K230985 · Planmeca Oy · OAS · Dec 28, 2023 · Radiology
Device Facts
Record ID
K230985
Device Name
Planmeca Viso
Applicant
Planmeca Oy
Product Code
OAS · Radiology
Decision Date
Dec 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Endodontic image quality
—
—
—
—
—
Study conducted with eleven patients comparing images with no denoising and AI denoising
3 (dental professionals)
Indications for Use
Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital X-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
Device Story
Planmeca Viso is a CBCT imaging system for dento-maxillo-facial, cervical spine, and ENT regions. It captures 3D volumes via hundreds of sample images taken from different directions; produces 2D panoramic and cephalometric images via tomosynthesis. System includes X-ray unit, generator, collimator, and new X-ray tube/detectors. Data processed via dedicated 3D reconstruction hardware and viewed on external workstation using Planmeca Romexis software. Operated by healthcare professionals in clinical settings. Includes AI-based denoising for endodontic image processing. Output provides diagnostic support for clinicians to assess patient anatomy. Benefits include improved image quality and diagnostic accuracy through updated IGZO detector technology and denoising algorithms.
Clinical Evidence
Bench testing performed using Sedentex and DIN 6868 phantoms compared subject and predicate devices. Clinical evaluation of image quality conducted by dental professionals using human phantom images. AI Denoising feature evaluated in a study with 11 patients by three dental professionals comparing images with and without denoising; results confirmed diagnostic image quality.
Technological Characteristics
CBCT imaging system; 120kV max voltage; Tungsten anode; IGZO-based flat panel detectors; Nvidia Jetson Nano module for interface; Ethernet connectivity. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-63, 62366-1, and 62304. Software-based 3D reconstruction and AI denoising.
Indications for Use
Indicated for pediatric and adult patients requiring 2D/3D digital X-ray or 3D optical imaging of dento-maxillo-facial, cervical spine, and ENT regions for diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Planmeca Viso, Computed Tomography System (K181576)
{0}------------------------------------------------
December 28, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Planmeca Oy Niina Vuorikallas Director, Quality & Regulatory Affairs Asentajankatu 6 Helsinki. 00880 Finland
Re: K230985
Trade/Device Name: Planmeca Viso Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: October 17, 2023 Received: November 3, 2023
Dear Niina Vuorikallas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gabriela M. Rodal Digitally signed by for Gabriela M. Rodal -S -2
Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K230985
Device Name
Planmeca Viso
#### Indications for Use (Describe)
Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital X-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
#### Manufacturer
Planmeca Oy Asentajankatu 6 00880 Helsinki, Finland Phone: +358 20 7795 500 Fax: +358 20 7795 555 Contact person: Niina Vuorikallas
#### U.S. designated agent
Planmeca U.S.A. Inc. 2600 Forbs Ave. Hoffman Est, IL 60192 UNITED STATES Contact person: Glen Kendrick
Date Prepared: December 27, 2023
#### II. DEVICE
| Name of Device: | Planmeca Viso |
|-----------------------|------------------------------------------------------------|
| Common or Usual Name: | Computed Tomography X-ray System |
| Classification Name: | Computed Tomography X-ray System (CT) (21 CFR<br>892.1750) |
| Regulatory Class: | II |
| Product Code: | OAS |
#### III. PREDICATE DEVICE
Primary predicate device, #1, Planmeca Viso, Computed Tomography System 510(k): K181576 Regulation number: 892.1750 Regulatory Class: II Product Code: OAS
Predicate device #2, HDX DENTRIα. Computed Tomography System 510(k): K160140 Regulation number: 892.1750 Regulatory Class: II Product Code: OAS
Predicate device #3, DigiX FDX, Stationary X-ray system 510(k): K223060 Regulation number: 892.1680 Regulatory Class: II Product Code: KPR
{4}------------------------------------------------
Predicate device #2 is only used for X-ray tube comparison. Outside of figure 1, when referred to predicate device, the reference is made to primary predicate device, #1.
