GE INNOVA/INNOVA IGS/DISCOVERY IGS/OPTIMA ANGIOGRAPHIC, FLUOSCOPIC X-RAY SYSTEMS WITH CATHLAB FRONTIERS SOLUTIONS
K122457 · GE Healthcare · OWB · Jan 2, 2013 · Radiology
Device Facts
Record ID
K122457
Device Name
GE INNOVA/INNOVA IGS/DISCOVERY IGS/OPTIMA ANGIOGRAPHIC, FLUOSCOPIC X-RAY SYSTEMS WITH CATHLAB FRONTIERS SOLUTIONS
Applicant
GE Healthcare
Product Code
OWB · Radiology
Decision Date
Jan 2, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathlab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathlab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: (1) communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices, (2) communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface, (3) interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.
Device Story
System comprises GE angiographic/fluoroscopic X-ray systems integrated with 'Cathlab Frontiers' solutions; enables interoperability with other FDA-cleared imaging/data medical devices in cathlab environments. Input: patient/exam data and imaging signals from integrated devices. Transformation: Cathlab Frontiers implements communication protocols for automatic synchronization of patient/exam info, remote control of integrated device functions via X-ray system UI, and display of integrated device output on X-ray system monitors. Used in cathlabs/ORs by clinicians; simplifies end-to-end workflow. Output: synchronized data and integrated visual displays. Benefits: improved clinical workflow efficiency and centralized control/viewing of multi-modality data during interventional/surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, module verification, system integration testing, performance/safety verification, and simulated use validation.
Technological Characteristics
Solid-state X-ray imaging; digital flat-panel detector. Cathlab Frontiers provides communication protocols (data exchange, device control) and display interfaces for third-party medical devices. Complies with voluntary standards for radiation-emitting products and electrical/electromagnetic safety.
Indications for Use
Indicated for patients newborn to geriatric requiring fluoroscopic and rotational imaging for cardiovascular, vascular, non-vascular diagnostic/interventional procedures, and image-guided surgery.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Innova Solid State X-ray Imager Fluoroscopic X-ray System (K113034)
GE Discovery IGS angiographic, fluoroscopic X-Ray System (K113403)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2016
Ms. Dorai Subramaniam Regulatory Affairs Leader 283, RUE DE LA MINIERE BUC. 78530 FRANCE
Re:K122457
Trade/Device Name: GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray Systems with Cathlab Frontiers solutions Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, IZI Dated: December 12, 2012 Received: December 19, 2012
Dear Ms. Dorai Subramaniam:
This letter corrects our substantially equivalent letter of January 2, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oaks
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known): K122457
### Device Name: GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray systems with Cathlab Frontiers Solutions
Indications for Use:
The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures.
The OR table is suitable for interventional and surgical procedures.
The Cathlab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution.
The Cathlab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing:
- (1) communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices,
- (2) communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface,
- (3) interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.
Prescription Use X AND/OR Over-The-Counter Use. (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Janine M. Morris -S 2013.01.02 17:38:37 -05'00'
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K122457 510(k)
Page 1 of .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. There are decorative swirls around the circle, giving it a dynamic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.
# Section 5: 510(k) Summary
GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray systems with Cathlab Frontiers solutions
5.1 510(k) Summary
5-2
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white.
# GE Healthcare 510(k) Premarket Notification Submission
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 10 August 2012 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>GE Medical Systems, SCS<br>283 rue de la Miniere<br>Buc, FRANCE, 78530<br>T: +33-01-30-70-42-07 |
| Primary Contact Person: | Dorai Subramaniam<br>Regulatory Affairs Leader<br>GE Healthcare, (GE Medical Systems, SCS)<br>T: +91-80-4088-3769<br>Email: dorai.subramaniam@ge.com |
| Secondary Contact Person: | Carol Alloian<br>Regulatory Affairs Leader<br>GE Healthcare -Americas,<br>9900 W innovation drive<br>Wauwatosa, WI, USA, 53226-4856<br>T: (847) 244-8327<br>F: (847) 589-8524<br>Email: carol.alloian@ge.com |
| Device:<br>Trade Name: | GE Innova/Innova IGS/Discovery IGS/Optima angiographic,<br>fluoroscopic X-ray systems with Cathlab Frontiers solutions<br>Innova 4100-IQ, Innova 3100-IQ, Innova 2100-IQ, Innova 2121-<br>IQ, Innova 3131-IQ, Innova IGS 540, Innova IGS 530, Innova<br>IGS 520, Innova IGS 620, Innova IGS 630, Discovery IGS 730,<br>Optima CL320i, Optima CL323i. |
| Common/Usual Name: | Interventional fluoroscopic x-ray system, angiographic x-ray system |
| Regulation Description: | Image-intensified fluoroscopic x-ray system |
| Regulation number: | 892.1650 |
