← Product Code [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB) · K234109

# Ziehm Solo FD (K234109)

_Ziehm Imaging GmbH · OWB · Jan 26, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K234109

## Device Facts

- **Applicant:** Ziehm Imaging GmbH
- **Product Code:** [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB.md)
- **Decision Date:** Jan 26, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

## Device Story

Mobile C-arm fluoroscopic X-ray system; utilizes X-ray generator and flat panel detector to capture real-time anatomical images. X-rays penetrate patient; detector converts signals to digital images. System includes mobile stand with mono-block generator, heat management, and CAN BUS control. Optional viewing station (monitor cart) provides remote touch control. Operated by physicians/clinicians in surgical/interventional settings. Provides real-time visualization, digital subtraction, and cine loop acquisition to assist in localizing pathology and guiding surgical procedures. Benefits include improved visualization of anatomical structures during minimally invasive interventions.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation testing conducted per 21 CFR 820 design controls. Compliance with electrical safety (ANSI/AAM ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and X-ray performance standards (21 CFR 1020.30-32, IEC 60601-1-3, IEC 60601-2-54) confirmed. Software testing performed per FDA guidance; cybersecurity improvements implemented.

## Technological Characteristics

Mobile C-arm fluoroscopic X-ray system. Components: mono-block X-ray generator, flat panel detector, CAN BUS control system, integrated laser positioning. Connectivity: wireless features and interfaces. Software: Ubuntu 20.04 OS. Standards: ANSI/AAM ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, 21 CFR 1020.30-32.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Ziehm Solo FD ([K231669](/device/K231669.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 26, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Ziehm Imaging GmbH c/o Tsvetelina Milanova Specialist Regulatory Affairs Lina-Ammon-Strasse 10 Nuremberg, 90471 GERMANY

Re: K234109

Trade/Device Name: Ziehm Solo FD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB , JAA, OXO Dated: December 23, 2023 Received: December 27, 2023

Dear Tsvetelina Milanova:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### Indications for Use

Submission Number (if known)

K234109

Device Name

Ziehm Solo FD

#### Indications for Use (Describe)

The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging.

The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures.

The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summ

Prepared on: 2024-01-18

### Contact Details

21 CFR 807.92(a)(1)

| Applicant Name                     |                                                          | Ziehm Imaging GmbH                            |                     |                     |
|------------------------------------|----------------------------------------------------------|-----------------------------------------------|---------------------|---------------------|
| Applicant Address                  |                                                          | Lina-Ammon-Strasse 10 Nuremberg 90471 Germany |                     |                     |
| Applicant Contact Telephone        |                                                          | +4991166067581                                |                     |                     |
| Applicant Contact                  |                                                          | Ms. Tsvetelina Milanova                       |                     |                     |
| Applicant Contact Email            |                                                          | Zie-Regulatory@ziehm.com                      |                     |                     |
| Device Name                        |                                                          |                                               |                     | 21 CFR 807.92(a)(2) |
| Device Trade Name                  |                                                          | Ziehm Solo FD                                 |                     |                     |
| Common Name                        |                                                          | Image-intensified fluoroscopic x-ray system   |                     |                     |
| Classification Name                |                                                          | Interventional Fluoroscopic X-Ray System      |                     |                     |
| Regulation Number                  |                                                          | 892.1650                                      |                     |                     |
| Product Code(s)                    |                                                          | OWB, JAA, OXO                                 |                     |                     |
| Legally Marketed Predicate Devices |                                                          |                                               | 21 CFR 807.92(a)(3) |                     |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) |                                               |                     | Product Code        |
| K231669                            | Ziehm Solo FD                                            |                                               |                     | OWB                 |
|                                    |                                                          |                                               |                     |                     |

## Device Description Summary

Ziehm Solo FD uses X-ray imaging to visualize the human anatomy. The X-ray tube in the generator produces X-rays that penetrate the patient and then hit a special detector that converts them into digital images. This is done under the user and at the direction of a physician who determines the specific clinical procedure. This visualization assists the physician in localizing pathological areas or during surgical procedures. The device enables real-time image acquisition of in vivo surgical procedures and post-operative results.

The Ziehm Solo FD consist of one mobile unit, the Mobile Stand. Optionally the device can be ordered with a Viewing Station (Monitor Cart). The Mobile Stand incorporates a small compact design making the C-arm in relation to the patient easier for the operator. The generator with X-ray tube, advanced heat management system, X-ray control and collimators are assembled in one housing in a mono-block generator. The system control is handled via CAN BUS control system.

The mechanical C-Profile supports the flat panel detector and an integrated laser positioning device. The optional available Viewing Station (Monitor Cart) provides a remote touch Solo Center that duplicates the touch Solo Center mounted on the Mobile Stand.

The comparison of the predicate device and the modified device shows that the scientific and technical characteristics of the Ziehm Solo FD are substantially equivalent as those of the Ziehm Solo FD predicate device.

# Intended Use/Indications for Use

## 21 CFR 807.92(a)(5)

CFR 807.92(a)(4)

The Ziehm Solo FD is intended for use in providing mediatic populations, using pulstions, using pulsed and continuous fluoroscopic imaging.

The device provides contactless fluoroscopic image capture, temporarily storing, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, neurologic, neurologic, vascular, cardiac, angiographic, critical care, and emergency

{4}------------------------------------------------

#### room fluoroscopy procedures.

The visualization of such anatomical structures assists the clinical outcome. This device does not support direct radiographic film exposures and is not in performing mammography. The system is not intended for use near MRI systems.

### Indications for Use Comparison

### 21 CFR 807.92(a)(5)

The indications for use of the predicate and modified device are equivalent.

### Technological Comparison

The predicate and proposed devices have technological characteristic. The key modification refers to an updated release of the software, which incorporates an operating system upgrade from Ubuntu 16.04 to Ubuntu 20.04.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The design of the modified Ziehm Solo FD was completed in accordance with Ziehm Imaging GmbH Quality Management System Design Controls, 21 CFR 820 and applicable standards. Verification and Validation testing were successfully conducted on the device in compliance with FDA requirements as stated in the following documentation.

No new tests regarding electrical safety according to ANSI/AAM ES60601-1 and regarding electromagnetic compatibility according to IEC 60601-1-2 were performed. The device is compliant with both standards.No new tests have been done according to Guidance's "Radio Frequency Wireless Technology in Medical Devices" and "Design Considerations and Premarket Submissions Recommendations for Interoperable Medical Devices". Neither the wireless features nor the interfaces of the device affect the safety and effectiveness.

The modified Ziehm Solo FD complies with 21 CFR 1020.30-32 Federal Performance Standards for X-Ray Fluoroscopic equipment and with relevant safety standards such as EC 60601-1-3, IEC 60601-2-54.No new non-clinical image comparison with sets of images with the modified device and the predicate have been performed, the device equivalent regarding image quality.Software testing was performed as required by "Content of Premarket Submissions for Device Software Functions". The Cybersecurity of the device has been improved.

21 CFR 807.92(a)(6)

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K234109](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K234109)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
