ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
K141979 · Philips Medical Systems Netherland BV · OWB · Aug 19, 2014 · Radiology
Device Facts
Record ID
K141979
Device Name
ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Applicant
Philips Medical Systems Netherland BV
Product Code
OWB · Radiology
Decision Date
Aug 19, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room. Allura Xper FD10 is compatible with specified magnetic navigation systems. Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR.
Device Story
Modular angiographic X-ray system; performs vascular, cardiovascular, neurovascular, and non-vascular imaging/interventions. Inputs: X-ray radiation; processed via ClarityIQ technology using XRES4 noise reduction algorithms to reduce quantum noise. Outputs: Fluoroscopic/angiographic images for clinical guidance. Used in hospitals/hybrid ORs by physicians (radiologists, cardiologists, surgeons). System includes C-arm configurations, X-ray detectors (FD15"), and OR tables. Healthcare providers use real-time imaging to guide minimally invasive procedures (stents, embolizations, etc.), improving procedural accuracy and patient outcomes.
Clinical Evidence
Bench testing only. No clinical trials performed. Substantial equivalence supported by system verification/validation testing, conformance to IEC 60601-2-43, IEC 60601-2-28, ISO 14971, and IEC 62304 standards, and sample clinical images reviewed by a board-certified radiologist to confirm image quality for neuroradiological procedures.
Technological Characteristics
Modular angiographic X-ray system; includes FD15" (PX2630SV) X-ray detector. Features ClarityIQ technology with XRES4 noise reduction. Connectivity: Networked workstation. Software: Windows Embedded Standard (WES) 2009. Standards: IEC 60601-2-43, IEC 60601-2-28, ISO 14971, IEC 62304.
Indications for Use
Indicated for human patients requiring vascular, cardiovascular, neurovascular, and non-vascular imaging, diagnostic, and interventional procedures (e.g., angiography, PTCA, stent placement, embolization, thrombolysis, pacemaker implantation, EP, drainage, biopsy, vertebroplasty). Compatible with hybrid OR environments and specific magnetic navigation systems.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name in a circular arrangement around a stylized symbol. The symbol features three overlapping profiles facing right, representing the department's focus on people and health. The text is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 19, 2014
Philips Medical Systems Netherland, B.V. % Jeanette Becker Regulatory Affairs Manager Veenpluis 4-6 5684 PC Best The Netherlands
Re: K141979
> Trade/Device Name: Allura Xper FD series / Allura Xper Or Table series Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA Dated: July 16, 2014 Received: July 21, 2014
Dear Ms. Becker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
Allura Xper series and Allura Xper OR Table series
Indications for Use (Describe)
The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform:
• Vascular, cardiovascular and neurovascular imaging applications, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.
· Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).
· Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
Additionally:
· The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room.
- · Allura Xper FD10 is compatible with specified magnetic navigation systems.
· Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | July 10, 2014 | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684-PC, Best, The Netherlands<br>Establishment Registration Number: 3003768277 | |
| Contact: | Mrs. Jeanette Becker<br>Regulatory Affairs Manager<br>+31 6 113 86380 | |
| Device Trade<br>Names: | Allura Xper FD series<br>Allura Xper OR Table series | |
| Classification | Classification Name: | Image-intensified fluoroscopic x-ray<br>system |
| | Classification Regulation: | 21 CFR 892.1650 |
| | Classification Panel: | Radiology |
| | Device Class: | II |
| | Primary product code: | OWB |
| | Secondary product code: | JAA |
| Predicate Device #1 | Device Name | Allura Xper FD series |
| | Manufacturer | Philips Medical Systems Nederland<br>B.V. |
| | 510(k) number | K133292 |
| | Classification Regulation | 21 CFR 892.1650 |
| | Device Class: | II |
| | Product Code: | Primary code: OWB<br>Subsequent code: JAA |
| Predicate Device #2 | Device Name | Allura Xper OR Tables Series |
| | Manufacturer | Philips Medical Systems Nederland<br>B.V. |
| | 510(k) number | K133292 |
| | Classification Regulation | 21 CFR 892.1650 |
| | Device Class: | II |
| | Product Code: | Primary code: OWB<br>Subsequent code: JAA |
