The Ziehm Vision RFD is intended for use in providing medical imaging for adult and pediatric populations, using pulsed and continuous fluoroscopic digital imaging, as well as digital subtraction and cine image capture during diagnostic interventional and surgical procedures where intraoperative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for computer aided surgery procedures and whenever the clinician benefits from the high degree of geometric imaging accuracy, and where such fluoroscopic, cine and DSA imaging is required. The visualization of such anatomical structures assists the clinician in the clinical outcome. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.
Device Story
Mobile fluoroscopic C-arm system consisting of a mobile stand (C-arm with X-ray generator and flat panel detector) and a monitor cart/workstation. Captures X-ray images for real-time visualization during interventional and surgical procedures. Operated by physicians/clinicians in ORs, clinics, or emergency settings. Transforms X-ray inputs into digital images displayed on integrated LCD monitors. Supports digital subtraction angiography (DSA) and cine capture. Provides high-resolution anatomical visualization to assist in localization of pathology, surgical guidance, and assessment of clinical outcomes. Benefits patients by enabling minimally invasive procedures through precise intraoperative imaging. Includes optional wireless footswitch and various network interfaces for DICOM image storage and peripheral connectivity.
Clinical Evidence
Bench testing only. Non-clinical image comparison and dose testing were performed using anatomical simulation phantoms representative of adult and pediatric populations. A radiologist assessed image sets, concluding that image quality is comparable to the predicate device. Compliance with electrical safety (ANSI/AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and X-ray performance standards (21 CFR 1020.30-32) was demonstrated.
Technological Characteristics
Mobile C-arm X-ray system. Detectors: 8-inch/12-inch a-Si, CMOS, and new 12-inch IGZO flat panel detectors with Cesium-Iodide (CsI) scintillators. Generator: 20kW, 25kW, or 30kW mono-block. Connectivity: DICOM wired/wireless, USB, DVD. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54. Software: Application-Oriented Anatomical Programs (AOAP) and Anatomical Marking Tool (AMT).
Indications for Use
Indicated for adult and pediatric populations requiring fluoroscopic, cine, and digital subtraction angiography (DSA) imaging during diagnostic, interventional, and surgical procedures (e.g., cardiology, orthopedics, neurosurgery, urology, emergency care). Contraindicated for mammography and use near MRI systems.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 20, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
Ziehm Imaging GmbH % Stefan Fiedler Director Quality Management & Regulatory Affairs Lina-Ammon-Strasse 10 Nuremberg, Bavaria 90471 GERMANY
Re: K231692
Trade/Device Name: Ziehm Vision RFD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: JAA, OWB, OXO Dated: October 23, 2023 Received: October 23, 2023
Dear Stefan Fiedler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K231692
Device Name
Ziehm Vision RFD
## Indications for Use (Describe)
The Ziehm Vision RFD is intended for use in providing medical imaging for adult and pediatric populations, using pulsed and continuous fluoroscopic digital imaging, as well as digital subtraction and cine image capture during diagnostic interventional and surgical procedures where intraoperative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for computer aided surgery procedures and whenever the clinician benefits from the high degree of geometric imaging accuracy, and where such fluoroscopic, cine and DSA imaging is required. The visualization of such anatomical structures assists the clinician in the clinical outcome.
This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the company name "ziehmimaging" on the right. The text is in a teal color, with the "imaging" portion being a darker shade of teal.
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center –W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
K231692
November 20, 2023
In accordance with 21 CFR §807.92 the following 510(k) summary information is provided:
| Submitter Address: | Ziehm Imaging GmbH<br>Lina-Ammon-Strasse 10<br>90471 Nuremberg<br>Germany<br>Phone: +49 911 66067 0<br>Fax: +49 911 66067 390 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person<br>Agent: | Kevin Bridgman<br>VP Quality Assurance & Requlatory Affairs<br>Ziehm-Orthoscan, Inc.<br>14555 N 82nd St<br>Scottsdale AZ, 85260 |
| Secondary Contact<br>Person: | Stefan Fiedler<br>Director QM/RA<br>Ziehm Imaging GmbH<br>Phone: +49 911 66067 219<br>Fax: +49 172 8651712 |
| Device (Trade Name): | Ziehm Vision RFD |
| Common /Usual<br>Names: | Mobile Fluoroscopic C-Arm |
| Regulation: | 21CFR 892.1650 |
| Regulation Description: | Image-intensified fluoroscopic x-ray system |
| Product Code: | JAA; OWB; OXO |
| Classification: | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The graphic features a stylized human figure inside a globe-like structure. The text is in a teal color and appears to be a sans-serif font.
