LessRay® with Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Device Story
LessRay with Tracking is a software application that interfaces with a fluoroscope via video cable to enhance real-time imaging. It captures fluoroscopic video feeds through a frame grabber, processing them on a standalone computer. The system combines current low-dose images with a prior high-dose 'Baseline' image to reduce noise and graininess. When surgical instruments are present, the software renders metallic tools translucent or toggles their visibility to reveal underlying anatomy. The 'Tracking' feature integrates with external 6-DOF tracking systems (optical or electromagnetic) to assist C-arm technicians in repositioning the fluoroscope to match previous anatomical views. The system provides visual cues to ensure the C-arm is centered over the correct anatomy before additional X-rays are taken. Used in clinical settings, it allows physicians to maintain image quality while utilizing lower radiation doses, potentially reducing patient radiation exposure. The device is passive and does not control the fluoroscope.
Clinical Evidence
Bench testing only. Performance was validated through regression/image recognition testing, verification of glyph tracking, and accuracy testing of the NDI Vicra Optical Tracking System and Patriot M Electromagnetic Tracking System (including dynamic metal distortion). Fluoroscope navigation was validated. Image quality improvement was evaluated by human observers in a side-by-side visual comparison of 30 image pairs.
Technological Characteristics
Software-based image processing system running on Windows 7/8.1. Hardware requirements: Intel Core 2 Duo CPU, NVIDIA Quadro 4000 GPU, 8GB RAM, 256GB HDD, Aver Media H339 or Elgato frame grabber. Integrates with 6-DOF tracking systems (optical/electromagnetic) via mounting hardware. Passive device; does not control fluoroscope. Algorithm: Summation of prior full-dose baseline images with subsequent low-dose images.
Indications for Use
Indicated for use in any application where a fluoroscope is used to aid in the diagnosis and treatment of disease in patients undergoing fluoroscopic procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2014
SafeRay Spine, LLC % Ms. Calley Herzog Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 ALEXANDRIA VA 22314
Re: K142243
Trade/Device Name: LessRay with Tracking Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, LLZ Dated: October 17, 2014 Received: October 20, 2014
Dear Ms. Herzog:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142243
Device Name LessRay® with Tracking
Indications for Use (Describe)
LessRay® with Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the LessRay with Tracking is provided below.
| Device Common Name: | Image Processing System |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | LessRay® with Tracking |
| Submitter: | SafeRay Spine, LLC<br>5102 Durham Chapel Hill Blvd., Suite 203<br>Durham, NC 27707<br>Phone: 877-814-7509 |
| Contact: | Calley Herzog<br>Consultant<br>Biologics Consulting Group, Inc.<br>Phone: 720-883-3633<br>Fax: 703-548-7457<br>Email: cherzog@bcg-usa.com |
| Classification Regulation: | 21 CFR 892.1650, Class II |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Panel: | Radiology |
| Primary Product Code: | OWB - interventional fluoroscopic x-ray system |
| Secondary Product Codes: | LLZ - system, image processing, radiological<br>JAA – system, x-ray, fluoroscopic, image-intensified |
| Date Prepared: | October 17, 2014 |
| Predicate Device: | K132970, LessRay® |
## Indication for Use:
LessRay® with Tracking is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
## Device Description:
LessRay is a software application which can be interfaced to a fluoroscope with a video cable. The images produced by the fluoroscope are transmitted to a frame grabber in the computer running LessRay® where the images are enhanced and then displayed. When used in connection with the low dose and/or pulse setting on the fluoroscope, the user can improve the quality (clarity, contrast, noise level, and usability') of a noisy (low-quality) image. Using this system, much of the graininess of low radiation dose images can be eliminated. This allows for greater utility of low dose imaging. -
{4}------------------------------------------------
LessRay® works by combining the current image being taken with a prior "Baseline" image of the same anatomy. The initial image, taken at regular radiation dose settings, serves as a baseline to which images taken at lower dose radiation settings can be compared and enhanced.
In procedures where instruments are introduced into the field of view, LessRay® images provide information that is not available with conventional images. By merging an image of interest containing tools with the unencumbered Baseline image, the radio dense metallic tools and implants that obscure the underlying anatomy can be rendered partially translucent in the merged image. In addition, by alternating the new image with the Baseline a user can make these metallic tools or implants disappear and reappear from the image, revealing the anatomy blocked by the tool.
