K200663 · Controlrad, Inc. · OWB · Jun 24, 2020 · Radiology
Device Facts
Record ID
K200663
Device Name
ControlRad Trace Model 9
Applicant
Controlrad, Inc.
Product Code
OWB · Radiology
Decision Date
Jun 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The ControlRad™ Trace Model 9, when used with OEC® 9900 Elite, is indicated to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures while reducing patients and clinicians' radiation exposure (Dose Area Product) outside of the ControlRad Trace's region of interest (ROI) as compared to OEC® 9900 Elite non-collimated image area. 1 The ControlRad Trace semi-transparent filter should not be used in lieu of the OEC® 9900 Elite's collimators, as they block the most radiation, but can help physicians balance dose visualize structures outside the ROI when it is considered clinically advantageous to do so. Clinications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Device Story
System consists of tablet, semi-transparent filter, hardware, and software modules installed on GE OEC 9900 Elite fluoroscopic X-ray system. Clinician selects Region of Interest (ROI) via tablet GUI; system controls filter positioning to reduce radiation outside ROI while maintaining full image quality within ROI. Filter provides 61-97% radiation reduction in peripheral areas. Used in OR/clinic by physicians to balance dose reduction with peripheral visualization. Output includes live fluoroscopic images with ROI overlay and DAP metrics. Benefits include reduced radiation exposure for patients and clinicians.
Clinical Evidence
Bench testing only. Evaluated Air Kerma impact, DAP calculation accuracy (within ±35%), DAP reduction (≥50% at 50kVp), filter attenuation, and comparative image quality (inside/outside ROI). Clinical simulations confirmed peripheral context utility. Usability testing verified touchscreen operation with sterile gloves. Mechanical testing confirmed compliance with IEC 60601-1 for instability.
Technological Characteristics
System integrates with GE OEC 9900 Elite. Components: Tablet, semi-transparent filter, hardware/firmware modules, communication interface. Filter reduces radiation outside ROI by 61-97%. Connectivity: Wireless/wired communication with host X-ray system. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 62304, ISO 14971, 21 CFR 1020.30-32.
Indications for Use
Indicated for patients undergoing diagnostic, surgical, and interventional fluoroscopic and spot-film imaging procedures, including cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
June 24, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
ControlRad, Inc. % Linda Braddon, Ph.D. Regulatory Consultant Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 WOODSTOCK GA 30188
Re: K200663
Trade/Device Name: ControlRad™ Trace Model 9 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: April 9, 2020 Received: April 10, 2020
Dear Dr. Braddon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K200663
Device Name
#### ControlRad™ Trace Model 9
Indications for Use (Describe)
The ControlRad™ Trace Model 9, when used with OEC® 9900 Elite, is indicated to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures while reducing patients and clinicians' radiation exposure (Dose Area Product) outside of the ControlRad Trace's region of interest (ROI) as compared to OEC® 9900 Elite non-collimated image area. 1 The ControlRad Trace semi-transparent filter should not be used in lieu of the OEC® 9900 Elite's collimators, as they block the most radiation, but can help physicians balance dose visualize structures outside the ROI when it is considered clinically advantageous to do so. Clinications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
1 Relative to open Field of View (FOV), the ControlRad Trace Model 9 reduces at least 50% of the Dose Area Product at 50 kVp and ROI with width and length that are smaller than 1/3 the diameter of the full FOV.
Type of Use (Select one or bothas applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ControlRad. The logo is in green and white. The word "CONTROL" is in dark green, and the word "RAD" is in light green. There is a green circle around the "RAD" part of the logo.
# 510(k) SUMMARY
# ControlRad, Inc's ControlRadTM Trace Model 9
## K200663
| Applicant's name: | ControlRad, Inc.<br>275 Scientific Dr NW<br>Suite 1100<br>Norcross, Georgia 30092, USA<br>P: 1-800-522-5148 |
|-------------------|-------------------------------------------------------------------------------------------------------------|
|-------------------|-------------------------------------------------------------------------------------------------------------|
- Contact Person: Linda Braddon, Ph.D. Secure BioMed Evaluations 7828 Hickory Flat Hwy Suite 120 Woodstock, GA 30188 770-837-2681 Regulatory@SecureBME.com
Date Prepared: June 15, 2020
# Subject Device:
| Device Name | ControlRad™ Trace Model 9 |
|---------------------|---------------------------------------------|
| Common Name | Interventional Fluoroscopic X-ray System |
| Regulation | 21 CFR §892.1650 |
| Classification Name | Image-intensified fluoroscopic x-ray system |
| Class | II |
| Panel | Radiology |
| Product Code(s) | Primary: OWB<br>Secondary: OXO, JAA |
## Predicate Device:
| Device Name | ControlRad™ Trace Model 8 (K183109) |
|---------------------|---------------------------------------------|
| Common Name | Interventional Fluoroscopic X-ray System |
| Regulation | 21 CFR §892.1650 |
| Classification Name | Image-intensified fluoroscopic x-ray system |
| Class | II |
| Panel | Radiology |
| Product Code(s) | Primary: OWB<br>Secondary: OXO, JAA |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ControlRad. The logo consists of the word "CONTROL" in dark teal, followed by "RAD" in a lighter green. To the right of "RAD" is a circular design with a dark teal lower half and a light green upper half. The letters "RAD" are integrated into the circular design, and there is a trademark symbol to the upper right of the logo.