| Predicate Device: | K203428 - Ziehm Vision RFD |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Decision Date: | 03/17/2021 |
| Regulation: | 21CFR 892.1650 |
| Regulation Description<br>Name: | Image-intensified fluoroscopic x-ray system |
| Product Code: | JAA; OWB; OXO |
| Summary of<br>Technological<br>Characteristics | The Ziehm Vision RFD employs X-rays as its imaging technology for visualizing human anatomy. The X-ray tube in the generator produces X- rays, guided toward the patient under control of the user at the direction of a physician who determines the specific clinical procedure. The images from the system assist the physicians in visualizing the patient's anatomy. This visualization helps to localize regions of pathology and for surgical procedures. The device provides both real-time image capture and post capture visualization and of in vivo surgical procedures and post-surgical outcomes. The Ziehm Vision RFD mobile fluoroscopy system is a flat panel detector (EPD) and fluoroscopic X-ray imaging system consisting of two mobile units. |
at panel detector (FPD) and fluoroscopic X-ray imaging system consisting of two mobile units: a Mobile Stand (C-Arm) and a Monitor Cart/Workstation. The Mobile Stand is comprised of a mono-block high voltage generator, X-ray control, and a C-Profile which is "C" shaped and supports the X-ray generator, and the image receptor Flat Panel Detector (FPD).
The mobile stand supports the optional wireless footswitch for optimum positioning for the surgeon by removing the cable on the floor.
The Monitor Cart is a mobile platform that connects to the Mobile Stand by a cable, and which integrates the LCD flat panel display monitors, image processing, user controls and image recording devices. Interfaces provided for optional peripheral devices such as external monitors, thermal video printers, wireless video display, wireless video server, injector connection and image storage devices (USB, DVD) and DICOM fixed wired and wireless network interfaces.
The proposed modified device Ziehm Vision RFD C-arm employs the same fundamental control, and substantially equivalent scientific technology as that of our predicate device Ziehm Vision RFD C-arm (K203428).
The radiation control, X-Ray monoblock generator, power supplies as well as our advanced imaging system are the same as the predicate device Ziehm Vision RFD C-arm (K203428).
Software architecture design is substantially equivalent to that of the predicate device Ziehm Vision RFD C-arm (K203428).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic with a stylized human figure inside, accompanied by the text "ziehmimaging" in a teal color. The text is written in a sans-serif font, with the letters gradually becoming darker from left to right.
The primary modification of the C-Arm includes a 12 inch IGZO (Indium gallium zinc oxide) flat panel detector (FPD) which have substantially equivalent technology characteristics as our predicate device Ziehm Vision RFD C-arm (K203428). The new 12 inch IGZO FPD is an addition to already introduced a-Si and CMOS FPD's. The flat panel detectors have the same outer product design of the housing, both devices use safety shielding for radiation suppression and use solid state x-ray image receptors (SSXI / FPD) 8inch and 12 inch aSi, 8 inch and 12 inch CMOS and the only difference to the predicate Ziehm Vision RFD is the new 12 inch IGZO panel.
The comparison of the predicate device and the modified devices shows that the scientific and technical characteristics of the Ziehm Vision RFD are substantially equivalent as those of the Ziehm Vision RFD predicate device (K203428).
- Intended Use The Ziehm Vision RFD is a mobile C-arm providing image data by means of a non-invasive x-ray technique during medical procedures and stores them temporarily. The Ziehm Vision RFD is intended for use in all medical indications requiring fluoroscopy. The Ziehm Vision RFD is intended for use to provide image data specifically but not limited in the field of interventional radiology and cardiology as well as in cardiac surgery and in hybrid applications. The system is intended for use with human beings of any age. It is the physician's responsibility to decide whether to use the system with infants, children and adipose patients. The system is intended for use with human bodies covering such structures but not limited to the following, e.q. organs, tissue, bones, implants depending on the medical indication. These devices are not intended for use in performing mammographic exposures. The systems are not intended for use near MRI systems.
- Indications for Use: The Ziehm Vision RFD is intended for use in providing medical imaging for adult and pediatric populations, using pulsed and continuous fluoroscopic digital imaging, as well as digital subtraction and cine image capture during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurology, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumber regions of the spine, and joint fractures of the upper and lower extremities, and where digital image data is required for computer aided surgery procedures and whenever the clinician benefits from the high degree of geometric imaging accuracy, and where such fluoroscopic, cine and DSA imaging is required. The visualization of such anatomical structures assists the clinician in the clinical outcome.
