K183101 · Philips Medical Systems Nederland B.V. · OWB · Dec 31, 2018 · Radiology
Device Facts
Record ID
K183101
Device Name
Zenition 50
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OWB · Radiology
Decision Date
Dec 31, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Zenition 50 device is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
Device Story
Mobile diagnostic X-ray imaging and viewing system comprising C-arm stand and Mobile View Station (MVS). Inputs: X-ray radiation; outputs: fluoroscopic images for clinical visualization. Operated by trained healthcare professionals in OR or clinical settings. System architecture utilizes PC platform with migrated software-based image processing. Features include touch-screen interface, ClearGuide color coding, and manual outlining tools. Output aids surgeons in real-time guidance during interventional procedures. Benefits include improved workflow, enhanced security, and optimized X-ray tube heat management. Device provides visual feedback to clinicians to support surgical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including verification and validation against international consensus standards (IEC 60601-1, 60601-2-43, 60601-2-54, 62304, 62366) and FDA guidance documents.
Technological Characteristics
Mobile C-arm X-ray system. Components: iXion HF generator, RTM 780 H X-ray tube, iXion Monoblock V. Connectivity: wireless network, DICOM. Software: PC-based platform, migrated image processing. Standards: IEC 60601-1 (safety), IEC 60601-2-43/54 (X-ray performance), IEC 62366 (usability), IEC 62304 (software lifecycle). Cybersecurity: disk encryption, FIPS 140-2, audit trails, white listing.
Indications for Use
Indicated for radiological guidance and visualization during diagnostic, interventional, and surgical procedures (orthopedic, neuro, abdominal, vascular, thoracic, cardiac) in healthcare facilities. Patient population: all patients except neonates (birth to one month).
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 31, 2018
Philips Medical Systems Nederland BV Sumit Kumar Regulatory Approbation Officer Veenpluis 4-6 BEST, NL 5684PC
Re: K183101
Trade/Device Name: Zenition 50 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: November 1, 2018 Received: November 7, 2018
Dear Sumit Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| | See PRA Statement below. |
| 510(k) Number (if known) | K183101 |
| Device Name | Zenition 50 |
| Indications for Use (Describe) | The Zenition 50 device is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. |
| Applications: | <ul><li>Orthopedic</li><li>Neuro</li><li>Abdominal</li><li>Vascular</li><li>Thoracic</li><li>Cardiac</li></ul> |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC
{4}------------------------------------------------
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
Date Prepared: November 1, 2018
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Primary<br>Contact<br>Person: | Sumit Kumar<br>Regulatory Approbation Officer<br>Phone: +91 7028039288<br>E-mail: sumit.kumar_3@philips.com | |
| Secondary<br>Contact<br>Person: | Sonavane, Neena Sandeep<br>Regulatory Affairs Manager<br>Phone: +91 8446020023<br>E-mail: neena.sonavane@philips.com | |
| Device: | Trade Name: | Zenition 50 |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | 90-Radiology |
| | Device Class: | Class II |
| | Primary Product Code: | OWB |
| | Secondary Product Code: | JAA, OXO |
| Primary<br>Predicate<br>Device: | Trade Name: | BV Pulsera |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K010435 |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | 90-Radiology |
| | Device Class: | Class II |
| | Product Code: | OWB; JAA; OXO |
| Reference<br>device: | Trade Name: | Veradius Unity |
| | Manufacturer | Philips Medical Systems Nederland B.V |
| | 510(k) Clearance: | K142708 |
| | Classification Name: | Image-intensified fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | 90-Radiology |
| | Device Class: | Class II |
| | Product Code: | OWB; JAA; OXO |
| Device description: | The proposed Zenition 50 is a mobile, diagnostic X-ray imaging and viewing system. It is designed for medical use in healthcare facilities where X-ray imaging is needed. The system comprises of two main components: the C-arm stand and a Mobile View Station(MVS) | |
| Indications for Use: | The proposed Zenition 50 is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device.<br>The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.<br>The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.<br>Applications:<br>Orthopedic Neuro Abdominal Vascular Thoracic Cardiac | |
| Fundamental Scientific Technology: | The proposed Zenition 50 employs the same basic construction and fundamental scientific technology as the currently marketed and predicate BV Pulsera. The technology used in the development of the major components of the proposed Zenition 50, which includes X-ray generator, X-ray tube housing assembly, and Image detection system is identical to the currently marketed and predicate BV Pulsera. See the below table for the comparison of the major components of the proposed Zenition 50 and the currently marketed and predicate BV Pulsera. | |
| | The outcome of this comparison demonstrates that the minor differences in the technological | |
{5}------------------------------------------------
characteristics do not affect the safety or effectiveness of the Zenition 50 when compared to the currently marketed and predicate BV Pulsera.