#### Reference Device:
GE Healthcare Surgery, OEC® 9900 Elite (K122234) Device Name Common Name Interventional Fluoroscopic X-ray System Regulation 21 CFR §892.1650 Classification Name Image-intensified fluoroscopic x-ray system Class II Radiology Panel Product Code(s) Primary: OWB Secondary: OXO, JAA
## Indications for Use
The ControlRad™ Trace Model 9, when used with OEC® 9900 Elite, is indicated to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures while reducing patients and clinicians' radiation exposure (Dose Area Product) outside of the ControlRad Trace's region of interest (ROI) as compared to OEC® 9900 Elite noncollimated image area.1 The ControlRad Trace semi-transparent filter should not be used in lieu of the OEC® 9900 Elite's collimators, as they block the most radiation, but can help physicians balance dose reduction with the need to visualize structures outside the ROI when it is considered clinically advantageous to do so. Clinical applications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
1 Relative to open Field of View (FOV), the ControlRad Trace Model 9 reduces at least 50% of the Dose Area Product at 50 kVp and ROI with width and length that are smaller than 1/3 the diameter of the full FOV.
#### Device Description
The ControlRad™ Trace Model 9 consists of the following main components: ControlRad Trace Tablet, ControlRad Trace Filter, ControlRad Hardware, ControlRad Software and Firmware Modules and ControlRad Communication Interface all installed on the OEC 9900 Elite. The ControlRad™ Trace Model 9 is a system used to assist trained clinicians which is used to provide X-ray images when the clinician performs a medical procedure while reducing the patients and clinicians' radiation exposure (Dose Area Product) outside of the ControlRad Trace's region of interest (ROI) as compared to the OEC 9900 Elite non-collimated image area. The ControlRad™ Trace Model 9 can help physicians balance dose reduction with the need to visualize structures outside the ROI when it is considered clinically advantageous to do so.
#### Technological Characteristics
The ControlRad™ Trace Model 9 consists of the following main components: ControlRad Trace Tablet, ControlRad Trace Filter, ControlRad Hardware, ControlRad Software and Firmware Modules and ControlRad Communication Interface all installed on the OEC 9900 Elite.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ControlRad. The logo consists of the word "CONTROL" in a bold, sans-serif font, followed by the word "RAD" in a similar font but with a green color. To the right of "RAD" is a stylized green circle, partially enclosing the letters "RAD", with a darker green section at the bottom. There is a trademark symbol to the right of the word "RAD".
The ControlRad" Trace Model 9 components are installed semi-permanently in the cleared GE Healthcare Surgery's OEC® 9900 Elite (K122234) and operate in parallel to the GE Healthcare Surgery's OEC® 9900 Elite. The removal of the ControlRad™ components will restore the device to OEM specifications.
The ControlRad™ Trace Model 9 components provide the following functionalities:
- The CR Trace Tablet provides the user operational control of the ControlRad™ Trace . Model 9 device via a Graphical User Interface ("GUI"). The CR Trace Tablet enables the clinician to select a Region of Interest ("ROI") on the image displayed on the CR Trace Tablet, which is the same image that is displayed on the GE Healthcare Surgery's OEC® 9900 Elite's live monitor.
- . The CR Trace Filter is installed on top of the GE Healthcare Surgery's OEC® 9900 Elite's collimator. The CR Trace Filter does not affect or modify the functionality of the collimator. The CR Trace Filter is a semi-transparent filter which reduces the X-ray radiation outside the clinician-selected ROI, typically by 61% to 97%. The actual dose reduction achieved will depend upon specific imaging parameters such as OEC collimator settings, the kVp and the percentage of the OEC non-collimated image covered by the ControlRad Trace Filter.
- The ControlRad Hardware, Software and Firmware Modules control the ControlRad Trace . Filter's positioning, which is determined by the location of the clinician-selected ROI and perform image processing.
- The ControlRad Communication Interface provides communication between the various . components of the ControlRad™ Trace Model 9 and between the ControlRad™ Trace Model 9 and the GE Healthcare Surgery's OEC® 9900 Elite.