#### DEVICE DESCRIPTION IV.
The Planmeca Viso -X-ray unit uses cone beam computed tomography (CBCT) to produce three-dimensional (3D) images of the maxillofacial and ENT anatomies. Two dimensional (2D) images are produced with tomosynthesis method (panoramic) imaging) as well as conventional 2D radiography (cephalometric imaging, 2D views). In CBCT a cylindrical volume of data is captured in one imaging procedure. The data consists of several hundred sample images which are taken from different directions to cover a certain pre-programmed target area. These samples are used for 3D reconstruction (using a dedicated 3D reconstruction hardware) that can be viewed in three dimensions using separate workstation and Planmeca Romexis software.
#### INDICATIONS FOR USE V.
Same as predicate device.
Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dentomaxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
## General
Both devices are 3D CBCT imaging system indicated for imaging dento-maxillofacial-. cervical spine- and ENT-regions and offer a variety of 3D fields of view suitable for each application.
Both systems also provide 2D imaging capabilities for panoramic and cephalometric imaging. In general there are only a few changes in the device.
# Integrated detector
The X-ray detectors used are of same size as previously. Viso G5 and ProCeph use new X-ray detectors with similar qualities as the predicate device. Only difference between Viso G7 predicate and subject devices is the interface. Predicate device uses proprietary interface, and subject device uses standard ethernet interface. This has some effects on the device hardware, namely the grabber PCB, which is being replaced with a commercial Nvidia Jetson Nano module and suitable interface PCB. Otherwise they are of similar qualities as before and the produced 3D images provide similar diagnostic value. See fig 2 for technical comparison between devices.
# X-ray unit
The dimensions and materials are identical to the predicate device.
# X-ray tube
A new x-ray tube has been introduced to the device. It is of similar quality as predicate devices' fixed anode x-ray tubes, and it has maximum tube voltage of 120kV. Previously used fixed anode x-ray tubes may also be used with this device. No other changes for the tube head. Predicate device #2 operates a similar tube manufactured by the same company, the only difference being the target angle and thus the heat characteristics. See fig 1 for technical comparison on X-ray tube properties.
fig 1. X-ray tube comparison
{5}------------------------------------------------
| | Subject device | Primary predicate<br>device, #1 | Predicate device #2 |
|--------------------------------------------|----------------------------|---------------------------------|----------------------------|
| Tube type | OPX/105-10 | SXR 130-10-0.5 SC | OPX/105 |
| Nominal voltage | 120 kV | 130 kV | 110 kV |
| Focal spot | 0.5 mm | 0.5 mm | 0.5mm |
| Filament characteristics | 3.5 ÷ 6.0 V<br>2.5 ÷ 4.0 A | N/A | 3.5 ÷ 6.0 V<br>2.5 ÷ 4.0 A |
| Anode material | Tungsten | Tungsten | Tungsten |
| Target angle | 10° | 10° | 5° |
| Anode heat storage<br>capacity | 40000 J | 31500 J | 30000 J |
| Maximum anode cooling<br>rate | 500 W | 310 W | 250 W |
| Nominal anode input<br>power at 0.1 s (DC) | 1900 W | N/A | 2000 W |
| Inherent filtration | 0.5 mm Al | 1.1 mm Al | 0.5 mm Al |
### X-ray generator
The device uses the same x-ray generator as before.
# X-ray collimator
The device uses the same x-ray collimator as before.
### Software
Software changes made to accommodate new hardware (see integrated detectors and new xray tube).
