K183144 · Shanghai United Imaging Healthcare Co., Ltd. · OWB · Dec 19, 2019 · Radiology
Device Facts
Record ID
K183144
Device Name
uMC 560i
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
OWB · Radiology
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The uMC 560i is intended to provide digital imaging of patients during diagnostic and surgical procedures. Examples of clinical application may include, but are not limited to: orthopedic, pain therapy, neurological, pediatric examinations and critical care. The visualization of such anatomical structures assists the clinical decisions. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
Device Story
Mobile C-arm X-ray system; utilizes CMOS flat panel detector to capture X-ray images of anatomical structures. System components include C-arm, monoblock, anti-scatter grid, collimator, exposure controls, monitor cart, and workstation. Operated by clinicians in surgical or diagnostic settings. Transforms X-ray input into digital images displayed on 34" TFT flat screen. Features uFree dose management technology for radiation optimization. Assists clinicians in real-time visualization during procedures; supports orthopedic, pain therapy, neurological, and critical care applications. Benefits include high-quality imaging for surgical guidance and diagnostic assessment.
Clinical Evidence
Clinical evaluation provided fluoroscopy and static images acquired in a clinical environment. Images were reviewed by a board-certified clinical reviewer, confirming diagnostic quality across various exposure modes, parameters, and anatomical regions.
Technological Characteristics
Mobile C-arm X-ray system; stationary anode X-ray tube; 40-110 kV range; 3.5 kW max power. CMOS flat panel detector (XINEOS 2222HS); 151.8 μm pixel size; 1416 x 1420 matrix. Includes uFree dose management technology. Complies with IEC 60601-1, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-43, 60601-2-54, 60601-1-2, and IEC 62304. Biocompatibility per ISO 10993-5/10.
Indications for Use
Indicated for digital imaging of adult and pediatric patients during diagnostic and surgical procedures, including orthopedic, pain therapy, neurological, and critical care applications. Not for mammography or direct radiographic film exposures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2019
Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang OM & RA VP No. 2258 Chengbei Rd., Jiading Industrial District Shanghai, Shanghai 201807 CHINA
Re: K183144
Trade/Device Name: uMC 560i Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: November 27, 2018 Received: November 29, 2018
Dear Shumei Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting
{1}------------------------------------------------
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183144
Device Name uMC 560i
#### Indications for Use (Describe)
The uMC 560i is intended to provide digital imaging of patients during diagnostic and surgical procedures. Examples of clinical application may include, but are not limited to: orthopedic, pain therapy, neurological, pediatric examinations and critical care. The visualization of such anatomical structures assists the clinical decisions. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.
# 510 (k) SUMMARY
K183144
- 1. Date of Preparation: November 26, 2019
## 2. Sponsor Identification
### Shanghai United Imaging Healthcare Co.,Ltd.
No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China Establishment Registration Number: 3011015597 Contact Person: Xin GAO Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
### 3. Identification of Proposed Device
Trade Name: uMC 560i Common Name: Mobile C-arm X-Ray System Model(s): uMC 560i Regulatory Information Classification Name: Image-intensified fluoroscopic x-ray system Classification: Product Code: OWB, OXO, JAA Regulation Number: 21 CFR 892.1650 Review Panel: Radiology
- 4. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K153244 Device Name: Cios Fusion Manufacturer: Siemens AG Regulatory Information Classification Name: Image-intensified fluoroscopic x-ray system Classification: II Product Code: OWB, OXO Regulation Number: 21 CFR 892.1650 Review Panel: Radiology Reference Device
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold, sans-serif font. The logo is simple and modern, and the use of bold fonts gives it a strong and confident look.
510(k) Number: K181560 Device Name: Cios Alpha Manufacturer: Siemens AG
#### Regulatory Information
Classification Name: Image-intensified fluoroscopic x-ray system Classification: II Product Code: OWB. OXO Regulation Number: 21 CFR 892. 1650 Review Panel: Radiology
#### 5. Device Description
uMC 560i is the first mobile C-arm X-ray system developed and produced by UIH Medical XR Business Unit. It is designed to meet medical institutions' demands. To cover the most surgical imaging needs, it's X-ray imaging of the anatomical structures of patient during clinical applications may include, but are not limited to: orthopedic, pain therapy, neurological, pediatric examinations and critical care. uMC 560i consists of C-arm, Monoblock, Anti-scatter Grid, CMOS flat panel detector, Collimator system, Exposure handswitch, Exposure footswitch, Connects cables for Monitor cart and C-arm mobiles stand. Monitor cart, Touchable control panel, HD monitor, Examination workstation. The powerful system performance brings a safe, fluent and efficient operation experience.
