ARCO FP-S is a mobile X-ray device with a Flat Panel detector used for radiological guidance and visualization during diagnostic, interventional and surgical procedures. In particular, the device is to be applied during othopaedic, neuro, abdominal, vascular, thoracic and cardiac procedures. ARCO FP-S device can be used on all patients except paediatric patients within the limit of the device.
Device Story
Mobile C-arm X-ray system for fluoroscopy and radiography; used in OR/clinic for surgical guidance. Inputs: X-ray attenuation through patient tissue; detected by flat panel detector (PaxScan 2121DXV or 3030DXV). System generates X-rays (5 kW generator, rotating anode tube); converts to electrical signals for image processing. Output: Real-time fluoroscopic or static radiographic images displayed on 27" Quad HD monitor. Operated by clinicians (physicians/technicians) via control panel, footswitch, or remote. Supports DSA and Road Mapping modes. Benefits: Provides high-quality visualization for minimally invasive procedures; aids clinical decision-making through real-time anatomical guidance. Includes wireless monitor trolley for image duplication and DICOM/PACS connectivity.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety, EMC, and radiologist evaluation of images acquired by the new digital receptor panels to confirm diagnostic quality.
Technological Characteristics
Mobile C-arm; 5 kW rotating anode X-ray tube (Rhenium-Tungsten-Molybdenum); PaxScan 2121DXV/3030DXV flat panel detectors; 27" Quad HD LCD monitor; Windows 10 OS. Connectivity: DICOM, Wi-Fi, Ethernet, NFC login. Sterilization: N/A (surface cleaning). Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-43, 60601-2-54, IEC 60825-1, IEC 62304, ETSI EN 301 489-17.
Indications for Use
Indicated for radiological guidance and visualization during diagnostic, interventional, and surgical procedures (orthopedic, neuro, abdominal, vascular, thoracic, cardiac). For all patients except pediatric patients.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System (K200022)
Submission Summary (Full Text)
{0}------------------------------------------------
August 13, 2021.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
A.T.S. Applicazione Tecnologie Speciali S.R.L % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
## Re: K211777
Trade/Device Name: ARCO FP-S Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO Dated: June 2, 2021 Received: June 9, 2021
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211777
Device Name ARCO FP-S
Indications for Use (Describe)
ARCO FP-S is a mobile X-ray device with a Flat Panel detector used for radiological guidance and visualization during diagnostic, interventional and surgical procedures.
In particular, the device is to be applied during othopaedic, neuro, abdominal, vascular, thoracic and cardiac procedures. ARCO FP-S device can be used on all patients except paediatric patients within the limit of the device.
Type of Use (Select one or both, as applicable)
| | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--|-------------------------------------------------------------|
| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K211777
Image /page/3/Picture/1 description: The image shows a logo with the letters "ATS" in a stylized, bold, sans-serif font. The letters are a dark blue color against a light blue background. The letters are connected to each other, forming a single, unified shape.
## A.T.S. APPLICAZIONE TECNOLOGIE SPECIALI S.R.L.. VIA A. VOLTA, 10 -TORRE DE' ROVERI (BG) 24060 ITALY -TEL. +39.035.584311 -FAX +39.035.580220 Date Prepared: July 19, 2021
#### 1. Administrative Information
Submitter:
Submission contact person: Eng. Livia PILLITTERI, QMS & Regulatory Affairs Manager
#### Identification:
| Identification: | ARCO FP-S |
|--------------------|---------------------------------------------|
| Regulation Number: | 21 CFR 892.1650 |
| Regulation Name: | Image-intensified fluoroscopic x-ray system |
| Regulatory Class: | Class II |
| Product Code: | OWB, OXO |
This is a MODIFIED device. The digital x-ray receptor panels have changed.