### CT reconstruction algorithm
Device uses same CT reconstruction algorithm as before, no significant changes are made to the algorithm.
| fig 2. technical comparison | | |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Technical specification | Subject device | Primary predicate device, #1 |
| Classification | | |
| FDA product class | OAS, class II | OAS, class II |
| RoHS | 2011/65/EU | 2011/65/EU |
| IEC 60601-1 | Class I, type B | Class I, type B |
| CISPR 11 | Class B | Class B |
| IP Classification | IPX0 | IPX0 |
| Applied parts<br>(according to IEC 60601-1: 2012) | | |
| Patient supports | As shown in section Patient<br>supports in user's manuals | As shown in section Patient<br>supports in user's manuals |
| Patient handles | | |
| Generator<br>(according to IEC 60601-2-7:<br>1998) | | |
| | Resonant-mode, DSP-controlled, 80 - 160 kHz | Resonant-mode, DSP-controlled, 80 - 160 kHz |
| X-ray tube | | |
| | D-059SBR or SXR 130-10-0.5 SC or OPX 105-10 | D-059SBR or SXR 130-10-0.5 SC |
| Focal spot size<br>(according to IEC 60336: 2005) | | |
| | 0.5 x 0.5 mm | 0.5 x 0.5 mm |
| Filtration | | |
| 3D | Total 2.5 mm Al + 0.2mm / 0.5 mm Cu | Total 2.5 mm Al + 0.2mm / 0.5 mm Cu |
| Pan (SmartPan) / ProCeph | Total 2.5 mm Al | Total 2.5 mm Al |
| Tube housing front cover<br>quality equivalent filtration<br>(not included in the specified<br>total filtration) | 0.3 mm Al @ 70 kV / HVL 2.6 mm Al | 0.3 mm Al @ 70 kV / HVL 2.6 mm Al |
| Anode voltage | | |
| 3D | 80 - 120 kV ±5% | 80 - 120 kV ±5% |
| Pan (SmartPan) | 60 - 84 kV ±5% | 60 - 84 kV ±5% |
| ProCeph | 60 - 84 kV ±5% | 60 - 84 kV ±5% |
| Anode current | | |
| 3D | D-059SBR: 2-12.5 mA ±10%<br>SXR 130-10-0.5 SC: 2-16mA ±10%<br>OPX 105-10: 2-16mA ±10% | D-059SBR: 1-12.5 mA ±10%<br>SXR 130-10-0.5 SC: 1-16mA ±10% |
| Pan (SmartPan) | D-059SBR: 2-14 mA ±10%<br>SXR 130-10-0.5 SC: 2-16mA ±10%<br>OPX 105-10: 2-16mA ±10% | D-059SBR: 1-14 mA ±10%<br>SXR 130-10-0.5 SC: 1-16mA ±10% |
| ProCeph | D-059SBR: 14 mA ±10%<br>SXR 130-10-0.5 SC: 16mA ±10%<br>OPX 105-10: 14mA ±10% | D-059SBR: 14 mA ±10%<br>SXR 130-10-0.5 SC: 16mA ±10% |
| mAs range | | |
| | min. / max. as indicated ±(10% + 0.2 mAs) | min. / max. as indicated ±(10% + 0.2 mAs) |
| Dose range and accuracy | | |
| | Dose range min. / max. as indicated on system user interface<br>Accuracy of dosimetric indication (DAP, CTDI): ±40% | Dose range min. / max. as indicated on system user interface<br>Accuracy of dosimetric indication (DAP, CTDI): ±40% |
| Linearity of radiation output | | |
| | < 0.1 | < 0.1 |
| Exposure time | | |
| 3D | Pulsed, effective 1.5 - 36 s as<br>indicated ±10% | Pulsed, effective 1.5 - 36 s as<br>indicated ±10% |
| Pan (SmartPan) | 2.5 - 15.6 s as indicated ±10% | 2.5 – 15.6 s as indicated ±10% |
| ProCeph | 0.1 - 1.6s as indicated ±10% | 0.1 - 1.6s as indicated ±10% |
| SID | | |
| 3D / Pan (SmartPan) | 700 mm | 700 mm |
| Ceph | 1700 mm (66.9 in.) | 1700 mm (66.9 in.) |
| Magnification | | |
| 3D | 1.40 - 1.71 | 1.40 - 1.71 |
| Pan (SmartPan) | 1.32 | 1.40 |
| Ceph | 1.13 | 1.13 |
| Duty cycle for height adjustment | | |
| | 25 s ON / 400 s OFF | 25 s ON / 400 s OFF |
| Line voltage | | |
| | 100 - 220 V~ / 50 - 60 Hz<br>230 - 240 V~ / 50 Hz | 100 - 220 V~ / 50 - 60 Hz<br>230 - 240 V~ / 50 Hz |