| Device Description: | The Allura Xper family consists of the Allura Xper FD series and<br>the Allura Xper OR Table series and is identified as Allura Xper FD<br>R8.2.1. The Allura Xper FD R8.2.1 is a modular angiographic X-ray<br>system, based on a set of components that can be combined into<br>different single and biplane configurations to provide specialized<br>angiography. Combined with a qualified, compatible OR table, the<br>Allura Xper FD R8.2.1 can also be used for imaging in the Hybrid<br>OR. The Allura Xper FD R8.2.1 is optionally provided with<br>ClarityIQ technology, which utilizes the advanced XRES4 noise<br>reduction algorithms to reduce quantum noise in X-ray images. | |
| Indications for Use: | The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform:<br>Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolization and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP). Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room. Allura Xper FD10 is compatible with specified magnetic navigation systems. Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular, Non Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR. | |
| Technology: | The Allura Xper FD R8.2.1 has the same technological characteristics compared to the predicate devices. Modifications implemented in the Allura Xper FD R8.2.1 include: The addition of a FD15" (PX2630SV) X-ray detector, which has been cleared on July 10, 2014, as part of Philips' Veradius system with 510(k) number K133819, A Low Load Uninterruptable Power Supply (UPS) has been added as an option to the system, and Software updates for: Use of FD15" (PX2630SV) X-ray detector, Change in windows Host OS from XP to Windows Embedded Standard (WES) 2009, Elimination of unexpected C-arm movement during startup of the system, Enhancement of an overlay functionality on an XtraVision 3D-workstation for 3D-image guidance software applications, Support of non-US language keyboard layouts, and Sorting Photo files of a Biplane system when they | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The text is a vibrant blue color, contrasting against a white background. The letters are evenly spaced and appear to be a logo or brand name.
Based on the information provided in this premarket notification, the Allura Xper FD R8.2.1 is considered substantially equivalent to the
System (PACS).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
currently marketed and predicate devices in terms of:
- Design and functionality
- · Indications for use
- · Fundamental Scientific Technology
- · Performance specifications and testing
#### Non-clinical Performance Data:
The Allura Xper R8.2.1 complies with the following international and FDA recognized consensus standard and FDA Guidance Documents:
- IEC 60601-2-43. ●
- IEC 60601-2-28, ●
- ISO 14971: ●
- IEC 62304,
- . FDA Guidance document entitled, "Guidance for the Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.
- . FDA Guidance document entitled, "General Principals of software Validation; Final Guidance for Industry and FDA Staff" issued January 11, 2002.
- FDA Guidance document entitled: "Guidance for the submission of 510(k)'s for solid state X-ray Imaging Devices" issued August 6, 1999
The test results demonstrate that the Allura Xper FD R8.2.1.
- Complies with the aforementioned international and FDA-● recognized consensus standards and/or FDA guidance documents
- . Meets the acceptance criteria and is adequate for its intended use.
Therefore, the Allura Xper FD R8.2.1 is substantially equivalent to the currently marketed and predicate devices in terms of safety and effectiveness.
The subject of this premarket submission, the Allura Xper FD Clinical Performance Data: R8.2.1., did not require clinical studies to support substantial equivalence. Sample clinical images accompanied by a signed statement of a board certified radiologist are included in this 510(k) premarket notification to demonstrate that the images made with the FD15" (PX2630SV) X-ray detector of the Allura Xper FD R8.2.1 are of adequate quality for use in neuroradiological procedures.
Conclusion: The Allura Xper FD R8.2.1 is substantially equivalent to the predicate devices in terms of design features, fundamental scientific technology, indications for use and safety and effectiveness.
- The modifications of the Allura Xper FD R8.2.1 are within ● the design controls and specifications;
- . A system verification test was performed to ensure that the modifications are properly introduced;
- Verification and validation testing was conducted to ensure the proper introduction of the individual modifications
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered horizontally and vertically in the image. The background is plain white.
listed;
- . Sample clinical images as well as conformance to IEC standards and guidance documents were provided. All of the aforementioned components and tests were used to support substantial equivalence of the Allura Xper FD R8.2.1.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.