K171800 · Ge Hualun Medical Systems Co. , Ltd. · OXO · Jul 12, 2017 · Radiology
Device Facts
Record ID
K171800
Device Name
OEC Elite MiniView
Applicant
Ge Hualun Medical Systems Co. , Ltd.
Product Code
OXO · Radiology
Decision Date
Jul 12, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OEC Elite™ MiniView™ (mobile mini C-arm) is intended to provide fluoroscopic images of the patient during diagnostic or therapeutic treatment/surgical procedures of the limbs/extremities and shoulders.
Device Story
Mobile mini C-arm system for real-time fluoroscopic imaging of limbs, extremities, and shoulders. System comprises C-arm with CsI(Tl)-CMOS flat panel detector, X-ray source monoblock, and mobile workstation with image processing software and display monitors. C-arm supports vertical, horizontal, orbital, and rotational positioning; features electronically controlled braking system. Operated by physicians/surgeons in clinical settings (e.g., OR, emergency medicine). Provides real-time visualization to guide diagnostic or therapeutic procedures. Subject device adds optional 2.4GHz wireless footswitch for X-ray control, maintaining identical functionality to wired version. Wireless option reduces cable clutter and improves placement flexibility.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including design verification and validation, and compliance with IEC 60601-1, IEC 60601-2-54, and IEC 60601-2-43 standards. Wireless coexistence testing performed per FDA guidance.
Technological Characteristics
Mobile mini C-arm; CsI(Tl)-CMOS flat panel detector; high-voltage generator; X-ray monoblock. Optional 2.4GHz wireless footswitch using proprietary private protocols. Compliant with IEC 60601-1 Ed. 3, IEC 60601-2-54, and IEC 60601-2-43. Electronically controlled braking system.
Indications for Use
Indicated for real-time general fluoroscopic visualization of patients of all ages to aid physicians and surgeons during diagnostic or therapeutic treatment/surgical procedures of the limbs, extremities, and shoulders, including orthopedics and emergency medicine.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 12, 2017
GE HUALUN MEDICAL SYSTEMS CO. LTD. % Ms. Lifeng Wang Regulatory Affairs Leader No1 Yong Chang North Road Beijing Economic Technological Development Zone Beijing 100176 Beijing CHINA
Re: K171800
Trade/Device Name: OEC Elite™ MiniView™ Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO, JAA Dated: June 15, 2017 Received: June 16, 2017
Dear Ms. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K171800
Device Name OEC Elite™ MiniView™
#### Indications for Use (Describe)
The OEC Elite MiniView (mobile mini C-Arm) is designed to provide physicians with real time general fluoroscopic visualization of patients of all ages. It is intended to aid physicians and surgeons during diagnostic or therapeutic treatment/surgical procedures of the limbs/extremities and shoulders including, but not limited to, orthopedics and emergency medicine.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/7 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. The circle has small, wave-like embellishments around its perimeter, giving it a slightly decorative appearance. The logo is presented in a blue color.
## GE Healthcare
510(k) Premarket Notification Submission- OEC Elite™ MiniView™
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the requirements of 21 CFR 807.92 the following summary of information is provided:
| Date:<br>Submitter: | June 16, 2017<br>GE HUALUN MEDICAL SYSTEMS CO. Ltd.<br>No.1 Yong Chang North Road, Beijing Economic<br>Technological Development Zone<br>Beijing 100176,China |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/<br>Manufacturing Location | GE HUALUN MEDICAL SYSTEMS CO. Ltd.<br>No.1 Yong Chang North Road, Beijing Economic<br>Technological Development Zone<br>Beijing 100176,China |
| Primary Contact Person: | Lifeng Wang<br>Regulatory Affairs Manager<br>GE Healthcare<br>+86 10 57083145<br>Lifeng.Wang@ge.com |
| Secondary Contact Person: | Helen Peng<br>Regulatory Affairs Director<br>GE Healthcare<br>+1 262 424-8222<br>Hong.Peng@ge.com |
| Device Trade Name: | OEC Elite™ MiniView™ |
| Common/Usual Name: | Fluoroscopic X-Ray System, Mobile<br>Mini Mobile C-Arm, Mini C-Arm |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Device Class | Class II |
| Classification regulation: | 21CFR 892.1650 |
| Primary Product Code: | OXO |
| Secondary Product Code: | JAA |
| Predicate Device(s): | K160131, OEC Elite MiniView |
| Device Description: | The OEC Elite™ MiniView™ is a mobile fluoroscopic mini C-<br>arm system that provides fluoroscopic images of patients of all<br>ages during diagnostic, treatment, and surgical procedures of the<br>shoulders, limbs, and extremities. The system consists of a C-arm<br>attached to an image processing workstation. A CsI(TI) -CMOS<br>flat panel detector and the identical X-ray source monoblock are<br>used for image acquisition. |