| Line current | | |
| | 8 - 17 A | 8 - 17 A |
| Input power | | |
| Stand by | 150 VA | 150 VA |
| Exposure | 1800 W | 1800 W |
| | | |
| Line harmonics | | |
| | Cos better than 0.9 | Cos better than 0.9 |
| Max. permissible apparent<br>impedance of supply mains | | |
| | 0.5 Ohms (100 VAC) | 0.5 Ohms (100 VAC) |
| Max. continuous heat dissipation | | |
| | 250 W | 250 W |
| Internal fuse(s) | | |
| User replaceable<br>- 1 fuse on permanently<br>installed X-ray units | 100 - 220 V~ / 16A FF H 500 V<br>230 - 240 V~ / 8A FF H 500 V | 100 - 220 V~ / 16A FF H 500 V<br>230 - 240 V~ / 8A FF H 500 V |
| Type | 195100 ELU | 195100 ELU |
| External fuse(s) | | |
| | 100 - 220 V ~ / 16A min. - 20A<br>max. T 250 V<br>230 - 240 V ~ / 10A min. - 20A<br>max. T 250 V | 100 - 220 V ~ / 16A min. - 20A<br>max. T 250 V<br>230 - 240 V ~ / 10A min. - 20A<br>max. T 250 V |
| Battery | | |
| | Lithium battery: 3V, CR2032<br>Panasonic / Varta | Lithium battery: 3V, CR2032<br>Panasonic / Varta |
| Max. weight | | |
| Base unit | 165 kg (364 lb) | 165 kg (364 lb) |
| ProCeph | 20 kg (44 lb) | 20 kg (44 lb) |
| Environmental requirements | | |
| Transport: | | |
| Temperature | -20°C - +60°C (-4°F - +140°F) | -20°C - +60°C (-4°F - +140°F) |
| Relative humidity | 10 - 90% RH (non-condensing) | 10 - 90% RH (non-condensing) |
| Air pressure | 700 - 1060 hPa | 700 - 1060 hPa |
| Storage: | | |
| Temperature | -10°C - +50°C (+14°F - +122°F) | -10°C - +50°C (+14°F - +122°F) |
| Relative humidity | 10 - 90% RH (non-condensing) | 10 - 90% RH (non-condensing) |
| Air pressure | 700 - 1060 hPa | 700 - 1060 hPa |
| Operating: | | |
| Temperature | +10°C - +30°C (+50°F - +86°F) | +10°C - +30°C (+50°F - +86°F) |
| Relative humidity | 10 - 90% RH (non-condensing) | 10 - 90% RH (non-condensing) |
| Air pressure | 800 - 1060 hPa | 800 - 1060 hPa |
| Max. altitude | 2000 m (1.25 miles) | 2000 m (1.25 miles) |
| Image properties | | |
| ProCeph: | | |
| Panel type | Varex 2530C Gen5 | Varex 2530P |
| Flat panel pixel size | 131 μm | 139 μm |
| Flat panel active surface | 302 x 249 mm (11.89 x 9.80 in.) | 302 x 249 mm (11.89 x 9.80 in.) |
| DQE (0) | 64 % | 32 % |
| MTF | > 51 % @ 1lp/mm | > 48 % @ 1lp/mm |
| 3D: | | |
| Panel type | Viso G5: Varex 1616Z<br>Viso G7: Varex 2530DX | Viso G5: Varex 1616PT<br>Viso G7: Varex 2530PX |
| Flat panel pixel size | Viso G5: 105 μm<br>Viso G7: 139 μm | Viso G5: 127 μm<br>Viso G7: 139 μm |
| Flat panel active surface | Viso G5: 161.3 x 161.3 (6.35 x 6.35<br>in.)<br>Viso G7: 247.7x 301.1 mm (9.75 x<br>11.85 in.) | Viso G5: 157.5 x 157.5 mm (6.20 x<br>6.20 in.)<br>Viso G7: 299.7x 246.3 mm (11.80<br>x 9.70 in.) |
| DQE (0) | Viso G5: 78 % (1x1)<br>Viso G7: 70 % (1x1) | Viso G5: 70 % (1x1)<br>Viso G7: 70 % (1x1) |
| MTF | Viso G5: > 59 % @ 1lp/mm (1x1)<br>Viso G7: > 48 % @ 1lp/mm (1x1) | Viso G5: > 48 % @ 1lp/mm (1x1)<br>Viso G7: > 50 % @ 1lp/mm (1x1) |
| Voxel sizes | 75 μm, 150 μm, 300 μm, 450 μm,<br>600 μm | 75 μm, 150 μm, 300 μm, 450 μm,<br>600 μm |
| Pan (SmartPan): | | |
| Panel type | Viso G5: Varex 1616Z<br>Viso G7: Varex 2530DX | Viso G5: Varex 1616PT<br>Viso G7: Varex 2530PX |
| Flat panel pixel size | Viso G5: 105 μm<br>Viso G7: 139 μm | Viso G5: 127 μm<br>Viso G7: 139 μm |