Substantially equivalent device: K182086
| Manufacturer: | A.T.S. APPLICAZIONE TECNOLOGIE SPECIALI S.R.L. |
|--------------------|------------------------------------------------|
| Trade/Device Name: | ARCO FP |
| Regulation Number: | 21 CFR 892.1650 |
| Regulation Name: | Image-intensified fluoroscopic x-ray system |
| Regulatory Class: | Class II |
| Product Code: | OWB, OXO |
REFERENCE Device: (New digital x-ray receptor panels) K200022 submitted by Livermoretech Trade/Device Name: FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA, IZI
- 2. Indications for Use: ARCO FP-S is a mobile X-ray device with a Flat Panel detector used for radiological guidance and visualization during diagnostic, interventional and surgical procedures. In particular, the device is to be applied during orthopedic, neuro, abdominal, vascular, thoracic and cardiac procedures. ARCO FP-S device can be used on all patients except pediatric patients within the limit of the device.
- 3. Device description: ARCO FP-S is a mobile C-arm with the solid-state image detector "Flat Panel" It will be realized in 2 models, depending on the size of the detector: Model: ARCO FP-SR21
{4}------------------------------------------------
Dual focus 0.3/0.6 mm, rotating anode X-Ray tube, generator nominal power 5 KW. Detector size 21x21 cm, 1024x1024 pixel.
## Models: ARCO FP-SR30
Dual focus 0.3/0.6 mm, rotating anode X-Ray tube, generator nominal power 5 KW. Detector size 30x30 cm, 1536 x 1536 pixel.
Image /page/4/Picture/3 description: The image shows two medical devices, a C-arm and a mobile monitor stand. The C-arm is a medical imaging device used in surgical procedures, and it has a large, C-shaped arm that houses an X-ray tube and detector. The image also has labels indicating the stand, footswitch, and remote control. The mobile monitor stand is a device used to display medical images during surgical procedures.
Photo: ARCO FP-S
OPTIONAL STAND: Wireless Monitor
The system consists of:
# C-arm Stand:
- MECHANICAL STAND
- X-RAY MONOBLOC, rotating anode
- FLAT PANEL DETECTORS: 2121cm or 3030cm
- MONITOR LCD 27" QUAD HD format 16:9
- CONTROL PANEL
- X-RAY COLLIMATOR
- SYSTEM CONTROLLER
- X-RAY HAND SWITCH
- ARTICULATED ARM FOR MONITOR SUPPORT
- ANTISCATTERING GRID
## Options:
- FOOTSWITCH, two-foot pedal
- LASER LOCALISER
- Dose Area Meter,
- DSA Packages
{5}------------------------------------------------
- I.R. remote controls.
- DICOM Classes
- EXTERNAL VIDEO OUTPUT-Wireless transmission
- DICOM DATA Wi-Fi transmission to Ethernet/PACS
- NFC LOGIN
- Wireless monitor trolley, model TROTTER W, for duplicating the image of the monitor on board.
# Technology Employed
ARCO FP-S is a mobile C-arm unit for fluoroscopy and radiography designed for a wide rangeof surgical applications.
The normal use allows for imaging in the following modes:
- Low Dose Fluoroscopy -
- -High Quality Fluoroscopy
- Digital radiography (Snapshot) l
- -Fluoroscopy in Road Mapping mode (optional)
- -Fluoroscopy in DSA mode (optional)
The image acquisitions employ X-rays emitted by an X-ray generator which can produce upto 5 kW. The X-rays, attenuated in different ways depending on the tissue they have to pass through, leave an impression on a detector and then they are converted into electrical signal, which forms the image. The main differences from the predicate device are:
The image detectors are new. The Pixium detectors (used in the predicate device) were replaced with Paxscan (used with the subject device, cleared under K200022).
Only the 5 kW power generator is now available.
The software has been upgraded from the predicate software functionality remains the same. The original software (SYSTEMA dRF- software ISIX) and the modified software (SYSTEMA DRF-S software) are FUNCTIONALLY IDENTICAL.
The workstation image display monitor is being upgraded to a higher resolution.