| | The C-arm supports the high-voltage generator, X-ray tube, X-ray<br>controls, collimator, and the FPD. The C-arm is capable of<br>performing linear (vertical, horizontal, orbital) and rotational<br>motions that allow the user to position the X-Ray imaging<br>components at various angles and distances with respect to the<br>patient extremity anatomy to be imaged. The C and support arm<br>are mechanically balanced allowing for ease of movement and<br>capable of being "locked" in place using an electronically<br>controlled braking system. The workstation is a stable mobile<br>platform that supports the C-arm, image display monitor(s),<br>image processing equipment/software, recording devices, data<br>input/output devices and power control systems. |
| | The primary technology change for the subject device compared<br>to the unmodified predicate OEC Elite MiniView, was to add an<br>optional wireless footswitch which provides identical<br>functionalities as the wired footswitch to control X-ray on and<br>off. |
| Intended Use: | The OEC Elite™ MiniView™ (mobile mini C-arm) is intended to<br>provide fluoroscopic images of the patient during diagnostic or<br>therapeutic treatment/surgical procedures of the limbs/extremities<br>and shoulders. |
| Indications for Use | The OEC Elite™ MiniView™ (mobile mini C-Arm) is designed<br>to provide physicians with real time general fluoroscopic<br>visualization of patients of all ages. It is intended to aid<br>physicians and surgeons during diagnostic or therapeutic<br>treatment/surgical procedures of the limbs/extremities and<br>shoulders including, but not limited to, orthopedics and<br>emergency medicine. |
| Technology: | The modified OEC Elite™ MiniView™ employs the same<br>fundamental scientific technology as the unmodified predicate<br>device. The proposed device will provide an optional off-the-shelf<br>wireless foot switch uses 2.4GHz radio frequency technology<br>with proprietary private protocols. The introduction of this<br>wireless footswitch does not change any existing hardware or<br>software of the predicated device. The new option is equivalent to<br>the wired footswitch on the predicate in that the functionalities<br>are identical. The wireless option provides placement flexibility<br>and reduces cable clutter. |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Testing: |
| | The modified OEC Elite™ MiniView™ system with the wireless<br>footswitch was developed under the GE Healthcare's design<br>controls processes and overall quality management system. |
| | The system has been tested by an NRTL and certified compliant<br>with the IEC 60601-1 Ed. 3 series, including IEC60601-2-54 and<br>IEC 60601-2-43. All applicable 21CFR Subchapter J<br>performance standards are met. |
| | Risk management activities including using risk analysis to<br>identify any potential issues incorporating the wireless footswitch<br>were performed. These issues were reviewed and mitigated with<br>inherent safe designs and labeling. The mitigations were verified<br>and validated as a part of the design verification and validation<br>testing that has been executed with acceptable results. This<br>testing also includes the EMC and coexistence testing per the<br>"Radio Frequency Wireless Technology in Medical Devices"<br>Guidance for Industry and Food and Drug Administration Staff<br>issued on August 14, 2013 (Wireless Guidance). |
| | Clinical Testing:<br>The subject of this premarket submission, OEC Elite MiniView<br>with the addition of the wireless footswitch, did not require<br>clinical studies to support substantial equivalence. |
| | Design verification and validation testing was performed to<br>confirm that the safety and effectiveness of the devices has not<br>been affected. The test plans and results have been executed with<br>acceptable results. |
| Conclusion: | The OEC Elite MiniView incorporates a wireless footswitch for<br>the user to initiate and terminate x-ray exposures as an option.<br>The addition of the optional wireless footswitch to the system<br>does not raise new questions on safety and effectiveness. The<br>modified subject device has the same technological characteristics<br>and perform as well as the unmodified predicate device and other<br>similar devices that incorporate a wireless footswitch currently on<br>the market. |
| | Based on the successful verification and validation testing,<br>additional engineering bench testing per the FDA Wireless<br>guidance, conformance to standards, and development under GE's<br>Quality Management System, we believe that the modified OEC<br>MiniView is of comparable type and substantially equivalent to<br>the predicate device OEC Elite MiniView (K160131), and<br>therefore is safe and effective for its intended use. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by swirling, water-like shapes, also in blue, which add a dynamic and fluid element to the design. The overall design is simple, clean, and recognizable, reflecting GE's long-standing presence and reputation.
# GE Healthcare
# 510(k) Premarket Notification Submission- OEC Elite™ MiniView™
{5}------------------------------------------------
# GE Healthcare
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are decorative flourishes around the letters, giving the logo a classic and recognizable appearance.
# 510(k) Premarket Notification Submission- OEC Elite™ MiniView™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.