This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use in all MRI environments.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The text is in a teal color and appears to be a sans-serif font.
## Device Comparison Table
The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
| Model | Modified<br>Ziehm Vision RFD | Predicate<br>Ziehm Vision RFD<br>(K203428) | Comparable Properties<br>Substantial<br>Equivalence Discussion |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K231692 | K203428 | - |
| Classification | Class II | Class II | Identical |
| Product Code | JAA (system, x-ray,<br>fluoroscopic, image-<br>intensified) | JAA (system, x-ray,<br>fluoroscopic, image-<br>intensified) | Identical |
| Application / Indications for Use | | | |
| Indications for<br>Use | The Ziehm Vision RFD is<br>intended for use in<br>providing medical<br>imaging for adult and<br>pediatric populations,<br>using pulsed and<br>continuous fluoroscopic<br>digital imaging, as well<br>as digital subtraction<br>and cine image capture<br>during diagnostic<br>interventional and<br>surgical procedures<br>where intra-operative<br>imaging and<br>visualization of complex<br>anatomical structures of<br>both lower and higher<br>contrast density are<br>required. Such<br>procedures may include<br>but are not limited to<br>those of interventional<br>cardiology, heart<br>surgery, hybrid<br>procedures,<br>interventional radiology,<br>interventional | The Ziehm Vision RFD is<br>intended for use in<br>providing medical imaging<br>for<br>adult and pediatric<br>populations, using pulsed<br>and continuous<br>fluoroscopic<br>digital imaging, as well as<br>digital subtraction and cine<br>image capture during<br>diagnostic interventional<br>and surgical procedures<br>where intra-operative<br>imaging and visualization<br>of complex anatomical<br>structures of both lower<br>and<br>higher contrast density are<br>required. Such procedures<br>may include but are<br>not limited to those of<br>interventional cardiology,<br>heart surgery, hybrid<br>procedures, interventional<br>radiology, interventional<br>angiography,<br>electrophysiology,<br>pediatrics, endoscopic | Identical |
| Model | Modified<br>Ziehm Vision RFD | Predicate<br>Ziehm Vision RFD<br>(K203428) | Comparable Properties<br>Substantial<br>Equivalence Discussion |
| | angiography,<br>electrophysiology,<br>pediatrics, endoscopic,<br>urological,<br>gastroenterology,<br>orthopedic, maxillofacial<br>surgery, neurology,<br>neurosurgery, critical<br>care, emergency room<br>procedures, and those<br>procedures visualizing<br>structures of the<br>cervical, thoracic, and<br>lumber regions of the<br>spine, and joint<br>fractures of the upper<br>and lower extremities,<br>and where digital image<br>data is required for<br>computer aided surgery<br>procedures and<br>whenever the clinician<br>benefits from the high<br>degree of geometric<br>imaging accuracy, and<br>where such fluoroscopic,<br>cine and DSA imaging is<br>required. The<br>visualization of such<br>anatomical structures<br>assists the clinician in<br>the clinical outcome.<br>This device does not<br>support direct<br>radiographic film<br>exposures and is not<br>intended for use in<br>performing<br>mammography. The<br>system is not intended<br>for use in all MRI<br>environments. | urological,<br>gastroenterology,<br>orthopedic, maxillofacial<br>surgery, neurology,<br>neurosurgery, critical care,<br>emergency room<br>procedures, and those<br>procedures visualizing<br>structures of<br>the cervical, thoracic, and<br>lumber regions of the<br>spine, and joint fractures<br>of<br>the upper and lower<br>extremities, and where<br>digital image data is<br>required for<br>computer aided surgery<br>procedures and whenever<br>the clinician benefits from<br>the high degree of<br>geometric imaging<br>accuracy, and where such<br>fluoroscopic,<br>cine and DSA imaging is<br>required. The visualization<br>of such anatomical<br>structures assists the<br>clinician in the clinical<br>outcome.