K170800 · Nu Vasive, Incorporated · OWB · Aug 18, 2017 · Radiology
Device Facts
Record ID
K170800
Device Name
NuVasive LessRay with Enhanced Tracking
Applicant
Nu Vasive, Incorporated
Product Code
OWB · Radiology
Decision Date
Aug 18, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
NuVasive® LessRay® with Enhanced Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Device Story
LessRay is a software application interfacing with a fluoroscope via video cable to enhance real-time imaging. It captures fluoroscopic video, applies image processing to reduce noise and improve contrast/clarity, and displays the enhanced output on a secondary monitor. The system supports C-arm tracking to guide positioning relative to prior baseline images and includes instrument tracking, where a tracker attached to an instrument allows the system to track its location and aid in positioning relative to baseline x-rays. Used in clinical settings by technicians and physicians to facilitate navigation and instrument placement. By enabling high-quality visualization from low-dose or pulsed fluoroscopy, the device aims to reduce radiation exposure while maintaining diagnostic utility. The system is passive, relying on the fluoroscope's video output without transmitting signals back to the imaging hardware.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included instrument tracking accuracy, GUI performance, and regression testing for image alignment and registration. Validation confirmed that instrument tracking allows for instrument relocalization with fewer x-ray images, greater accuracy, and less time compared to conventional fluoroscopy.
Technological Characteristics
Software-based image processing system. Interfaces with fluoroscope via video cable and frame grabber. Uses averaging algorithms for contrast/brightness enhancement and random noise reduction. Supports electromagnetic or optical tracking systems for 6 DOF C-arm and instrument tracking. Hardware requirements: Intel Core 2 Duo, NVIDIA Quadro 4000, 8GB RAM, Windows 7/8.1. Passive display system; does not transmit signals to the fluoroscope.
Indications for Use
Indicated for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease. Applicable to patients requiring fluoroscopic imaging during interventional procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol composed of three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2017
NuVasive, Incorporated % Manthan Damani Senior Regulatory Affairs Specialist 7475 Lusk Boulevard SAN DIEGO CA 92121
Re: K170800
Trade/Device Name: NuVasive® LessRay® with Enhanced Tracking Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ and JAA Dated: July 25, 2017 Received: July 26, 2017
Dear Manthan Damani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K170800
Device Name
NuVasive® LessRay® with Enhanced Tracking
Indications for Use (Describe)
NuVasive® LessRay® with Enhanced Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to reguirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf shape is divided into two parts, with the top part being purple and the bottom part being gray.
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### Submitted by: A.
Manthan J. Damani Senior Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
Date Prepared: August 17, 2017
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® LessRay® with Enhanced Tracking |
|----------------------------|---------------------------------------------|
| Common or Usual Name: | Image processing system |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Device Class: | Class II |
| Classification: | 21 CFR § 892.1650 |
| Product Code: | OWB, LLZ, JAA |
#### C. Predicate Devices
The subject device is substantially equivalent to the predicate device LessRay® with Tracking (K142243).
#### D. Device Description
LessRay is a software application which can be interfaced to a fluoroscope with a video cable. The images produced by the fluoroscope are transmitted to a frame grabber in the computer running LessRay where the images are enhanced and then displayed. When used in connection with the low dose and/or pulse setting on the fluoroscope, the user can improve the quality (clarity, contrast, noise level, and usability-) of a noisy (low-quality) image. Using this system, much of the graininess of low radiation dose images can be eliminated. This allows for greater utility of low dose imaging.2 LessRay with Tracking provided the additional feature of being able to interface LessRay with a tracking system in order to aid the C-arm technician in positioning the fluoroscope between the various views of the patient necessary for the intervention. LessRay with Tracking ensures that the fluoroscope is centered over the correct anatomy prior to taking any additional x-ray images.
NuVasive LessRay with Enhanced Tracking has additional capability of instrument tracking to aid the user in positioning an instrument using prior baseline x-rays. A tracker is attached to the instrument and as the instrument moves, the tracking system connected to LessRay tracks the location of the instrument. NuVasive LessRay with Enhanced Tracking uses this information to aid the user in positioning the instrument.
{5}------------------------------------------------
NuVasive LessRay with Enhanced Tracking is designed per recommendations provided in the following FDA guidance documents:
- Guidance for the Content of Premarket Submissions for Software Contained in ● Medical Devices
- Off-The-Shelf Software Use in Medical Devices ●
- Content of Premarket Submissions for Management of Cybersecurity in Medical ● Devices
#### Indications for Use E.
NuVasive® LessRay® with Enhanced Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
#### F. Comparison of Technological Characteristics with Predicate Device
As was established in this submission, the subject NuVasive LessRay with Enhanced Tracking is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent to its predicate device through comparison in areas including design, labeling/intended use, and function.