| Flat panel active surface | Viso G5: 8.4 x 161.3 mm (0.33 x<br>6.35 in.)<br>Viso G7: 8.9/17.8 x 166.8 mm<br>(0.35/0.7 x 6.57 in.) | 8 - 25 x 146 mm (0.31 - 0.98 x<br>5.74 in.) |
| Operating requirements for<br>ProFace program | | |
| Optimum colour temperature | Approx. 6500 Kelvin | Approx. 6500 Kelvin |
| Even and uniform lighting | | |
| No bright lights | | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### PERFORMANCE DATA VII.
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
No changes in patient contact parts compared to primary predicate device. Electrical safety and electromagnetic compatibility (EMC)
{10}------------------------------------------------
CB and EMC testing was conducted as per IEC standards 60601-1+A1:2012+A2:2020, 60601-1-2+A1:2020, 60601-1-3+A1:2013+A2:2021, 60601-1-6+A1:2013+A2:2020, 60601-2-63+A1:2017+A2:2021,62366-1+A1:2020 and 62304+A1:2015
# Software Verification and Validation Testing
Software verification and validation was performed as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance of Premarket Submissions for Software Contained in Medical Devices.". No issues have been identified during the verification and validation process.
### Bench testing
Performance testing in laboratory environment was performed with Viso (primary predicate device) and Viso (subject device) of the Sedentex and DIN 6868 Phantoms. The data was compared in order to see that the performance of the device remains substantially similar to that of the primary predicate device.
### Clinical evaluation
The main measure of clinical performance for Planmeca Viso is determined to be image quality. This has been evaluated by the product relevant team of professionals who evaluated human phantom images which were determined to be suitable for the intended purpose and indications for use of the device. These results are substantially equivalent to the evaluation performed on primary predicate device.
AI Denoising feature of the endodontic image processing has been evaluated based on its ability to produce diagnostic image quality in its intended application. This was evaluated in a study performed by three dental professionals in a study conducted with eleven patients, by comparing images with no denoising and AI denoising.
### Testing summary
The above testing confirms that Planmeca Viso is safe and effective in its intended use.
# VIII. CONCLUSIONS
The performance data of the subject device and the predicate device is equal or close to equal. The performance testing provides data to back this similarity in the devices. The subject device performs equally or better in all the testing scenarios.
The IGZO technology is proven to bring improvements in electronic noise, image lag and noise-equivalent quanta. The IGZO technology has been used in several FDA approved devices at least since 2019.
The tube specifications regarding eg. focus size, anode angle and tube voltage are the same as previously approved Canon D-059SB X-ray tube. The OPX 105-10 X-ray tube is driven with the same hardware as our other previously approved X-ray tubes.
The comparison of characteristics supports substantial equivalence. Planmeca Viso is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.