<br>This device does not<br>support direct radiographic<br>film exposures and is not<br>intended for use in<br>performing<br>mammography. The<br>system is not intended<br>for use near MRI systems. | |
| Model | Modified<br>Ziehm Vision RFD | Predicate<br>Ziehm Vision RFD<br>(K203428) | Comparable Properties<br>Substantial<br>Equivalence Discussion |
| Digital<br>Radiography<br>(Snapshot) /<br>Operating Values | • Variant 20kW<br>kV range: 40 - 120 kV<br>mA range: up to 200<br>mA<br>• Variant 25kW<br>kV range: 40 - 120 kV<br>mA range: up to 250<br>mA<br>• Variant 30 kW<br>kV range: 40 - 120 kV<br>mA range: up to 300<br>mA | • Variant 20kW<br>kV range: 40 - 120 kV<br>mA range: up to 200<br>mA<br>• Variant 25kW<br>kV range: 40 - 120 kV<br>mA range: up to 250<br>mA<br>• Variant 30 kW<br>kV range: 40 - 120 kV<br>mA range: up to 300 mA | Identical |
| Beam Limiter/ Collimator | | | |
| Collimator<br>System | Asymmetrical Collimator:<br>• Devices with 12inch<br>and 8inch cm Flat Panel<br>Detectors:<br>• Dedicated pre-<br>collimator for FPD<br>• Collimator Rotation:<br>+/- 90°<br>• Iris and Asymmetric<br>SlotCollimator:<br>- 50 - 307 mm<br>diameter (12 inch FPD<br>CMOS & IGZO)<br>-50 - 298 mm<br>diameter (12 inch FPD<br>aSi)<br>- 50-205 mm diameter<br>(8inch FPD CMOS)<br>• Virtual Collimation<br>without radiation | Asymmetrical Collimator:<br>• Devices with 12inch and<br>8inch cm Flat Panel<br>Detectors:<br>• Dedicated pre-<br>collimator for FPD<br>• Collimator Rotation:<br>+/- 90°<br>• Iris and Asymmetric<br>SlotCollimator:<br>- 50 - 307 mm<br>diameter (12 inch FPD<br>CMOS)<br>-50 - 289 mm diameter<br>(12 inch FPD aSi)<br>- 50-205 mm diameter<br>(8inch FPD CMOS)<br>- 50-198 mm diameter<br>(8inch FPD aSi)<br>• Virtual Collimation<br>without radiation | The collimator system of<br>the IGZO is the same as<br>the already introduced<br>CMOS FPD. |
| Image Detector | | | |
| Detector<br>Technology | Variant aSi FPD:<br>• Type:<br>Amorphous<br>Silicon<br>Flat Panel<br>Detector (aSi)<br>• Scintillator:<br>Cesium-<br>Iodide (CsI)<br>Variant CMOS FPD: | Variant aSi FPD:<br>• Type: Amorphous Silicon<br>Flat Panel Detector (aSi)<br>• Scintillator:<br>Cesium-<br>Iodide (CsI)<br>Variant CMOS FPD:<br>• Type: CMOS Flat Panel | The new 12 inch IGZO<br>(FPD) is an additional flat<br>panel type beside the<br>already established FPDs<br>based on aSi (amorphous<br>Silicon) and CMOS<br>(Complementary Metal<br>Semi-conductor) |
| Model | Modified<br>Ziehm Vision RFD | Predicate<br>Ziehm Vision RFD<br>(K203428) | Comparable Properties<br>Substantial<br>Equivalence Discussion |
| | • Type: CMOS Flat Panel<br>Detector<br>• Scintillator: Cesium-<br>Iodide (CsI)<br><br>Variant IGZO FPD:<br>• Type: IGZO Flat Panel<br>Detector<br>• Scintillator: Cesium-<br>Iodide (CsI) | • Scintillator: Cesium_ | technology. IGZO FPDs<br>are using the aSi/IGZO<br>sensor technology |
