← Product Code [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB) · K250241

# Cios Select (K250241)

_Siemens Medical Solutions USA, Inc. · OWB · Nov 4, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K250241

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB.md)
- **Decision Date:** Nov 4, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care, and in Veterinary Medicine. The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of an intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.

## Device Story

Mobile X-ray system for surgical environments; consists of mobile acquisition unit (C-arm, X-ray generator, flat panel detector) and monitor cart (TFT displays, digital image processing). Inputs: X-ray radiation; Outputs: digital fluoroscopic/radiographic images. Operated by healthcare professionals in OR, trauma, endoscopy, ICU, and ambulatory settings. System provides real-time visualization for intraoperative procedures; aids clinical decision-making by enabling precise positioning of implants, catheters, and probes. Benefits include improved procedural accuracy and low-dose imaging options for pediatric patients.

## Clinical Evidence

No clinical data. Bench testing only. Performance evaluated via unit, subsystem, and system integration testing; electrical safety (IEC 60601-1); EMC (IEC 60601-1-2); and imaging performance (dose assessment/image quality). Human factors/usability validation performed.

## Technological Characteristics

Mobile C-arm X-ray system. Components: floor stand, C-arm, high-frequency generator, X-ray tube, flat panel detector (PaxScan2121DX, Mercu0909F), collimator, TFT displays. Connectivity: USB storage. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-43/54 (X-ray performance), ISO 14971 (risk management). Software: VA21(F) version.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Cios Select (VA21) Flat Panel ([K223410](/device/K223410.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 4, 2025

Siemens Medical Solutions USA, Inc
% Patricia D. Jones
Regulatory Affairs Professional
40 Liberty Boulevard
MALVERN PA 19355

Re: K250241
Trade/Device Name: Cios Select
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-Intensified Fluoroscopic X-Ray System
Regulatory Class: Class II
Product Code: OWB, OXO, JAA
Dated: September 30, 2025
Received: September 30, 2025

Dear Patricia D. Jones:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250241 - Patricia D. Jones
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250241 - Patricia D. Jones
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250241  |   |
|  Device Name Cios Select  |   |
|  Indications for Use (Describe) The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care, and in Veterinary Medicine. The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of an intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

# 510(k) Summary: Cios Select

K250241

Company: Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard, 65-1A
Malvern, PA 19355

Date Prepared: January 22, 2025
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

# 1. General Information:

Importer / Distributor:
Siemens Medical Systems USA, Inc.
40 Liberty Boulevard, 65-1A
Malvern, PA 19355
Establishment Registration Number: 2240869

Manufacturing Site:
Siemens Shanghai Medical Equipment Ltd.
278 Zhou Zhu Road, Shanghai
201318, China
Establishment Registration Number: 3003202425

Manufacturing Site:
Siemens Healthineers AG.
ROENTGENSTRASSE 19-21
95478 KEMNATH, Germany
Establishment Registration Number: 3002466018

# 2. Contact Person:

Patricia D. Jones
Regulatory Affairs Specialist
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Phone: (678) 575-8832
Email: patricia.jones@siemens-healthineers.com

CONFIDENTIAL

{5}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

## 3. Device Name and Classification:

Trade Name: Cios Select
Classification Name: Image-Intensified Fluoroscopic x-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Codes: OWB, OXO, JAA

## 4. Legally Marketed Predicate Device

Trade Name: Cios Select
510(k) Clearance: K223410
Clearance Date: December 7, 2022
Classification Name: Image-intensified fluoroscopic x-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1650
Device Class: Class II
Product Code: OWB
Subsequent Product Codes: JAA, OXO
Total Product Life Cycle: All products Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any applicable issues.

## 5. Device Description:

The Cios Select (VA21F) Mobile X-ray System is designed for the surgical environment. The Cios Select (VA21F) is a modification of the Cios Select (VA21) Flat Panel originally cleared under Premarket Notification K223410 on December 7, 2022.

The Cios Select consists of two major units:

The Siemens Healthineers Cios Select mobile fluoroscopy C-arm system is an X-ray imaging system consisting of two mobile units: a mobile acquisition unit and a monitor cart as the image display station.

