K213113 · Orthoscan, Inc. · OXO · Oct 21, 2021 · Radiology
Device Facts
Record ID
K213113
Device Name
Orthoscan Tau Mini C-Arm
Applicant
Orthoscan, Inc.
Product Code
OXO · Radiology
Decision Date
Oct 21, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Orthoscan TAU Mini C-arm is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including when imaging limbs/extremities, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
Device Story
Mobile fluoroscopic mini C-arm system; provides real-time X-ray imaging for diagnostic and surgical procedures. Inputs: X-ray radiation passing through patient anatomy; captured by flat panel detector (FPD). Transformation: X-ray signals converted to digital images via image processing board; enhanced by software (OrthoTouch application). Output: Fluoroscopic images displayed on workstation monitor for physician review. Used in OR, clinic, and emergency settings by physicians/technicians. Clinical decision-making: assists in orthopedic/surgical guidance; enables real-time visualization of anatomy/implants. Benefits: improved image clarity at equal or lower dose levels compared to predicate; supports pediatric imaging workflows.
Clinical Evidence
No clinical data; bench testing only. Performance evaluated using anthropomorphic (PMMA) and anatomical simulation phantoms. A board-certified radiologist performed a comparative assessment of image sets from the subject and predicate devices, concluding the subject device provides slight improvement in image quality at equivalent patient dose rates.
Indicated for general fluoroscopic visualization (pulsed or continuous) for patients of all ages, including pediatric populations, during diagnostic, surgical, and critical emergency care procedures. Anatomical regions include limbs, extremities, and shoulders. Used in hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 21, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Orthoscan, Inc. % Mr. Kevin Bridgman Director of Regulatory Affairs and Quality Assurance 14555 N. 82nd Street SCOTTSDALE AZ 85260
Re: K213113
Trade/Device Name: Orthoscan Tau Mini C-Arm Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OXO, JAA, MQB Dated: September 21, 2021 Received: September 27, 2021
Dear Mr. Bridgman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
. for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213113
Device Name Orthoscan TAU Mini C-arm
### Indications for Use (Describe)
The Orthoscan TAU Mini C-arm is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including when imaging limbs/extremities, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
| Type of Use (Select one or both, as applicable) | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|--|--|
| <span style="font-size:14px;"> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size:14px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K213113
Image /page/3/Picture/1 description: The image shows the logo for Orthoscan. The logo consists of a stylized letter "C" in blue, followed by a vertical blue line. To the right of the line is the word "ORTHOSCAN" in black, with each letter evenly spaced.
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center -W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
## Volume 005 Special 510(k) Summary
# Special 510 (k) Premarket Notification Submission- Orthoscan, Inc. TAU Mini C-Arm
Date: September 21, 2021
In accordance with the requirements of 21 CFR §807.92 the following Special 510(k) summary of information is provided:
| Submitter Address: | Orthoscan, Inc.<br>14555 N 82nd St.<br>Scottsdale, AZ 85260<br>Phone: (480) 503-8010<br>Fax: (480) 503-8011 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Kevin Bridgman<br>14555 N 82nd St.<br>Scottsdale, AZ 85260<br>Cell: (909) 262-9930<br>Phone: (480) 503-8010<br>Fax: (480) 503-8011 |
| Secondary Contact<br>Person: | Brenda Sparks<br>14555 N 82nd St.<br>Scottsdale, AZ 85260<br>Phone: (480) 503-8010<br>Fax: (480) 503-8011 |
{4}------------------------------------------------
# ORTHOSCAN
## Proposed Device:
Device Trade Name: 510(k) Number: Common /Usual Names:
Device Class: Classification(s): Classification Names: Device: Product Code:
## 21CFR 807.92(a)(2)
21 CFR 807.92(a)(3)
Orthoscan, Inc. TAU Mini C-Arm Unknown at this time Fluoroscopic X-Ray System, Mobile Mini Mobile C-arm, Mini C-arm Class II 21CFR 892.1650 image-intensified fluoroscopic x-ray system, mobile Image-intensified fluoroscopic x-ray system. OXO, JAA, MQB
## Predicate Device:
Device Identification: Classification(s): Device Class: Classification Names: Regulation Description: Product Codes:
General Description:
Orthoscan, Inc. TAU Mini C-Arm (K183220) 21CFR 892.1650 Class II Image-intensified fluoroscopic x-ray system, mobile Image-intensified fluoroscopic x-ray system OXO, JAA, MQB
The proposed modifications to Orthoscan, Inc. TAU Mini C-Arm series (which we will refer to internally and in this submittal as Orthoscan TAU 2.0, for distinction from predicate Orthoscan TAU) retain identical function as the predicate TAU Mini C-arm (K183220) as a mobile fluoroscopic mini C-arm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment and surgical procedures involving anatomical regions such as but not limited to that of extremities, limbs, shoulders, knees, and Hips. The system consists of C-arm support attached to the image workstation.
