Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
Device Story
Planmeca Viso is a CBCT imaging system for dento-maxillo-facial, cervical spine, and ENT regions. It captures hundreds of 2D sample images from multiple angles using a fixed anode X-ray tube and flat panel detector. Data is processed via a separate 3D reconstruction PC to generate 3D volumes. The system also performs 2D panoramic imaging (tomosynthesis) and cephalometric imaging. Operated by healthcare professionals in clinical settings, the device allows for patient positioning (seated or standing) with an open design. Output images are viewed on a separate workstation using Planmeca Romexis software. Clinicians use these images to support diagnostic decision-making. Benefits include high-resolution 3D visualization of anatomical structures for improved diagnostic accuracy.
Clinical Evidence
Bench testing only. Performance bench testing compared image quality of Planmeca Viso against predicate and reference devices, demonstrating substantial equivalence.
Technological Characteristics
CBCT imaging system; fixed anode X-ray tube (0.5mm focal spot); Varex Imaging flat panel detectors (aSi with CsI scintillator); 16-bit bit depth; tube voltage 60-120 kV; tube current 1-16 mA; voxel size 75-600µm; 300-600 projections over 210°/360° rotation. Connectivity via Planmeca Romexis software workstation.
Indications for Use
Indicated for pediatric and adult patients requiring 2D/3D digital x-ray or 3D optical imaging of dento-maxillo-facial, cervical spine, and ENT regions for diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of a human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
September 13, 2018
Planmeca Oy Lars Moring Regulatory Affairs Specialist Asentajankatu 6 FI-00880 Helsinki FINLAND
Re: K181576
Trade/Device Name: Planmeca Viso Regulation Number: 21 CFR 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: OAS Dated: August 15, 2018 Received: August 17, 2018
Dear Lars Moring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Rodgers
Digitally signed by Jeffrey J. Ballyns -S DN· c=US o=US Government ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=20005697 25, cn=Jeffrey J. Ballyns -S Date: 2018.09.13 16:30:25 -04'00'
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181576
Device Name Planmeca Viso
Indications for Use (Describe)
Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dento-maxillo-facial, cervical spine and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510K) SUMMARY
#### I. SUBMITTER
### MANUFACTURER
Planmeca Oy Asentajankatu 6 FI-00880 Helsinki, Finland Phone: +358 20 7795 500 Contact person: Lars Moring
## U.S. DESIGNATED AGENT
Planmeca USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 894 2300 Contact person : Bob Pienkowski
Date prepared: September 7, 2018
#### PRODUCT, CLASSIFICATION NAME II.
Name of the device: Common or Usual Name: Classification name: Requlatory Class: Product Code: 510(k) number:
Planmeca Viso Computed tomography x-ray system Computed tomography x-ray system (21 CFR 892.1750) -OAS #K181576
#### III. PREDICATE DEVICE
Planmeca ProMax 3D Max #K160506
#### IV. REFERENCE DEVICE
Planmeca ProMax 3D Mid #K103689 RAYSCAN A-EXPERT 3D, #K142247
## V. DEVICE DESCRIPTION
The Planmeca Viso -X-ray unit uses cone beam computed tomography (CBCT) to produce three-dimensional (3D) images of the maxillofacial and ENT anatomies. Two dimensional (2D) images are produced with tomosynthesis method (panoramic) imaging) as well as conventional 2D radiography (cephalometric imaging, 2D views). In CBCT a cylindrical volume of data is captured in one imaging procedure. The data consists of several hundred sample images which are taken from different directions to cover a certain pre-programmed target area. These samples are used for 3D reconstruction (using a separate 3D reconstruction PC) that can be viewed in three dimensions using separate workstation and Planmeca Romexis software.
## VI. INDICATIONS FOR USE
Planmeca Viso is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images as well as three-dimensional (3D) optical images of the dento-maxillofacial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.
{4}------------------------------------------------
The Indications for Use (IFU) for the subject device are the same as those for the predicate device.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both devices are 3D CBCT imaging system indicated for imaging dento-maxillofacial-, cervical spine- and ENT-regions and offer a variety of 3D fields of view suitable for each application. Both systems also provide 2D imaging capabilities, predicate device only for panoramic and Planmeca Viso both for panoramic and cephalometric imaging.