The mobile acquisition unit is comprised of the X-ray control, the C-arm which supports the single-tank high-frequency generator/X-ray tube assembly, the flat panel detector, and user controls.

The monitor cart connects to the acquisition unit by a cable. It integrates the TFT flat panel displays, Digital Imaging Processing System, user controls and image storage devices (USB).

Page 2 of 8

{6}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

The following modifications were made to the predicate device the Cios Select (VA21) Flat Panel mobile X-ray System cleared under Premarket Notification K223410 on December 7, 2022. Siemens Medical Solutions USA, Inc. submits this Special 510(k) to request clearance for the Subject Device the Cios Select (VA21F) system in comparison to the Predicate Device.

1. Updated Software from VA21(E) to VA21(F) to support the following hardware changes.
a. Add one new Detector
b. Add one new Grid
2. Updated 510(k) Information

## 6. Indications for Use:

The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine.

The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of an intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.

## 7. Substantial Equivalence:

The Cios Select (VA21F) system is substantial equivalent to the legally marketed predicates listed in Table 1 below:

Table 1: Predicate Device Comparable Properties for Subject Device Modifications:

|  Predicate Device Name and Manufacturer | 510(k) Number | Clearance Date | Comparable Properties  |
| --- | --- | --- | --- |
|  Predicate Device | K223410 | 12/7/2022 | • Indications for use
• X-ray technology
• Image processing
• Mechanical design
• Cybersecurity
• Software
• Collimator
• PC Hardware
• Detector  |
|  Cios Select (VA21) Flat Panel
Siemens.  |   |   |   |

Page 3 of 8

{7}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

# 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:

The Cios Select (VA21F) System is designed as a set of components (floor stand, C-arm, X-ray tube and housing, flat panel detector, digital imaging system, collimator, generator etc.) that is combined to provide a specialized angiography system. Components used with Cios Select (VA21F) System are either commercially available with current Siemens systems or include modifications to existing components. Technological differences between the Subject Device and the Predicate Device is provided in the Table 2 below for all modifications.

Table 2: Summary of Comparison of Technological Characteristics

|  Modifications | Subject Device
Cios Select (VA21F) | Predicate Device
Cios Select (VA21)
Flat Panel
K223410 | Comparable Properties  |
| --- | --- | --- | --- |
|  1. Updated Cios Select system software from VA21(E) to VA21(F) to support the following hardware changes, with no new software features added.  |   |   |   |
|  a. Detector | Detector
PaxScan2121DX (11020805)
Mercu0909F (11514190)
K200218 | Detector
PaxScan2121DX (11020805) | The new detector has the same functionality as cleared in the Predicate Device Cios Select (VA21) Flat Panel K223410 and does not raise any new or effectiveness issue.  |
|  b. Grid | Grid | Grid | The new grid has the similar functionality as cleared in the Predicate Device Cios Select (VA21) Flat Panel K223410 and does not raise any new or effectiveness issue.  |

# 9. Nonclinical Performance Testing:

Non-clinical tests were conducted for the Cios Select (VA21F) with Flat Panel Detector during product development. The Siemens Cios Select (VA21F) with Flat Panel Detector has been tested to meet the requirements for conformity to multiple industry standards. Performance testing confirmed, that the Siemens Cios Select (VA21F) system complies with the following 21 CFR Federal Performance Standards.

Code of Federal Regulations Title 21 Subchapter J- Radiological Health, applicable sections include:

- 1020.30 Diagnostic X-Ray Systems and their major components

Page 4 of 8

{8}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

- 1020.32 Fluoroscopic Equipment
- 1040.10 Laser products

The Cios Select (VA 21F) system was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility:

- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
- IEC 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION
- IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION
- IEC 60825-1:2014
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION
- IEC 60601-2-28 Edition 3.0 2017-06
- IEC 60601-2-43 Edition 3.0 2022-12
- IEC 60601-2-54 Edition 2.0 2022-09
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION
- ISO 14971:2019

Provided in Table 3 is a list of relevant Guidance Documents that were used in the development of the 510(k) content.

