PreXion3D Expedition is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height: these height and weight measurements approximately correspond to that of an average 5 year old.
Device Story
Scanner generates X-ray data; console manages data/operation. Rotation arm orbits patient head; X-ray generator emits beams through patient to flat panel detector. Console performs image reconstruction (3D CT, panoramic, cephalometric). Used in clinical settings by healthcare professionals. Output: 2D/3D digital images for diagnostic support. Benefits: provides anatomical visualization for dental/ENT/maxillofacial assessment; aids growth/maturity evaluation via carpus imaging.
Clinical Evidence
Bench testing only. Performance verified via biocompatibility (ISO 10993), electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-1-3/6), EMD (IEC 60601-1-2), and dental X-ray equipment standards (IEC 60601-2-63). Software verified per IEC 62304. Clinical image validation confirmed clinical validity.
Technological Characteristics
Computed tomography X-ray system (OAS). Components: X-ray generator, flat panel detector, console. Energy: High-voltage transformer (100-240V to kV range). Connectivity: Networked console for image analysis/printing. Standards: IEC 60601-2-63, IEC 61223-3-4/5, ISO 10993, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for adult and pediatric patients requiring 2D (panoramic, cephalometric) or 3D digital X-ray imaging of dental, oral, maxillofacial, ENT, and neck regions. Cephalometric imaging includes hand/wrist for growth assessment. Contraindicated for patients <21 kg and <113 cm height.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 8, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
PreXion Corporation % Mr. Kenji Tanaka Manager 1-14-1. Kandasuda-cho Chiyoda-ku, Tokyo 101-0041 JAPAN
Re: K232166
Trade/Device Name: PreXion3D Expedition Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: July 21, 2023 Received: July 21, 2023
Dear Mr. Tanaka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K232166
Device Name
PreXion3D Expedition
Indications for Use (Describe)
PreXion3D Expedition is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height: these height and weight measurements approximately correspond to that of an average 5 year old.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| K232166 | 510(k) Summary | Prepared on: 2023-09-08 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | PreXion Corporation | |
| Applicant Address | 1-14-1, Kandasuda-cho Chiyoda-ku Tokyo 101-0041 Japan | |
| Applicant Contact Telephone | +81-3-5297-7553 | |
| Applicant Contact | Mr. Kenji Tanaka | |
| Applicant Contact Email | px-ra@prexion.co.jp | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | PreXion3D Expedition | |
| Common Name | X-ray, Tomography, Computed, Dental | |
| Classification Name | Computed tomography x-ray system | |
| Regulation Number | 21 CFR 892.1750 | |
| Product Code | OAS | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K222150 | PreXion3D Expedition | OAS |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| PreXion3D Expedition consists of a scanner, which is used for generating X-ray and detecting<br>image data, and a console, which is used for operating the scanner and managing the data. The | | |
scan data acquired by the scanner will be transferred to the Console. PreXion3D Explore Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating crosssection diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed. X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan), the tomographic image (CT scan, Panoramic scan) and Cephalometric Scan.
The operating principle of the device is as follows.
<X-ray generation principle>
X-rays are generated by the conversion of electron kinetic energy.
Part of the kinetic energy which is generated when electrons moving at high speed are
decelerated inside matter becomes the conversion source.
Use a high-voltage transformer to boost the commercial voltage (100 to 240 V) to direct current
high voltage (several tens of kV) and apply it to the X-ray tube to accelerate the X-ray tube's
thermal electrons, and then the X-ray will be generated.
The change in the voltage (tube voltage) and current (tube current) applied to the X-ray tube brings the following features.
• The higher the X-ray tube voltage is, the greater the penetration strength of X-rays is.
· The higher the current (tube current) is, the more the X-ray dose is.
{4}------------------------------------------------
With the consideration of the above features, X-ray devices are designed to be able to control the X-ray dose and strength according to the intended use.
<CT Scan principle and Panoramic Scan principle>
X-ray photography is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed.
X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan) and the tomographic image (CT scan, Panoramic scan).
#### <Software>
PreXion3D Expedition consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating cross-section diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed.
X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan), the tomographic image (CT scan, Panoramic scan) and Cephalometric Scan.
# Intended Use/Indications for Use
PreXion3D Expedition is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric maging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
# Indications for Use Comparison
### 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
The device has the same indications for use in comparison to the predition/P04A). The subject device of PreXion3D Expedition is intended to produce two digital panoramic and cephalometric images, and three dimensional digital x-ray images of the dental(oral), maxillofacial, and Throat) region at the direction of healthcare professionals as diagnostic support for adult
and pediatric patients.
Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
# Technological Comparison
The different technological characteristics are summarized in the points. The difference is in the FPD, part of the software, and the maximum X-ray field. First, we have confirmed that there are no problems with the FPD according to SSXI guidance, and the report is available as an attachment to this application. Second, the software only changes the protocol of the interface due to the change in the FPD, and we have confirmed that there are no problems with the software in the system test. Finally, the maximum X-ray irradiation field has been confirmed in the IEC 60601-2-63 test report. Other than the above, the intended use/ appearance/design/materials/ labeling for precautions and warnings/operating principle/energy source/manufacturer, etc. are all the same. Based on these points, we claim the substantial equivalence of this device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
As part of demonstrating safety and effectiveness of PreXion and in showing substantial equivalence to the predicate device, PreXion Corporation completed a number of nonclinical performance tests. The PreXion3D Expedition meets all the requirements for overall design, biocompatibility, performance, and electrical safety results confirming that the the design inputs and specifications for the device.
### 21 CFR 807.92(a)(6)
{5}------------------------------------------------
The PreXion3D Expedition passed all the testing in accordance with internal requirements, national standards, and international standards to support substantial equivalence of the subject device.
The PreXion3D Expedition passed all the testing in accordance with internal requirements, and international standards shown below to support substantial equivalence of the subject device:
-Biocompatibility testing per ISO 10993-1, ISO 10993-5 and ISO 10993-10
- -Electrical safety testing per ANSI/AAMI ES 60601-1, IEC 60601-1-3 and IEC 60601-1-6
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2
- Dental extra-oral X-ray equipment testing per IEC 60601-2-63
- Software verification and validation IEC 62304
- Acceptance testing of X-ray equipment per IEC 61223-3-4 and IEC 61223-3-5
- Storage and Transport Testing per ISO 4180
- IEC 62366-1:2015,AMD1:2020 Usability engineering to medical devices
- Software Documentation per: "Guidance for the Content of Premarket Submissions for
Software Contained in Medical Devices
- Establish the substantial equivalence of an SSXI to a previously cleared conventional
- radiographic SSXI per: "Guidance for the Submission of 510(k)'s for Solid State X-ray
Imaging Devices"
- Cybersecurity Activities per: "Cybersecurity-for-Networked-Medical-Devices-Containing-Offthe-Shelf-(OTS)-Software-Gor Industry", "Content of Premarket Submissions for
Management of Cybersecurity in Medical Devices"
We conducted the clinical image validation and as a result, it was confirmed that they are clinically valid.
The results of the above non-clinical studies indicate that the device is substantially equivalent to the predicate device in safety, efficacy, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.