## Principles of Operation
The GE Healthcare Surgery's OEC® 9900 Elite provides an image that its boundaries are defined by the OEC® 9900 Elite's collimator, i.e. the image FOV is defined by the OEC non-collimated region. The image FOV size is not affected or modified by the ControlRad™ Trace Model 9.
Within the OEC® 9900 Elite non-collimated image region, when using a clinician selected Region of Interest ("ROI") on the ControlRad Trace Tablet, the ControlRad Trace Filters reduce radiation exposure outside the ROI. The resulting image has two parts:
- . The image inside the clinician-selected ROI (unfiltered radiation area in the FOV), which has at least the same image quality in the ROI as the GE Healthcare Surgery's OEC® 9900 Elite (K122234); and
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ControlRad. The logo consists of the word "CONTROL" in a dark teal sans-serif font, followed by "RAD" in a lighter green color. To the right of "RAD" is a circular design with two shades of green, creating a stylized "C" shape. The letters and the circular design are aligned horizontally.
- . The image outside the clinician-selected ROI (filtered radiation area in the FOV), a lowerdose processed image which provides peripheral image context to the ROI.
The GE Healthcare Surgery's OEC® 9900 Elite's collimator, when used, also reduces radiation emission. However, that collimator reduces radiation emission by blocking the delivery of radiation to the area covered by the collimator. As a result, the GE Healthcare Surgery's OEC® 9900 Elite's image FOV is limited to the non-collimated region. The ControlRad Trace Filter can be used along with the GE Healthcare Surgery's OEC® 9900 Elite's collimator to further reduce radiation emissions, and the additional radiation reduction provided by the ControlRad Trace Filter will be outside the clinician-selected ROI and within the un-collimated region/image FOV.
The clinician has the option not to use the CR Trace Filter. In this case, the GE Healthcare Surgery's OEC® 9900 Elite operates as if the CR Trace Filter was not present.
## Comparison of Technological Characteristics with the Predicate devices
The ControlRad™ Trace Model 9 for use with GE Healthcare Surgery's OEC® 9900 Elite has the same intended use and the same indications for use as the cleared predicate. The ControlRad™ Trace Model 9 is identical in construction to the predicate with the only modifications being the design differences to work with OEC® 9900 Elite. The performance data demonstrates that the ControlRad™ Trace Model 9 is at least as safe and effective as the predicate and reference devices and is substantially equivalent to the predicate and reference devices. A comparison table of technological characteristics of the ControlRad™ Trace Model 9 device for use with OEC® 9900 Elite compared to those of the predicates is provided below:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ControlRad. The word "CONTROL" is in a bold, sans-serif font in a dark teal color. To the right of the word "CONTROL" is the word "RAD" in a lighter green color, with a circle around it that is also green. The letters "RAD" are also in a bold, sans-serif font.
| Device<br>Feature<br>510(k) Number | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 9<br>(Subject Device)<br>K200663 | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 8<br>(K183109)<br>(Predicate Device)<br>K183109 | GE Healthcare<br>Surgery's<br>OEC 9900 Elite<br>(K122234)<br>(Reference)<br>K122234 | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Device Class | Class II | Class II | Class II | |
| Product Codes | Primary: OWB<br>Secondary: OXO, JAA | Primary: OWB<br>Secondary: OXO, JAA | Primary: OWB<br>Secondary: OXO, JAA | |
| Regulation Number | 21 CFR §892.1650 | 21 CFR §892.1650 | 21 CFR §892.1650 | |
| Indications for use | The ControlRad™ Trace Model<br>9, when used with OEC® 9900<br>Elite, is indicated to provide<br>fluoroscopic and spot-film<br>imaging of the patient during<br>diagnostic, surgical, and<br>interventional procedures while<br>reducing patients and clinicians'<br>radiation exposure (Dose Area<br>Product) outside of the<br>ControlRad Trace's region of<br>interest (ROI) as compared to<br>OEC® 9900 Elite non-<br>collimated image area.1 The<br>ControlRad Trace semi-<br>transparent filter should not be<br>used in lieu of the OEC® 9900<br>Elite's collimators, as they<br>block the most radiation, but<br>can help physicians balance<br>dose reduction with the need to<br>visualize structures outside the<br>ROI when it is considered<br>clinically advantageous to do<br>so. Clinical applications may<br>include cholangiography.<br>endoscopic, urologic,<br>orthopedic, neurologic,<br>vascular, cardiac, critical care<br>and emergency room<br>procedures.