INSIGHT - FD MINI C-ARM FLUOROSCOPIC IMAGING SYSTEM
Applicant
Hologic, Inc.
Product Code
OXO · Radiology
Decision Date
Apr 6, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide the physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.
Device Story
Portable Mini C-Arm fluoroscopic imaging system; utilizes CMOS flat detector technology for rectangular imaging; replaces conventional circular image intensifier. Operated by medical personnel in diagnostic, surgical, and post-operative settings for extremity imaging (hand, wrist, forearm, shoulder, foot, ankle, knee). Hardware includes X-ray source assembly with manual collimator rotation (+/-90 degrees). Software synchronizes collimator rotation with display orientation to maintain rectangular view. Provides real-time fluoroscopic visualization to physicians; assists in surgical guidance and diagnostic assessment; enables improved visualization of anatomy compared to circular intensifiers.
Clinical Evidence
Bench testing only. Performance characterized via non-clinical imaging studies comparing the new flat panel detector system to the prior image-intensified system. Metrics included resolution, contrast detail, uniformity, and motion. Compliance with 21 CFR 1020.30 and 1020.32 radiation protection standards verified.
Technological Characteristics
Portable Mini C-Arm; CMOS flat panel detector; X-ray source with manual collimation; +/-90 degree collimator rotation synchronization; rectangular field of view; complies with 21 CFR 1020.30 and 1020.32.
Indications for Use
Indicated for patients requiring general fluoroscopic visualization during surgical orthopedic, podiatry, critical care, emergency care, and light anatomy imaging procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System (K051025)
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 5 -510(k) Summary
K120388
APR - 6 2012
### Manufacturing Establishment and Contact Information 1
- 1.1 Manufacturer Name and Address:
Hologic, Inc. 35 Crosby Drive Bedford, MA 01730
#### Establishment Registration Number: 1.2
1221300
#### Name, Title, and Telephone Number of Contact: 1.3
Name: Eileen M. Boyle Regulatory Affairs Specialist II Title: (781) 999-7781 Phone: (866) 652-8674 Fax : Email: eileen.boyle@hologic.com
#### 2 Device Identification
#### 2.1 Device Trade Name:
Hologic® InSight - FD Mini C-Arm Fluoroscopic Imaging System
#### Common/Usual Name 2.2
Fluoroscopic Imaging System
#### 2.3 Proposed Intended Use:
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide the physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.
{1}------------------------------------------------
#### 3 Device Classification
3.1 Classification:
Class II
#### Classification Name and Rule 3.2
Image-Intensified Fluoroscopic X-Ray System, 21 CFR 892.1650
- Classification Panel 3.3
Radiology
#### 3.4 Product Code
охо
#### 3.5 Predicate Devices
- 510(k) No: K051754 ● Trade Name: Orthoscan, OrthoScan HD (which includes OrthoScan HD with Flat Detector) SE Date: 8/9/2005 Manufacturer: Orthoscan, Inc.
- 510(k) No: K113708 . Trade Name: Orthoscan Mobile DI SE Date: 1/5/2012 Manufacturer: Orthoscan, Inc.
- 510(k) No: K051025 . Trade Name: Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System SE Date: 5/13/2005 Manufacturer: Hologic, Inc.
Page 2 of 4
{2}------------------------------------------------
#### 4.0 Device Description
The Hologic® InSight-FD System , is a portable Mini C-Arm fluoroscopic imaging system with Flat Detector technology used by medical personnel in a variety of diagnostic, surgical and post-operative procedures for imaging extremities, including the hand, wrist, forearm, shoulder, foot, ankle and knee.
The hardware enhancements for the InSight-FD System exist primarily within the detector assembly and collimation. The InSight-FD System offers CMOS Flat Detector technology which provides a rectangular viewing area in comparison t the circular view with the conventional image intensifier. For user convenience, additional buttons were added to the X-ray source assembly control panel from the available functions.
The software enhancements for the InSight-FD System synchronize the rotation of the X-ray collimator with a manual rotation of +/-90 degrees. The synchronization enables the user to rotate the display image in a rectangular image as compared to the circular view of the image intensifier.
#### 5 Performance Testing
Software verification and validation test was performed following the recommendations in the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The purpose of the testing was to verify the incorporation and functionality of the flat panel detector on the Insight-FD System, that the accompanying software changes did not negatively impact the functionality, and that the functions and features tested meet their specific acceptance criteria.
Nonclinical and clinical information was provided following the recommendations in the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices." This included information to characterize the solid state detector. Pairs of images were acquired to test the radiographic and fluoroscopic parameters {e.g., resolution, contrast detail, uniformity and motion) of the new Insight-FD with flat panel detector in comparison with the prior image intensified system.
Testing was also performed to demonstrate that the new Insight-FD with flat panel detector continues to comply with the applicable regulatory performance standards for radiation protection, including: 21 CFR 1020.30 (Diagnostic x-ray systems and their major components) and 21 CFR 1020.32 (Fluoroscopic equipment).
Testing was successfully conducted and demonstrated that the Hologic InSight-FD System meets all of its functional requirements and performance specifications.
{3}------------------------------------------------
### Substantial Equivalence
б
7
The InSight-FD System is substantially equivalent to commercially available devices used for fluoroscopic imaging. The predicate devices selected for comparison are the OrthoScan- HD and the OrthoScan–HD with Flat Detector (K051754), OrthoScan Mobile DI (K113708) and the Fluoroscan InSight Mini C-Arm Fluoroscopic Imaging System (K051025). The predicate devices provide substantially equivalent or the same intended use, features and functions as the proposed device.
### Conclusion
The features and functions on the InSight-FD System are substantially equivalent to those of the indicated commercially distributed devices: OrthoScan-HD with Flat Detector (K051754), OrthoScan Mobile DI (K113708) and Insight (K051025) with regard to intended use, performance, safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Eileen M. Boyle Regulatory Affairs Specialist II Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730
2012
Re: K120388
Trade/Device Name: InSight-FD Mini C-Arm Fluoroscopic Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO Dated: March 12, 2012 Received: March 22, 2012
### Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device · Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE FORM
510(k) Number (if known):
K120388
### Device Name: InSight - FD Mini C-Arm Fluoroscopic Imaging System
Indications for Use:
The Fluoroscan InSight is a Mini C-Arm Fluoroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including, but not limited to surgical orthopedic and podiatry use, critical and emergency care procedures, and light anatomy imaging situations.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device
Evaluation and safety
510(k) K120388
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.