Table 4: FDA Guidance Documents

|  FDA Guidance Document and Effective Date  |   |
| --- | --- |
|  1. | Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket Notification Submissions 510(k)
Document issued on October 2, 2017  |
|  2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policy for 510(k)s
Document issued on September 13, 2019  |
|  3. | Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s
Document issued on September 13, 2019  |
|  4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a change to an existing device.
Document issued on October 25, 2017  |
|  5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]
Document Issued on July 28, 2014  |
|  6. | Guidance for Industry and FDA Staff: Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices
Document issued on September 1, 2016  |
|  7. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in Medical Devices
Document issued on September 27, 2019  |
|  8. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submission for Device Software Functions  |

{9}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

|   | Document issued on June 14, 2023  |
| --- | --- |
|  9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability Engineering to Medical Devices.
Document issued February 3, 2016  |
|  10. | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging Device Premarket Notifications.
Document issued on November 28, 2017  |
|  11. | Guidance for Industry and FDA Staff: Cybersecurity in Medical Devices:Quality System Considerations and Content of Premarket Submission
Document issued on September 27, 2023  |
|  12. | Guidance for Industry and FDA Staff: Appropriate Use of Voluntary Consensus Standards in Premarket Submission for Medical Devices
Document issued on September 14, 2018  |
|  13. | Guidance for Industry and FDA Staff: Medical Device Accessories - Describing Accessories and Classification Pathways
Document issued on December 20, 2017  |
|  14. | Guidance for Industry and FDA Staff: Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions
Document issued on December 20, 2019  |
|  15. | Guidance for Industry and FDA Staff: Electronic Submission Template for Medical Device 510(k) Submission
Document issued on October 2, 2023  |
|  16. | Guidance for Industry and FDA Staff: Electromagnetic Compatibility (EMC) of Medical
Document issued on June 6, 2022  |
|  17. | Guidance for Industry and FDA Staff: Radio Frequency Wireless Technology in Medical Devices
Document issued on August 14, 2007  |

The modifications described in this Premarket Notification are supported with verification and validation testing.

## Verification and Validation:

Software Documentation for a Basic Documentation Level of Concern software per FDA's Guidance Document "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the Subject Device Cios Select VA21(F) during product development.

The Risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.

Page 6 of 8

{10}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

Bench testing in the form of Unit, Subsystem and System Integration testing was performed to evaluate the performance and functionality of the new features, hardware, and software updates. All testable requirements in the Engineering Requirements Specifications keys, Subsystem Requirements Specifications keys, and the Risk Management Hazard keys have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process. The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.

Electrical safety and EMC testing were conducted on the Cios Select, consisting of the acquisition unit (C-arm system) and the image processing and display station. The system complies with the IEC 60601-1, IEC 60601-2-43 and IEC 60601-2-54 standards for safety and the IEC 60601-1-2 standard for EMC.

The Cios Select VA21(F) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment.

Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. The required cybersecurity information is provided.

Additional engineering bench testing was performed including: the non-clinical testing identified in the guidance for submission of 510(k)s for Solid State X-Ray Imaging Devices (SSXI); demonstration of system performance; and an imaging performance evaluation - X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards was performed with acceptable results.

## Summary:

Performance tests were conducted to test the functionality of the Cios Select (VA21F). These tests have been performed to assess the functionality of the Subject Device. Results of all conducted testing and clinical assessments were found acceptable and do not raise any new issues of safety or effectiveness.

Page 7 of 8

{11}

510(k) Cios Select (VA21F) System

SIEMENS Healthineers

## 10. General Safety and Effectiveness Concerns:

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.

Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical and mechanical hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.

## 11. Conclusion as to Substantial Equivalence:

The Cios Select (VA21F) has the same indications for use, operating environment, and mechanical design as the predicate. Siemens concludes via the documentation provided in the 510(k) submission that the Cios Select (VA21F) does not introduce any new potential safety risks and is substantially equivalent and performs as well as the predicate devices.

Page 8 of 8

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K250241](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K250241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
