Garion Image-Intensified Fluoroscopic X-Ray System, Mobile
Applicant
Sg Healthcare Co, Ltd.
Product Code
OWB · Radiology
Decision Date
Oct 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
Device Story
Garion is a mobile fluoroscopy C-arm system providing real-time fluoroscopic and spot-film X-ray images. Input consists of X-ray radiation attenuated by the patient, captured by a high-resolution flat panel detector (FPD). The system includes a high-voltage generator, X-ray tube, collimator, and image processing unit. Operators (trained healthcare professionals) use the device in ORs or clinical settings. The image processing unit performs real-time enhancements including positive/negative inversion, zoom, rotation, noise reduction, and edge enhancement. Output is displayed on a monitor for clinical visualization, aiding in surgical guidance and diagnostic procedures. Dual laser pointers assist in patient positioning to minimize unnecessary radiation exposure. An attachable grid allows for low-dose examination options. The device facilitates clinical decision-making by providing high-quality, real-time anatomical visualization during procedures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and compliance with international safety and performance standards (IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, and IEC 62366) and FDA radiation safety standards (21 CFR 1020.32, 1020.33).
Technological Characteristics
Mobile C-arm fluoroscopy system. Components: rotating anode X-ray tube (10° target angle, 0.3/0.6 focal size), Cesium Iodide amorphous silicon FPD (9"x9"), high-voltage generator, and image processing unit. Connectivity: digital image processing and display. Sterilization: N/A (device is for use in sterile/non-sterile environments). Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, IEC 62366, 21 CFR 1020.32/33.
Indications for Use
Indicated for radiological guidance and visualization during diagnostic, interventional, and surgical procedures for all patients except neonates (birth to one month).
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 6, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
SG Healthcare CO. LTD. % Daniel Kamm Principal Engineer Kamm & Assosciates 8870 Ravello Ct NAPLES FL 34114
Re: K222080
Trade/Device Name: Garion Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB Dated: July 8, 2022 Received: July 15, 2022
## Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222080
Device Name Garion
Indications for Use (Describe)
The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-family: Arial;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary, 510(k) K222080
# Submitter: SG Healthcare Co., Ltd. 10 Yangnyeong-ro, Moga-myeon, Icheon-si, Gyeonggi-do, 17408 Seoul, Republic of Korea Tel: +82 2 449 6908 Fax: +82 2 449 6909 www.sghealthcare.com Contact: Byungju Kang, bjkang@sghealthcare.com Date Prepared: September 14, 2022
#### 1. Identification of the Device:
Trade/Device Name: Garion Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: II Product Code: OWB
## 2. Equivalent legally marketed device: K183040, Philips Medical Systems International B.V.
Trade/Device Name: Zenition 70 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB
#### 3. Reference Device:
The digital x-ray image acquisition panel employed has been cleared for the same intended use within: K200218 Trade/Device Name: Digiscan FDX Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, OXO
### 4. Indications for Use
The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
### 5. Description of the Device:
GARION is mobile fluoroscopy system is designed to provdie fluoroscopic and spot-film images of thepatient during diagnostic, surgical procedures.
This device is a digital X-ray radiographic equipment that consists of high voltage generator, X-ray control unit, X-ray tube unit, Collimator, image processing unit. The image processing unit consists of medical image detector, power supply, medical image collecting unit and relevant software.
{4}------------------------------------------------
- Fine-Tuned Movement
- 5kW FPD Surgical C-arm
• Garion is equipped with a high resolution FPD. Its composition, along with othe components such as positive/negative inversion, realtime zoom & rotation,as well as advanced noise reduction and optimized contrast/brightness, providing clear and enhanced image quality.
- Attachable Grid
The Grid can be attached and detached, allowing low-dose examination for each part.
- Dual Laser Pointer
Using the dual laser pointer, it is possible to set the patient's position without unnecessary exposure.
7. Comparison Table with the Predicate Device for technological characteristics:
· Powerful Digital Image, Processing and Display
The combination of Dynamic FPD means that Garion provides a excellent quality on screen image. Our advanced image-processing technology enables you to utilize various functions such as image flip (horizontal & vertical), image rotation (360°), realtime zoom, noise reduction, and realtime edge enhancement, etc.
