The ORCA is a small image intensified fluoroscopic C-Arm, for imaging extremities in orthopedic applications. The ORCA Operator's manual contains the following statement: "CAUTION: Federal Law (USA) restricts this device to sale and use by or on the order of a physician."
Device Story
ORCA is a mobile image-intensified fluoroscopic C-Arm for orthopedic extremity imaging. Device utilizes an X-ray source (stationary anode, 0.3 mm focal spot) to generate fluoroscopic images; system features digital image rotation, 6-inch image intensifier, and 17-inch monitor. Operated by physicians in clinical settings for real-time visualization of extremities. Output consists of fluoroscopic images displayed on a monitor, supporting clinical decision-making during orthopedic procedures. System includes image memory/processing capabilities (512x512x12 or 1024x1024x12) and interfaces for printers, VCRs, and DICOM 3 connectivity. Benefits include low operator radiation exposure and enhanced positioning flexibility (270-degree orientation) compared to predicate. Device complies with 21 CFR 1020.30-32 and various IEC safety standards.
Clinical Evidence
Bench testing only. Device performance verified against 21 CFR 1020.30-32 and IEC 601 series standards. Radiation exposure metrics provided: operator exposure < 2 mR/hr; skin entrance dose < 5 R/min, which is 2x below the 10 R/min regulatory limit.
Indicated for fluoroscopic visualization of patient extremities in orthopedic applications. Restricted to use by or on the order of a physician.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
OEC Medical Systems Inc. Series 6600 Digital Mobile C-Arm (K951765)
Submission Summary (Full Text)
{1}
LUNAR
K960907
313 W. BELTLINE HIGHWAY
MADISON, WI 53713
(608) 274-2663
MAY 20 1996
# 9.0 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92(c).
Contact Person: Gary Syring
LUNAR Corporation
313 West Beltline Highway
Madison, WI 53713
Phone: (608) 274-2663
Fax: (608) 274-0853
Date: February 23, 1996
Device/Trade Name: ORCA
Common Name: Ortho C-Arm
Classification Name: Image Intensified Fluoroscopic X-Ray System
21CFR 892.1650
Predicate Device: K95 1765
OEC Medical Systems Inc.
Series 6600 Digital Mobile C-Arm
# 9.1 DESCRIPTION OF THE DEVICE:
The ORCA is a small image intensified fluoroscopic C-Arm, for imaging extremities in orthopedic applications. The ORCA is substantially equivalent to the Series 6600 Digital Mobile C-Arm, manufactured by OEC Medical Systems, Inc.
# 9.2 SUMMARY OF TECHNICAL CHARACTERISTICS
The technical characteristics of the ORCA are substantially equivalent to the Series 6600 Digital Mobile C-Arm. A comparison table of technical characteristics follows.
Summary of Safety, Effectiveness Page 1 of 3
Page 2/
{2}
| Feature | Predicate OEC MINI 6600 | ORCA |
| --- | --- | --- |
| Intended Use: | Provide physician with general fluoroscopic visualization of the patient's extremities | Provide physician with general fluoroscopic visualization of patient's extremities |
| X-Ray Source: | Stationary Anode
0.25 mm focal spot
75 KVp/0.1 mA (7.5w)
40 cm SID | Stationary Anode
0.3 mm focal spot
80 KVp/0.7 mA (56w)
45 cm SID |
| Image Information: | Motorized Image Rotation
4 or 6 inch image intensifier
16 inch monitor | Digital Image Rotation
6 inch image intensifier
17 inch monitor |
| Fluoroscopy: | Manual Control -
(combined KVp/mA)
Automatic Exposure Rate Control | Manual Control -
(combined KVp/mA)
Automatic Exposure Rate Control |
| Image Memory/Processing: | 640 X 510 X 10 | 512 X 512 X 12
1024 X 1024 X 12 |
| Image Handling Interface: | Printer
VCR Option
DICOM 3
Parallel Port | Printer
VCR Option
DICOM 3
Parallel Port |
| C-Arm Specification: | 30 cm Opening (with variance)
9 cm Source to Skin
Distance (with variance)
Manual mechanical positioning
200 degrees of orientation | 35 cm Opening (with variance)
10 cm Source to Skin
Distance (with variance)
Manual mechanical positioning
270 degrees of orientation |
| Scatter Radiation to Operator: | < 5 mR/Hr | < 2 mR/Hr |
| Power Requirements: | 110VAC, 1500 W | 110VAC, 600 W |
The x-ray technique factors of ORCA are higher than those of the predicate. However, the resulting extremity radiation exposure is below the 10 R/min limit of 21 CFR 1020.32.
The ORCA is designed to meet U.S. standards 21 CFR 1020.30 through 1020.32 with a variance for source to skin distance. The ORCA is designed to meet applicable International standards; IEC 601 - 1, IEC 601 - 1-1, IEC 601 - 1-2, IEC 601 - 1-3, IEC 601 - 2-7, IEC 601 - 2-28, IEC 601 - 2-32.
The typical operator radiation exposure is quite low, at less than 0.5 mR/hr @ 50 cm for Hand and less than 2.0 mR/hr @ 50 cm for Knee. The skin entrance dose of less than 5 R/min is a factor of 2x below the 10 R/min limit of 21 CFR 1020.32.
Summary of Safety, Effectiveness Page 2 of 3
{3}
Summary of Safety, Effectiveness Page 3 of 3
Page 23
## 9.3 CONCLUSION
No new safety or effectiveness questions are raised by the ORCA device.

Gary Syring
Printed Name
Director, Regulatory Affairs
Title
{4}
HUMAN SERVICES, INC.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
NOV 17 2011
Mr. Gary Syring
Director, Regulatory Affairs
Lunar Corporation
313 W. Beltline Highway
MADISON WI 53713
Re: K960907
Trade/Device Name: Mobile Orthopedic C-Arm
Regulation Number: 21 CFR 892.1650
Regulation Name: Image intensified fluoroscopic x-ray system, mobile
Regulatory Class: II
Product Code: OXO
Dated: February 29, 1996
Received: March 5, 1996
Dear Mr. Syring:
This letter corrects our substantially equivalent letter of May 20, 1996.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please Note: CDRH does not evaluate information related to contact liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}
comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health’s (CDRH’s) Office of Compliance. Also, please note the regulation entitled, “Misbranding by reference to premarket notification” (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely Yours,
Mary S. Pastel, Sc.D.
Director
Division of Radiological Devices
Office of In Vitro Diagnostic Device
Evaluation and Safety
Center for Devices and Radiological Health
Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K960907
Device Name: ORCA
Indications For Use:
The ORCA is a small image intensified fluoroscopic C-Arm, for imaging extremities in orthopedic applications. The ORCA Operator's manual contains the following statement:
"CAUTION: Federal Law (USA) restricts this device to sale and use by or on the order of a physician."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seyman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K960907
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.