K131263 · Philips Medical Systems Nederland B.V. · OWB · Jul 24, 2013 · Radiology
Device Facts
Record ID
K131263
Device Name
XPERGUIDE
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OWB · Radiology
Decision Date
Jul 24, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
XperGuide is an extension of XperCT, which assists in percutaneous interventions, such as biopsies, drainages, etc., by providing needle path planning and image guidance by superimposing live fluoroscopic images of the needle on a cross sectional image of the targeted anatomy. The XperGuide Ablation option assists with planning the position of multiple needles by visualizing the ablation coverage of the lesion.
Device Story
XperGuide is a software medical device for percutaneous interventions (biopsies, drainages). It inputs live 2D fluoroscopic images from Philips Allura Xper FD X-Ray systems and 3D cross-sectional anatomy data (from XperCT, CT, or MR). The software overlays live 2D needle images onto 3D planning volumes to provide real-time needle position feedback. An ablation option visualizes combined ablation zones for multiple needles. Used in interventional suites by physicians; operated via control room keyboard/mouse or exam room Xper module. Output displayed on monitors for guidance during procedures. Benefits include improved needle placement accuracy and procedural planning.
Clinical Evidence
Bench testing only. Software verification and validation testing performed per IEC 62304, IEC 62366, and ISO 14971. Results demonstrate compliance with requirement specifications and acceptance criteria.
Technological Characteristics
Software-based medical device hosted on Philips Interventional Workspot platform. Integrates with digital X-ray imaging systems via real-time link. Features 2D/3D image fusion, needle path planning, and ablation zone visualization. Complies with IEC 62304, IEC 62366, and ISO 14971 standards.
Indications for Use
Indicated for patients undergoing percutaneous interventions, including biopsies and drainages, requiring needle path planning and image guidance.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## K131263
Page 1 og 3
| | JUL 2 4 2013 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | This 510(k) Summary of Safety and Effectiveness has been prepared in accordance with Code of Federal Regulations, Title 21 CFR, Part 807.92. |
| Date Prepared: | April 25, 2013 |
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6, 5684PC Best, The Netherlands<br>Establishment registration number: 3003768277 |
| Contact Person (Owner): | Dr. Jos van Vroonhoven<br>Phone: +31 40 2791111 Fax: +31 40 2769100 |
| Proposed Device: | |
| Trade name: | XperGuide |
| Classification regulation: | 21 CFR, Part 892.1650 |
| Classification name: | Image-intensified fluoroscopic x-ray system |
| Classification panel: | Radiology |
| Device classification: | Class II |
| Primary Product Code: | OWB (interventional X-ray system) |
| Secondary Product Code: | LLZ (radiological image processing system) |
| Primary Predicate Device 1: | |
| Trade name: | Allura Xper FD20 X-Ray Imaging Systems |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| FDA clearance: | * K033737 (December 9. 2003) |
| Classification regulation: | 21 CFR. Part 892.1650 |
| Classification name: | Image-intensified fluoroscopic x-ray system |
| Classification panel: | Radiology |
| Device classification: | Class II |
| Product Code: | OWB (interventional X-ray system) |
| Primary Predicate Device 2: | |
| Trade name: | Allura Xper FD OR Table Series |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| FDA clearance: | K102005 (August 9, 2010) |
| Classification regulation: | 21 CFR. Part 892.1650 |
| Classification name: | Image-intensified fluoroscopic x-ray system |
| Classification panel: | Radiology |
| Device classification: | Class II |
| Product Code: | OWB (interventional X-ray system) |
{1}------------------------------------------------
K131263
page 2063
Note: The primary predicate devices 1 and 2 (Allura Xper FD20 X-Ray Imaging Systems and Allura Xper FD OR Table Series) are collectively referred to as "Philips Allura Xper FD X-Ray Imaging Systems" in this 510(k) Summary of Safety and Effectiveness.
## Secondary Predicate Device:
| Trade name: | Innova Vision Applications |
|----------------------------|---------------------------------------------|
| Manufacturer: | GE Healthcare |
| FDA clearance: | K092639 (December 2, 2009) |
| Classification regulation: | 21 CFR, Part 892.2050 |
| Classification name: | Picture archiving and communications system |
| Classification panel: | Radiology |
| Device classification: | Class II |
| Product Code: | LLZ (radiological image processing system) |
XperGuide is a software medical device intended to assist the physician during Device Description: percutaneous interventions by providing live 3D needle image guidance. XperGuide overlays live 2D fluoroscopic images on a 3D reconstruction of the anatomy. XperGuide provides real-time feedback on needle position with respect to the planned path. In addition, the XperGuide has an ablation option, which visualizes the combined ablation zone of multiple planned needles.