#### 6. Indications for Use
The uMC 560i is intended to provide digital imaging of patients during diagnostic and surgical procedures. Examples of clinical application may include, but are not limited to: orthopedic, pain therapy, neurological, pediatric examinations and critical care. The visualization of such anatomical structures assists the clinician with clinical decisions. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
## 7. Performance Data
#### Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electric for basic safety and essential performance (IEC 60601-1:2005, MOD).
- > IEC 60601-1-3 Edition 2.1 2013-04 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized letter "U" that is split in the middle by a vertical line. The logo is simple and modern.
Standard: Radiation protection in diagnostic X-ray equipment.
- > IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- > IEC 60601-2-28 Edition 2.0 2010-03 Medical electrical equipment -Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis.
- > IEC 60601-2-43 Edition 2.1 2017-05 CONSOLIDATED VERSION Medical electrical equipment - Part 2-43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures.
- A IEC 60601-2-54 CONSOLIDATED VERSION Edition 1.1 2015-04 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
- > IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cvcle processes.
- A IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
# Clinical Test Conclusion
The fluoroscopy sequences images and static images produced by the uMC 560i in clinical environment accompanied by exposure mode, exposure parameters and anatomical regions were provided in the clinical evaluation. Both dynamic and static images were reviewed by a board-certified clinical reviewer with a statement indicating that images are of diagnostic quality.
## 8. Substantially Equivalent (SE) Comparison
| Table 1 Comparison of Device Functionality and Specification | | | |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ITEM | Proposed Device<br>uMC 560i | Predicate Device<br>Siemens Cios Fusion<br>(K153244) | Remark |
| <b>General</b> | | | |
| Mobile<br>Fluoroscopic<br>C-arm | Yes | Yes | Same |
| Product Code | OWB, OXO | OWB, OXO | Same |
| Class | II | II | Same |
| Indications for<br>Use | The uMC 560i is intended to<br>provide digital imaging of<br>patients during diagnostic and<br>surgical procedures.<br>Examples of clinical<br>application may include, but | The Cios Fusion is a mobile X-<br>Ray System designed to<br>provide X-ray imaging of the<br>anatomical structures of patient<br>during clinical applications.<br>Clinical applications may | Note 1 |
| | are not limited to: orthopedic, pain therapy, neurological, pediatric examinations and critical care. The visualization of such anatomical structures assists the clinician with clinical decisions. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. | include but are not limited to: interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients. | |
| Specifications | | | |
| Imaging Modes | Pulsed Fluoroscopy Spot Fluoroscopy Digital Radiography | Pulsed Fluoroscopy Continuous Fluoroscopy Digital Radiography | Note 2 |
| X-ray Tube | Stationary Anode | Stationary Anode | Same |
| X-ray Generator and Tube Housing Assembly Monoblock Technology | Yes | Yes | Same |
| KV Range | 40 kV – 110 kV | 40 kV – 110 kV | Same |
| Max Power Output | 3.5 kW | 2.3 kW | Note 3 |
| Pulsed Fluoroscopy Current | 0.1 mA – 30 mA | 3 mA - 25 mA | Note 4 |
| Displays | 34" TFT Flat Screen Color Display Panel | 19" TFT Flat Screen Color Display Panels | Note 5 |
| Collimator | Yes, Pd, Rectangular and Slot Collimator | Yes, Pd, Rectangular and Slot Collimator | Same |
| SID | 107.5 cm | 102 cm | Note 6 |
| Removable Anti-scatter Grid | Yes | Yes | Same |
| Dose Area Product | Yes | Yes | Same |
| Monitor Cart | Mandatory Monitor Cart | Mandatory Monitor Cart | Same |
| User Interface | Yes, Touch Panel | Yes, Touch Panel | Same |
| Dose Optimization | uFree Dose Management Technology | Siemens CARE Program | Note 7 |
| DICOM Function | Yes | Yes | Same |
| 2D Image Post<br>Processing | Yes | Yes | Same |
| <b>Safety</b> | | | |
| Electrical<br>Safety | AAMI ANSI ES60601-<br>1:C1:2009/(R)2012 and<br>A2:2010/(R)2012 | Comply with AAMI/ANSI ES<br>60601-1(IEC 60601-1:2005) | Same |
| EMC | Comply with IEC60601-1-2 | Comply with IEC60601-1-2 | Same |
| Biocompatibili<br>ty | Comply with ISO10993-5,<br>ISO10993-10 | Comply with ISO10993-5,<br>ISO10993-10 | Same |
Table 1 Comparison of Device Functionality and Specification
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING". To the right of the text is a stylized letter "U" that is split into two sections by a horizontal line.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is also in a bold, sans-serif font.
