The X-ray unit system is a diagnostic imaging system which consists of multiple image acquisition modes; panoramic, cephalometric, and CBCT (Cone Beam Computed Tomography). X-ray unit system is used for dental radiographic examination and diagnosis of teeth, jaw, oral structures and skull. The device is to be operated and used by dentists and other legally qualified professionals.
Device Story
Diagnostic imaging system for dental/craniofacial applications; utilizes X-ray tube and flat panel detectors (CMOS/TFT) to acquire panoramic, cephalometric, and 360-degree rotational CBCT image sequences. Operated by dentists/qualified professionals in clinical settings. Transforms X-ray inputs into 2D/3D anatomical reconstructions. Output viewed via Triana or Theia software for diagnostic assessment of teeth, jaws, and skull. Facilitates surgical planning and treatment monitoring. Benefits include high-resolution visualization of oral/craniofacial structures for improved diagnostic accuracy.
Clinical Evidence
Bench testing and clinical image evaluation. Imaging performance of new detectors (FXDD-0909GA, FXDD-1012CA) validated using phantoms (spatial resolution, contrast-to-noise ratio, geometric distortion, etc.). Clinical images evaluated by US board-certified oral surgeon to confirm diagnostic adequacy. Software (Theia) validated per EN 62304 and FDA guidance.
Technological Characteristics
Computed tomography X-ray system. Materials: Plastic enclosure. Sensing: CMOS and TFT flat panel X-ray detectors. Energy: X-ray tube (60-90 kV, 4-12 mA). Connectivity: DICOM compliant. Software: Triana or Theia. Power: 100-240V, 50/60Hz. Form factor: Floor-standing dental imaging unit with optional cephalometric arm.
Indications for Use
Indicated for dental radiographic examination and diagnosis of teeth, jaw, oral structures, and skull in patients requiring panoramic, cephalometric, or CBCT imaging. Operated by dentists and qualified professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 16, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Genoray Co., Ltd. % Ms. Kaitlynn Min Business Development Manager Genoray America, Inc. 147 E. Bristol Lane ORANGE CA 92865
## Re: K200469
Trade/Device Name: PAPAYA 3D Premium & PAPAYA 3D Premium Plus Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS, MUH Dated: August 12, 2020 Received: August 13, 2020
Dear Ms. Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200469
Device Name PAPAYA 3D Premium & PAPAYA 3D Premium Plus
Indications for Use (Describe)
The X-ray unit system is a diagnostic imaging system which consists of multiple image acquisition modes; panoramic, cephalometric, and CBCT (Cone Beam Computed Tomography). X-ray unit system is used for dental radiographic examination and diagnosis of teeth, jaw, oral structures and skull. The device is to be operated and other legally qualified professionals.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit 5 510(k) Summary
## K200469
Date of Summary Preparation: February 21, 2020
#### 1. Submitter and US Official Correspondent
Submitter: GENORAY Co., Ltd. Address: 512, 560, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-Do, Korea Telephone No .: +82-31-740-4100 Fax: +82-31-740-4195
### Official Correspondent (U.S): Kaitlynn Min - Business Manager
Correspondent: GENORAY America Inc. Address: 147 E. Bristol Lane, Orange, CA 92865 USA Telephone No .: +1-855-436-6729 Fax: +1-714-786-8919 Email: kaitlynn@genorayamerica.com
#### 2. Establishment Registration Number
: 3005843418
#### 3. Device Information
- Product/Trade Name: PAPAYA 3D Premium & PAPAYA 3D Premium Plus
- 510(k) Number: K200469
- Common Name: Computed Tomography X-ray System
- Classification Name: Computed Tomography X-ray System
- Classification panel: Radiology
- Regulation number: 21 CFR 892.1750
- Device Class: Class II
- Primary Product code: OAS
- Secondary product code: MUH
### 4. Predicate Device (Equivalent Legally Marketed Device)
PAPAYA 3D Plus (K150354, GENORAY Co., Ltd)
#### 5. Description of the Device
The proposed device PAPAYA 3D Premium Plus the computed tomography x-ray system which consists of image acquisition modes; panorana, cephalometric, and computed tomography. The difference between PAPAYA 3D Premium Plus is only optional of the cephalometric detector. It designed for dental radiography of the oral and cranionarial as teeth, jaws and oral structures. The device with cephalometric detector is named PAPAYA 3D Premium Plus and the device without cephalometric detector is named PAPAYA 3D Premium.
