ENCOMPASS Eagle 3D CBCT/Panoramic/Cephalometric Dental X-Ray
Applicant
Panoramic Corp.
Product Code
OAS · Radiology
Decision Date
Feb 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
ENCOMPASS Eagle 3D is intended to acquire two-dimensional digital panoramic and cephalometric radiographies, and multi-field of view 3D computed tomography images of dento-maxillo-facial region for the purpose of advanced diagnosis at the direction of qualified dental healthcare professionals
Device Story
ENCOMPASS Eagle 3D is a 3-in-1 dental imaging system; performs tomography, digital panoramic, and cephalometric profiles. Device uses CCD/CMOS sensor technology with scintillator; features three movement axes (two orthogonal, one rotational) for complex imaging profiles around dental arch. System includes radiographic emission compensation for spinal region; reconstructs dental arch into plane images. Operated by dental healthcare professionals in clinical settings. Device provides various imaging modes (e.g., standard, low dose, high definition) to prioritize diagnostic needs or dose reduction. Output consists of digital radiographic images used by dentists for advanced diagnosis; aids in clinical decision-making by providing detailed anatomical visualization of dento-maxillo-facial region.
Clinical Evidence
Clinical evaluation involved comparing images acquired by the subject device and the predicate device. A licensed dentist reviewed the images and determined that the ENCOMPASS Eagle 3D images are comparable in diagnostic ability, sharpness, and quality to those produced by the predicate device.
Indicated for patients requiring 2D digital panoramic/cephalometric or 3D CBCT imaging of the dento-maxillo-facial region for advanced diagnosis by qualified dental healthcare professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 4, 2016
Panoramic Corp. % Daniel Kamm, P.E. Principal Consultant Kamm & Associates 8870 Ravello Court NAPLES FL 34114
Re: K152489
Trade/Device Name: ENCOMPASS Eagle 3D CBCT/Panoramic/Cephalometric Dental X-ray Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: January 18, 2016 Received: January 22, 2016
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Robert Oolo**
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152489
Device Name
ENCOMPASS Eagle 3D CBCT/Panoramic/Cephalometric Dental X-Ray
Indications for Use (Describe)
ENCOMPASS Eagle 3D is intended to acquire two-dimensional digital panoramic and cephalometric radiographies, and multi-field of view 3D computed tomography images of dento-maxillo-facial region for the purpose of advanced diagnosis at the direction of qualified dental healthcare professionals
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K152489
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date prepared: January 18, 2016
1. Company and Correspondent making the submission: Name - Panoramic Corp. Address - 4321 Goshen Rd., Fort Wayne, IN 46818 Telephone - 800-654-2027 Contact - Doug Pack
2. Device:
Device trade name, ENCOMPASS Eagle 3D Device common name. CBCT/Panoramic/Cephalometric Dental X-Ray Regulation number, 21 CFR §892.1750 Regulation name, Computed tomography x-ray system Classification: 2 Product code) OAS
- 3. Predicate Device: The Encompass 3D is substantially equivalent to the following legally marketed device: K122199, Prexion3d Eclipse made by The Yoshida Dental Mfg. Co., Ltd. 21 CFR§ 892.1750, Class 2 Product code OAS.
- 4. Description: The ENCOMPASS Eagle 3D is a complete 3-in-1 system for dental imaging capable of make Tomography, Digital Panoramic Profiles and Digital Cephalometric Profiles. The machine uses a CCD sensor technology with the traditional scintillator technologies and auto image processing that speeds up the diagnostic and improves the clinic workflow. The equipment has three movement axes (two in orthogonal directions and one rotational) making it possible to execute elaborate imaging profiles. It features a complex profile movement around the dental arch and radiographic emission compensation in the spinal region, when necessary reconstructing the dental arch into a plane image. Each individual profile prioritizes a set of characteristics improving diagnostic capabilities. For example, the standard panoramic prioritizes image layer width, constant vertical magnification and homogeneous exposure along the whole image. Likewise, the low dosage profile prioritizes the reduction of dosage (time and anodic current). The Encompass Eagle 3D complies with applicable aspects of 21CFR 1020.30,31, and 33.
