PreXion3D Explorer PRO is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment. This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
Device Story
PreXion3D Explorer PRO is a dental/maxillofacial X-ray system. It consists of a scanner and a console. The scanner uses a rotating arm with an X-ray generator and a flat panel detector (FPD) to capture X-ray absorption data while rotating around the patient's head. The console receives this data to perform image reconstruction (3D CT, panoramic, or cephalometric). The device is operated by healthcare professionals in a clinical setting. The system uses filtered-back-projection for CT and panoramic reconstruction; cephalometric images are output as 2D data. The output images support diagnostic decision-making for dental and ENT conditions. The device benefits patients by providing high-resolution diagnostic imaging for treatment planning and growth assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1 series), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation (IEC 62304).
Indicated for adult and pediatric patients (≥21 kg, ≥113 cm) requiring 2D (panoramic, cephalometric) or 3D digital X-ray imaging of dental, oral, maxillofacial, ENT, and neck regions. Cephalometric imaging includes hand/wrist for growth/maturity assessment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
July 22, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The Department of Health & Human Services seal is a stylized representation of a human figure. The FDA acronym is in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font.
PreXion Corporation % Hiroaki Takahashi General Manager 1-14-1. Kandasuda-cho Chiyoda-ku, Tokyo 101-0041 JAPAN
# Re: K221525
Trade/Device Name: PreXion3D Explorer PRO Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: May 24, 2022 Received: May 26, 2022
## Dear Hiroaki Takahashi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221525
Device Name PreXion3D Explorer PRO
#### Indications for Use (Describe)
PreXion3D Explorer PRO is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for PreXion. The logo features the word "PreXion" in a stylized font, with the "IT" in the tagline "make IT visible" in blue. Above the text is a blue and white swoosh design, adding a modern and dynamic element to the logo.
# 510(k) Summary
# PreXion3D Explorer PRO
# K221525
# 1. Submission Sponsor
PreXion Corporation 1-14-1, Kandasuda-cho, Chiyoda-ku, Tokyo 101-0041 Japan Hiroaki Takahashi General Manager, Quality Assurance & Regulatory Division Email: px-ra@prexion.co.jp Office number: +81-3-5297-7551
# 2. Submission Correspondent
Same as above
# 3. Date Prepared
May 24th, 2022
## 4. Device Identification
Trade/Proprietary Name: PreXion3D Explorer PRO
- Common/Usual Name: OAS: Computed Tomography X-Ray System
- Regulation Number: OAS: 21 CFR 892.1750
- Product Code: OAS
- Device Class: Class II
- Classification Panel: Radiology
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Prexion. The word "Prexion" is written in a light gray, italicized font. Below the word "Prexion" is the phrase "make IT visible" in a smaller font, with the "IT" in blue. Above and to the right of the word "Prexion" is a blue swoosh design.
# 5. Legally Marketed Predicate Device(s)
Predicate Device: K203784, PreXion3D Explorer PRO (Model: P03A), PreXion Corporation
#### 6. Indication for Use Statement
PreXion3D Explorer PRO is intended to produce two-dimensional digital x-ray images including panoramic and cephalometric image, and three-dimensional digital x-ray images of the dental, oral, maxillofacial region, ENT (Ear, Nose and Throat) and neck region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
This device is not intended for use on patients less than approximately 21 kg (46 lb) in weight and 113 cm (44.5 in) in height; these height and weight measurements approximately correspond to that of an average 5 year old.
#### 7. Device Description
PreXion3D Explorer PRO consists of a scanner, which is used for generating X-ray and detecting image data, and a console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Explore Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating crosssection diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed. X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm). The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan), the tomographic image (CT scan, Panoramic scan) and Cephalometric Scan.
The operating principle of the device is as follows.
<X-ray generation principle>
X-rays are generated by the conversion of electron kinetic energy.
Part of the kinetic energy which is generated when electrons moving at high speed are decelerated inside matter becomes the conversion source.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for PreXion. The logo features the word "PreXion" in a light gray, elegant font. Below the name is the tagline "make IT visible" in a smaller, bolder font, with the "IT" in blue. Above and to the right of the name is a stylized graphic of two curved shapes, one in a light gray and the other in a gradient of light to dark blue, creating a sense of depth and movement.
Use a high-voltage transformer to boost the commercial voltage (100 to 240 V) to direct current high voltage (several tens of kV) and apply it to the X-ray tube to accelerate the X-ray tube's thermal electrons, and then the X-ray will be generated.
