K220409 · Siemens Medical Solutions USA, Inc. · OWB · Jun 29, 2022 · Radiology
Device Facts
Record ID
K220409
Device Name
ARTIS pheno (VE21) System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Jun 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis. The ARTIS family include also the software option DynaCT with following indications for use: DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
Device Story
ARTIS pheno (VE21) is a multi-axis robotic angiography system for interventional radiology, cardiology, and hybrid surgery. System inputs include X-ray data acquired via C-arm, flat panel detector, and X-ray tube. The system transforms these inputs into 2D/3D images using integrated post-processing software. Operated by healthcare professionals in ORs or clinics, the system provides visualization via ceiling-mounted displays and control consoles. The system supports DynaCT for 3D reconstruction. Output allows clinicians to perform image-guided interventions, surgery, and treatment follow-up. Benefits include high-end imaging capabilities for complex procedures, including pediatric and obese patient populations. The system features flexible isocenter positioning and optional anti-microbial coating.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing, including electrical safety (AAMI ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and software verification and validation. Bench testing confirmed that all system specifications met acceptance criteria.
Technological Characteristics
Multi-axis robotic floor stand with C-arm, X-ray tube, flat panel detector, and high-voltage generator (Polydoros ACX). Connectivity includes integrated display systems and control consoles. Software version VE21 includes NOMSIE DSA (IQ) for contrast customization. Complies with IEC 62304 (software), ISO 10993-1 (biocompatibility), and various IEC 60601 series standards for medical electrical equipment. Optional anti-microbial coating.
Indications for Use
Indicated for single and biplane diagnostic imaging and interventional procedures, including pediatric and obese patients. Procedures include cardiac, neuro, general, rotational, and multipurpose angiography, and whole-body radiographic/fluoroscopic procedures. Supports image-guided surgery and spatial data synthesis. DynaCT option indicated for 3D reconstruction of 2D angiographic images for hard/soft tissue diagnosis, surgical planning, and treatment follow-up.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
June 29, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
Siemens Medical Solutions USA, Inc. % Patricia Jones Regulatory Affairs Professional 40 Liberty Boulevard MALVERN PA 19355
# Re: K220409
Trade/Device Name: ARTIS pheno (VE21) System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA Dated: June 15, 2022 Received: June 17, 2022
## Dear Patricia Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220409
Device Name ARTIS pheno (VE21) System
### Indications for Use (Describe)
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general anglography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for Xray imaging, surgery and interventions.
ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.
The ARTIS family include also the software option DynaCT with following indications for use:
DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots.
# 510(k) Summary: ARTIS pheno (VE21)
Siemens Medical Solutions USA, Inc. Company: 40 Liberty Boulevard, 65-1A Malvern, PA 19355
Date Prepared: June 28, 2022
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
### 1. General Information: Importer / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Reqistration Number: 2240869
## Manufacturing Site:
Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### Contact Person: 2.
Ms. Patricia D. Jones Requlatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65 Malvern, PA 19355 Phone: (678) 575-8832 Email: patricia.jones@siemens-Healthineers.com
- 3. Device Name and Classification:
Trade Name: Classification Name: Classification Panel: Regulation Number: Device Class: Product Codes:
# ARTIS pheno (VE21) Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II OWB. JAA
#### 4. Legally Marketed Primary Predicate Device
Trade Name: 510(k) Clearance Clearance Date Classification Name: Classification Panel: Requlation Number: Device Class:
ARTIS pheno (VE20) K201156 June 29, 2020 Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots arranged in a circular pattern.
Product Code: Subsequent Product Code : OWB, JAA
Total Product Life Cycle:
All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.
#### 5. Device Description:
The ARTIS pheno (VE21) system is a multi-axis system, specifically designed to meet the growing demands of high-end imaging for interventional radiology, interventional cardiology, minimally invasive and hybrid surgery procedures. The stand allows positioning in angular, orbital, lateral, longitudinal, and vertical directions, leveraging the flexible isocenter. The ARTIS pheno (VE21) system is equipped with a robotic multi-axis floor stand, C-arm, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post processing. The ARTIS pheno is partially coated with an optional anti-microbial coating. syngo Application Software is optional, available for the support of dedicated clinical workflows.
The ARTIS pheno (VE21) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical angiography, multipurpose angiography, rotational angiography, and radiographic/fluoroscopic procedures.
The following components are configured to create the floor configuration:
- (1) Floor stand with C-arm, X-ray tube assembly and FD
- (2) Patient table
- (3) Display ceiling suspension with displays
- (4) Footswitch for releasing radiation
(5) Control console for controlling the stand, patient table, collimator, and imaging system
Images and operating elements are displayed on screens. Different display variants are used to visualize image and information content. Panoramic display configurations or large displays can be used, configurable to visualize multiple images and information content in various layouts.
Post processing can be done in the exam room or in the control room that offers monitors as well, with a footswitch location in the exam room or the control room. The ARTIS pheno (VE21) System is capable of 2D and 3D imaging.
