K171194 · Canon, Inc. · OWB · May 24, 2017 · Radiology
Device Facts
Record ID
K171194
Device Name
AS-10, CXDI-401RF
Applicant
Canon, Inc.
Product Code
OWB · Radiology
Decision Date
May 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The AS-10 / CXDI-401RF is indicated for use in generating fluoroscopic and radiographic images of human anatomy for angiography, diagnostic, and interventional procedures. The device is intended to replace spot-film devices is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Device Story
AS-10 / CXDI-401RF is a solid-state flat panel X-ray detector (FPD) for integration into dynamic and static imaging systems. Device intercepts X-ray photons via a scintillator, which emits visible light onto an amorphous silicon (a-Si) photo-detector array; signals are converted to digital values. Used in clinical settings (angiography, diagnostic, interventional) by professional operators (physicians, technicians). Device replaces spot-film and image intensifier technology. Output is digital image data provided to an integrated imaging system for clinical review and decision-making. Benefits include improved image acquisition capabilities (e.g., increased pixel count, binning modes) compared to predecessors.
Clinical Evidence
Bench testing only. Verification and validation testing performed against internal functional specifications. Non-clinical image comparisons (static and dynamic) conducted against predicate CXDI-50RF. Compliance demonstrated with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-54, 62304, 62366, ANSI/AAMI ES60601-1, and IEC 60825-1.
Technological Characteristics
Solid-state flat panel detector (FPD) using amorphous silicon (a-Si) and Cesium Iodide (CsI(Tl)) scintillator. 160 x 160 µm pixel pitch. 16-bit A/D conversion. Wired connectivity to image capture unit. Dimensions: 469 x 468 x 58 mm. Weight: 13 kg. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-54, 62304, 62366, and ANSI/AAMI ES60601-1.
Indications for Use
Indicated for generating fluoroscopic and radiographic images of human anatomy for angiography, diagnostic, and interventional procedures. Not for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Canon, Inc. – Medical Equipment Group % Diane Rutherford Submission Manager Ken Block Consulting 1201 Richardson Drive, Suite 160 Richardson, Texas 75080
Re: K171194
Trade/Device Name: AS-10, CXDI-401RF Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, MQB Dated: April 24, 2017 Received: April 24, 2017
Dear Diane Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara
For
Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K171194
Device Name AS-10 CXDI-401RF
### Indications for Use (Describe)
The AS-10 / CXDI-401RF is indicated for use in generating fluoroscopic and radiographic images of human anatomy for angiography, diagnostic, and interventional procedures. The device is intended to replace spot-film devices is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "Canon" in red font. The font is bold and slightly italicized. The word is centered and takes up most of the image space.
# 5. 510(k) SUMMARY
| Submitter: | Canon, Inc. - Medical Equipment Group<br>9-1, Imaikami-cho, Nakahara-ku<br>Kawasaki, Kanagawa 211-8501, Japan |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Canon, Inc. – Medical Equipment Group<br>Mr. Shinji Mori<br>Manager<br>9-1, Imaikami-cho, Nakahara-ku<br>Kawasaki, Kanagawa 211-8501, Japan<br>TEL: 81-3-3758-2111<br>FAX: 81-44-739-6493<br>mori.shinji@canon.co.jp |
| Date Prepared: | April 18, 2017 |
| Proposed Device | Manufacturer: Canon, Inc. - Medical Equipment Group<br>Trade Name: AS-10<br>Common Name: CXDI-401RF<br>Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Interventional Fluoroscopic X-ray System<br>Primary Product Code / Regulatory Standard: OWB<br>892.1650, Image-intensified fluoroscopic x-ray system<br>Subsequent Product Codes: JAA, MQB |
| Predicate Devices: | Clearance: K092439 dated November 30, 2009<br>Manufacturer: Virtual Imaging, A Canon USA Company<br>Trade Name: Model CXDI-50RF<br>Common Name: Fluoroscopic Digital X-Ray Receptor Panel<br>Classification Name: Image-Intensified Fluoroscopic X-Ray System<br>Product Code(s) / Regulatory Standard(s): OWB, JAA, IZI [SE Letter correction July 30, 2012]<br>892.1650, Image-intensified fluoroscopic x-ray system<br>892.1600, Angiographic x-ray system<br>[Originally cleared under MQB, 892.1680] |
| | Clearance: K162909 dated January 27, 2017<br>Manufacturer: Canon, Inc. - Medical Equipment Group<br>Trade Name: Model CSX-30<br>Common Name: Solid State X-ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray system<br>Primary Product Code MQB<br>Regulatory Standard: 892.1680, Stationary X-ray System<br>Subsequent Product Code JAA<br>Regulatory Standard: 892.1650, Image-intensified fluoroscopic x-ray system |
| Device Description: | The AS-10 / CXDI-401RF is a solid state x-ray imager. It intercepts x-ray photons and<br>the scintillator of the AS-10 / CXDI-401RF emits visible spectrum photons that<br>illuminate an array of photo-detectors that create electrical signals. After the electrical<br>signals are generated, it is converted to digital value. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Canon" in red. The font is bold and slightly italicized. The background is white.
- Indications for Use: The AS-10 / CXDI-401RF is indicated for use in generating fluoroscopic and radiographic images of human anatomy for angiography, diagnostic, and interventional procedures. The device is intended to replace spot-film devices. The device is also intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Summary of Comparison with the predicate devices shows the technological characteristics of the Technological AS-10 / CXDI-401RF is substantially equivalent to the predicate devices. The flat Characteristics: panel detector units are functionally identical.