The change in the voltage (tube voltage) and current (tube current) applied to the X-ray tube brings the following features.
- The higher the X-ray tube voltage is, the greater the penetration strength of X-rays is.
- The higher the current (tube current) is, the more the X-ray dose is.
With the consideration of the above features, X-ray devices are designed to be able to control the X-ray dose and strength according to the intended use.
<CT Scan principle and Panoramic Scan principle>
X-ray photography is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed.
X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm).
The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan) and the tomographic image (CT scan, Panoramic scan).
। Cephalometric Exposure
Based on cephalometric radiography, a plain radiographic image of the properly positioned "patient's head" between X-ray generator and flat panel detector with stable magnification ratio, can be acquired. Also, the image of "patient's hand" can be obtained.
## <Software>
PreXion3D Explorer PRO consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Explorer PRO Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating cross-section diagram, etc.), and output the image to a printer.
X-ray image data is acquired while the rotation arm is rotating around the secured "patient's head" at a constant speed.
X-rays, which are emitted from X-ray generator (built in one side of rotation arm), pass through a patient and are detected by the flat panel detector (built in the other side of rotation arm).
The detected X-ray absorption data is used to process image reconstruction on the Console to create the 3D image (CT scan), the tomographic image (CT scan, Panoramic scan) and Cephalometric Scan.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for PreXion. The logo features the word "PreXion" in a stylized font, with the "IT" in the tagline "make IT visible" highlighted in blue. Above the text is a swooping, abstract graphic in shades of blue and gray, adding a modern and dynamic element to the overall design.
For CT and Panoramic scan, the detected data is reconstructed using filtered-back-projection method. In the case of Cephalometric scan, the acquired 2D data is output as it is.
The software is unchanged from the predicates in terms of function.
- Software Level of Concern -
The software level of concern for the PreXion3D Explorer PRO is Moderate. The rationale is as follows:
Even prior to mitigation of hazards, there is no risk of serious injury or death associated with this software. Excessive X-ray irradiation can be considered as a risk of serious injury, but it is limited in scope by a hardware timer. A hardware activated buzzer indicating exposure and an emergency switch are also provided. As such, there is a risk of injury associated with the software, but not serious injury.
# 8. Substantial Equivalence Discussion
The following table compares the PreXion3D Explorer PRO to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| | Subject Device | Predicate Device | | |
|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Manufacturer | PreXion Corporation | PreXion Corporation | Comparison<br>to Predicate | |
| Trade Name | PreXion3D Explorer<br>PRO | PreXion3D Explorer<br>PRO | | |
| Model Name | P03B | P03A | N/A | |
| 510(k) Number | K221525 | K203784 | N/A | |
| Product Code | OAS | OAS | Same | |
| Regulation Number | OAS: 21 CFR 892.1750 | OAS: 21 CFR 892.1750 | Same | |
| Regulation Name | OAS: Computed<br>tomography x-ray<br>system | OAS: Computed<br>tomography x-ray<br>system | Same | |
| Device Classification<br>Name | X-Ray, Tomography,<br>Computed, Dental | X-Ray, Tomography,<br>Computed, Dental | Same | |
| PreXion3D Explorer PRO | | | | |
