K980322 · GE Medical Systems · OXO · Apr 24, 1998 · Radiology
Device Facts
Record ID
K980322
Device Name
ACCENT 16, ACCENT 22
Applicant
GE Medical Systems
Product Code
OXO · Radiology
Decision Date
Apr 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The ACCENT systems are mobile image-intensified fluoroscopic x-ray systems, intended for use in general surgical procedures such as orthopedics, traumatology, visceral surgery, pacemaker implants, urology, etc. or other interventional procedures.
Device Story
Mobile C-arm fluoroscopic X-ray system; includes X-ray generator, tube, image intensifier with CCD camera, and monitor cart with digital image processing/storage. Used in surgical settings (OR, clinic) by physicians/technicians for real-time imaging during orthopedics, traumatology, visceral surgery, and pacemaker implants. Provides live fluoroscopic visualization to guide interventional procedures; assists clinicians in anatomical visualization and device placement. Benefits include intraoperative imaging capability, reducing need for patient transport. System features include motorized C-arm movements (tilting, rotation) and optional laser alignment/dose area product monitoring.
Clinical Evidence
Bench testing only. Compliance with UL 2601-1 and IEC 601-1 standards for electrical/mechanical safety and 21 CFR 1020.30-32 for radiation safety.
Technological Characteristics
Mobile C-arm fluoroscopic X-ray system. Materials compliant with UL 2601-1 and IEC 601-1. Energy source: 100-240V, 50/60Hz. Max tube potential: 110 kV. Features: CCD video camera, digital image processing/storage, beam-limiting device, optional laser alignment, and dose area product meter. Hardware redundancies for X-ray emission and C-arm motion.
Indications for Use
Indicated for patients undergoing general surgical procedures including orthopedics, traumatology, visceral surgery, pacemaker implants, urology, or other interventional procedures requiring mobile image-intensified fluoroscopic x-ray imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
STENOSCOP 2, Model 6000
STENOSCOP 2, Model 9000
Submission Summary (Full Text)
{1}------------------------------------------------
K980 322
APR 2 4 1998
# 510(k) SUMMARIES OF SAFETY AND EFFECTIVENESS
# IDENTIFICATION OF THE PRODUCT
ACCENT, MODELS 16 and 22 NAME: MOBILE X-RAY SYSTEM
GE MEDICAL SYSTEMS - EUROPE MANUFACTURER: Loncin, Belgium
GE MEDICAL SYSTEMS DISTRIBUTOR: Milwaukee, Wi
### INDICATIONS FOR USE:
The ACCENT systems are mobile image-intensified fluoroscopic x-ray systems, intended for use in The ACCENT systems are noble intension instruction instructions of the orgery, pacemaker
general surgical procedures such as orthopedics, traumatology, visceral surgery, pace general surgiour procedures or other interventional procedures.
### DEVICE DESCRIPTION
The ACCENT system is composed of:
a mobile C-Arm that includes:
- an image intensifier with CCD video camera .
- an x-ray generator and controller ●
- an x-ray generator and comfolio.
an x-ray tube mounted in the same housing as the x-ray generator ●
- a beam-limiting device ●
- optional: a Laser beam alignment device .
- optional: a dose area product meter and display .
and a monitor cart that includes:
- TV monitors ●
- r v Thonkols
an image digital processing and storage system, may be of different models .
- optional: VCR and video printers ●
| Materials: | All construction and materials are compliant with UL 2601-1, and IEC 601-1 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | equipment standards. |
| | |
| Design: | The design meets or exceeds standards set forth in UL 2601-1 and IEC 601-<br>1. There are hardware redundancies to prevent single point failures (such as<br>x-ray emission and C-Arm motion) |
| Energy Source: | Single phase, plus ground, 100-108-120-200-208-220-228-240 V, 50/60<br>Hz.<br>The supply voltage is set at the installation. |
| Exposure levels: | Maximum rated peak tube potential: 110 kV<br>Fluoroscopic Entrance Exposure Rate (EER): less than 10 R/min.<br>Radiography: 160 mAs maximum, manual setting<br>Systems comply with the requirements of 21CFR 1020.30 - 31 - 32 |
| Mechanical features: | - mobile: forth and back, interlocked lateral travel<br>- C-Arm: forth and back, tilting (wig-wag), rotations |
{2}------------------------------------------------
## MARKETING HISTORY:
The ACCENT systems perform the same functions and replace the units listed below:
STENOSCOP 2, Model 6000 ACCENT 16: STENOSCOP 2, Model 9000 ACCENT 22:
# ADVERSE EFFECTS ON HEALTH:
The potential hazards (x-ray, mechanical and electrical) are controlled by compliance with the label The potential hazards (X-ray, mechanical and electived) are ochinations of other in the labeling.
federal x-ray standard, adherence to industry standards (UL 2601-1) and wari
#### CONCLUSIONS
The ACCENT systems are substantially equivalent to the above listed devices. The ACCENT systems are substantians of operation are very similar.
The energy source, exposure levels and principles of operation are very similar. The energy source, exposure levels and pinciples of operation in the courte.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a central emblem. The emblem depicts a stylized image of three human profiles facing right, with flowing lines extending from their heads, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Mr. Mike Chilbert Quality, Safety, & Regulatory Engineer Technology & Product Assurance Group GE Medical Systems P.O. Box 414, W-709 MILWAUKEE WI 53201
NOV 1 7 2011
Re: K980322
Trade/Device Name: Accent 16 and 22 (Mobile X-ray System) Regulation Number: 21 CFR 892.1650 Regulation Name: Image intensified fluoroscopic x-ray system, mobile Regulatory Class: II Product Code: OXO Dated: January 22, 1998 Received: January 27, 1998
Dear Mr. Chilbert:
This letter corrects our substantially equivalent letter of April 24, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please Note: CDRH does not evaluate information related to contact liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely Yours,
Michael D. O'Hannon for
Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
avice Name: __ Accent Mobile X-RAY System_
Indications For Use:
The ACCENT systems are mobile image-intensified fluoroscopic x-ray systems, intended for use in general surgical procedures such as orthopedics, traumatology, visceral surgery, pacemaker implants, urology, etc. or other interventional procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Slavid C. Segnor
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devi
510(k) Number K980322
> Prescription Use_ (Per 21 CFR 801.109)
OR
ﺎ.
Over-The-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.