The operating system is upgraded (Windows 7 is no longer supported).
| Comparison Chart | PROPOSED DEVICE<br>ARCO FP-S | PREDICATE DEVICE<br>ARCO FP K182086 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATIONS FOR USE | ARCO FP-S is a mobile X-ray device used for<br>radiological guidance and visualization<br>during diagnostic, interventional and<br>surgical procedures. In particular, the<br>device is to be applied during orthopedic,<br>neuro, abdominal, vascular,thoracic and<br>cardiac procedures. ARCO FP-S device can<br>be used on all patients except pediatric<br>patients, within the limits ofthe device. | ARCO FP is a mobile X-ray device used for<br>radiological guidance and visualization<br>during diagnostic, interventional and<br>surgical procedures. In particular, the<br>device is to be applied during orthopedic,<br>neuro, abdominal, vascular,thoracic and<br>cardiac procedures. ARCO FP device can be<br>used on all patients except pediatric<br>patients, within the limits ofthe device. |
| CLASS | 2 | 2 |
# 4. Technological characteristics: Comparison Table
{6}------------------------------------------------
| Comparison Chart | PROPOSED DEVICE<br>ARCO FP-S | PREDICATE DEVICE<br>ARCO FP K182086 |
|--------------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Product code | OWB; OXO | OWB: OXO |
| Photo | Image: PROPOSED DEVICE ARCO FP-S | Image: PREDICATE DEVICE ARCO FP K182086 |
| PERFORMANCE SPECIFICATION | | |
| A. C-arm | | |
| C-Arm Weight | 310 kg | 280 kg |
| B. Power Supply | | |
| Supply type | Single phase (live/neutral,separate<br>earth) | Single phase (live/neutral,separate earth) |
| Input Voltage | 100, 110, 120, 130, 210, 220, 230, 240 V<br>@50-60Hz | 100, 110, 120, 130, 210, 220, 230, 240 V<br>@50-60Hz |
| X-ray Generator<br>Maximum power | 5kW | 5kW; 20kW |
| C. X-ray tube | | |
| Type | Rotating anode | Rotating anode |
| Nominal voltage | 120 kV | 120 kV |
| Nominal focal<br>spot valuesinput<br>power | 0.3 mm = 6.0kW<br>0.6 mm = 25 kW | 0.3 mm = 6.0kW<br>0.6 mm = 25 kW |
| Maximum heat<br>dissipation | 750 W | 750 W |
| Maximum<br>anode heat | 225 kJ | 225 kJ |
{7}------------------------------------------------
| Comparison Chart | PROPOSED DEVICE<br>ARCO FP-S | PREDICATE DEVICE<br>ARCO FP K182086 | | | |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------|---------------------------------|--|
| content | | | | | |
| Anode material | Rhenium-Tungsten-Molybdenum | Rhenium-Tungsten-Molybdenum | | | |
| Filtration | Possible Filters To Apply:<br>- none<br>- 2mmAl<br>- 1mmAl + 0,1mmCu<br>- 1mmAl + 0,2mmCu | Possible Filters To Apply:<br>- none<br>- 2mmAl<br>- 1mmAl + 0,1mmCu<br>- 1mmAl + 0,2mmCu | | | |
| D. Detector | | | | | |
| | PaxScan<br>2121DXV | PaxScan<br>3030DXV | PIXIUM 2121S | PIXIUM 3030S | |
| Active<br>detector<br>size /<br>Sensitive<br>area | 1024x1024 pixel<br>205x205 mm | 1516x1516 pixel<br>294x294 mm | 1344 x 1344 pixel<br>207x207 mm | 1956 x 1956 pixel<br>301x301 mm | |
| Pixel size | 205 μm | 194 μm | 154 μm | 154 μm | |
| Cooling | Passive | | Passive | | |
| Detective | 0 –80% | 0 –80% | 0 –77% | 0 –77% | |
| Quantum<br>Efficiency,<br>RQA5<br>[ Lp/mm@2uGy ] | 1.0 - 65% | 1.0 - 65% | 1.0 - 56% | 1.0 – 57% | |
| | 2.0 - 40% | 2.0 - 44% | 2.0 - 46% | 2.0 - 48% | |
| Modulation<br>Transfer<br>Efficiency<br>(MTF)<br>[ Lp/mm@2uGy ] | 1.0 - 50% | 1.0 – 55% | 1.0 – 59% | 1.0 – 59% | |
| | 2.0 – 22% | 2.0 – 23% | 2.0 – 29% | 2.0 – 29% | |
| | Nyquist - 15% | Nyquist - 15% | Nyquist-11% | Nyquist-11% | |
| Maximum frame per<br>second | 30 | | 25 | | |
| | E. Anti-Scattering Grid | | | | |
| Material | Aluminum | | Aluminum | | |
| Lines/cm | 80 | | 80 | | |
| Grid focus distance | 100 cm | | 100 cm | | |
{8}------------------------------------------------
| Comparison Chart | PROPOSED DEVICE<br>ARCO FP-S | PREDICATE DEVICE<br>ARCO FP K182086 |