| Technological characteristics comparison of the currently marketed predicate device, BV<br>Pulsera versus the proposed Zenition 50 | | | |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component<br>/feature | Currently<br>market BV<br>Pulsera | Proposed Zenition 50 | Conclusion |
| X-ray<br>Generator | iXion HF<br>Generator<br>Model: 10359400 | iXion HF Generator<br>Model: 10359400 | Identical. |
| X-ray tube | Model: RTM 780<br>H (Type RO-0306) | Model : RTM 780 H<br>(Type RO-0306) | Identical. |
| X-ray housing assembly | iXion Monoblock IV with X-ray tube RO-0306 | iXion Monoblock V with X-ray tube RO-0306 | |
| | | | Similar with minor change in<br>monoblock design for providing<br>more clearance to the floor.<br>The monoblock design has been<br>updated to reduce the height with<br>3.5cm, and thus improves clearance<br>between the tank and the floor which<br>resulting in more space to move. In<br>addition, the filtration has been<br>reduced to lower the amount of<br>power necessary for the same dose.<br>The updated monoblock is evaluated<br>in risk management. The change<br>does not introduce new risks. This<br>change was shown to be compliant<br>with the international and FDA<br>recognized standards IEC60601-1,<br>IEC60601-2-43 and IEC60601-2-54<br>for basic safety and essential<br>performance. Hence, this change<br>does not impact the safety and<br>effectiveness of the Device. Thus,<br>demonstrating substantial<br>equivalence. |
| Image Intensifier 9" | 23HRC | 23HRC | Identical. |
| Image Intensifier 12" | 31GG | 31GG | Identical |
| II Laser alignment tool | Z-LAD 9 Inch<br>Z-LAD 12 Inch | Z-LAD 9 Inch<br>Z-LAD 12 Inch | Identical |
| Beam Limiting Device | CoRa<br>9890 010 23201 | Collimator IITV<br>459801200241 | A new collimator has been designed<br>such that it reuses the design of<br>existing reference device Veradius<br>Unity (K142708) FD15 (PX2630Sv)<br>collimator with the only exception<br>the square fixed diaphragm is<br>removed to match the format for<br>IITV9" and IITV12". There is no<br>change in the functional<br>specification compared to predicate<br>device. All other parts of the<br>collimator are identical.<br>This change was shown to be<br>compliant with the international and<br>FDA recognized |
| | | | |
| Laser<br>Alignment<br>tool | Tube Laser<br>Aiming Device | Tube Laser Aiming<br>Device | essential performance. Hence, this<br>does not impact safety or<br>effectiveness of the device Thus,<br>demonstrating substantially<br>equivalent. |
| | | | Changed to fit in new collimator.<br>Functional specification are similar<br>compared to predicate device. This<br>tube laser unit is reused from the<br>reference device Veradius Unity<br>(K142708).<br>Hence, this change do not impact<br>the safety or effectiveness of the<br>Device. Thus, demonstrating<br>substantial equivalence |
| Mobile C-arm<br>Stand | BV Pulsera Stand | Pulsera R3.1 Stand | The mechanical interfaces have<br>been updated for the updated<br>monoblock and the stand<br>touchscreen. These changes are<br>assessed in risk management, and<br>they did not introduce new risks.<br>The changes was also shown to be<br>compliant with the FDA recognized<br>standard IEC60601-1 for basic<br>safety and essential performance.<br>Therefore, the change does not<br>impact safety and effectiveness.<br>Thus, demonstrating substantial<br>equivalence. |
| Mobile<br>Viewing<br>Station | MVS BV Family<br>R2 | MVS BV Family R3 | The MVS has been changed to<br>match the updated interfaces from<br>the System architecture and room<br>interface changes. These changes<br>are assessed in risk management,<br>and they did not introduce new risks.<br>The changes was also shown to be<br>compliant with the FDA recognized<br>standard IEC60601-1 for basic<br>safety and essential performance.<br>Therefore, the change does not<br>impact safety and effectiveness.<br>Thus, demonstrating substantial<br>equivalence |
| System<br>architecture | Current System<br>architecture | Improved system<br>architecture with PC<br>platform<br>• Solved obsolescence | These changes were assessed in risk<br>management, and they did not<br>introduce new risks. This change<br>was shown to be compliant with the<br>international and FDA recognized |
| | | | |