Both systems are based on fixed anode X-ray tube and a flat panel detector of the same technology. Pixel sizes of the detector differ but voxel sizes on resulting 3D images are the same." Tube voltage range with Planmeca Viso is wider and ranges up to 120kV, providing users with option to utilize even higher X-ray penetration when necessary.
| | Planmeca Viso | Predicate device Planmeca ProMax<br>3D Max |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Planmeca Viso is a system intended to<br>produce two-dimensional (2D) and<br>three-dimensional (3D) digital x-ray<br>images as well as three-dimensional<br>(3D) optical images of the dento-<br>maxillo-facial, cervical spine and ENT<br>(Ear, Nose, and Throat) regions at the<br>direction of healthcare professionals as<br>diagnostic support for pediatric and<br>adult patients. | The Planmeca ProMax 3D Max or<br>Maximity is a system intended produce<br>two-dimensional (2D) and three-<br>dimensional (3D) digital X-ray images o<br>the dento-maxillo-facial, cervical spine<br>and ENT (Ear, Nose and Throat)<br>regions at the direction of healthcare<br>professionals as diagnostic support for<br>pediatric and adult patients. |
| 3D Imaging | | |
| Technical<br>characteristics | • CBCT (cone beam computed<br>tomography)<br>• 300 - 600 projections over 210° / 360°<br>rotation<br>• Reconstruction time 30 s - 6 min<br>• Voxel size 75-600µm | • CBCT (cone beam computed<br>tomography)<br>• 300 - 600 projections over 210° / 360°<br>rotation<br>• Reconstruction time 30 s - 6 min<br>• Voxel size 75-600µm |
| Field Of View (FOV) | Ø30x30mm - Ø200x170mm (G5) /<br>Ø300x300mm (G7) | Ø42x48mm - Ø230x260mm |
| 2D Imaging | | |
| Technology | • SmartPan panoramic imaging,<br>tomosynthesis<br>• Multiview panoramic imaging,<br>tomosynthesis<br>• Cephalostat | • SmartPan panoramic imaging,<br>tomosynthesis<br>• Multiview panoramic imaging,<br>tomosynthesis<br>• No cephalostat |
| Imaging system<br>performance<br>specifications | | |
| X-ray detector | • Varex Imaging 16x16cm flat panel<br>(G5) / 25x30cm flat panel detector<br>(G7)<br>• aSi with Csl scintillator<br>• 127µm pixel pitch (G5) / 139µm pixel<br>pitch (G7)<br>• Bit depth 16bit<br>• Varian PaxScan 2530C<br>30.2(w)x24.9(h)cm<br>• GADOX scintillator | • Varex Imaging 25x20cm flat panel<br>detector<br>• aSi with Csl scintillator<br>• 127µm pixel pitch<br>• Bit depth 15bit |
### Table 1: Comparison with predicate device
{5}------------------------------------------------
| X-ray tube and<br>generator | | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | • Fixed anode X-ray tube focal spot size 0.5 mm<br>• Tube voltage 60-120 kV<br>• Tube current 1-16 mA<br>• Exposure time 0.1–36 sec (pulsed) | • Fixed anode X-ray tube with 0.6mm focal spot<br>• Tube voltage 66-96 kV<br>• Tube current 1-12,5 mA<br>• Exposure time 3–24 sec (pulsed) |
| Other | | |
| Workflow | • Open positioning, doctor facing the patient<br>• Patient seated or standing | • Open positioning, doctor facing the patient<br>• Patient seated or standing |
| Imaging Software | • Planmeca Romexis | • Planmeca Romexis |
Cephalometric scanning feature and detector are identical to those in previously cleared reference devices. The PaxScan 2530C, used for cephalometric imaging in the subject device, was included as a cephalometric detector in a previously cleared 510(k) for the RAYSCAN a-Expert 3D (K142247).
## VIII. PERFORMANCE DATA
Results from performance bench testing demonstrate that Planmeca Viso produces substantially equivalent image quality compared to the predicate and reference devices.
## IX. CONCLUSIONS
The comparison of characteristics supports substantial equivalence. Planmeca Viso is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.