rainbow MCT is a computed tomography x-ray system intended to produce 3D and panoramic diagnostic images of the maxillofacial areas for treatment planning for adult and pediatric patients. The device is operated and used by physicians, dentists, and x-ray technicians. Rainbow 3D Image Viewer software functions for acquiring, saving, searching, displaying, diagnosing and sending digital X-ray image data in dental practices and clinics.
Device Story
Cone beam CT X-ray system; generates sectional images of teeth, nasal cavity, and temporomandibular joint. X-ray source and detector rotate around patient; 2D images reconstructed into 3D volumetric view via mathematical algorithm. System components: X-ray generator, detector, collimator, main frame, rotation unit, PC, and monitor. Used in dental practices/clinics by physicians, dentists, and x-ray technicians. Rainbow 3D Image Viewer software manages acquisition, storage, display, and transmission of digital X-ray data. Output viewed on monitor for diagnostic and treatment planning purposes; assists clinicians in visualizing maxillofacial anatomy.
Clinical Evidence
Bench testing only. Performance evaluated per IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63, IEC 60601-1-2, IEC 61223-3-4, and IEC 61223-3-5. Imaging performance (MTF, DQE, pixel resolution) compared against predicate and reference devices. No clinical data required.
Technological Characteristics
Computed tomography x-ray system; 60-100 kV tube voltage; 4-12 mA tube current; 0.5x0.5 mm focal spot; 2.5 mm Al filtration. Uses DTX3024 detector. Connectivity via DICOM 3.0. Software: Rainbow 3D ImageViewer. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric patients requiring 3D and panoramic diagnostic imaging of the maxillofacial area for treatment planning.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
April 16, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Dentium Co., Ltd (ICT Branch) % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 7707 Fannin St. Ste 200, V111 HOUSTON TX 77054
Re: K200270
Trade/Device Name: rainbow MCT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: March 10, 2021 Received: March 15, 2021
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200270
Device Name rainbow MCT
### Indications for Use (Describe)
rainbow MCT is a computed tomography x-ray system intended to produce 3D and panoramic diagnostic images of the maxillofacial areas for treatment planning for adult and pediatric patients. The device is operated and used by physicians, dentists, and x-ray technicians.
Rainbow 3D Image Viewer software functions for acquiring, saving, searching, displaying, diagnosing and sending digital X-ray image data in dental practices and clinics.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table><tr><td><span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span></td></tr><tr><td><span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span></td></tr></table> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> |
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> | | |
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="color: black;"> </span> </span> </span> <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> | | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K200270
This 510(k) is being submitted in accordance with the requirements of 21 CFR §807.92.
| 1. | Date Summary Prepared: | March 10,2021 |
|----|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Submitter's Identification: | |
| | Submitter's Name :<br>Submitter's Address: | Dentium Co., Ltd (ICT Branch)<br>76, Changnyong-daero 256beon-gil, Yeongtong-gu,<br>Suwon-si, Gyeonggi-do, 16229<br>Republic of Korea |
| | Submitter's Telephone: | ++82-70-7098-6932 |
| | Contact person: | Mr. Sang Woo Lee (swlee1@dentium.com) |
| | Official Correspondent:<br>(U.S. Designated agent) | Dave Kim (davekim@mtech-inc.net) |
| | Address:<br>Telephone: | 7707 Fannin St. Ste 200, Houston, TX 77054<br>+1- 713-467-2607 |
#### 3. Device:
| Trade /Proprietary Name: | rainbow MCT |
|--------------------------|-------------------------------------|
| Device: | X-Ray, Tomography, Computed, Dental |
| Regulation Description: | Computed tomography x-ray system. |
| Review Panel: | Radiology |
| Product Code: | OAS |
| Regulation Number | 892.1750 |
| Device Class | II |
#### 4. Predicate Device:
| Legally Marketed Predicate Device Information: | |
|------------------------------------------------|-------------------------------------|
| 510(k) Number: | K181432 |
| Trade /Proprietary Name: | ProVecta 3D Prime with VistaSoft |
| Device: | X-Ray, Tomography, Computed, Dental |
| Regulation Description: | Computed tomography x-ray system. |
| Review Panel | Radiology |
| Product Code | OAS |
| Regulation Number | 892.1750 |
| Device Class | II |
{4}------------------------------------------------
#### 5. Reference Device:
Legally Marketed Predicate Device Information:
| 510(k) Number: | K102196 |
|---------------------------|----------------------------------|
| Trade / Proprietary Name: | PaX-Zenith3D |
| Regulation Description: | Computed tomography x-ray system |
| Review Panel | Radiology |
| Product Code | OAS |
| Regulation Number | 892.1750 |
| Device Class | II |
#### 6. Device Description:
- rainbow MCT is a cone beam CT X-ray device for generating sectional images of dental images such as tooth, nasal cavity and temporomandibular joint. this is a medical diagnostic equipment designed to generate sectional images by placing X-ray source opposite to the imaging detector unit and rotating it around a patient. 2D images of the region of interest are reconstructed using a mathematical algorithm in 3 dimensional volumetric view and displayed on the computer monitor.