| | | hardware and software by moving to state of the art computing hardware and corresponding interconnects. Migrated system software to the new computing hardware Migrated image processing from hardware to software implementation Integrated internal distributed components related to power functions into the mains control unit to simplify the design Integrated wireless network connection design to simplify configuration and setting to work by service engineers | standards IEC60601-1, IEC60601-2-43 and IEC60601-2-54 for basic safety and essential performance, IEC60601-1-2 for Electromagnetic compatibility, IEC62366 for application of usability engineering and the FDA Performance Standard for Diagnostic X-Ray Systems and Their Major Components (21 CFR 1020.30 / 1020.32). Image processing performance is compared by means of an image quality performance comparison with the predicate device and found equal or improved, see <b>Appendix A38</b> .<br>This change does not affect safety or effectiveness of the device. Thus, demonstrating substantial equivalence. |
| ClearGuide and color coding | Not Present | Introduction of ClearGuide and color coding | These visual aids are ease-of-use improvements and improve communication in the OR between surgeon and operator. These features are reused from the reference device <i>Veradius Unity</i> (K142708).<br>The changes do not introduce new risks or change any of the existing risks compared to the predicate device.<br>This change does not affect safety or effectiveness of the device. Thus, demonstrating substantial equivalence. |
| C-arm Stand user interface | C-arm Stand user interface based on hard keys and a small monochrome display | C-arm Stand touch screen user interface | The hard-key based interface on the predicate device has been replaced with a touch-screen based stand user interface This feature is reused from the reference device <i>Veradius Unity</i> (K142708). This change was shown to be compliant with the international and FDA recognized standards IEC60601-1, IEC60601- |
| | | | 2-43 and IEC60601-2-54 for basic<br>safety and essential performance,<br>IEC60601-1-2 for Electromagnetic<br>compatibility,<br>IEC62366<br>for<br>application of usability engineering.<br>This change does not affect safety or<br>effectiveness of the device. Thus,<br>demonstrating<br>substantial |
| Manual<br>Outline | Not Present | Manual outlining | equivalence<br>With this so-called Manual Outline<br>function, it is possible to draw lines<br>and dots on an overlay of the clinical<br>image with help of the finger, stylus<br>pen or mouse. It is not related to<br>diagnostics. This feature is reused<br>from the reference device Veradius<br>Unity (K142708).<br>The changes do not introduce new<br>risks or change any of the existing<br>risks compared to the predicate<br>device. |
| | | | This change does not affect safety or<br>effectiveness of the device. Thus,<br>substantial<br>demonstrating<br>equivalence. |
| Wireless<br>footswitch | Not Present | Wireless footswitch | Next to the wired footswitch and<br>optional wireless footswitch can be<br>used with the system. This wireless<br>footswitch is reused from reference<br>device Veradius Unity (K142708).<br>The change does not introduce new<br>risks, but only minor update to<br>existing risks compared to the<br>predicate device.<br>This change does not affect safety or<br>effectiveness of the device. Thus,<br>demonstrating<br>substantial<br>equivalence. |
| DICOM<br>connectivity | Current DICOM<br>connectivity | Improved DICOM<br>connectivity workflow<br>· Easier selection of<br>patient data for<br>export<br>• Introduced<br>unattended network<br>transfer of export<br>jobs<br>• Integrated workflow<br>for export to local | The improvements are ease-of-use<br>improvements, fully compliant with<br>the DICOM standard and do not<br>impact clinical functionality. The<br>changes are shown to be compliant<br>with the international and FDA<br>recognized standard IEC62366 for<br>application of usability engineering.<br>Therefore, this change does not<br>impact safety or effectiveness of the |
| | | media (USB and<br>DICOM DVD)<br>• Improved workflow<br>for multimodality<br>viewer functionality<br>• Improved DICOM<br>transfer speed | demonstrating<br>device.<br>Thus,<br>substanti…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.