| Indications for use: | PreXion3D Explorer PRO<br>is intended to produce<br>two dimensional digital<br>panoramic and<br>cephalometric images,<br>and three-dimensional<br>digital x-ray images of<br>the maxillofacial, and<br>ENT (Ear, Nose and<br>Throat) and neck region<br>at the direction of<br>healthcare professionals<br>as diagnostic support for<br>adult and pediatric<br>patients.<br>Cephalometric imaging<br>also includes the hand<br>and wrist to obtain<br>carpus images for<br>growth and maturity<br>assessment.<br>This device is not<br>intended for use on<br>patients less than<br>approximately 21 kg (46<br>lb) in weight and 113 cm<br>(44.5 in) in height; these<br>height and weight<br>measurements<br>approximately<br>correspond to that of an<br>average 5 year old. | PreXion3D Explorer PRO<br>is intended to produce<br>two dimensional digital<br>panoramic and<br>cephalometric images,<br>and three-dimensional<br>digital x-ray images of<br>the maxillofacial, and<br>ENT (Ear, Nose and<br>Throat) and neck region<br>at the direction of<br>healthcare professionals<br>as diagnostic support for<br>adult and pediatric<br>patients.<br>Cephalometric imaging<br>also includes the hand<br>and wrist to obtain<br>carpus images for<br>growth and maturity<br>assessment.<br>This device is not<br>intended for use on<br>patients less than<br>approximately 21 kg (46<br>lb) in weight and 113 cm<br>(44.5 in) in height; these<br>height and weight<br>measurements<br>approximately<br>correspond to that of an<br>average 5 year old. | Same | |
| Patient/User<br>Characteristics | | | | |
| Target Population | Children aged 6 (except<br>infants) to elderly | Children aged 6 (except<br>infants) to elderly | Same | |
| Anatomical Site | The dental, oral,<br>maxillofacial region<br>ENT (Ear, Nose and<br>Throat) and neck region | The dental, oral,<br>maxillofacial region<br>ENT (Ear, Nose and<br>Throat) and neck region | Same | |
| Users | Health care<br>professionals | Health care<br>professionals | Same | |
| Technological<br>Characteristics and<br>Performance | | | | |
| Patient Contact<br>Material | CHIN REST:<br>polycarbonate<br>Forehead Holder:<br>silicone rubber<br>HANDLE GRIP:<br>silicone rubber | CHIN REST:<br>polycarbonate<br>Forehead Holder:<br>silicone rubber<br>HANDLE GRIP:<br>silicone rubber | Same | |
| Sterility | Non-sterile | Non-sterile | Same | |
| X-ray<br>Generati<br>on<br>Device | Tube<br>Voltage | 90-110KV | 90-110KV | Same |
| | Pulse<br>Exposure<br>function | Yes | Yes | Same |
| | Tube<br>Current | 1-5.3mA | 1-5.3mA | Same |
| | Focal Spot<br>Size | 0.3mm x 0.3mm | 0.3mm x 0.3mm | Same |
| Collimator Size | | CT scan (Face) /CT-<br>Panorama Scan (Face):<br>33.6mm x 27mm<br>CT scan (Arch): 20.6mm<br>x 20.8mm<br>CT scan (Full): 23mm x<br>27mm<br>CT scan (Teeth) :<br>10.5mm x 10.2mm<br>Panoramic Scan: 1mm x<br>24.2mm<br>Cephalometric Exposure:<br>22.0mm x 17.6mm | CT scan (Face) /CT-<br>Panorama Scan (Face):<br>Same<br>CT scan (Arch): Same<br>CT scan (Full): Same<br>CT scan (Teeth): Same<br>Panoramic Scan: Same<br>Cephalometric Exposure:<br>23.6mm x 18.7mm | Similar<br>(Same except<br>for<br>Cephalometric<br>Exposure) |
| | Detector | FPD (TFT) | FPD (TFT) | Same |
| X-ray<br>Image<br>Capturin<br>g Device | Pixel Size | 248 µm x248µm (With<br>binning)<br>(CT, CT-Panoramic,<br>Panoramic)<br>124 µm x124µm<br>(Without binning)<br>(CT, CT-Panoramic,<br>Panoramic, Ceph) | 248 µm x248µm (With<br>binning)<br>(CT, CT-Panoramic,<br>Panoramic)<br>124 µm x124µm<br>(Without binning)<br>(CT, CT-Panoramic,<br>Panoramic, Ceph) | Same |
| | Pixel<br>Number | 1024x1280(With<br>binning)<br>(CT, CT-Panoramic) | 1024x1280(With<br>binning)<br>(CT, CT-Panoramic) | Same |
| | | 2560x2048 (Without<br>binning)<br>(CT, CT-Panoramic,<br>Ceph) | 2560x2048 (Without<br>binning)<br>(CT, CT-Panoramic,<br>Ceph) | Same |
| | | 1900 x 120<br>(Panoramic) | 1900 x 120<br>(Panoramic) | Same |
| | | 2560 x 2048<br>(Cephalometric) | 2560 x 2048<br>(Cephalometric) | Same |
| | Size of<br>Area<br>Receiving<br>X-ray | 253.95mm x 317.44mm<br>(CT, CT-Panoramic)<br>230mm x 15mm<br>(Panoramic) | 253.95mm x 317.44mm<br>(CT, CT-Panoramic)<br>230mm x 15mm<br>(Panoramic) | Same |
| | | | | Image: Prexion3D logo with text "make IT visible" |
| | | 239mm x 302mm<br>(Ceph) | 253.95mm x 317.44mm<br>(Ceph) | Smaller |
| | Number of<br>Bits | 16bits (CT, Panorama, Ceph) | 16bits (CT, Panorama, Ceph) | Same |
| Scanner | SID/SOD | 700mm/ 420mm (CT, CT-Panoramic, Panoramic)<br>1000mm / 840mm (Ceph)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.