The changes to the Orthoscan TAU series of Mini C-arm X-ray systems represent a modification of our presently legally marketed device Orthoscan TAU mini C-Arm K183220. The proposed modifications to the predicate encompass the implementation of an optional IGZO 15 cm x 15 cm Flat Panel Detector (FPD) in the 15x12cm and 15x15cm device detector sizes, a new LINUX based operating system and related software, image processing board revisions and a revised Power Manager Board for AC to DC conversion that will distribute 24Vdc via a medical grade DC power supply. The proposed device incorporates software architecture and other improvements that
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Orthoscan. The logo consists of a stylized letter "C" in blue, followed by a vertical blue line. To the right of the line is the word "ORTHOSCAN" in black sans-serif font.
replicate the features and functions of the predicate device and improve image clarity without increasing dose levels.
For both the predicate and proposed device, the following are unchanged; Identical C-arm support and mechanical connections, balancing, locking, rotations, work-station platform, monitor display and main user interface controls, touch screen interface, selectable imaging, X-ray technique control, entry of patient information, wired or wireless footswitch operation, interface connection panel and DICOM fixed wire and wireless network interfaces.
The proposed modifications to Orthoscan, Inc. TAU Mini C-Arm series do not change the identical intended use. Intended use is the same as the predicate device TAU (K183220) Mini C-Arm.
Indications for Use: The Orthoscan TAU Mini C-arm is designed to provide physicians with general fluoroscopic visualization, usinq pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders; at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
> The proposed modifications to Orthoscan, Inc. TAU Mini C-Arm series do not change the indications for use. The indications for use are the same as the predicate device TAU (K183220) Mini C-Arm.
Technology: The changes to proposed device Orthoscan TAU 2.0 series of Mini C-arm X-ray systems represent a modification of our presently legally marketed device Orthoscan TAU Mini C-Arm K183220. The proposed device will provide an option in supplier of detectors. The optional IGZO Solid-State X-ray Imagers (SSXI) provide newer technology in the Flat Panel Detectors (FPD). The inclusion of IGZO detector size 15cmx15cm in the 15cmx15cm sized Orthoscan device will also be fitted into the smaller active field
Intended Use:
{6}------------------------------------------------
model 15cmx12cm sized device with minor mechanical modifications.
The proposed modified TAU 2.0 will be equipped with a revised Power Manager Board for AC to DC conversion that will distribute 24vdc via a new medical grade DC power supply. This supply will power all electronic sub-assemblies. The AC to DC conversion will provide intrinsic value through risk reduction such as leakage, while standardizing distribution of 24Vdc.
With the introduction of a Linux based operating system (OS) an open source highly secure OS, and revisions to the image processing board, software architecture design has been improved and functions are identical to that of the predicate device Orthoscan, Inc. TAU Mini C-arm (K183220).
Summary of Technological Characteristics: The comparisons of the proposed modifications to Orthoscan, Inc. TAU Mini C-Arm demonstrate that the scientific and technology characteristics indicate substantial equivalence to the predicate device Orthoscan, Inc. TAU Mini C-Arm (K183220). The following table provides a comparison of the technology characteristics.