Other systems and software syngo Application Software, syngo X Workplace, Sensis, and or third-party systems may also be integrated into the ARTIS pheno (VE21) screen configuration. Different screen configurations and layouts are possible in the examination room and in the control room.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
The 510(k) submission, Subject Device "ARTIS pheno with software version VE21" will support the following modifications made to the Subject Device in comparison to the Predicate Device:
| Table 1. | Modifications for ARTIS pheno (VE21) System |
|----------|---------------------------------------------|
|----------|---------------------------------------------|
| Software/Hardware modifications specific to New System Software VE21 | | |
|----------------------------------------------------------------------|--|--|
| 1. Updated system Software from VE20 to VE21 | | |
| A. NOMSIE DSA (IQ Customization of Contrast in DSA images | | |
| B. Optional New Generator | | |
| 2. Updated 510(k) Information | | |
### 6. Indications for Use:
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography, and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e., patient extremities. This does not include projection radiography.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery, and interventions.
ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.
The ARTIS family also include the software option DynaCT with following indications for use:
DynaCT is an X-ray imaging software option, which allows the reconstruction of twodimensional images acquired with a standard angiographic C-arm device into a threedimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures, and treatment follow-up.
#### Substantial Equivalence: 7.
The ARTIS pheno (VE21) system is substantial equivalent to the legally marketed predicate listed in the Table 2 below:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.
| Predicate Device Name<br>and Manufacturer | 510(k) Number | Clearance<br>Date | Comparable Properties |
|------------------------------------------------|---------------|-------------------|---------------------------------------------------------------------------------------|
| <b>Primary Predicate</b><br>ARTIS pheno (VE20) | K201156 | 06/29/2020 | • Indications for use<br>• Software Version<br>VE20<br>• Image Quality<br>• Generator |
Table 2: Predicate Device Comparable Properties for Subject Devic
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The ARTIS pheno (VE21) System is designed as a set of components (floor stand, C-arm, X-ray tube and housing, flat panel detector, digital imaging system, collimator, generator etc.) that is combined to provide a specialized angiography system. Components used with ARTIS pheno (VE21) System are either commercially available with current Siemens systems or include modifications to existing components. Technological differences between the Subject Device and the Predicate Device is provided in the Table 3 below for all modifications.
| Modifications | Subject Device<br>ARTIS pheno (VE21) | Predicate Device<br>ARTIS pheno (VE20)<br>K201156 | Comparison<br>Results |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| New System<br>Software<br>Changes | 1. Updated system Software<br>from VE20 to VE21<br>A. NOMSIE DSA (IQ)<br>Customization of<br>Contrast in DSA<br>Images<br>B. Optional New<br>Generator<br>(Polydoros ACX) | System Software Version<br>VE20<br>Classical edge enhancement<br>with filters and automatic<br>windowing was used.<br>Generator (Polydoros<br>A100G) | Comparable:<br>System software VE21 was<br>updated to support<br>NOMSIE DSA (IQ)<br>Customization of Contrast<br>in DSA Images and a new<br>Generator. System<br>Software modifications<br>conforms to "Guidance for<br>the content of Premarket<br>submission for software in<br>Medical Devices". Provided<br>is all required software<br>testing. Provided in this<br>submission are Bench Test<br>Summaries. These<br>software and hardware<br>changes do not raise any<br>new risks or any issues<br>regarding the safety nor<br>effectiveness of the device.<br>All test results met all<br>acceptance criteria. |
| Update 510(k)<br>Information | 2. Updated 510(k) information for the Primary Predicate Device is provided. | | |
Table 3: Summary of Comparison of Technological Characteristics
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.
#### 9. Nonclinical Performance Testing:
Non-clinical tests were conducted for the ARTIS pheno (VE21) during product development.
The ARTIS pheno (VE21) was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance, and Electromagnetic Compatibility:
- AAMI ANSI ES60601-1:2005/(R)2012 ●
- IEC 60601-1-2:2014
- IEC 60601-1-3:2013
- IEC 60601-1-6:2010/A1:2013 ●
- IEC 60825-1:2007 .
- TR 60878:2015
- IEC 62304:2015
- IEC 80001-1:2010 ●
- IEC 60601-2-28:2017
- IEC 60601-2-43:2017
- IEC 60601-2-54:2009/A1:2015 ●
- ISO 10993-1:2009 ●
- ISO 14971:2007 ●
- IEC 62366-1:2015 ●
The modifications described in this Premarket Notification are supported with verification and validation testing.
## Verification and Validation:
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on ARTIS pheno System software (VE21) during product development.
The Risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.
ARTIS pheno System software (VE21) was tested and found to be safe and effective for intended users, uses, and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual. Customer employees are adequately trained in the use of this equipment.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a circular pattern.
Siemens conforms to the cybersecurity requirementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 is the hospital. Provided in the Software Section is the required cybersecurity information.
## Summary:
Performance tests were conducted to test the functionality of ARTIS pheno (VE21) System. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness.
### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
## 11. Conclusion as to Substantial Equivalence:
The predicate devices were cleared based on non-clinical supportive information and clinical images and data. Similar non-clinical test results demonstrate that the ARTIS pheno (VE21) System acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Devices that is currently marketed for the same intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.