|---------------------------------|--------------------------------------------------|-------------------------------------|
| Ratio | 8:1 | 8:1 |
| F. Workstation monitor | | |
| Size / Type | 21.7" / LCD | 21.5" / LCD |
| Resolution | 2560 x 1440 (4K) | 1920 x 1080 |
| Nominal light output | 350 cd/m² | 1000 cd/m² |
| G. Control panel monitor | | |
| Size / Type | 12.5" / LCD touch | 12.5" / LCD touch |
| Resolution | 1920 x 1080 | 1920 x 1080 |
| H. Workstation Operating system | | |
| Windows version | Windows 10 (Windows 7 is no longer<br>supported) | Windows 7 or Windows 10 |
5. Non clinical testing: In order to demonstrate the EMC and Electrical safety of the proposed device ARCO FP-S, the manufacturer A.T.S. Applicazione Tecnologie Speciali Srl. has performed safety tests incompliance with the standards listed here below:
- IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basicsafety and essential performance;
- IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements forbasic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests;
- IEC 60601-1-3: Medical Electrical Equipment - Part 1-3: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation ProtectionIn Diagnostic X-Ray Equipment;
- IEC 60601-1-6: Medical Electrical Equipment - Part 1-6: General Requirements ForBasic Safety And Essential Performance - Collateral Standard: Usability;
- IEC 60601-2-43: Medical Electrical Equipment - Part 2-43: Particular RequirementsFor The Safety And Essential Performance Of X-Ray Equipment For Interventional Procedures;
- IEC 60601-2-54: Medical Electrical Equipment - Part 2-54: Particular RequirementsFor The Basic Safety And Essential Performance Of X-Ray Equipment For Radiography And Radioscopy.
- IEC 60825-1: Safety of laser products - Part 1: Equipment classification andrequirements
- IEC 62304: Medical device software - Software life cycle processes
- ETSI EN 301 489-17: ElectroMagnetic Compatibility (EMC) standard for radio equipment and services - Part 17: Specific conditions for Broadband Data Transmission Systems; Harmonised Standard for Electro Magnetic Compatibility
In order to demonstrate the performance of the ARCO FP-S, the following quality testing has been performed by A.T.S. Applicazione Tecnologie Speciali Srl:
- Radiologist's Images evaluation. A Board Certified Radiologist reviewed the still and moving images ●
{9}------------------------------------------------
acquired by the newer digital receptor panels and found them to be of good and acceptable diagnostic quality.
The software remains at a MODERATE LEVEL OF CONCERN, and software was validated according to the FDA Software Guidance at that level.
The following FDA guidance was employed in regard to cybersecurity concerns: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff. Our labeling and internal procedures were updated to reflect the recommendations of this guidance.
6. Clinical testing. Clinical Images are not necessary to establish substantial equivalence to the predicate. Adequate bench testing results should clearly indicate device safety and effectiveness.
# 7. Substantial Equivalence Discussion.
The ARCO FP-S performs the same functions as the predicate using the same technological methods to produce and store diagnostic x-ray images. In all material aspects, the A.T.S. APPLICAZIONE TECNOLOGIE SPECIALI S.R.L. devices are substantially equivalent to each other. The detector technological characteristics (resolution, MTF and DQE performance, see comparison table above) all point to substantial equivalence
# 8. Substantial Equivalence Conclusion:
After analyzing bench test results, risk analysis, and standards compliance, it is the conclusion of A.T.S. APPLICAZIONE TECNOLOGIE SPECIALI S.R.L.. that the ARCO FP-S is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.