CANON DIGITAL FLUOROSCOPY AND RADIOGRAPHY, MODEL CXDI-50RF
K092439 · Virtual Imaging, Inc. · OWB · Nov 30, 2009 · Radiology
Device Facts
Record ID
K092439
Device Name
CANON DIGITAL FLUOROSCOPY AND RADIOGRAPHY, MODEL CXDI-50RF
Applicant
Virtual Imaging, Inc.
Product Code
OWB · Radiology
Decision Date
Nov 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The CXDI-50RF is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Device Story
The Canon CXDI-50RF is a portable digital X-ray receptor panel designed for clinical environments. It utilizes a LANMIT (Large Area New MIS Sensor and TFT) sensor to capture X-ray radiation, directly converting it into high-resolution digital signals. The device supports both static and dynamic (moving) image acquisition. It is intended to replace traditional image intensifier technology in fluoroscopic applications. Operated by healthcare professionals, the device provides digital images that assist clinicians in vascular, cardiac, and interventional procedures. By providing direct digital output, it facilitates real-time visualization of human anatomy, supporting diagnostic accuracy and procedural guidance.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including performance testing and software validation, to demonstrate safety and effectiveness and substantial equivalence to predicate devices.
Technological Characteristics
Portable digital X-ray receptor panel; LANMIT (Large Area New MIS Sensor and TFT) sensor technology; direct conversion of X-rays to digital signals; supports static and dynamic imaging; complies with US Performance Standard for radiographic equipment.
Indications for Use
Indicated for human patients requiring fluoroscopic imaging for vascular angiography, cardiology, diagnostic, and interventional procedures. Not indicated for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K092439
# 510(k) SUMMARY
NOV 3 0 2009
Image /page/0/Picture/3 description: The image shows the logo for Virtual Imaging, a Canon USA company. The logo features a stylized "V" shape on the left, followed by the words "VIRTUAL IMAGING" in bold, sans-serif font. Below the logo is the text "A CANON USA Company" in a smaller font size.
Image /page/0/Picture/4 description: The image shows the word "Canon" in a stylized font. The letters are outlined and filled with a pattern that resembles a mosaic or a textured surface. The overall design gives the word a three-dimensional appearance.
### 720 S. Powerline Road, Suite E Deerfield Beach, FL 33442 954-428-6191 {Office}
## 954-428-6195 (Fax)
Page 1 of 2, 510(k) Summary: Canon Dynamic/Static DR Fluoroscopic Digital X-Ray Receptor Panel
Model CXDI-50RF
| Date Prepared | July 30, 2009 |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary prepared by: | Chris Duca, Chief Operating Officer |
| Device Name | Canon Dynamic/Static DR<br>Fluoroscopic Digital X-Ray Receptor Panel |
| Trade Name | Model CXDI-50RF |
| Common Name | Fluoroscopic Digital X-Ray Receptor Panel |
| Classification | Class: II<br>Product Code: MQL OWB, JAA, I2I<br>Regulation: 21 CFR § 892.1650 |
| Identification of<br>Predicate Devices and<br>Summary of Substantial<br>Equivalence | Siemens AXIOM Luminos dRF, K062623<br>GE Innova 4100 K023178/K091658<br>Canon CXDI-50C K060433 |
| Device Description | The Canon Dynamic/Static DR CXDI-50RF is a portable digital<br>radiography that can take images of any part of the body.<br>It directly converts the X-ray images captured by the LANMIT<br>(Large Area New MIS Sensor and TFT) sensor into a high-resolution<br>digital images. The instrument is suited for use inside a patient<br>environment. This unit converts the X-rays into digital signals. The<br>unit can acquire still and moving images. |
{1}------------------------------------------------
#### Page 2 of 2, 510(k) Summary: Canon Dynamic/Static DR Fluoroscopic Digital X-Ray Receptor Panel Model CXDI-50RF
| Intended Use and Indications | The CXDI-50RF is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications. |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics and Substantial Equivalence | Comparison with the predicate shows the technological characteristics of the CXDI-50RF are equal to or better than the predicate device. The units are functionally identical. |
| Performance Testing/Data | Tests were performed on the device which demonstrated that the device is safe and effective, performs comparably to and is substantially equivalent to the predicate device.<br>Tests include: Performance testing and Software Validation. Electrical safety and Electromagnetic Compatibility testing has been performed. The unit complies with the US Performance Standard for radiographic equipment. |
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Virtual Imaging, Inc. % Mr. Daniel Kamm, P.E. Submission Correspondent Kamm & Associates 8726 Ferrara Ct. NAPLES FL 34114
JUL 3 0 2012
Re: K092439
Trade/Device Name: Canon Dynamic/Static DR Model CXDI-50RF Fluoroscopic Digital X-Ray Receptor Panel
Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA and IZI Dated: October 28, 2009 Received: November 3, 2009
Dear Mr. Kamm:
This letter corrects our substantially equivalent letter of November 30. 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting of
{3}------------------------------------------------
#### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Mdrri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known): K092439 Device Name: Canon Dynamic/Static DR Model CXDI-50RF Fluoroscopic Digital X-Ray Receptor Panel
The CXDI-50RF is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Prescription Use _X_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Page 1 of 1
ductive, Abdominal
10(k) **Number**