6. Safety and Effectiveness, comparison to predicate device. This device has the same indications for use and very similar technological characteristics as the predicate device. This can be readily seen in the comparison table presented below. Many of the characteristics compared are the same, while other differences are insignificant in nature. The digital receptor panel chosen has already been cleared (reference device) for the same intended use.
| Detailed comparison chart | | | | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------------------------------|
| | Predicate device Zenition 70 K183040 | Proposed device Garion | SE | |
| Indications for Use | The Zenition 70 device is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. | The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. | SAME | |
| X-ray Tube | Anode Type<br>Rotating Anode | Rotating Anode | SAME | |
| | Anode target angle<br>10° | 10° | SAME | |
| | Focal size<br>0.3/0.6 | 0.3/0.6 | SAME | |
| Fluoroscopic Mode | kV range<br>40-120 kV | 40-125 kV | Similar.<br>Slightly<br>higher | |
| | | Predicate device Zenition 70 K183040 | Proposed device Garion | SE |
| | | | | peak<br>kVp |
| | mA range | 0.5-60 mA | 0.1-100 mA | Similar,<br>slightly<br>higher<br>mA |
| | Pulse<br>Fluoro | YES | YES | SAME |
| | ABS<br>function | YES | YES | SAME |
| FPD<br>Image<br>Detect-<br>or | Make | Trixell's PX2121S | IRAY 's, Mercu 0909F | -- |
| | Model | 8.15" x 8.15" | 9" x 9" | Similar,<br>slightly<br>larger<br>image. |
| | Scintilla-<br>tor | Cesium lodide | Cesium lodide | SAME |
| | Detector<br>type | amorphous silicon detector | amorphous silicon detector | SAME |
| | Active<br>detector<br>size | 207×207 mm | 228.6 mm x228.6 mm | Similar,<br>slightly<br>larger<br>image |
| | Total pixel<br>matrix | 1344 x 1344 | 1024×1024 | Similar |
| | Pixel pitch | 154 μm | 205 μm | Similar |
| | A/D<br>Convers-<br>ion | 16 bit | 16 bit | SAME |
| | MTF (1.0<br>lp/mm) | 0.59 | 0.64 | Similar,<br>slightly<br>better. |
| | DQE (0<br>lp/mm) | 0.77 | 0.77 | SAME |
| | Manu-<br>facturer | Philips | SG HealthCare Co. Ltd. | -- |
| | SID | 993mm | 980mm | Similar |
| C-arm | Range of<br>C-arm Rail<br>Rotation | ±200° | ±180° | Similar |
| | Range of<br>the Linear<br>FR-arm<br>Move-<br>ment | 200 mm | 200mm | SAME |
| | Range of<br>the Liner<br>T-arm<br>Move-<br>ment | 490 mm | 400 mm | Similar |
| | | Predicate device Zenition 70 K183040 | Proposed device Garion | SE |
| | Range of<br>swing-arm<br>Rotation | $\pm 10°$ | $\pm 15$ | Similar,<br>slightly<br>better. |
| Collimator | | Motor control / rotation | Motor control / rotation | SAME |
| Photo | | Image: Predicate device Zenition 70 K183040 | Image: Proposed device Garion | Similar |
# Detailed comparison chart
{5}------------------------------------------------
{6}------------------------------------------------
# 8. Summary of non-clinical testing: We employed applicable IEC standards:
IEC60601-1, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
IEC60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
IEC60601-1-3, Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
IEC60601-1-6, Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance -Collateral Standard: Usability
IEC60601-2-28 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
IEC60601-2-54 Medical electrical equipment Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
IEC62366 Medical device – Application of usability engineering to medical devices
In addition, the device complies with the FDA Radiation Safety Performance Standard for Diagnostic X-Ray Systems and Their Major Components (21CFR 1020.32, 1020.32, 1020.33); Small Entity Compliance Guide. In addition, labeling was developed and information provided in accordance with this FDA Guidance Document: Pediatric Information for X-ray Imaging Device Premarket Notifications, Guidance for Industry and Food and Drug Administration Staff. Labeling also includes reference to the Image Gently website (http://www.imagegently.org/). Also we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff
9. Summary of clinical testing: Not required for a finding of substantial equivalene..
10. Conclusion: After analyzing bench tests, and safety testing data, SG Healthcare concludes that the "Garion" is as safe and effective as the predicate device "Zenition 70" has few technological differences and has identical indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.