The proposed device are composed of flat panel x-ray detectors which are based on CMOS, and TFT detector types and divided in to CT, panoramic and cephalometric radiography, and x-ay tube. CMOS, and TFT detectors are used to capture scamed image for obtaining diagnostic information for craniofacial surgery or other treatments. And it also provides 3D the anatomic stuctures by acquiring 3609rotational image sequences of oral and craniofacial area.
#### 6. Indications for use (intended use)
The X-ray unit system is a dagnostic maging system which consists of multiple mage acquisition modes; panorams, cephalometric, and CBCT (Cone Beam Computed Tomography). X-ray unit system is used for dental radiographic examination and diagnosis of teeth, aw, oral stuctures and skull. The device is to be operated and used by dentists and other legally qualified professionals.
{4}------------------------------------------------
## 7. Substantial equivalence chart
| Criteria | Proposed Device | | Predicate Device |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | PAPAYA 3D Premium &<br>PAPAYA 3D Premium Plus | | PAPAYA 3D Plus |
| Manufacturer | GENORAY Co., Ltd. | | GENORAY Co., Ltd. |
| 510(k) No. | K200469 | | K150354 |
| Figure | Image: PAPAYA 3D Plus | | |
| Indications<br>for use | The X-ray unit system is a diagnostic imaging system which consists of<br>multiple image acquisition modes; panoramic, cephalometric, and CBCT<br>(Cone Beam Computed Tomography). X-ray unit system is used for dental<br>radiographic examination and diagnosis of teeth, jaw, oral structures and<br>skull. The device is to be operated and used by dentists and other legally<br>qualified professionals. | | PAPAYA 3D Plus is a digital panoramic, cephalometric and tomographic<br>extra-oral X-ray system,<br>indicated for use in:<br>(i) Producing panoramic X-ray images of the maxillofacial area, for<br>diagnostic examination of dentition (teeth), jaws and oral structures; and<br>(ii) Producing radiographs of jaws, parts of the skull and carpus for the<br>purpose of cephalometric examination, when equipped with the<br>cephalometric arm;<br>(iii) Producing tomographic images of the oral and maxillofacial structure,<br>for diagnostic examination of dentition (teeth), jaws, oral structures and<br>some cranial bones if equipped with CBCT option.<br>The system accomplishes tomographic exam by acquiring a 360-degree<br>rotational X-ray sequence of images and reconstructing a three-dimensional<br>matrix of the examined volume, producing two-dimensional views of this<br>volume and displaying both two dimensional images and three-dimensional<br>renderings |
| Performance Specification | Panoramic | Panoramic<br>Cephalometric | Panoramic<br>Cephalometric |
| | Computed tomography | Computed tomography | Computed tomography |
| Input Voltage | 100-240 V~, 50/60Hz | 100-240 V~, 50/60Hz | 100-120 V~, 50/60Hz |
| Tube Voltage | 60-90 kV | 60-90 kV | 60-90 kV |
| Tube Current | 4-12 mA | 4-12 mA | 4-12 mA |
| Focal Spot Size | 0.5 mm | 0.5 mm | 0.5 mm |
| Exposure Time | Panorama: max 17 sec.<br>CBCT: max 24sec | Panorama: max 17 sec.<br>Cephalometric: max 15.5 sec.<br>CBCT: max. 24sec | Panorama: max 17sec.<br>Cephalometric: max 12sec.<br>CBCT: max. 15sec. |
| Image Receptor | Panoramic image receptor:<br>- Extor-P (K150354)<br>CBCT image receptor:<br>- DualRay-S (K150354)<br>- FXDD-0606CA (K182805)<br>- FXDD-0909GA | Panoramic image receptor:<br>- Extor-P (K150354)<br>Cephalometric image receptors:<br>- Extor-C (K150354)<br>- FXDD-1012CA<br>CBCT image receptors:<br>- DualRay-S (K150354)<br>- FXDD-0606CA (K182805)<br>- FXDD-0909GA | Panoramic image receptor:<br>- Extor-P<br>Cephalometric image receptor:<br>- Extor-C<br>CBCT image receptor:<br>- DualRay-S |
| Image processing Software | Triana (K103182), Theia (optional) | Triana (K103182), Theia (optional) | Triana (K103182) |