| Specification<br>/<br>Property | Predicate Device<br>LessRay with Tracking (K142243) | Subject Device<br>NuVasive LessRay with Enhanced<br>Tracking | Discuss<br>ion |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Intended Use<br>/ Indications<br>for Use | LessRay® with Tracking is intended for<br>use in any application where a<br>fluoroscope is incorporated to aid in<br>diagnosis and treatment of disease. | NuVasive® LessRay® with Enhanced<br>Tracking is intended for use in any<br>application where a fluoroscope is<br>incorporated to aid in diagnosis and<br>treatment of disease. | Same |
| Device Class | II | II | Same |
| Product Code | OWB, JAA, LLZ | OWB, JAA, LLZ | Same |
| Regulation<br>Number<br>(21CFR) | §892.1650 | §892.1650 | Same |
| Device<br>Classification<br>Name | Interventional Fluoroscopic X-Ray<br>System | Interventional Fluoroscopic X-Ray<br>System | Same |
| Specification / Property | Predicate Device<br>LessRay with Tracking (K142243) | Subject Device<br>NuVasive LessRay with Enhanced Tracking | Discussion |
| Device<br>Functionalities (image<br>acquisition, enhancement,<br>and display) | - Software based device used to<br>provide computer display systems<br>interfaced to a fluoroscope through<br>a video cable. The images<br>produced by the fluoroscope are<br>transmitted through a cable to a<br>frame capture board in the<br>computer where the images are<br>enhanced and then displayed on the<br>monitor.<br>- Enhanced images are displayed on<br>a computer monitor at the same<br>time that the corresponding<br>original image is displayed on the<br>fluoroscope monitor(s).<br>- Serves only as an image display<br>which is in addition to the<br>fluoroscope's standard image<br>display device. Device is passive,<br>in that the operation depends only<br>on the video output of the<br>fluoroscope, and it does not<br>transmit any signals or images to<br>the fluoroscope. | - Software based device used to<br>provide computer display systems<br>interfaced to fluoroscope through a<br>video cable. The images produced<br>by the fluoroscope are transmitted<br>through a cable to a frame capture<br>board in the computer where the<br>images are enhanced and then<br>displayed on the monitor.<br>- Enhanced images are displayed on<br>a computer monitor at the same<br>time that the corresponding<br>original image is displayed on the<br>fluoroscope monitor(s).<br>- Serves only as an image display<br>which is in addition to the<br>fluoroscope's standard image<br>display device. Device is passive,<br>in that the operation depends only<br>on the video output of the<br>fluoroscope, and it does not<br>transmit any signals or images to<br>the fluoroscope. | Same |
| Algorithms | - Image quality improvement using<br>averaging algorithm<br>- Contrast and brightness<br>enhancement with simultaneous<br>reduction of random noise | - Image quality improvement using<br>averaging algorithm<br>- Contrast and brightness enhancement<br>with simultaneous reduction of<br>random noise | Same |
| Compatible<br>hardware<br>platforms | Any computer that meets the following<br>minimum specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro 4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media H339 or<br>Elgato<br>Operating System: Windows 7 or 8.1 | Any computer that meets the following<br>minimum specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro 4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media H339 or<br>Elgato<br>Operating System: Windows 7 or 8.1 | Same |
| Specification / Property | Predicate Device<br>LessRay with Tracking (K142243) | Subject Device<br>NuVasive LessRay with Enhanced<br>Tracking | Discussion |
| C-arm<br>Tracking | - When tracking is enabled, will<br>automatically choose the Baseline<br>when the fluoroscope is near the<br>location and orientation that the<br>Baseline was initially taken.<br><br>- When tracking is enabled, requires | - When tracking is enabled, will<br>automatically choose the Baseline<br>when the fluoroscope is near the<br>location and orientation that the<br>Baseline was initially taken.<br><br>- When tracking is enabled, requires | Same |
| | hardware components in order to<br>mount the off-the-shelf tracking<br>hardware to the C-arm and to the<br>operating table. | hardware components in order to<br>mount the off-the-shelf tracking<br>hardware to the C-arm and to the<br>operating table. | |
| | - When tracking is enabled, requires<br>the use of an off-the-shelf tracking<br>system in order to track the 6 DOF<br>location of the C-arm relative to the<br>operating table. | - When tracking is enabled, requires<br>the use of an off-the-shelf tracking<br>system in order to track the 6 DOF<br>location of the C-arm relative to the<br>operating table. | |
| | - When tracking is enabled, visual<br>cues are provided which help guide<br>the user in positioning the C-arm<br>back to where a prior Baseline was<br>taken. | - When tracking is enabled, visual<br>cues are provided which help guide<br>the user in positioning the C-arm<br>back to where a prior Baseline was<br>taken. | |
| Tracking<br>options | Electromagnetic or optical | Electromagnetic or optical | Same |
{6}------------------------------------------------
#
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray. The leaf shape is on the left side of the logo, and the word "NUVASIVE" is on the right side.
Speci
# Traditional 510(k) Submission NuVasive® LessRay® with Enhanced Tracking
#### G. Performance Data
Instrument
Tracking
No
Nonclinical testing was performed to demonstrate that the subject NuVasive LessRay with Enhanced Tracking is substantially equivalent to the predicate device. The following testing was performed:
Enhanced Tracking has additional
capability of instrument tracking to aid
the user in positioning an instrument
using prior baseline x-rays.
- . Verification of Instrument Tracking to confirm that subject device allows the user to position the instrument back to where it was located when the desired image is taken.
- Verification of Graphical User Interface (GUI) to confirm that the GUI performs ● according to specifications.
- Validation of Instrument Tracking to confirm that using subject device instruments ● can be relocalized with less number of x-ray images, greater accuracy, and less time than when using conventional fluoroscopy.
Additio
nal
function
ality
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf shape is partially purple and partially gray. The logo is simple and modern.
- Regression Testing to verify image alignment, glyph tracking, and image registration . performance functionalities.
The results demonstrate that the subject NuVasive LessRay with Enhanced Tracking is substantially equivalent to the predicate.
#### H. Conclusions
The subject NuVasive LessRay with Enhanced Tracking has been shown to be substantially equivalent to legally marketed predicate devices for their intended use.
<sup>1</sup> As evaluated by a human observer in a side by side visual comparison of 30 image pairs with and without LessRay processing.
<sup>2</sup> In clinical practice, the amount of improvement achieved when a Pulsed and/or Low Dose image is processed with LessRay is dependent on the clinical task, patient size, anatomical location, and clinical practice. The dose should be set at a level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.