 x 1600<br>(portrait) supporting DVI-I (both digital and analog interface) and<br>DisplayPort (digital interface).            |
| Intended Use:            | 21.3 inch (54 cm) Co lor 2M pixel LCD M<br>onitor, CCL210<br>(CL21210) is intended to be used in displaying<br>and viewin g<br>medical images for diagnosis by trained medical practitioners. It is<br>not meant to be used in digital mammography. |
| Substantial Equivalence: | CCL210 (CL21210) shares the same technical characteristics,<br>application and intended use with our predicate device CCL2 08<br>(K092728)                                                                                                          |

{4}------------------------------------------------

# Technical Specification

- 1. Luminance uniformity
	- [SPEC] Less than 30% based on AAPM-TG18 4.4. Refer to actual Luminance uniformity data
- 2. Pixel Defects / Fault [SPEC] Class II or more. ISO13406-2
- 3. Artifacts
	- · phase/clock issues flicker
	- miscellaneous including ringing, ghosting, image sticking [SPEC] By visible check, no flicker, ringing, ghosting and image sticking
- 4. Chromaticity Measurement of 5%, 50%, 95% Level [SPEC] Refer to actual data.
- 5. Chromaticity

[SPEC] Delta (u', v') ≤ 0.01 measured at 80% Lmax based on AAPM-TG18 4.8.4 Refer to Chromaticity actual data

- 6. Power On Luminance Drift [SPEC] Refer to actual data.

{5}------------------------------------------------

|                                     | CCL208 (CDL2013A)                                                                                                  | CCL210 (CL21210)                                                                                                                  |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                       | K092728                                                                                                            | Not Known                                                                                                                         |
| Display Area                        | Horizontal: 408.0mm, Vertical: 306.0mm                                                                             | Horizontal: 432.0mm, Vertical: 324.0mm                                                                                            |
| Input Signal                        | Mini D-sub 15-pin connector (same connector)<br>DVI-I 29-pin connector                                             | DVI-I 29-pin connector, DisplayPort connector                                                                                     |
| Maximum Resolution                  | 1200 x 1600 at portrait display<br>1600 x 1200 at landscape display                                                | 1200 x 1600 at portrait display<br>1600 x 1200 at landscape display                                                               |
| Pixel Pitch                         | 0.255 mm x 0.255mm                                                                                                 | 0.270 mm x 0.270 mm                                                                                                               |
| Scanning Frequency                  | Horizontal: 30 – 75kHz<br>Vertical: 55 – 60Hz                                                                      | Horizontal: 30 – 75kHz<br>Vertical: 55 – 60Hz                                                                                     |
| Maximum Image Clock                 | 162MHz                                                                                                             | 162MHz                                                                                                                            |
| Maximum Luminance                   | 120 cd/m² DICOM calibrated<br>300 cd/m² typ. As LCD component                                                      | 250 cd/m² DICOM calibrated<br>440 cd/m² typ. As LCD component                                                                     |
| Luminance Calibration<br>(Optional) | Software<br>Photo Sensor (optional): X-Rite Chroma 5                                                               | Software<br>Photo Sensor (optional): X-Rite Chroma 5                                                                              |
| Contrast Ratio                      | 800:1                                                                                                              | Typ 1500:1                                                                                                                        |
| Serial Communication                | DDC-ci                                                                                                             | USB                                                                                                                               |
| Safety Standards                    | Medical Safety: MET (for US)/MET-C (for C<br>anada),<br>IEC60601-1, IEC60601-1-2, IEC60950-1, FC<br>VCCI-B, MDD/CE | Medical Safety: ANSI/AAMI ES60601-1, CAN/CSA C22.2 No.<br>60601-1, FCC (Class B), MDD/CE, VCCI-B (Class B), ICES-003<br>(Class B) |
| Weight & Dimension                  | Net: 10.7kg<br>453(w) x 462 - 523(H) x 220(D) mm<br>Packed: 15.5kg<br>470(w) x 685(H) x 345(D) mm                  | Net: 10.7kg<br>473(w) x 476.4 – 537.9(H) x 220(D) mm<br>Packed: 15.5kg<br>470(w) x 675(H) x 347(D) mm                             |
| Power Supply                        | 100-240V AC, 50/60Hz                                                                                               | 100-240V AC, 50/60Hz                                                                                                              |

# bstantial Equivalence Compari

.210 (CL21210) employs a different LCD panel, power supply, and main board from predicate device CCL208 (K0927 ifferences:

CL210 (CL21210) employs the same tilt stand as that employed on the predicate device CCL208 (K09278). Th
zel and back enclosure of CCL210 (CL21210) are slightly larger but cl Similarities

{6}------------------------------------------------

CCL210 (CL21210) can be considered to have equivalent display performances to those of the predicate device CCL208 (K092728) due to the following reasons:

- a. The active area of CCL210 (CL21210) differs from that of the predicate device CCL208 (K092728) in size: 432.0mm x 324.0mm for CCL210 and 408.0mm x 305.0mm for the predicate device. Both devices, however, have the same maximum display resolution of 1600 x 1200.
- b. The DICOM calibrated luminance value of the predicate device CCL208 (K092728) is 120cd/m2, whereas that of CCL210 (CL21210) is set to a higher value of 250cd/m2. In regard to the maximum luminance, CCL210 (CL21210) also achieves superior value of 400cd/m2 compared to the predicate device CCL208 (K092728) with 300cd/m2. Additionally CCL210 (CL21210) realizes high luminance at constantly stable level by adopting LED backlight, which lasts longer than conventional CCFL backlights do.
- c. The LED backlight was newly employed instead of CCFL backlight because it is mercury-free, consumes less power and deteriorates more slowly. We have not recognized any adverse effects of the LED backlight on the quality of displayed images. Refer to "Technical Data" where several image quality characteristics of the proposed device are compared with those of the predicate device.
- d. The both devices display images in accordance with DICOM GSDF by default utilizing the factory calibrated display mode stored in lookup tables inside of them.
- e. The predicate device CCL208 (K092728) supports DVI-D (digital interface) and D-Sub (analog interface), whereas CCL210 (CL21210) supports DVI-I (both digital and analog interface) and DisplayPort (digital interface).

As for the maintenance, the same QC software is used for both devices. Both devices have Backlight Sensor to stabilize the luminance.

The overall design of the CCL210 (CL21210) was validated in accordance with internationally recognized Safety and EMC standards by third-party certifiers. Besides, JVC KENWOOD Corporation performed a range of system and performance tests to ensure that the CCL210 (CL21210) performs in accordance with its specifications. None of the tests revealed behaviors inconsistent with the expected performance.

## Conclusion

The 2M pixel Color LCD Monitor, CCL210 (CL21210) is substantially equivalent to the predicate device with respect to technical characteristics, application and intended use. The specifications of the primary component employed by the proposed device are the same to those of the predicate device and other differences have been independently validated. Any differences between the devices do not affect safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K151134](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/PGY/K151134)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