- The system is composed of X-ray generator, X-ray detector, X-ray collimator, main frame, rotation unit, PC and Monitor, etc. in compliance with US performance standard and regulatory requirement.
#### 7. Indications for use:
rainbow MCT is a computed tomography x-ray system intended to produce 3D and panoramic diagnostic images of the maxillofacial areas for treatment planning for adult and pediatric patients. The device is operated and used by physicians, dentists, and x-ray technicians.
Rainbow 3D Image Viewer software features functions for acquiring, saving, searching, displaying, diagnosing and sending digital X-ray image data in dental practices and clinics.
{5}------------------------------------------------
### 8. Summary of the technological characteristics of the device compared to the predicate devices:
| Descriptive Information | | Rainbow MCT<br>K200270 | K181432, ProVecta 3D Prime<br>with VistaSoft (DÜRR DENTAL) |
|-----------------------------------------------------------------------------------------------|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | - rainbow MCT is a computed<br>tomography x-ray system intended<br>to produce 3D and panoramic<br>diagnostic images of the<br>maxillofacial areas for treatment<br>planning for adult and pediatric<br>patients. The device is operated<br>and used by physicians, dentists,<br>and x-ray technicians.<br>Rainbow 3D Image Viewer software<br>features functions for acquiring,<br>saving, searching, displaying,<br>diagnosing and sending digital X-ray<br>image data in dental practices and<br>clinics. | ProVecta 3D Prime is computed<br>tomography x-ray unit intended to<br>generate 3D and panoramic X- ray<br>images in dental radiography for<br>adult and pediatric patients. It<br>provides diagnostic details of the<br>maxillofacial areas for a dental<br>treatment. The device is operated<br>and used by physicians, dentists, and<br>x-ray technicians.<br>Not intended for mammography use.<br>The VistaSoft software features<br>functions for recording, displaying,<br>analyzing, diagnosing, managing and<br>sending digital or digitized video and<br>X-ray images in dental practices and<br>specialist dental clinics. Not intended<br>for mammography use |
| Image Acquisition Modes | | Panoramic and computed<br>tomography | Panoramic and computed<br>tomography |
| Input Voltage | | AC 100-240 V, 50/60 Hz | AC 200-240V |
| Tube Voltage | | 60~100 kV | 50-99 KV |
| Tube Current | | 4~12 mA | 4~16mA |
| Focal Spot Size | | 0.5 x 0.5 mm | 0.5 mm |
| Exposure Time | | Max. 19 s | Max. 16.4s |
| Slice Width | | 0.1 mm min. | 0.1 mm min. |
| Total Filtration | | 2.5 mm Al | 2.8 mm Al |
| Chin Rest | | Bite block, chin rest and headrest | Bite block, chin rest and headrest |
| Mechanical | | Compact design | Compact design |
| Electrical | | LDCP logic circuit (Low Dark Current<br>Processing) | LDCP logic circuit (Low Dark Current<br>Processing) |
| Software | | Rainbow 3D ImageViewer, DICOM<br>3.0 Format compatible | VistaSoft, DICOM 3.0 compatible |
| Anatomical Sites | | Maxillofacial | Maxillofacial |
| Image<br>Receptor<br>Note: CT and | Panoramic | DTX3024 | Xmaru1404CF |
| | CBCT | DTX3024 | Xmaru1404CF |
| panoramic<br>image<br>performance<br>is identical<br>because the<br>sensors are<br>identical. | MTF@ 1<br>lp/mm | 50% | 53% |
| | DQE @ 0.5<br>lp/mm | 63% | 64% |
| | | | |