> The AS-10 / CXDI-401RF is a successor model of the CXDI-50RF; however, the AS-10 / CXDI-401RF is for integration only like the CSX-30. The differences between the AS-10 / CXDI-401RF and the predicates CXDI-50RF and CSX-30 are primarily overall dimensions, increased pixels, and increased image size.
> The differences between the AS-10 / CXDI-401RF and the predicates are primarily increased size. Increased size results in larger dimensions, larger image area, more pixels, and weight differences over the predicate. An additional binning mode has been added to the AS-10 / CXDI-401RF in an effort to improve the marketability. The proposed indications for use statement differs slightly in wording from the predicates, but does not impact the intended use. The proposed and both predicate devices are indicated for fluoroscopic and radiographic use.
> In the table below, the predicates are identified as a) CXDI-50RF or b) CSX-30. If no letter precedes the predicate information, the information applies to both predicates.
| | New Device:<br>AS-10 / CXDI-401RF<br>K171194 | Predicate Devices:<br>a) CXDI-50RF K092439<br>b) CSX-30 K162909 | |
|------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------------|
| Application | Fluoroscopy and Spot Radiology | Fluoroscopy and Spot Radiology | Identical |
| Technology | Flat panel detector:<br>Scintillator and a-Si | Flat panel detector:<br>a) Scintillator and a-Si | Identical |
| Scintillator | CsI(TI) | CsI(TI) | Identical |
| Pixel Pitch | 160 x 160 µm | 160 x 160 µm | Identical |
| Pixels | 2,688 x 2,688 (≈ 7.2 million) | a) 2,208 x 2,688 (≈ 5.9 million)<br>b) 2,496 x 1,856 (≈ 4.6 million) | Modified |
| Image Size | 430 x 430 mm | a) 350 x 430 mm<br>b) 399 x 297 mm | Modified |
| Overall Dimensions | 469 x 468 x 58 mm | a) 493 x 503 x 26 mm<br>b) 470 x 363 x 82.5 mm<br>(Except tubes protrusions) | Modified |
| Weight | 13 kg | a) 5.7 kg<br>b) 19.0 kg | Modified |
| Acquisition Mode<br>(Binning mode) | Up to 15 fps (1x1)<br>Up to 30 fps (2x2)<br>Up to 30 fps (3x3) | a) Up to 15 fps (1x1)<br>b) Up to 30 fps (2x2) | Identical<br>Identical<br>Modified |
| DQE<br>@ 1 µGy in 0 lp/mm,<br>RQA5 | 0.75 | a) 0.6<br>b) 0.79 | Modified<br>Modified |
| Spatial Resolution<br>[MTF@2cycle/mm,<br>RQA5] | 0.28 | a) 0.3<br>b) 0.22 | Modified<br>Modified |
| A/D Conversion | 16-bit | a) 14-bit<br>b) 16-bit | Modified<br>Identical |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly italicized. The word is easily readable and takes up most of the image.
Summary of Tests were performed on the device which demonstrated that the device is safe and Non-Clinical /Test effective, performs comparably to and is substantially equivalent to the predicate devices. Data: Tests included verification/validation testing to internal functional specifications (including software) and non-clinical image comparisons involving flat panel display static and dynamic images taken with the new device and the predicate device CXDI-50RF. Documentation was provided demonstrating compliance of the AS-10 / CXDI-401RF to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, including results of verification/validation_plus_traceability of_verification/validation_tests_to_software requirements and software risk hazards. Documentation was provided demonstrating that the AS-10 / CXDI-401RF complies with the FDA requirements stated in Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices. The evaluations of the AS-10 / CXDI-401RF compared to the CXDI-50RF, show the AS-10 / CXDI-401RF to be equivalent to the CXDI-50RF. Testing confirmed that the AS-10 / CXDI-401RF complies with the U.S. Performance Standard for radiographic equipment and with relevant voluntary safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-54, 62304, and 62366 as well as ANSI/AAMI ES60601-1. Testing also confirmed compliance to relevant voluntary safety standard IEC 60825-1. Together, these verification/validation activities successfully demonstrated that the AS-10 / CXDI-401RF correctly performs as designed, has been validated for its intended use, and raises no new questions regarding either safety or effectiveness when compared to the predicate devices. Therefore, the verification/validation testing conducted supports a determination of substantial equivalence for the AS-10 / CXDI-401RF device. Integration: The AS-10 / CXDI-401RF is a hardware only FPD for integration into dynamic and static imaging systems. The AS-10 / CXDI-401RF connects to the image capture unit (compliant with IEC 60601-1 or IEC 60950-1) through a wired connection and does not connect directly to an x-ray generator. The Installation Manual provides detailed instructions and information for safe and effective system integration including: system configuration; electrical, mechanical, and cooling requirements; installation conditions; interfaces (image, command and service, and power input). Integrators are expected to adhere to the instructions and other information in the published Installation Manual. Conclusion: Canon, Inc. - Medical Equipment Group considers the AS-10 / CXDI-401RF to be substantially equivalent to the predicate devices listed above. This conclusion is based on the same intended use and similar principles of operation, functional design, and established medical use for the flat panel detector.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.