K232710 · Yian Medical Technology (Haining) Co., Ltd. · OAS · Jan 9, 2024 · Radiology
Device Facts
Record ID
K232710
Device Name
Dental Cone-beam Computed Tomography
Applicant
Yian Medical Technology (Haining) Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Jan 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Dental Cone-beam Computed Tomography (Model:iDT901X1) is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
Device Story
Device (iDT901X1) performs CBCT, panoramic, and cephalometric radiography; inputs X-ray cone-beam radiation through dental/maxillofacial anatomy; transforms signals via FPD (TFT/CsI) receptor and image processing workstation into digital images. Used in clinical settings by healthcare professionals; output displayed on monitor for diagnostic assessment. Facilitates clinical decision-making by providing anatomical visualization; benefits patients through diagnostic support for dental/oral/maxillofacial conditions. System comprises frame, X-ray generator, receptor, and software workstation.
Clinical Evidence
Clinical evaluation based on review of sample clinical images by a US radiation-board certified radiologist. Radiologist independently scored essential image quality items across various scanning modes for both adult and pediatric cases. Results indicated acceptable overall image quality for all cases and image types.
Technological Characteristics
CBCT X-ray system; FPD (TFT) with CsI scintillator; 60-120kV tube voltage; 1-20mA tube current; 0.5mm focal spot. Connectivity via DICOM. Software workstation for image acquisition/processing. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-63, and ISO 10993 series for biocompatibility.
Indications for Use
Indicated for adult and pediatric patients (≥21 kg, ≥113 cm) requiring 2D (panoramic, cephalometric) or 3D digital X-ray imaging of the dental, oral, and maxillofacial regions for diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 9, 2024
Yian Medical Technology (Haining) Co., Ltd. % Lili Yan Director of Regulatory Affair Center Yian Medical Technology (Haining) Co., Ltd 1st Floor Area 1, 2nd Floor Area 1, Building A, No. 2 Caohejing Road, Haining Economic Development Zone JIAXING, ZHEJIANG 314400 CHINA
Re: K232710
Trade/Device Name: Dental Cone-beam Computed Tomography Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: December 14, 2023 Received: December 26, 2023
Dear Lili Yan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232710
Device Name Dental Cone-beam Computed Tomography
#### Indications for Use (Describe)
Dental Cone-beam Computed Tomography (Model:iDT901X1) is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo for Yian Medical. The logo consists of a stylized letter "N" in green, followed by the Chinese characters for "Yian Medical" in gray. Below the Chinese characters is the English text "YIAN MEDICAL" in gray.
# 510(k) Summary K232710
| Date Prepared: | Jun 30, 2023 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Yian Medical Technology (Haining) Co., Ltd<br>1st Floor Area 1, 2nd Floor Area 1, Building A, No.<br>2 Caohejing Road, Haining Economic<br>Development Zone, Haichang Street, Haining City,<br>Jiaxing City, Zhejiang Province, China |
| Contact Person: | Lili YAN<br>Director of Regulatory Affair Center<br>Yian Medical Technology (Haining) Co., Ltd<br>Tel: +86-0573-89739736 |
E-mail: lili.yan@yian-medical.com
### Identification of the Device:
| Proprietary/Trade Name: | / |
|-------------------------|--------------------------------------|
| Device Name: | Dental Cone-beam Computed Tomography |
| Classification Name: | Computed tomography x-ray system |
| Regulatory Number: | 21 CFR Part 892.1750 |
| Product Code: | OAS |
| Device Class: | Class II |
| Review Panel: | Radiology |
### Identification of the Legally Marketed Predicate Device:
| Trade Name: | PreXion3D Explorer PRO, Model P03B |
|--------------------------|------------------------------------|
| Classification Name: | Computed tomography x-ray system |
| Regulatory Number: | 21 CFR Part 892.1750 |
| Product Code: | OAS |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Submitter/510(k) Holder: | PreXion Corporation |
| Clearance: | K221525 (cleared July 22, 2022) |
# Device Description:
( Product Name: Dental Cone-beam Computed Tomography, The product Model:iDT901X1) uses cone-beam computed tomography (CBCT) through X-ray cone-beam, panoramic radiography, cephalometric radiography to produce images of
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for Yian Medical. The logo consists of a stylized letter "N" in green and gray, followed by the Chinese characters "奕安医疗" in gray. Below the Chinese characters, the words "YIAN MEDICAL" are written in gray.
the dental, oral and maxillofacial areas to provides diagnostic details for the medical facilities. This product consists of Frame, X-ray generator (including Integrated X-ray source tube head, X-ray tube,Collimator), Image receptor and Image processing system (including Computer, Monitor, Software workstation).
