← Product Code [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB) · K142126

# EP NAVIGATOR REL. 5.0 (K142126)

_Philips Medical Systems Nederland B.V. · OWB · Dec 9, 2014 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K142126

## Device Facts

- **Applicant:** Philips Medical Systems Nederland B.V.
- **Product Code:** [OWB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB.md)
- **Decision Date:** Dec 9, 2014
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Intended Use

EP navigator is intended to provide navigation support for cardiovascular devices for heart rhythm disorders, such as catheters and guidewires, by superimposing 3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over intra-procedural X-ray images of the same anatomy.

## Device Story

EP navigator Rel. 5.0 is a software medical device operating on a PC-based platform; used in clinical settings to assist physicians during cardiovascular procedures. Input data includes previously acquired 3D anatomical images (CT, MRI, or 3D rotational scans) and real-time intra-procedural X-ray images. The software segments 3D datasets, registers them with X-ray images, and superimposes the 3D anatomy and landmarks onto the live X-ray view. Features include visual marker placement (tag points), EndoView for internal volume visualization, manual 2D measurements, and anatomical-based viewing planes. It communicates with Philips X-ray systems to enable Automatic Position Control (APC). By providing a 3D roadmap of the patient's cardiac anatomy overlaid on live fluoroscopy, the device aids navigation of catheters and guidewires, potentially improving procedural accuracy and efficiency for heart rhythm disorder treatments.

## Clinical Evidence

No clinical data; bench testing only. Verification and validation testing, including system verification and assessment of individual modifications, demonstrated compliance with IEC 62304, IEC 62366, ISO 14971, and NEMA PS 3.1-3.20 standards.

## Technological Characteristics

PC-based software medical device. Integrates with Philips X-ray systems. Features: 3D image segmentation (CT, MRI, rotational scan), registration, and superimposition. Connectivity: DICOM compliant. Standards: IEC 62304 (software lifecycle), IEC 62366 (usability), ISO 14971 (risk management), NEMA PS 3.1-3.20 (DICOM).

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- EP navigator Rel. 3.0 ([K101311](/device/K101311.md))
- HeartNavigator Rel. 2.0 ([K140138](/device/K140138.md))

## Reference Devices

- Allura Xper FD series ([K133292](/device/K133292.md))
- Allura Xper OR Table series ([K133292](/device/K133292.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Philips Medical Systems Nederland B.V. % Ms. Liselotte Kornmann Regulatory Affairs Manager Veenpluis 4-6 Best, 5684 PC Netherlands

December 9, 2014

Re: K142126 Trade/Device Name: EP Navigator rel. 5.0 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ Dated: November 14, 2014 Received: November 17, 2014

Dear Ms. Kornmann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K142126

Device Name

EP navigator

Indications for Use (Describe)

EP navigator is intended to provide navigation support for cardiovascular devices for heart rhythm disorders, such as catheters and guidewires, by superimposing 3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over intra-procedural X-ray images of the same anatomy.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.