### Table 2 Comparison of Flat Panel Detector Specification
| ITEM | Proposed Device | Reference Device | Remark |
|---------------------------------------|-------------------------------------------|------------------------------------------------------------------------------|--------|
| | uMC 560i | Siemens Cios Alpha (K181560) | |
| Specifications of Flat Panel Detector | | | |
| Image Receptor | CMOS Flat Panel Detector<br>XINEOS 2222HS | CMOS Flat Panel Detector<br>XINEOS 3030HS or XINEOS 2222HS | Same |
| DQE | 75% | 75% (XINEOS 2222HS)<br>72% (XINEOS 3030HS) | Same |
| Dynamic Range | 96dB | 96dB | Same |
| Modulation Transfer Function (MTF) | 60% @1lp/mm | 60% @1lp/mm(XINEOS 2222HS)<br>58% @1lp/mm(XINEOS 3030HS) | Same |
| Digitization Depth | 16bit | 16bit | Same |
| Field of View | Square, 21.5 cm ×21.5 cm | Square, 20 cm ×20 cm (XINEOS 2222HS)<br>Square, 30 cm ×30 cm (XINEOS 3030HS) | Note 8 |
| Pixel Size | 151.8 μm | 152 μm | Same |
| Matrix Size | 1416 × 1420 | 1416 × 1420 (XINEOS 2222HS)<br>1952×1952 (XINEOS 3030HS) | Same |
# Table 3 Justifications of Differences
| Note ID | Justification |
|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note 1 | uMC 560i and the predicate devices are all used for providing digital<br>imaging of adults and pediatric during diagnostic and surgical procedures.<br>The difference is that uMC 560i is not intended to be used during<br>interventional, which does not affect the clinical application of uMC 560i.<br>The difference does not affect safety and effectiveness. |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" symbol, which is dark blue and appears to be made up of two vertical lines and a curved base.
| Note 2 | Compared to predicate devices Siemens Cios Fusion (K153244),<br>uMC 560i does not provide continuous fluoroscopy, whose fundamental principle and clinical application are the same with pulsed fluoroscopy. The difference does not affect safety and effectiveness. uMC 560i also provides spot fluoroscopy, whose fundamental principle is the same with pulsed fluoroscopy. The difference between spot fluoroscopy and pulsed fluoroscopy is that radiation stops automatically when meeting target grey level, but the image quality is the same with LIH of pulsed fluoroscopy. The difference does not affect safety and effectiveness. |
|--------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note 3 | 3.5 kW has the ability to provide higher mA at lower kV levels allowing shorter pulse widths at variable frame rates to increase the image quality of moving objects by reducing or eliminating the movement during each single pulse. The difference in maximum power output does not affect safety and effectiveness. |
| Note 4 | The pulsed fluoroscopy current of predicate devices can be covered by uMC 560i's in clinical application. And the maximum current of uMC 560i is larger than predicate devices, which will improve the image quality of large patient. The difference in pulsed fluoroscopy current does not affect safety and effectiveness. |
| Note 5 | The display with larger size will improve user experience, which does not affect safety and effectiveness. |
| Note 6 | The slight difference of SID does not affect clinical application. The difference does not affect safety and effectiveness. |
| Note 7 | uMC 560i and the predicate devices all provide dose optimization method. Even though the principle is different, the effectiveness of uFree can be proven by Section 21 Software Design Description (Chapter 3.3.8). The difference does not affect safety and effectiveness. |
| Note 8 | uMC 560i and the reference device (K181560) both use the same flat panel detector. uMC 560i has a larger effective imaging area than Cios Alpha. The slight difference of Field of View does not affect clinical application. The difference does not affect safety and effectiveness. |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in bold. The color of the text and the "U" shape is a dark teal.
# 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device has same intended use, similar performance, equivalence safety and effectiveness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness.
The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.