| Size of Imaging<br>Volume (cm) | | DTX3024: Max. 10x8, 23x21 | Xmaru1404CF : Max. 10x8.5 |
| Pixel Resolution | CBCT | DTX3024: 5 lp/mm | 2.5 lp/mm - 4x4 binning |
| | Panoramic | DTX3024: 5 lp/mm | 2.5 lp/mm - 4x4 binning |
| Pixel Size | CBCT | DTX3024 :100 μm | Xmaru1404CF : 99 μm -<br>2x2 binning<br>198 μm- 4x4 binning |
| | Panoramic | DTX3024:100 μm | Xmaru1404CF : 99 μm -<br>2x2 binning<br>198 μm- 4x4 binning |
# Summary of the Technological Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
#### 9. Rainbow MCT detectors VS. PaX-Zenith 3D Detectors
| | | Rainbow MCT<br>(K200270) | Pax-Zenith3D (K102196) | | |
|--------------------------------------------------------|--------------------|------------------------------|--------------------------------|--------------------------------|-------------|
| Imaging<br>detector for<br>CT and<br>panoramic<br>mode | CBCT | DTX3024 (10x8, 23<br>x 21cm) | Xmaru2430CF<br>(FOV 24 x 19cm) | Xmaru1524CF<br>(FOV 15 x 16cm) | |
| | Panoramic | DTX3024 | | | Xmaru1501CF |
| Imaging<br>Performance | MTF@ 1<br>lp/mm | 50% | 53% | 52% | 50% |
| | DQE @ 0.5<br>lp/mm | 63% | 64% | 45% | 60% |
| Size of Imaging<br>Volume (cm) | | | | | |
| Pixel Resolution | CBCT | 5 lp/mm - 1x1 | 2.5 lp/mm - 4x4<br>binning | | |
| | Panoramic | 5 lp/mm - 1x1 | | | 5 lp/mm |
| Pixel Size | CBCT | 100 μm | 200 μm- | 200 μm- | |
| | Panoramic | 100 μm | | | 100 μm |
# 10. Discussion of Similarities and Differences:
Rainbow MCT dental computed tomography X-ray system described in this 510(k) is similar to the predicate device in its indications for use, performance, materials, and safety characteristics.
The differences include the digital X-ray imagers and image viewing software. The subject device offers FOV 10x8 & 23x21 cm whereas the predicate device and the reference device provide FOV 10.8.5 and 24x19 cm, respectively. Performance testing was conducted for the subject device to access whether or not the parameter required for functionalities related to imaging properties of the dental X-ray device meets the designated acceptance criteria. The MTF, DQE and pixel resolution of the subject device performed similar or better than those of the predicate and reference device.
All test results were satisfactory.
# 11. Non-Clinical Data and Performance Testing
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(2005, AMD 1: 2012), IEC 60601-1-3 (2008 + A1: 2013), IEC 60601-2-63 (2012,041: 2017) )were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2 (2014).
rainbow MCT meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
{8}------------------------------------------------
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed. Acceptance test according to IEC 61223-3-4 and IEC 61223-3-5 was performed. All test results were satisfactory.
- 12. Clinical Data: Not required for a finding of substantial equivalence.
# 13. Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate device in terms of technology, performance and indications for use, Dentium Co., Ltd concludes that the rainbow MCT is substantially equivalent to ProVecta 3D Prime with VistaSoft (K181432), the predicate device, and Pax-Zenith3D (K102196), the reference device as described herein.
The differences between the new device and the predicate device shown in the comparison table above do not raise any new questions about safety and effectiveness and so we consider it substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.