#### Indications for Use:
Dental Cone-beam Computed Tomography (Model:iDT901X1) is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
# Standards:
- A IEC60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment Part 1-2: A General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC /TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment Part > 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- > IEC 60601-1-3 Edition 2.2 2021-01 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- > IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- > IEC 62366-1 Edition 1.1 2020-06 Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
- > IEC 60601-2-63 Edition 1.1 2017-07 CONSOLIDATED VERSION Medical electrical equipment - Part 2-63: Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for Yian Medical. The logo consists of a stylized letter "N" in green and gray, followed by the text "YIAN MEDICAL" in gray. Above the text "YIAN MEDICAL" are three Chinese characters, also in gray.
- > IEC 61223-3-4:2000 Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray equipment
- > IEC 61223-3-7 Edition 1.0 2021-12 Evaluation and routine testing in medical imaging departments - Part 3-7: Acceptance and constancy tests - Imaging performance of X-ray equipment for dental cone beam computed tomography
- > ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- A IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION - Medical device software - Software life cycle processes
- > ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- > ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- A ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization
- > ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
- > ISO 15223-1 Fourth edition 2021-07 Medical devices - Symbols to be used with medical device labels, labeling, and information to be supplied - Part 1: General requirements
- > ISO 20417: 2021 Medical devices - Information to be supplied by the manufacturer
- > ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
- > Health informatics - Digital imaging and communication ISO 12052:2017 in medicine (DICOM) including workflow and data management
- > IEC 60825-1:2014 Edition 3.0 Safety of laser products - Part 1: Equipment classification and requirements
# FDA Guidance Documents:
- Format for Traditional and Abbreviated 510(k)s Guidance issued on September A 13,2019
- A Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food and Drug Administration Staff JUNE 2022
- A Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff NOVEMBER 2017
- A Premarket Assessment of Pediatric Medical Devices Guidance for Industry and FDA Staff MARCH 2014
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for Yian Medical. The logo consists of a stylized letter "N" in green and gray, followed by the company name in Chinese characters and English text. The English text reads "YIAN MEDICAL" in gray, with the Chinese characters above it.
- > Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff JUNE 2023
- > Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff SEPTEMBER 2016
- A Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff SEPTEMBER 2023
# Performance standard:
- A 21CFR PART 1010 PERFORMANCE STANDARDS FOR ELECTRONIC PRODUCTS: GENERAL
- 21 CFR 1020.30: Diagnostic x-ray system and their major components A
- > 21 CFR 1020.31: Radiographic Equipment
- 21 CFR 1020.33 Computed tomography (CT) equipment A
# Comparison with Predicate Device:
The Dental Cone-beam Computed Tomography (Model:iDT901X1 ) and its predicate device, have the equivalent intended use, functions and similar physical characteristics, performance characteristics.
# Substantial Equivalence:
The comparison between the overall specifications of predicate device (PreXion3D Explorer PRO, model P03B) and the new device (Dental Cone-beam Computed Tomography (Model:iDT901X1) ) is shown in Table 1. Any differences between the predicate and the new device have no impact on safety or efficacy of the new device and do not raise any new potential or increased safety risks, and the new device is equivalent in performance to existing legally marketed devices :
| | Subject Device | Predicate Device |
|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Dental Cone-beam Computed<br>Tomography<br>(Model:iDT901X1 ) | P03B (K221525) |
| Manufacturer | Yian Medical Technology<br>(Haining) Co., Ltd. | PreXion Corporation |
