K240199 · Alphatec Spine, Inc. · OWB · May 22, 2024 · Radiology
Device Facts
Record ID
K240199
Device Name
IntraOp Alignment System
Applicant
Alphatec Spine, Inc.
Product Code
OWB · Radiology
Decision Date
May 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery. The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures. The IntraOp Alignment System is intended for use with patients aged 18 years and older.
Device Story
IntraOp Alignment System (IOA) is a software-based system for intraoperative spinal surgery; interfaces with mobile C-arm fluoroscope via video cable/frame grabber. Uses optical tracking sensor and reflective markers to track 6 DOF C-arm position/orientation in real-time. Transforms inputs by stitching multiple fluoroscopic images into long bi-planar radiographic images; performs distortion removal using C-arm calibration data. User identifies spine location in bi-planar views for scaling/depth correction. Provides measurement tools for spinal alignment assessment and surgical planning. Optional cloud connectivity for importing/exporting pre-operative/post-operative alignment parameters. Used in hospital operating rooms by trained healthcare professionals. Output displayed on monitor to assist intraoperative visualization, assessment, and surgical decision-making; benefits patient through improved alignment planning.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical performance testing, including verification of system accuracy, workflow, image stitching, C-arm tracking, spinal alignment measurements, and EMC/electrical safety testing.
Technological Characteristics
System includes optical tracking sensor, touchscreen computer, mobile cart, extension arm, and electronics box with frame grabber. Connectivity via Wi-Fi for cloud database access. Software-based image processing and stitching. Passive display system; does not transmit signals to fluoroscope. Optical tracking principle.
Indications for Use
Indicated for patients aged 18 years and older undergoing spinal surgery where a mobile C-arm fluoroscope is used for diagnosis and treatment. Assists healthcare professionals in viewing, storing, and measuring spinal alignment images and planning surgical procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Nuvasive® LessRay® with Enhanced Tracking (K170800)
{0}------------------------------------------------
May 22, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an eagle emblem, representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION."
Alphatec Spine, Inc. % Unnati Bhuptani Sr. Regulatory Affairs Specialist 1950 Camino Vida Roble CARLSBAD, CA 92008
Re: K240199
Trade/Device Name: IntraOp Alignment System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, LLZ Dated: January 24, 2024 Received: April 22, 2024
Dear Unnati Bhuptani:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240199
Device Name IntraOp Alignment System
### Indications for Use (Describe)
The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery.
The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures. The IntraOp Alignment System is intended for use with patients aged 18 years and older.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green, and the rest of the letters are dark blue. There is a green swoosh above the letters "tec". The letters are all lowercase.
# K240199 - 510(k) Summary
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| A. | SUBMITTER: | Alphatec Spine, Inc. |
|----|------------------------|-----------------------------------|
| | | 1950 Camino Vida Roble |
| | | Carlsbad, CA 92008 |
| | | Phone: (760) 432-9286 |
| | | Fax: (760) 431-0289 |
| | Contact Person: | Unnati Bhuptani |
| | | Sr. Regulatory Affairs Specialist |
| | | Alphatec Spine, Inc. |
| | | Contact Phone: (760) 356-6711 |
| | Date Summary Prepared: | January 24, 2024 |
#### B. DEVICE
| Trade Name: | IntraOp Alignment System |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Interventional Fluoroscopic X-ray System |
| Classification Name: | Image-Intensified Fluoroscopic X-ray System<br>(21 CFR §892.1650) |
| Regulatory Class: | Class II |
| Product Code: | OWB |
| Subsequent Codes: | JAA. LLZ |
#### C. LEGALLY MARKETED PREDICATE DEVICES
| Primary Predicate | |
|------------------------|-------------------------------------------------------------------|
| Trade Name: | Nuvasive® LessRay® with Enhanced Tracking |
| 510(k) number: | K170800 |
| 510(k) Clearance Date: | August 18, 2017 |
| Common or Usual Name: | Interventional Fluoroscopic X-ray System |
| Classification Name: | Image-Intensified Fluoroscopic X-ray System<br>(21 CFR §892.1650) |
| Regulatory Class: | Class II |
| Product Code: | OWB |
| Subsequent Codes: | JAA, LLZ |
| Additional Predicate | |
| Trade Name: | ATEC Alignment App |
| 510(k) number: | K211987 |
| 510(k) Clearance Date: | October 29, 2021 |
| Common or Usual Name: | System, Image Processing, Radiological |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is dark blue. There is a trademark symbol to the right of the "c".