The subject of this 510(k) is the addition of tracking capability to be used with LessRay LessRay can be interfaced with a tracking system in order to aid the C-arm technician in positioning the fluoroscope between the various views of the patient necessary for the intervention. LessRay® with Tracking ensures that the fluoroscope is centered over the correct anatomy prior to taking any additional x-ray images.
# Performance Data:
To establish the substantial equivalence of the modified LessRay® with Tracking to the predicate LessRay® device, the following performance tests were performed:
- Regression/Image Recognition Compared to LessRay v1.0 ●
- Verification of Glyph Tracking
- Tracking Accuracy with the NDI Vicra Optical Tracking System ●
- Tracking Accuracy with the Patriot M Electromagnetic Tracking System
- Tracking Accuracy with the Patriot M Electromagnetic Tracking System and Dynamic Metal Distortion
- . Collar Assembly Vertical Pull-Off Safety Test
- Validation of Fluoroscope Navigation
# Substantial Equivalence:
Based on the identical indication, similar technological characteristics, and results of performance testing, LessRay with Tracking is substantially equivalent to the previously cleared LessRay® device (K123226).
| | Proposed Device | Predicate Device |
|---------------------------|------------------------|--------------------|
| 510(k) Number | K142243 | K132970 |
| Device Name | LessRay® with Tracking | LessRay® |
| Submitter | SafeRay Spine, LLC | SafeRay Spine, LLC |
| Classification Regulation | 892.1650 | 892.1650 |
| Product Code | OWB, LLZ | OWB, LLZ |
#### Table 1: Device Comparison Table
{5}------------------------------------------------
| | Proposed Device | Predicate Device |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication | Indicated for use in any<br>application where a fluoroscope<br>is incorporated to aid in diagnosis<br>and treatment of disease. | Indicated for use in any<br>application where a fluoroscope<br>is incorporated to aid in diagnosis<br>and treatment of disease. |
| Compatible Hardware<br>Platforms | Any computer that meets the<br>following minimum<br>specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro 4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media<br>H339 or Elgato<br>Operating System: Windows 7 or<br>8.1 | Any computer that meets the<br>following minimum<br>specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro 4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media<br>H339 or Elgato<br>Operating System: Windows 7 or<br>8.1 |
| Software is run on a stand-<br>alone computer and monitor | Yes | Yes |
| Device is passive and doesn't<br>control the fluoroscope | Yes | Yes |
| Displays reduced noise images | Yes | Yes |
| For use during procedures that<br>involve fluoroscopy | Yes | Yes |
| Improves quality of low dose<br>images | Yes | Yes |
| Uses data from prior images to<br>improve the quality of<br>subsequent images | Yes | Yes |
| Algorithm used to improve<br>image quality | Summation of prior full dose<br>images with subsequent images | Summation of prior full dose<br>images with subsequent images |
| Provides visual cues which help<br>guide the user in positioning<br>the C-arm back to where a<br>prior Baseline was taken. | Yes | No |
| Automatically selects the<br>nearest Baseline as the C-arm<br>is being moved. | Yes | No |
| Requires a 6 DOF tracking<br>system in some configurations. | Yes | No |
{6}------------------------------------------------
| | Proposed Device | Predicate Device |
|-------------------------------------------------------------------------------------------------------------------|-----------------|------------------|
| Requires mounting hardware<br>used to mount the 6 DOF<br>tracking system to the C-arm<br>and the operating table. | Yes | No |
## Substantial Equivalence Summary
The addition of the optional tracking feature does not change the intended use of the device as a software application that enhances images from a fluoroscope. LessRay with Tracking provides the additional capability of aiding the user in acquiring fluoroscopic images by ensuring that the fluoroscope is centered over the correct anatomy prior to taking any additional x-ray images. The performance testing demonstrated substantial equivalence of LessRay with Tracking to the previously cleared LessRay device. Therefore, based on the identical indication, similar technological characteristics, and results of performance testing, LessRay® with Tracking is substantially equivalent to LessRay® as it was cleared in K123226 and K132970.
<sup>&#</sup>x27; As evaluated by a human observer in a side by side visual comparison of 30 image pairs with and without LessRay processing.
<sup>4</sup> In clinical practice, the amount of image quality improvement achieved when a Pulsed and/or Low Dose image is processed with LessRay is dependent on the clinical task, patient size, anatomical location, and clinical practice. The dose should be set at a level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.