XperGuide is an extension of XperCT, which assists in percutaneous Intended Use: interventions, such as biopsies, drainages, etc., by providing needle path planning and image guidance by superimposing live fluoroscopic images of the needle on a cross sectional image of the targeted anatomy. The XperGuide Ablation option assists with planning the position of multiple needles by visualizing the ablation coverage of the lesion.
- Technology: XperGuide is provided on the independent hosting software functionality platform of the currently marketed Philips Interventional Workspot (K121296, January 2. 2013). Live 2D fluoroscopy images and exam data are transmitted from the currently marketed and predicate Philips Allura Xper FD X-Ray Imaging Systems (K033737. K102005) to Interventional Workspot through a dedicated real-time link and are overlaid on a slab of the 3D planning volume. either created by the currently marketed XperCT (K060749. April 4. 2006) or from a previously acquired CT or MR data set. Live 2D/3D fluoroscopic overlay runs, together with planning information. snapshots and movies. can be stored in a local database for use in XperGuide and can be reviewed at any time. Live 2D/3D overlay images and the XperGuide user interface are shown on monitors in the control room and in the exam room with use of a video splitter. XperGuide can be operated from the control room (via keyboard and mouse) as well as in the exam room (via the Xper module of the currently marketed and predicate Philips Allura Xper FD X-Ray Imaging Systems).
{2}------------------------------------------------
The technology utilized in both XperGuide and the currently marketed and predicate Innova Vision Applications with TrackVision option (GE Healthcare, K092639) includes:
- · Execution on and connection to an independent software hosting platform;
- · Transmission of live 2D fluoroscopy images and exam data from a digital X-ray imaging system to the software hosting platform (workstation) through a dedicated real-time link;
- · Loading of previously acquired 3D data sets from the software hosting platform (workstation);
- · Fusion or superimposing live 2D fluoroscopy images on top of selected 3D cross-sectional images of the targeted anatomy;
- . Storage of live 2D/3D overlay runs and of fused photos and video clips in a local database;
- · Display of overlay images and user interface in the control room and in the exam room;
- · Operation of the software medical device from the control room and in the exam room.
Based on the information provided above, XperGuide is considered substantially equivalent to the currently marketed and predicate Innova Vision Applications with Track Vision option (GE Healthcare. K 092639) in terms of technology.
## Non-clinical Performance Data:
The proposed XperGuide software medical device complies with the following international FDA-recognized consensus standards:
- · IEC 62304 Medical device software Software life cycle processes (2006);
- IEC 62366 Application of usability engineering to medical devices (2007);
- · ISO 14971 Application of risk management to medical devices (2007).
Non-clinical verification and validation tests have been performed with regards to the requirement specifications and the risk management results. These tests specifically include software verification and validation conformance testing, The test results demonstrate that XperGuide meets the acceptance criteria, and is adequate for its intended use. Therefore, the XperGuide software medical device is substantially equivalent to the currently marketed and predicate Innova Vision Applications with Track Vision option (GE Healthcare, K092639) in terms of safety and effectiveness.
- XperGuide is substantially equivalent to the currently marketed and predicate Conclusion: Innova Vision Applications with Track Vision option (GE Healthcare, K092639) in terms of design features. fundamental scientific technology, indications for use. and safety and effectiveness.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
f
July 24, 2013
Philips Medical Systems Nederland B.V. % Dr. Jos van Vroonhoven Standardization Manager Veenpluis 4-6 5684 PC Best NETHERLANDS
Re: K131263
Trade/Device Name: XperGuide Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ Dated: April 25, 2013 Received: May 03, 2013
Dear Dr. Vroonhoven:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Dr. Vroonhoven
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDcvices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K131263 510(k) Number (if known):
Device Name: XperGuide
Indications for Use:
XperGuide is an extension of XperCT, which assists in percutaneous interventions, such as biopsies, drainages, etc.. by providing needle path planning and image guidance by superimposing live fluoroscopic images of the needle on a cross sectional image of the targeted anatomy.
The XperGuide Ablation option assists with planning the position of multiple needles by visualizing the ablation coverage of the lesion.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$m(h,7)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) ======================================================================================================================================================================= K131263
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.