<br>1Relative to open Field of View<br>(FOV), the ControlRad Trace<br>Model 9 reduces at least 50% of<br>the Dose Area Product at 50<br>kVp and ROI with width and<br>length that are smaller than 1/3<br>the diameter of the full FOV. | The ControlRad™ Trace Model<br>8, when used with OEC®<br>9800/OEC® 9800 Plus, is<br>indicated to provide<br>fluoroscopic and spot-film<br>imaging of the patient during<br>diagnostic, surgical, and<br>interventional procedures while<br>reducing patients and clinicians'<br>radiation exposure (Dose Area<br>Product) outside of the<br>ControlRad Trace's region of<br>interest (ROI) as compared to<br>the OEC® 9800/OEC® 9800<br>Plus non-collimated image<br>area.1 The ControlRad Trace<br>semi-transparent filter should<br>not be used in lieu of the<br>OEC® 9800/OEC® 9800 Plus'<br>collimators, as they block the<br>most radiation, but can help<br>physicians balance dose<br>reduction with the need to<br>visualize structures outside the<br>ROI when it is considered<br>clinically advantageous to do<br>so. Clinical applications may<br>include cholangiography,<br>endoscopic, urologic,<br>orthopedic, neurologic,<br>vascular, cardiac, critical care<br>and emergency room<br>procedures.<br>1Relative to open Field of View<br>(FOV), the ControlRad Trace<br>Model 8 reduces at least 50%<br>of the Dose Area Product at 50<br>kVp and ROI with width and<br>length that are smaller than 1/3<br>the diameter of the full FOV. | The OEC® 9900 Elite is<br>designed to provide<br>fluoroscopic and spot-film<br>images of the patient during<br>diagnostic, surgical and<br>interventional procedures.<br>Examples of clinical<br>application may include<br>cholangiography, endoscopy,<br>urologic, orthopedic,<br>neurologic, vascular, cardiac,<br>critical care and emergency<br>room procedures. | |
| X-ray Modulation | | | | |
| X-ray Modulation<br>Component | CR Trace Filter | CR Trace Filter | Iris Collimator and<br>Semitransparent<br>Leaf/Leaves Collimator | |
| Device | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 9<br>(Subject Device) | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 8<br>(K183109)<br>(Predicate Device) | GE Healthcare<br>Surgery's<br>OEC 9900 Elite<br>(K122234)<br>(Reference) | |
| Feature | | | | |
| X-ray Radiation<br>Modulation | Reduces X-ray radiation<br>outside the aperture/ROI<br>typically by 61% to 97% | Reduces X-ray radiation<br>outside the aperture/ROI<br>typically by 61% to<br>97% | Completely blocks X-<br>ray radiation outside the<br>aperture | |
| Aperture shape | Rectangular | Rectangular | Leaf/Leaves:<br>Rectangular-Like<br>(2 straight edges and 2<br>round edges)<br>Iris: Octagonal. | |
| Aperture Control | Set by the user using the<br>CR Trace Tablet | Set by the user using the<br>CR Trace Tablet | Set by the user using the<br>Collimator Control<br>buttons on GE<br>Healthcare Surgery's<br>OEC® 9900 Elite's c-<br>arm unit control panel | |
| Image Processing | | | | |
| Image Area<br>Processed | Image area outside the<br>ROI | Image area outside the<br>ROI | Entire Image | |
| Processing Bits | 12 bits | 12 bits | 12 bits | |
| Processing Rate | 30 fps | 30 fps | 30 fps | |
| Processing<br>Occurrence | Only when the CR<br>Trace Filter is engaged | Only when the CR<br>Trace Filter is engaged | At all times | |
| Image Layout<br>Information | Dose Area Product<br>(DAP) value and/or<br>percentage of DAP<br>reduction when using<br>ControlRad Trace Filter<br>and/or OEC Collimators<br>ROI frame border | Dose Area Product<br>(DAP) value and/or<br>percentage of DAP<br>reduction when using<br>ControlRad Trace Filter<br>and/or OEC Collimators<br>ROI frame border | Hospital, Physician and<br>Patient's name, date and<br>time, X-ray Generator's<br>voltage and current<br>settings, Brightness and<br>Contrast settings,<br>Magnification level,<br>Accumulated Exposure<br>Time per Examination,<br>Accumulated Air Kerma | |
| Parameters Accuracy Specifications | Dose Area Product<br>(DAP) Accuracy for<br>total x-ray field of<br>the ControlRad<br>Trace Filter and<br>OEC systems<br>combined* | *Overall: ±35%<br>For DAP reported by<br>ControlRad™ Trace<br>Model 9 | *Overall: ±35%<br>For DAP reported by<br>ControlRad™ Trace<br>Model 8 | Overall accuracy: ±40%<br>(with Iris Field and<br>Sutter Field > 5cm). |
| Electrical Requirements | Electrical<br>Requirements | 60 / 50 Hz; 120 VAC<br>(±10%), 15A | 60 / 50 Hz; 120 VAC<br>(±10%), 15A | 60 / 50 Hz; 120 VAC<br>(±10%), 15A |
| Device<br>Feature | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 9<br>(Subject Device) | ControlRad, Inc's<br>ControlRad™ Trace<br>Model 8<br>(K183109)<br>(Predicate Device)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.