| Classification Name: | Medical image management and processing system<br>(21 CFR §892.2050) |
|------------------------|----------------------------------------------------------------------|
| Regulatory Class: | Class II |
| Product Code: | LLZ |
| Reference Devices: | |
| Trade Name: | NuVasive® NuvaLine® |
| 510(k) number: | K192435 |
| 510(k) Clearance Date: | September 5, 2019 |
| Common or Usual Name: | System, Image Processing, Radiological |
| Classification Name: | Medical image management and processing system<br>(21 CFR §892.2050) |
| Regulatory Class: | Class II |
| Product Code: | LLZ |
#### D. DEVICE DESCRIPTION
The subject IntraOp Alignment (IOA) System consists of an optical tracking sensor, touchscreen computer, a mobile cart (stand) including an extension arm, and an electronics box with frame grabber, ports for video cables, power cables and external storage devices.
The subject IntraOp Alignment System is a software-based system that from one end interfaces with a mobile C-arm fluoroscope through a video cable and a frame capture device to receive fluoroscopic images as they are acquired, and from the other end connects to an optical position tracking sensor to track the position and orientation of a C-arm in real-time. By using subject IntraOp Alignment System along the side of mobile C-arm fluoroscopy equipment, the user of the system can combine multiple fluoroscopic images, as they are being acquired, into stitched long bi-planar radiographic images for intraoperative visualization and assessment. The user is required to identify the location of the spine in both bi-planar views to scale the image content and provide information regarding the depth of the operative anatomy. The user can then use the measurement tools available on the subject IntraOp Alignment System to visualize and assess the spinal alignment for intraoperative planning of the surgical procedure.
The subject IntraOp Alignment System has an optional capability to connect to a hospitalprovided Wi-Fi to import pre-operatively planned alignment parameters from an Alphatec cloud database. The subject IntraOp Alignment System also has the capability to export the post-operative alignment parameters to the cloud database.
The subject IntraOp Alignment System sits outside the sterile field and is provided non-The subject device is intended for use by trained healthcare professionals, and sterile. appropriately trained clinical and non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by trained non-clinical personnel.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "atec". The "a" is green and the "tec" is dark blue. There is a small trademark symbol in the upper right corner of the logo. The logo is simple and modern.
This submission seeks to establish that the IntraOp Alignment System is substantially equivalent to the primary predicate, Nuvasive® LessRay® with Enhanced Tracking (K170800), and additional predicate ATEC Alignment App (K211987) with respect to indications for use (IFU), intended use. technology, and performance specifications.
#### E. INDICATIONS FOR USE
The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery.
The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures.
The IntraOp Alignment System is intended for use with patients aged 18 years and older.
#### F. TECHNOLOGICAL COMPARISON TO PREDICATES
The subject device was compared to the predicate device in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, does not impact substantially equivalence, or safety and effectiveness. Table 1 below provided a detailed analysis of the substantial equivalence of the following device characteristics: indications for use, device functionalities. C-arm tracking, algorithms and accuracy specifications.
