K122644 · Siemens Medical Solutions USA, Inc. · OWB · May 16, 2013 · Radiology
Device Facts
Record ID
K122644
Device Name
ARTIS ZEE/ZEEGO WITH CSX-10 DETECTOR SW VC21
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
May 16, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational multipurpose angiography, whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. Artis zee / zeego can also support the acquisition triggered imaging for spatial data synthesis. The Artis zee and Artis zeego include also the software option DynaCT which identifies the Artis as a system with a C-arm CBCT functionality. DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other body structures for diagnosis, surgical planning, interventional internal procedures and treatment follow-up.
Device Story
Artis zee/zeego is a modular angiographic X-ray system utilizing a C-arm, X-ray tube, and CSX-10 flat panel detector. It captures 2D X-ray projections during diagnostic and interventional procedures in clinical or OR settings. Operated by healthcare professionals, the system supports image fusion, navigation, and spatial data synthesis. The DynaCT software option transforms 2D projections into 3D volumes (CBCT functionality) for hard/soft tissue visualization. Output is displayed for clinicians to guide surgical planning, interventions, and follow-up. Benefits include enhanced visualization for complex procedures and support for image-guided surgery.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness were established through hazard analysis, software verification, and validation testing.
Technological Characteristics
Modular angiographic system comprising C-arm, X-ray tube, generator, collimator, and CSX-10 flat panel detector. Utilizes X-ray energy for imaging. Supports 2D fluoroscopy/radiography and 3D CBCT reconstruction via DynaCT software. System is networked for image processing and display. Sterilization not applicable (non-invasive).
Indications for Use
Indicated for single and biplane diagnostic imaging and interventional procedures, including pediatric and obese patients. Procedures include cardiac, neuro, general, and rotational angiography, whole-body radiographic/fluoroscopic imaging, and image-guided surgery. DynaCT software is indicated for 3D reconstruction of 2D X-ray images for hard and soft tissue imaging, surgical planning, and treatment follow-up.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Artis zee, Artis zeego Angiographic X-ray Systems (K090745)
Submission Summary (Full Text)
{0}------------------------------------------------
K122644
SIEMENS
MAY 1 6 2013
## 510(k) Summary: Artis zee/zeego with CSX-10 Detector VC21
Company:
Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: May 14, 2013
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Systems, Inc. 51 Vallev Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869 Manufacturing Site: SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany Establishment Registration Number: 2240869
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist. Requlatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com
#### 3. Device Name and Classification: Trade Name:
Classification Name:
Classification Panel: Classification Regulation: Device Class: Product Code: Secondary Product Code:
Artis zee and Artis zeego - Modular Angiographic System Angiographic X-Ray System Image intensified fluoroscopic x-ray system Radiology 21 CFR §892.1650 Class II OWB, JAA, IZI JAK
{1}------------------------------------------------
# SIEMENS
- Legally Marketed Predicate Device 4. Trade Name:
510(k) Clearance Clearance Date Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
· Artis zee and Artis zeego - Modular Angiographic System K090745 June 18, 2009 Angiographic X-Ray System Radiology 21 CFR §892.1600 Class II 90 İZİ
#### 5. Device Description:
The Artis zee/zeego Modular Angiography System is designed as a set of components that may be combined into different configurations to provide specialized angiography systems. A new Flat Panel detector has been added to the systems.
The Artis zee / zeego Modular Angiography System with the new Flat Panel detector is substantially equivalent to the AXIOM Artis Modular Angiography System VC14 with all its components as described in the Device Description, Section 10 and the Substantial Equivalence Section 11.
#### 6. Indication for Use:
Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including. but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational multipurpose anqiography, whole bodv radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.
Artis zee / zeego can also support the acquisition triggered imaging for spatial data synthesis.
The Artis zee and Artis zeego include also the software option DynaCT which identifies the Artis as a system with a C-arm CBCT functionality.
{2}------------------------------------------------
DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other body structures for diagnosis, surgical planning, interventional internal procedures and treatment follow-up.
#### 7. Substantial Equivalence:
The Artis zee / zeego Modular Angiography System with CSX-10 Panel detector is a modification of a legally marketed device and substantial equivalent to Artis zee, Artis zeego SW VC14 as listed below. .
| 510(k) Number | Date of Clearance | Device Name |
|---------------|-------------------|---------------------------------------------------|
| K090745 | June 18, 2009 | Artis zee, Artis zeego Angiographic X-ray Systems |
### Summary of Technological Characteristics of the Subject Device as 8. Compared with the Predicate Device:
Artis zee / zeego Modular Angiography System is designed as a set of components (C-arm, X-ray tube and housing, flat detector, digital imaging system, collimator, generator etc.) that may be combined into different configurations to provide specialized angiography systems. Many of the components used with Artis zee / zeego are either commercially available with current Siemens systems or include minor modifications to existing components. New or modified features provided with Artis zee/zeego and the new Flat Panel detector are provided in the Device Description.
#### 9. General Safetv and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
{3}------------------------------------------------
SIEMENS
#### 10. Conclusion as to Substantial Equivalence:
The Artis zee / zeego Modular Angiography System with CSX 10 Flat Panel detector has the same indication for use as the predicate device Artis zee / zeego with software VC14. The new detector is designed to provide fluoroscopic and spot-film radiographic images of human anatomy during diagnostic, surgical and interventional procedures.
The functionality of Artis zee / zeego Modular Angiography System is similar to the predicate device. It is Siemens opinion, that the Artis zee / zeego Modular Angiography System with the new detector is substantially equivalent to the Artis zee / zeego Modular Angiography System (K090745).
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms or lines extending upwards, representing growth and vitality. The figure is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 16, 2013
Siemens Medical Solutions, Inc. % Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions 51 Valley Stream Parkway MALVERN PA 19355
Re: K122644
Trade/Device Name: Artis zee/zeego with CSX-10 Detector SW VC21 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB Dated: May 3. 2013 Received: May 6, 2013
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Jones
Enclosure
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sm.h.7)
. for
Janine M. Morris Director. Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Artis zee and Artis zeego with CSX 10 Detector VC21
Indications for Use:
Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuroangiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions
Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis.
The Artis zee and Artis zeego include also the software option DynaCT which identifies the Artis as a system with a C-arm CBCT functionality.
DynaCT is an x-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format.
DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.
| Prescription Use<br>__________ | AND/OR | Over-The-Counter Use<br>__________ |
|--------------------------------|--------|------------------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Smir. Fil
> (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K122644 510(k)________________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.