| Product Code: | OWB, JAA and IZI |
| Class: | II |
| Predicate Device(s): | K113034 : GE Innova Solid State X-ray Imager Fluoroscopic X-<br>ray System<br>K111209 : Integrated Innova - S5i System Option<br>K113403 : GE Discovery IGS angiographic, fluoroscopic X-Ray<br>System |
| Device Description: | The basis for this 510(k) submission is a modification of the<br>family of legally marketed GE angiographic, fluoroscopic X-ray<br>systems devices to expand their indications for use regarding the<br>generic integration interfaces for other imaging / data medical<br>devices used in interventional and surgical cathlab environments<br>and cleared/approved for commercial distribution. |
| | Same time with the intended use/indications for use expansion we<br>introduce the Integrated Vivid E9 solution. The Vivid E9 (BT12<br>version) integrates with Innova IGS 520, Innova IGS 530, and<br>Innova IGS 540 systems, integration consisting in displaying the<br>Vivid E9 system screen on Innova IGS Large Display Monitor<br>solution. |
| | For the Innova IGS Vivid E9 integration we apply the specific<br>qualifications criteria and processes described in the section<br>"Determination of Substantial Equivalence" below. |
| Intended Use: | The angiographic X-ray systems are indicated for use for patients<br>from newborn to geriatric in generating fluoroscopic and<br>rotational images of human anatomy for cardiovascular, vascular<br>and non-vascular, diagnostic and interventional procedures. |
| | Additionally, with the OR table, the angiographic X-ray systems<br>are indicated for use in generating fluoroscopic and rotational<br>images of human anatomy for image-guided surgical procedures. |
| | The OR table is suitable for interventional and surgical<br>procedures. |
| | The Cathlab Frontiers solutions are indicated for use in<br>conjunction with single plane and biplane GE angiographic X-ray<br>systems and imaging / data medical devices used in interventional<br>and surgical Cathlab environments and cleared for commercial<br>distribution. |
| | The Cathlab Frontiers solutions are integrated GE angiographic<br>X-ray and imaging / data medical devices that simplify the end- |
| | to-end clinical workflow in the Cathlab by implementing: |
| | (1) communication protocols for exchanging and automatically<br>synchronizing patient, exam, system, and image information<br>between the angiographic X-ray systems and the imaging / data<br>medical devices, |
| | (2) communication protocols for the control of imaging / data<br>medical device functions from the angiographic X-ray systems<br>user interface, |
| | (3) interfaces for displaying the imaging / data medical device<br>output on the monitor display solutions of the GE angiographic<br>X-ray systems. |
| Technology: | The GE Angiographic, Fluoroscopic X-ray Systems employs the<br>same Solid State X-ray Imaging and Digital Flat-Panel Detector<br>technology as its predicate devices. |
| Determination of | Summary of Non-Clinical Tests: |
| Substantial Equivalence: | The Integration has been evaluated for electrical and<br>electromagnetic safety. The GE Angiographic, Fluoroscopic X-<br>ray Systems and its applications and accessories comply with<br>voluntary standards and applicable performance standards for<br>radiation emitting products of this premarket submission. The<br>following quality assurance measures were applied to the<br>development of the system:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Testing on unit level (Module verification)<br>• Integration testing (System verification)<br>• Performance testing (Verification) |
| | • Safety testing (Verification)<br>• Simulated use testing (Validation) |
| | Summary of specific qualification of 3rd party systems<br>integrations:<br>The criteria to qualify the integration of 3rd party imaging/data<br>medical devices are based on the following aspects:<br>• Shall be "FDA cleared/approved device"<br>• Shall comply with one or more generic integration<br>interfaces offered by the GE angiographic, fluoroscopic<br>X-ray systems. |
| | Integration of devices shall be safe and effective and shall not raise different questions of safety and effectiveness than the predicate device(s). |
| | Summary of Clinical Tests: |
| | The subject of this premarket submission, Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray Systems with Cathlab Frontiers solutions, did not require clinical studies to support substantial equivalence as one of the criteria for integration rely on the fact that they are already cleared /approved devices by FDA. Integration does not introduce any new clinical information and the clinical information is pre-existing on the cleared or approved devices. |
| Conclusion: | GE Healthcare considers the GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray systems with Cathlab Frontiers solutions to be as safe, as effective, and performance is substantially equivalent to the predicate device(s). |
Section 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The circle has a stylized, swirling design around the letters, giving it a distinctive and recognizable appearance.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue.
# GE Healthcare 510(k) Premarket Notification Submission
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" written in a stylized script in the center. The circle is surrounded by a swirling pattern that resembles water or wind. The logo is simple and recognizable, and it is often used to represent the company's products and services.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.