- 5. Indications for use: ENCOMPASS Eagle 3D is intended to acquire two-dimensional digital panoramic and cephalometric radiographies, and multi-field of view 3D computed tomography images of dento-maxillo-facial region for the purpose of advanced diagnosis at the direction of qualified dental healthcare professionals.
- 6. Comparison with predicate devices: The PREXION 3D ECLIPSE consists of scanner intended to produce two-dimensional digital panoramic and cephalometric images, and three-dimensional digital X-ray images of the dento-maxillo-facial region at the direction of healthcare professionals as diagnostic support. The new device ENCOMPASS Eagle 3D is a digital capture type CBCT/Pan/Ceph system. Technologies employed by the predicates and our new device are nearly identical. A comparison table follows.
{4}------------------------------------------------
| ITEM | ENCOMPASS Eagle 3D K152489 | PREXION 3D ECLIPSE<br>K122199 | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Indications for use | ENCOMPASS Eagle 3D is intended to<br>acquire two-dimensional digital<br>panoramic and cephalometric<br>radiographies, and multi-field of view 3D<br>computed tomography images of dento-<br>maxillofacial region for the purpose of<br>advanced diagnosis at the direction of<br>qualified dental healthcare professionals. | PREXION 3D ECLIPSE is intended<br>to produce two dimensional<br>digital panoramic and<br>cephalometric images, and three-<br>dimensional digital X-ray images<br>of dento-maxillofacial region at<br>the direction of healthcare<br>professionals as diagnostic<br>support. | |
| X Ray<br>Generation<br>Device | Tube Voltage | 60-85 kV | 50-90 kV |
| | Tube Current | 4-8 mA | 1-4 mA |
| | Focal Spot Size | 0.5 mm | 0.2 mm |
| X Ray image<br>capturing device | Detector | CMOS (CT)<br>CCD (Panoramic)<br>CCD (Ceph) | FPD |
| | Pixel Size | 100 μm (CT)<br>108 μm (Panoramic)<br>108 μm (Ceph) | 200 μm (CT)<br>100 μm (Panoramic)<br>54 μm (Ceph) |
| | Pixel Number | 1316 x 1312 (CT)<br>64 x 1395 (Panoramic)<br>64 x 2040 (Ceph) | 640 x 656 (CT)<br>80 x 1312 (Panoramic)<br>128 x 4080 (Ceph) |
| | Size of Area<br>receiving X-Ray | 131.6 x 131.2 mm (CT)<br>6.9 x 151 mm (Panoramic)<br>6.9 x 221 mm (Ceph) | 128.1 x 131.3 mm (CT)<br>8 x 131.3 mm (Panoramic)<br>6.9 x 312 mm (Ceph) |
| | Number of Bit | 14 bits (CT,)<br>16 bits (Ceph, Panoramic) | 14 bits (CT, Panoramic)<br>16 bits (Ceph) |
| | SID/SOD | 634mm/400mm(CT)<br>564mm/400mm (Panoramic)<br>1681mm/1511mm (Ceph) | 620mm/400mm(CT, Panoramic)<br>1650 mm/1500 mm (Ceph) |
| Scanner | Dimension (WxDxH) | 1511mm x 1074mm x 1742mm | 1245mm x 1288mm x 2045mm<br>(CT, Panoramic)<br>1805mm x 1288mm x 2045mm<br>(Ceph) |
| | Weight | 115kg (CT, Panoramic)<br>152kg (Ceph) | 260kg (CT, Panoramic)<br>300kg (Ceph) |
| Imaging Mode | | CT scan, Panoramic scan, Cephalometric<br>radiography | CT scan, Panoramic scan,<br>Cephalometric radiography |
| Panoramic Scan Performance | | Standard Panoramic: 14 sec<br>TMJ Panoramic: 14 sec<br>Maxillary Sinus: 8 sec<br>Improved Orthogonality Panoramic: 14<br>sec | Standard Mode : 14 sec |
| ITEM | ENCOMPASS Eagle 3D K152489 | PREXION 3D ECLIPSE K122199 | |