| Differences<br>Features/Technology: | Modified Device Orthoscan,<br>Inc. TAU 2.0 Mini C-Arms<br>model#/sizes, serial<br>number prefix:<br>1000-0015, 15x12cm,<br>Serial No. sequence<br>5Txxx<br>1000-0016, 15x15cm,<br>Serial No. sequence<br>5Uxxx<br>1000-0017, 20x20cm,<br>Serial No. sequence<br>5Vxxx | Predicate Device Orthoscan,<br>Inc. TAU Mini C-Arms<br>model#/sizes, and serial<br>number prefix;<br>1000-0015, 15x12cm,<br>Serial No. sequence<br>5Nxxx<br>1000-0016, 15x15cm,<br>Serial No. sequence<br>5Pxxx<br>1000-0017, 20x20cm,<br>Serial No. sequence<br>5Rxxx<br>(K183220) | Comparison to Predicate,<br>Comments to Differences |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Codes | | | |
| Device Classification Name | image-intensified<br>fluoroscopic x-ray system,<br>mobile | image-intensified<br>fluoroscopic x-ray system,<br>mobile | Identical |
| Regulation Description | Image-intensified<br>fluoroscopic x-ray system. | Image-intensified<br>fluoroscopic x-ray system. | Identical |
| Classification Product Code | OXO | OXO | Identical |
| Subsequent Product Code | JAA | JAA | Identical |
| Regulation Number | 892.165 | 892.165 | Identical |
| Device Class | II | II | Identical |
| Non-Contact Device | Non-Contact | Non-Contact | Identical |
| 510(k) Panel Review | Radiology | Radiology | Identical |
| 510(K) Number | New Proposed Device -<br>unknown at this time | K183220 | New Proposed Device -<br>unknown at this time |
| Detector Specifications | | | |
| CMOS and IGZO Flat Panel<br>Detector/Image Receptor | CsI(T1)/ solid state X-ray<br>detector<br>20x20cm remains CMOS<br>same as predicate.<br>15x12cm and 15x15cm<br>will be CMOS or optional<br>IGZO | CsI(T1)/ solid state X-ray<br>detector<br>20x20cm, 15x15cm and<br>15x12cm are CMOS | Substantially Equivalent.<br>The introduction of the<br>optional IGZO technology<br>was found to be equal in<br>safety and effectiveness<br>including image quality<br>(Essential Performance).<br>IGZO sensor technology<br>demonstrates equal/<br>better image quality to<br>that of the predicate. The<br>IGZO detector features<br>the same scintillator<br>technology and equal or<br>improved active area,<br>pixel pitch, frame rate, X-<br>Ray energy, power<br>specifications and<br>mechanical dimensions,<br>and also has identical<br>image processing features<br>as the predicate.<br>IGZO based detectors are<br>manufactured using<br>processes that are similar<br>to CMOS detectors. The<br>slight differences in used<br>sensor glass technology<br>does not have influence<br>on safety & effectiveness<br>of the product and<br>provides slightly improved<br>image quality at equal<br>dose values as the<br>predicate. |
| Detector Resolution | TAU 2.0 2020 =2.0k x<br>2.2k<br>TAU 2.0 1515 =1.5 k x<br>1.5k<br>TAU 2.0 1512 =2.0 k x<br>1.5k | TAU 2020 =2.0k x 2.2 k<br>TAU 1515 =1.5k x 1.5k<br>TAU 1512 =2.0 x 1.5k | Identical |
| Field of View (Full) | TAU 2.0 2020 = 8" x 8"<br>TAU 2.0 1515 = 5.5" x<br>5.5"<br>TAU 2.0 1512 = 5.5" x<br>4.3" | TAU 2020 = 8" x 8"<br>TAU 1515 = 5.5" x 5.5"<br>TAU 1512 = 5.5" x 4.3" | Substantially Equivalent.<br>The 15x12cm model will<br>be mechanically modified<br>to house the 15x15cm<br>IGZO Flat Detector for a<br>useful array area<br>15x12cm. The difference<br>does not affect the safety<br>or efficacy of the device |