| Trade Name | / | PreXion3D Explorer<br>PRO |
| Product Code | OAS | OAS |
| Regulation | 21 CFR 892.1750 | 21 CFR 892.1750 |
| | Subject Device | Predicate Device |
| | Dental Cone-beam Computed<br>Tomography<br>(Model:iDT901X1) | P03B (K221525) |
| Number | | |
| Regulation Name | OAS: Computed<br>tomography x-ray<br>system | OAS: Computed<br>tomography x-ray<br>system |
| Device<br>Classification<br>Name | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental |
| Indications for<br>use | Dental Cone-beam Computed<br>Tomography<br>(Model:iDT901X1 ) is intended<br>to produce two-dimensional<br>digital x-ray images including<br>panoramic and cephalometric<br>image, and three-dimensional<br>digital x-ray images of the<br>dental, oral, maxillofacial<br>region, at the direction of<br>healthcare professionals as<br>diagnostic support for adult and<br>pediatric patients.<br><br>This device is not intended for<br>use on patients less than<br>approximately 21 kg (46 lb) in<br>weight and 113 cm (44.5 in) in<br>height; these height and weight<br>measurements approximately<br>correspond to that of an average<br>5 year old. | PreXion3D Explorer PRO is<br>intended to produce<br>two-dimensional digital x-ray<br>images including<br>panoramic and cephalometric<br>image, and three-dimensional<br>digital x-ray images of the<br>dental, oral, maxillofacial<br>region, ENT (Ear, Nose and<br>Throat) and neck region at the<br>direction of healthcare<br>professionals as diagnostic<br>support for adult and pediatric<br>patients. Cephalometric<br>imaging also includes the<br>hand and wrist to obtain<br>carpus images for growth and<br>maturity assessment.<br>This device is not intended for<br>use on patients less than<br>approximately 21 kg (46 lb)<br>in weight and 113 cm (44.5<br>in) in height; these height and<br>weight measurements<br>approximately<br>correspond to that of an<br>average 5 year old. |
| | Subject Device | Predicate Device |
| | Dental Cone-beam Computed<br>Tomography<br>(Model:iDT901X1) | P03B (K221525) |
| Target Population | Children aged 6 (except<br>infants) to elderly | Children aged 6 (except<br>infants) to elderly |
| Anatomical Site | The dental, oral,<br>maxillofacial region | The dental, oral,<br>maxillofacial region<br>ENT (Ear, Nose and Throat)<br>and neck region |
| Users | Health care professionals | Health care professionals |
| Patient Contact<br>Material | Meet ISO 10993 series standard | Meet ISO 10993 series<br>standard |
| Sterility | Non-sterile | Non-sterile |
| Tube<br>Voltage | 60kV~120kV | 90-110KV |
| Tube<br>Current | 1 mA~20 mA | 1-5.3mA |
| Focal Spot<br>Size | 0.5 | 0.3mm x 0.3mm |
| Pulse<br>Exposure<br>function | Yes | Yes |
| Power | Frequency: 50/60Hz<br>Voltage: 110-120VAC<br>Power: 3.0kVA | Frequency: 50/60Hz<br>Voltage: 110-240VAC<br>Power: 1.0KVA |
| Detector<br>material | FPD (TFT)<br>CsI | FPD (TFT)<br>CsI:Tl / Gd2 O2 S:Tb |
| Pixel Size | 200μm x 200μm (With binning)<br>(CT, Cephalometric)<br><br>100μm x 100μm ( Without<br>binning ) (CT, Panoramic) | 248 μm x248µm (With<br>binning)<br>(CT, CT-Panoramic,<br>Panoramic)<br><br>124 µm x124µm<br>(Without binning)<br>(CT, CT-Panoramic,<br>Panoramic, Ceph) |
| Pixel<br>Number | 1504*1248(With binning) (CT,<br>Cephalometric) | 1024x1280(With<br>binning)<br>(CT, CT-Panoramic) |
| | Subject Device | Predicate Device |
| | Dental Cone-beam Computed<br>Tomography<br>(Model:iDT901X1) | P03B (K221525) |
| | 800*2496(Without<br>binning)<br>(CT) | 2560x2048 (Without<br>binning)<br>(CT, CT-Panoramic,<br>Ceph) |
| | 3008*65(Without<br>binning)<br>(Panoramic) | 1900 x 120<br>(Panoramic) |
| | 1504*1248(With<br>binning)<br>(Cephalometric) | 2560 x 2048<br>(Cephalometric) |
| Size of<br>Area<br>Receiving<br>X-ray | 300.8mm x 249.6mm (CT,<br>CEPH) | 253.95mm x 317.44mm<br>(CT, CT-Panoramic) |
| | 80mm x 249.6mm (CT) | 230mm x 15mm<br>(Panoramic) |
| | 300.8mm x 6.5mm (PANO) | 239mm x 302mm<br>(Ceph) |
| spatial resolution | 5lp/mm | 4lp/mm |
| Number of<br>Bits | 16bits | 16bits |
| SID/SOD | SID: 750mm/490mm | 700mm/ 420mm (CT,<br>CT-Panoramic,<br>Panoramic)<br>1000mm / 840mm<br>(Ceph) |
| Dimension<br>(WxDxH) | Max.<br>1300mm*1272mm*2365mm. | 1,112 mm x 1,558 mm x<br>2330 mm (CT, CTPanoramic,<br>Panoramic)<br>1164 mm x 1690 mm x<br>2330 mm (with Ceph)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.