| Detector Sizes | 30cm x 30cm (12inch) aSi<br>(amorphous silicon TFT<br>technology)<br>• Size: 29.8 cm x 29.8<br>cm<br>• Detector matrix:<br>1,536 x 1,536 pixels<br>• Magnifier 1:<br>1,024 x 1,024 pixels<br>• Magnifier 2:<br>768 x 768 pixels<br>• Scintillator cesium<br>iodide<br>• Dynamic Range: 94 dB<br>• DQE: 77%<br>• MTF>50%@1lp/mm<br>• Pixel size 194µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>• System resolution<br>(Nyquist):<br>2.6 lp/mm | 30cm x 30cm (12inch) aSi<br>(amorphous silicon TFT<br>technology)<br>• Size: 29.8 cm x 29.8<br>cm<br>• Detector matrix:<br>1,536 x 1,536 pixels<br>• Magnifier 1:<br>1,024 x 1,024 pixels<br>• Magnifier 2:<br>768 x 768 pixels<br>• Scintillator cesium<br>iodide<br>• Dynamic Range: 94 dB<br>• DQE: 77%<br>• MTF>50%@1lp/mm<br>• Pixel size 194µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>• System resolution<br>(Nyquist):<br>2.6 lp/mm | The active pixel area of<br>the detector types are not<br>identical but are very<br>similar in image area of<br>approx. 12 inch x 12 inch.<br><br>IGZO detector has higher<br>number of pixel and<br>smaller pixel pitch<br>accordingly; therefor the<br>IGZO detector has better<br>resolution compared to<br>aSi detectors. This shows<br>the technical advantage<br>over the established aSi<br>systems.<br><br>The slight differences in<br>detector and pixel size<br>does not have influence on<br>safety and effectiveness of<br>the C-arm. |
| | Variant 31 cm x 31 cm<br>(12inch) CMOS<br>Technology:<br>• Size: 30.7 cm x 30.7<br>cm<br>• Detector matrix:<br>3,072 x 3,072 pixels<br>• Magnifier 1:<br>2,048x 2,048 pixels<br>• Magnifier 2:<br>1,536 x 1,536 pixels<br>• Dynamic Range:<br>- 1x1 binning:84 dB<br>- 2x2 binning: 95 dB | Variant 31 cm x 31 cm<br>(12inch) CMOS<br>Technology:<br>• Size: 30.7 cm x 30.7<br>cm<br>• Detector matrix:<br>3,072 x 3,072 pixels<br>• Magnifier 1:<br>2,048x 2,048 pixels<br>• Magnifier 2:<br>1,536 x 1,536 pixels<br>• Dynamic Range:<br>- 1x1 binning:84 dB<br>- 2x2 binning: 95 dB | |
| Model | Modified<br>Ziehm Vision RFD | Predicate<br>Ziehm Vision RFD<br>(K203428) | Comparable Properties<br>Substantial<br>Equivalence Discussion |
| | • Scintillator: cesium<br>iodide<br>• DQE: 75%<br>• MTF:<br>- 1 lp/mm: 55 %<br>- 2 lp/mm: 23 %<br>- 3 lp/mm: 10 %<br>- 4 lp/mm: 5 %<br>• Pixel size 100 µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>System resolution<br>(Nyquist):<br>5 lp/mm | • Scintillator: cesium<br>iodide<br>• DQE: 75%<br>• MTF:<br>- 1 lp/mm: 55 %<br>- 2 lp/mm: 23 %<br>- 3 lp/mm: 10 %<br>- 4 lp/mm: 5 %<br>• Pixel size 100 µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>System resolution<br>(Nyquist):<br>5 lp/mm | |
| | 20 cm x 20 cm (8inch) aSi<br>(amorphous silicon TFT<br>technology)<br>• Size: 19.9 cm x 19.9<br>cm<br>• Detector matrix:<br>1,024 x 1,024 pixels<br>• Magnifier 1:<br>768 x 768 pixels<br>• Magnifier 2:<br>512 x 512 pixels<br>• Scintillator cesium<br>iodide<br>• Dynamic Range: 94 dB<br>• DQE: 77%<br>• MTF>50%@1lp/mm<br>• Pixel size 194µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>• System resolution<br>(Nyquist):<br>2.6 lp/mm | 20 cm x 20 cm (8inch) aSi<br>(amorphous silicon TFT<br>technology)<br>• Size: 19.9 cm x 19.9<br>cm<br>• Detector matrix:<br>1,024 x 1,024 pixels<br>• Magnifier 1:<br>768 x 768 pixels<br>• Magnifier 2:<br>512 x 512 pixels<br>• Scintillator cesium<br>iodide<br>• Dynamic Range: 94 dB<br>• DQE: 77%<br>• MTF>50%@1lp/mm<br>• Pixel size 194µm<br>• Garyscale 65,536 gray<br>scale values (16bit)<br>• System resolution<br>(Nyquist):<br>2.6 lp/mm | |
| | Variant 21 cm x 21 cm<br>(8inch) CMOS<br>Technology:<br>• Size: 20.5 cm x 20.5<br>cm<br>• Detector matrix:<br>2,053x 2,053 pixels<br>• Magnifier 1:<br>1,536 x 1,536 pixels | Variant 20.5 cm x 20.5 cm<br>(8inch) CMOS Technology:<br>• Size: 20.5 cm x 20.5<br>cm<br>• Detector matrix:<br>2,048x 2,048 pixels<br>• Magnifier 1:<br>1,536 x 1,536 pixels<br>• Magnifier 2: | |
| Model | Modified<br>Ziehm Vision RFD…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.