| Field of View (Collimated Mag<br>Mode) | TAU 2.0 2020 = 4" x 4"<br>TAU 2.0 1515 = 4.3" x<br>4.3"<br>TAU 2.0 1512 = 4.3" x<br>3.3" | TAU 2020 = 4" x 4"<br>TAU 1515 = 4.3" x 4.3"<br>TAU 1512 = 4.3" x 3.3" | Identical to predicate |
| Detector Size | TAU 2.0 2020 = 20 x<br>20cm<br>TAU 2.0 1515 = 15 x<br>15cm<br>TAU 2.0 1512 = 15 x<br>15cm | TAU 2020 = 20 x 20 cm<br>TAU 1515 = 15 x 15 cm<br>TAU 1512 = 15 x 12 cm | Substantially Equivalent.<br>The TAU 2.0 (15x12cm)<br>model will be<br>mechanically modified to<br>house the 15x15cm IGZO<br>Flat Detector for a useful<br>array area 15x12cm. The<br>difference does not affect<br>the safety or efficacy of<br>the device. |
| Useful Array | TAU 2.0 2020 = 20 x<br>20cm<br>TAU 2.0 1515 = 15 x<br>15cm<br>TAU 2.0 1512 = 15 x<br>15cm | TAU 2020 = 20 x 20 cm<br>TAU 1515 = 15 x 15 cm<br>TAU 1512 = 15 x 12 cm | Substantially Equivalent.<br>The TAU 2.0 (15x12cm)<br>model will be<br>mechanically modified to<br>house the 15x15cm IGZO<br>Flat Detector for a useful<br>array area 15x12cm. The<br>difference does not affect<br>the safety or efficacy of<br>the device. |
| Pixel Spacing | TAU 2.0 2020 = 99<br>microns<br>TAU 2.0 1515 = same for<br>CMOS or 100microns for<br>IGZO detector<br>TAU 2.0 1512 = same for<br>CMOS or 100microns for<br>IGZO detector | TAU 2020 = 99 microns<br>TAU 1515 = 100 microns<br>TAU 1512 = 75 microns | Substantially Equivalent.<br>The difference for the<br>TAU 2.0 15x12 cm IGZO<br>detector pixel spacing<br>does not affect the<br>safety or efficacy of the<br>device. |
| Dynamic Range | TAU 2.0 same for CMOS<br>or >69dB for IGZO | TAU 2020 = 71 dB<br>TAU 1515 = 71 dB<br>TAU 1512 = 70 dB | Substantially Equivalent<br>The difference in dynamic<br>range for the IGZO<br>detector does not affect<br>the safety or efficacy of<br>the device. |
| DQE | TAU 2.0 2020 = 70%<br>TAU 2.0 1515 = same for<br>CMOS or 65% for IGZO<br>TAU 2.0 1512 = same for<br>CMOS or 65% for IGZO | TAU 2020 = 70%<br>TAU 1515 = 70%<br>TAU 1512 = 70% | Substantially Equivalent.<br>The difference in DQE for<br>the IGZO detector does<br>not affect the safety or<br>efficacy of the device. |
| A/D Conversion | 16 bit | 16 bit | Identical |
| Image Processing Features | | | |
| Startup time | 30 sec | 30 sec | Identical |
| Cine Loop Export | Yes | Yes | Identical |
| Fluoroscopy Frame Rate | 30/15/7.5/2 fps | 30/15/7.5/2 fps | Identical |
| Edge Enhancement | Yes | Yes | Identical |
| Post Process<br>Brightness/Contrast | Yes | Yes | Identical |
| Adaptive Noise Suppression | Automatic | Automatic | Identical |
| Manual Noise Suppression | 3 Modes | 3 Modes | Identical |
| AERC Automatic X-Ray<br>Technique Control | YES | YES | Identical |
| Adaptive Noise Filter | Noise reduction | Noise reduction | Identical |
| Save and Auto Store | YES | YES | Identical |
| Last image hold | YES | YES | Identical |
| Edge Enhancement | YES | YES | Identical |
| Cine Loop Frame Rate | 30 fps | 30 fps | Identical |
| Snapshot Capabilities | YES | YES | Identical |
| Post Processing<br>Brightness/Contrast Control | YES | YES | Identical |
| Image invert | YES | YES | Identical |
| Image Zoom | YES | YES | Identical |
| Manual Noise Suppression | 4 modes | 4 modes | Identical |
| Image Documentation: | | | |
| Wireless Communication (Wi-<br>Fi)/(WLAN) | Capable IEEE 802.11 | Capable IEEE 802.11 | Identical…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.