| Result | The proposed device, PAPAYA 3D Premium & PAPAYA 3D Premium Plus, was developed from the predicate device PAPAYA 3D Plus (K150354), and<br>modifications are changing of input power, addition of detector, addition of image processing software, and minor external design changes. Accordingly, the<br>characteristics of the proposed device is identical or similar to those of the predicate device regarding X-ray generation device characteristics including tube<br>voltage, tube current, focal spot size.<br>First of all, the input voltage (power rating) of Predicate device was 100-120V~, 50/60Hz, however due to changes of the power board, proposed device is<br>available to 100-240V~, 50/60Hz.<br>Second, the proposed device has same detector with predicate device which are DualRay-S (CT detector), Extor-C (cephalometric detector), Extor-P<br>(panoramic detector), and additionally proposed device has FXDD-1012CA (cephalometric detector), FXDD-0606CA (CT detector), and FXDD-0909GA (CT<br>detector). However cephalometric detectors (Extor-C, and FXDD-0909GA) are only available for PAPAYA 3D Plus and PAPAYA 3D premium Plus. Overall<br>PAPAYA 3D Premium has one option for panoramic detector, and three types of CT detectors and PAPAYA 3D Premium Plus has one option for panoramic<br>detector, two options for cephalometric detectors, and three types of CT detectors. Additionally, the FXDD-0606CA detector has been cleared under K182805,<br>RCT800.<br>Third, the predicate device can be used with image processing software, Triana, only. The proposed device can be used with same software, Triana, or newly<br>developed image processing software, Theia. The difference between Triana and Theia is only UI, and the Theia was developed for marketing purpose only.<br>Theia was validated throughout EN 62304:2006/AC: 2008.<br>Fourth, the minor external design of the proposed device has changed comparing predicate device. The patient supporting area can be foldable, the emergency<br>switch on the patient supporting area has been moved from downside to upside, the storage box has been added, and the plastic enclosure design has changed. | | |
{5}------------------------------------------------
#### 8. Safety, EMC and Performance data comparison to Predicate
Since electrical and mechanice of the proposed device las changed, such as input voltage, additional detector options, additional inage processing software option, external design, companing to predicate device the safety and EMC test was performed throughout IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-63, and IEC 60601-1-2 to minimize electrical, mechanical and radiation hazards
For the additional software, Theia, saves the patient inage data and offers an inquiry function in addition, supports the mage processing finction intended to obtain images by using the PAPAYA 3D Premium Plus and detector options for dagnosis. And has been validated according to NEMA PS 3.1-3.20. DICOM, FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical devices". FDA Guidance "Guidance for the content of premarket submissions for management of cyber security" to ensure substantial equivalence. The software, Theia, was consider level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.
For the newly added CBCT image receptor (FXDD-0909GA), and Cephalometric mage receptor (FXDD-1012CA) it has been validated according to FDA "Gudance for the submission of 510(k)'s for Solid State X-ray Imaging Devices" as clinical considerations, and the climcal images were evaluated by the US board-certified oral surgeon. Throughout the evaluation by oral surgeon, the proposed device with newly added detectors are well enough to diagnosable and meet its indications for use.
Furthernore, the imaging performance of newly added CBCT image receptor (FXDD-0909GA) such as ganty positioning accuracy, In-plane uniformity, Spatial Resolution section thickness, Nose, Contrast to Noise Rato, Geometric Distortion, and Metal Artifacts and Cephalometric image receptor (FXDD-1012CA) such as Line pair resolution was tested with phantoms to validate its performance. The detail information about the clinical image evaluation can be found in the "Attachment X Evaluation of PAPAYA 3D Premium & PAPAYA 3D Premium Plus mages" and the imaging performance test report can be found in the "Attachment IX Solid State X-ray Imaging Device".
Overall the modifications from the predicate device were completed in accordance GENORAY management system design controls. A risk analysis, validation testing and Declaration of Controls were submitted to support the substantial equivalence of the modified proposed device. The results of them indicate that the modified proposed device is as safe and effective as the predicate devices.
#### 9. Conclusion
Proposed device has the same indication for use as the predicate device have been no changes that impact ether the findamental technology or the indications for use. The proposed device with modifications outlined in this Premarket Notification is substantally equivalent to the currently commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.