{6}------------------------------------------------
| Attribute | Predicate Devices | | Reference Device | Subject Device | SE Discussion |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | |
| Indications for Use | NuVasive® LessRay® with Enhanced<br>Tracking is intended for use in any<br>application where a fluoroscope<br>incorporated to aid in diagnosis and<br>treatment of disease. | The ATEC Alignment App assists<br>healthcare professionals in viewing,<br>measuring, and storing images as well as<br>planning orthopedic surgeries. The app<br>allows the healthcare professional to<br>perform generic and specialty<br>measurements of the images, and to plan<br>surgical procedures. The app includes<br>tools for measuring anatomical<br>components for implant selection and<br>offers the possibility to share data among<br>ATEC Alignment App users. The app is<br>for planning use only and is not intended<br>for primary diagnostic use. | NuVasive NuvaLine is a medical device<br>software application intended to assist<br>healthcare professionals in capturing,<br>viewing, measuring, and storage and<br>distribution of spinal alignment assessment<br>images at various time points in patient<br>care. Online synchronization of the<br>database allows healthcare professionals<br>and service providers to conveniently<br>perform and review spinal alignment<br>assessments of images by featuring<br>measurement tools on various platforms.<br>Clinical judgment and experience are<br>required to properly use the software. | The IntraOp Alignment System is intended<br>for use in applications where a mobile C-<br>arm fluoroscope is incorporated to aid in<br>diagnosis and treatment during spinal<br>surgery.<br>The IntraOp Alignment System is also<br>intended to assist healthcare professionals<br>in viewing, storing and measuring spinal<br>alignment assessment images at various<br>time points during surgery as well as<br>planning spinal surgical procedures.<br>The IntraOp Alignment System is intended<br>for use with patients aged 18 years and<br>older. | The subject device indications for use are<br>same as the primary predicate,<br>Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) for the image<br>stitching functionality. The indications<br>for use for use are also identical to the<br>additional predicate ATEC Alignment<br>App (K211987) for the spinal alignment<br>assessments functionality. Completed<br>V&V testing successfully demonstrated<br>that the risks related to the differences in<br>the indications for use between the<br>subject device and primary predicate<br>have been mitigated and do not affect the<br>safety and effectiveness of the device.<br>Therefore, the subject device can be<br>considered substantially equivalent to the<br>primary predicate, Nuvasive® LessRay®<br>with Enhanced Tracking (K170800) and<br>additional predicate, ATEC Alignment<br>App (K211987). |
| Regulation Number | §892.1650 | §892.2050 | §892.2050 | §892.1650; | Same |
| Product Code | OWB, JAA, LLZ | LLZ | LLZ | OWB, JAA, LLZ | Same |
| Device Class | II | II | II | II | Same |
| Device Classification Name | Interventional Fluoroscopic X-Ray<br>System | System, Image Processing, Radiological | System, Image Processing, Radiological | Interventional Fluoroscopic X-Ray System | Same |
| Device Functionalities | Image Acquisition, Enhancement and<br>Display :<br>• Software based device used to<br>provide computer display systems<br>interfaced to fluoroscope through a<br>video cable. The images produced<br>by the fluoroscope are transmitted<br>through a cable to a frame capture<br>board in the computer where the<br>images are enhanced and then<br>displayed on the monitor.<br>• Enhanced images are displayed on a<br>computer monitor at the same time<br>that the corresponding original | Measurement:<br>• Spinal alignment assessments of<br>images | Measurement:<br>• Spinal alignment assessments of<br>images | Image Acquisition, Processing,<br>Stitching, and Display:<br>• Software based system that from<br>one end interfaces with a mobile C-<br>arm fluoroscope through a video<br>cable and a frame capture device to<br>receive fluoroscopic images as they<br>are acquired.<br>• As the images are being acquired,<br>the user of the system can combine<br>multiple fluoroscopic images, into<br>stitched long bi-planar radiographic<br>images for intraoperative | Substantially Equivalent. The image<br>stitching feature of the subject device is<br>similar in function to the primary<br>predicate. Nuvasive® LessRay® with<br>Enhanced Tracking (K170800). The<br>visualization and spinal alignment<br>assessment feature of the subject device is<br>similar in function to the additional<br>predicate, ATEC Alignment App<br>(K211987). |
| Attribute | Predicate Devices | | Reference Device | Subject Device | SE Discussion |
| | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | |
| | • image is displayed on the<br>fluoroscope monitor(s).<br>• Serves only as an image display<br>which is in addition to the<br>fluoroscope's standard image<br>display device. Device is passive, in<br>that the operation depends only on<br>the video output of the fluoroscope,<br>and it does not transmit any signals<br>or images to the fluoroscope. | | | visualization and spinal alignment<br>assessments.<br>• Serves only as an image display which<br>is in addition to the fluoroscope's<br>standard image display device. Device<br>is passive, in that the operation depends<br>only on the video output of the<br>fluoroscope, and it does not transmit<br>any signals or images to the<br>fluoroscope.<br>Spinal Alignment Measurements:<br>• After bi-planar images are acquired,<br>they are calibrated by the user using<br>the established tools available on<br>IntraOp Alignment System<br>software.<br>• The user can place measurement<br>tools for spinal alignment<br>assessments. | |
| C-arm Tracking | • When tracking is enabled, will<br>automatically choose the Baseline<br>when the fluoroscope is near the<br>location and orientation that the<br>Baseline was initially taken.<br>• When tracking is enabled, requires<br>hardware components in order to<br>mount the off-the-shelf tracking<br>hardware to the C-arm and to the<br>operating table.<br>• When tracking is enabled, requires<br>the use of an off-the-shelf tracking<br>system in order to track the 6 DOF<br>location of the C-arm relative to the<br>operating table.<br>• When tracking is enabled, visual<br>cues are provided which help guide<br>the user in positioning the C-arm<br>back to where a prior Baseline was<br>taken. | N/A | N/A | • IntraOp Alignment System uses off-<br>the-shelf tracking sensor and reflective<br>tracking markers to track the 6 DOF<br>location of the C-arm relative to<br>tracking sensor.<br>• The reflective tracking markers are<br>attached on the mobile C-arm and are<br>registered and calibrated on the IntraOp<br>Alignment System after its initial<br>application on the C-arm.<br>• The IntraOp Alignment System GUI<br>guides the user regarding the<br>placement of the C-arm over the<br>operative anatomy and tracks the C-<br>arm position relative to the tracking<br>sensor using the reflective markers. | Substantially Equivalent. |
| Tracking Options | • Electromagnetic or optical | N/A | N/A | • Optical | Substantially Equivalent |
| Algorithms | • Image quality improvement using<br>averaging algorithm | Patient parameters and calculations based<br>on published literature | Various spinal assessment algorithms | • Image quality improvement by<br>distortion removal using C-Arm<br>calibration data | Substantially Equivalent. The image<br>stitching algorithm of the subject device is<br>similar in function to the primary |
| Attribute | Predicate Devices | | Reference Device | Subject Device | |
| | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | SE Discussion |
| | • Contrast and brightness<br>enhancement with simultaneous<br>reduction of random noise | | | • Image stitching using C-Arm position<br>coordinates<br>• Depth correction based on user input<br>• Various spinal assessment algorithms | predicate, Nuvasive® LessRay® with<br>Enhanced Tracking (K170800). The<br>spinal alignment assessment algorithm of<br>the subject device is similar in function to<br>the additional predicate, ATEC<br>Alignment App<br>(K211987). |
## Table 1: Comparison for Substantial Equivalence
{7}------------------------------------------------
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green and has a swoosh-like extension that goes over the "t". The rest of the letters are in a dark blue color. There is a trademark symbol in the upper right corner of the "c".
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "atec". The "a" is green, and the rest of the letters are black. There is a trademark symbol in the upper right corner of the logo. There is a black line underneath the logo.
#### PERFORMANCE DATA G.
Nonclinical performance testing demonstrates that the subject IntraOp Alignment System meets the functional, system, and software requirements. The following testing was conducted:
- Verification testing to confirm that the system meets the accuracy specifications ●
- Verification of the workflow to confirm that the system performs according to ● specifications
- Verification testing to verify that the following features: image stitching, C-arm tracking and spinal alignment measurements meet performance specifications
EMC and Electrical Safety Testing of the IntraOp Alignment System was performed to ensure all functions of the system and its accessories are electrically safe and comply with recognized electrical safety standards.
## Clinical Information
Determination of substantial equivalence is not based on an assessment of clinical performance data.
#### H. CONCLUSION
The information provided in this submission demonstrates that the subject device does not pose additional risk to safety and effectiveness when compared to the predicate device. The subject device, IntraOp Alignment System, is substantially equivalent to the primary predicate, Nuvasive® with Enhanced Tracking (K170800), and additional predicate ATEC Alignment App (K211987).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.