| | Low Dose Panoramic: 11 sec<br>Child Panoramic: 10.5 sec<br>Bitewing: 7.6 sec<br>Improved Bitewing: 7.6 sec | | |
| Cephalometric Radiography | LA, PA, Carpus, Oblique:<br>6.6, 10, 11, 16.5 sec | LA, PA, Carpus:<br>8, 10, 12, 15 sec | |
| | Low Dose: 16.5 sec<br>Standard Dose: 20.5 sec<br>High Definition: 25.5 sec<br>Ultra High Definition: 32.0 sec | Light Mode: 8.7 sec<br>High Definition Mode: 8.7 sec<br>Ultra High Definition Mode: 17.4 sec<br>Wide Mode: 9.1 sec x 2 | |
| | Scan Time | | |
| | | | |
| CT Scan<br>Performance | Low Dose, Standard Dose, High Definition, Ultra High Definition:<br>φ50mm x H50 mm (0.08, 0.10, 0.13, 0.16 mm)<br>φ80mm x H60 mm (0.13, 0.16, 0.20, 0.25 mm)<br>φ80mm x H80 mm (0.13, 0.16, 0.20, 0.25 mm) | Light Mode, High Definition Mode, Ultra High Definition Mode:<br>Diameter 81mm, H75mm | |
| | FOV<br>(voxel size) | | |
| | Low Dose, Standard Dose, High Definition, Ultra High Definition:<br>φ80mm x H120 mm (0.16, 0.20, 0.25, 0.32 mm)<br>φ80mm x H160 mm (0.25, 0.32, 0.40, 0.50 mm) | Wide Mode:<br>Diameter 113mm, H72mm | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| MTF: 70% | 1.05 lp/mm | 1.2 lp/mm | |
| MTF: 40% | 2.2 lp/mm | 2.2 lp/mm | |
| DQE@0 lp/mm | 46.9% | 51.7% | |
| Photo | Image: ENCOMPASS Eagle 3D K152489 | Image: PREXION 3D ECLIPSE | |
| Operating temperature | +15°C to + 30°C | Not stated | |
| Mains voltage supply | 110/127/220 or 240 Vac<br>(50/60 HZ) | 230/240 Vac ±10% / 115<br>Vac (50/60 HZ) | |
| Power consumption | 1.4 kVA | Not Stated | |
| ITEM | ENCOMPASS Eagle 3D K152489 | PREXION 3D ECLIPSE<br>K122199 | |
| Operation Mode | CBCT, Panoramic or Cephalometric | SAME | |
| Generator type | High Frequency 100 kHz High | High Frequency 40 kHz | |
{5}------------------------------------------------
{6}------------------------------------------------
- 7. Safety, EMC, Biocompatibility (N/A) and Performance Data: Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2(2001). Compliance testing was performed for the applicable portions of the following X-Ray standards: IEC 60601-1-3/2001; IEC 60601-2-7/2001; 60601-2-28/2001; IEC 60601-2-32/2001. Performance testing: MTF and DOE measurements were taken and compared to the predicate. See table above. 3D performance testing was performed yielding the MTF curve and the NPS, noise power spectrum. Accuracy testing and software validation was performed. The Encompass Eagle 3D complies with applicable aspects of 21CFR 1020.30,31, and 33. All test results were satisfactory.
- 8. Clinical Data: In accordance with the FDA guidance document on Solid State X-Ray Imaging Devices, clinical images were acquired with both the new device and the predicate device. They were compared by a licensed dentist. The Images from the Encompass machine are comparable in diagnostic ability, sharpness and quality to images from the Predicate device.
- 9. Conclusion: In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Panoramic Corp. concludes that the ENCOMPASS Eagle 3D CBCT/Panoramic/Cephalometric X-Ray is safe and effective and substantially equivalent to predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.