# 510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.

| Date Prepared:                           | July 31, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                               |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                            | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                               |
| Contact Person:                          | Ms. Liselotte Kornmann, PhD<br>Regulatory Affairs Manager<br>Phone: +31 611621238<br>E-mail: Liselotte.Kornmann@philips.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                               |
| Device:                                  | Trade Name:<br>Common Name:<br>Device Class:<br>Classification Name:<br>Classification Regulation:<br>Classification Panel:<br>Primary Product Code:<br>Secondary Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | EP navigator Rel. 5.0<br>EP navigator<br>Class II<br>Image-intensified fluoroscopic x-ray system<br>21CFR §892.1650<br>Radiology<br>OWB (Interventional x-ray system)<br>LLZ (system, image processing, radiological)                                         |
| Primary1 Predicate<br>Device:            | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Allura Xper FD series<br>Allura Xper OR Table series<br>Philips Medical Systems Nederland B.V.<br>K133292 (March 05, 2014)<br>21 CFR, Part 892.1650<br>Image-intensified fluoroscopic x-ray system<br>Radiology<br>Class II<br>OWB (primary), JAA (secondary) |
| Secondary Predicate<br>Devices #1:       | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | EP navigator Rel. 3.0<br>Philips Medical Systems Nederland B.V.<br>K101311 (Sep 30, 2010)<br>21 CFR, Part 892.2050<br>Picture Archiving and Communications System<br>Radiology<br>Class II<br>LLZ                                                             |
| Secondary Predicate<br>Devices #2:       | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | HeartNavigator Rel. 2.0<br>Philips Medical Systems Nederland B.V.<br>K140138 (June 10, 2014)<br>21 CFR, Part 892.1650<br>Image-intensified fluoroscopic x-ray system<br>Radiology<br>Class II<br>OWB, LLZ                                                     |
| Device Description:                      | EP navigator Rel. 5.0 is a software medical device that contains image<br>processing algorithms that are executed on a PC-based hardware platform.<br>EP navigator Rel. 5.0 can perform the following functions:<br>view previously acquired 3D anatomical image data in a slice viewer, such<br>●<br>as CT, MRI, or 3D rotational scan,<br>segment previously acquired 3D anatomical image data,<br>●<br>register the segmented 3D dataset and landmarks with intra-procedural X-ray<br>●<br>images,<br>superimpose the segmented 3D dataset and landmarks on intra-procedural X-<br>●<br>ray images of the same anatomy,<br>position visual markers such as tag points, and<br>●<br>visualize the inside of the 3D volume (EndoView).<br>●                                                                                                                                                   |                                                                                                                                                                                                                                                               |
| Indications for Use:                     | EP navigator is intended to provide navigation support for cardiovascular devices<br>for heart rhythm disorders, such as catheters and guidewires, by superimposing<br>3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over<br>intra-procedural X-ray images of the same anatomy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                               |
| Fundamental<br>Scientific<br>Technology: | EP navigator Rel. 5.0 is intended to be used in combination with the primary<br>predicate device Allura X-ray system (K133292). While the secondary predicate<br>device EP navigator Rel. 3.0 (K101311) and EP navigator Rel. 5.0 share the<br>same technological characteristics, EP navigator Rel. 5.0 offers the following<br>enhancements to existing functions: (1) Support of the import and automatic<br>segmentation of MR datasets, (2) Reduced angular range of 3D EP rotational<br>scan and support of the automatic segmentation of a 3D EP rotational scan with<br>reduced angular angle, (3) Manual 2D measurements, (4) Anatomical-based<br>viewing planes, and (5) Communication with Philips X-ray system, including<br>Automatic Position Control (APC). The modifications represented by<br>EP navigator Rel. 5.0 do not affect the safety and effectiveness of the device. |                                                                                                                                                                                                                                                               |

<sup>1</sup> Note: the EP navigator Rel. 5.0 software medical device is an accessory to the currently marketed and primary predicate Allura Xper FD series and Allura Xper OR Table series K133292. The Allura Xper FD series and Allura Xper OR Table series will be abbreviated as Allura X-ray system in this 510(k) Premarket notification. Substantial equivalence demonstration to the primary predicate is not warranted as they are devices with which EP navigator Rel. 5.0 is to be used. Therefore, substantial equivalence is demonstrated to the secondary predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and centered in the image.

Therefore, EP navigator Rel. 5.0 is substantially equivalent to the currently marketed and secondary predicate EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of fundamental scientific technology.

Summary of Nonclinical Performance Data:

Non-clinical performance testing has been performed on EP navigator Rel. 5.0 and demonstrates compliance with the following International and FDArecognized consensus standards and FDA guidance document:

- IEC 62304 Medical device software Software life cycle processes (Ed. 1.0, . 2006).
- IEC 62366 Medical devices - Application of usability engineering to medical devices (Ed. 1.0, 2007),
- ISO 14971 Medical devices Application of risk management to medical ● devices (Ed. 2.0, corrected version, 2007),
- NEMA PS 3.1-3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2011), and
- . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005, document number 337).

Verification and validation tests have been performed to address intended use, the technical claims, requirement specifications and the risk management results. The modifications of the EP navigator Rel. 5.0 that are required to implement the enhancements are within the design controls and specifications; a system verification test was performed to ensure that the modifications are properly introduced; verification and validation testing was conducted to ensure the proper introduction of the individual modifications listed; sample clinical images as well as conformance to IEC standards and guidance documents were provided. All of these components and tests were used to support substantial equivalence of the subject device.

The test results in this 510(k) premarket notification demonstrate that EP navigator Rel. 5.0 complies with the aforementioned international and FDArecognized consensus standards and FDA guidance document, meets the acceptance criteria, and is adequate for its intended use. Additionally, all risks are sufficiently mitigated, no new risks are introduced and the overall residual risk is acceptable.

Therefore, EP navigator Rel. 5.0 is substantially equivalent to the currently marketed and secondary predicate devices EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of safety and effectiveness.

Summary of Clinical EP navigator Rel. 5.0 did not require clinical studies to demonstrate substantial Data: equivalence to the currently marketed and secondary predicate devices EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138).

The EP navigator Rel. 5.0 software medical device is substantially equivalent to Substantial Equivalence the currently marketed and secondary predicate devices EP navigator Rel. 3.0 Conclusion: (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with nonclinical performance testing, which complied with the requirements specified in

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The font is sans-serif and the background is white.

the international and FDA-recognized consensus standards. The non-clinical performance tests provided in this 510(k) premarket notification demonstrates that the proposed EP navigator Rel. 5.0 is as safe and effective as its predicate devices without raising any new safety and/or effectiveness concerns.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K142126](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OWB/K142126)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
