← Product Code [OAS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OAS) · K131695

# RAYSCAN A-EXPERT 3D (K131695)

_Ray Co., Ltd. · OAS · Nov 1, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OAS/K131695

## Device Facts

- **Applicant:** Ray Co., Ltd.
- **Product Code:** [OAS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OAS.md)
- **Decision Date:** Nov 1, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

RAYSCAN a-Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dento-maxillo-facial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Image is obtained using the standard narrow beam technique.

## Device Story

RAYSCAN a-Expert 3D is an extraoral dental X-ray system; performs panoramic, cephalometric, and 3D cone-beam computed tomography (CBCT) imaging. System utilizes a rotating C-arm with an integrated high-voltage generator, X-ray tube, and flat-panel detector. Inputs: X-ray projections at various angles. Processing: CBVT reconstruction of dentomaxillofacial volumes; 2D narrow-beam reconstruction for panoramic/cephalometric images. Outputs: 2D/3D dental images for viewing on external workstations. Used in dental clinics; operated by trained dental professionals. Provides diagnostic visualization of hard tissues (bones/teeth) to assist clinicians in implant planning, TMJ assessment, and orthodontic evaluation. Optional SMARTDent software supports image processing and archiving.

## Clinical Evidence

No clinical trials were conducted. Substantial equivalence supported by non-clinical performance testing (electrical, mechanical, and EMC safety per IEC 60601 series) and expert review of image comparisons between the subject and predicate devices. Compliance with FDA guidance for solid-state X-ray imaging devices was documented.

## Technological Characteristics

Extraoral X-ray system; 60-90kVp, 4-17mA. Detectors: Flat panel X-ray sensors (CT, Pano, One-shot Ceph, and optional CdTe Direct flat panel for Scan Ceph). Focal size: 0.5mm. Magnification: 1.11-1.39 depending on mode. Connectivity: 3D viewing stations. Standards: IEC 60601-1, 1-1, 1-2, 1-3, 2-7, 2-28, 2-32, 2-44. Software: SMARTDent (optional).

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- RAYSCAN α-Expert 3D ([K122981](/device/K122981.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

**NOV 01 2013**

K131695
Page 1 of 5

510(k) Submission- RAYSCAN a-Expe

## Special 510(k) Summary

The summary.of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

Date:

| APPLICANT      | RAY Co.,Ltd                                              |
|----------------|----------------------------------------------------------|
| ADDRESS        | #362-43 (218 Maeyoung St.) 3rd & 4th Floor,              |
|                | Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Manufacturer   | RAY Co.,Ltd                                              |
|                | #362-43 (218 Maeyoung St.) 3rd & 4th Floor,              |
|                | Wonchun-dong, Youngtong-gu, Suwon-si, Gyeonggi-do, Korea |
|                | TEL : +82-31-605-1000                                    |
|                | FAX : +82-2-6280-5534                                    |
| Contact Person | Yun-Jung HA / Manager                                    |
|                | e-mail : yunjung.ha@raymedical.co.kr                     |

## Device Name

Trade/Proprietary Name : RAYSCAN α~Expert 3D Common Name: Dental panoramic/tomography and cephalometric x-ray system

## Classification

x-ray, tomography, computed, dental (21 CFR 892.1750) Class : II Product code : OAS Panel : Radiology

## Predicate device

RAYSCAN α-Expert 3D(K122981)

{1}------------------------------------------------

## Description

RAYSCAN α-Expert 3D is a 3D computed tomography for scanning hard tissues such as bones and teeth. By rotating the c-arm that include the high voltage generator all-in-one xray tube and a detector on each end, a CBCT image of whole dentomaxillofacial is attained by recombining data from the same level that are scanned from different angles.

Panoramic image scanning function for attaining images of the entire or segmental teeth and cephalometric scanning option (One shot type & Scan type) for attaining the cephalic images are included. It allows to choose from two different types of CEPH detectors:

#### Detector Options:

Base: RAYSCAN α-3D: CT+PANO Option: RAYSCAN α-Multi 3D: CT+PANO+One-shot CEPH Option: RAYSCAN a-SM3D: CT+PANO+SCAN CEPH

SMARTDent software for processing and archiving is optional.

#### Indication for use

RAYSCAN a-Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dento-maxillo-facial 3D images.

The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations.

2D Image is obtained using the standard narrow beam technique.

{2}------------------------------------------------

| Parameter                    | RAYSCAN a-- Expert 3D [Modified]               | RAYSCANa-Expert 3D [K122981]                   |
|------------------------------|------------------------------------------------|------------------------------------------------|
| Common                       | Dental panoramic/tomography and                | Dental panoramic/tomography and                |
| Name                         | cephalometric x-ray system                     | cephalometric x-ray system                     |
| Indications                  | RAYSCAN &-Expert 3D panoramic x-ray            | RAYSCAN α-Expert 3D panoramic x-ray            |
| for use                      | imaging system with cephalostat, is an         | imaging system with cephalostat, is an         |
|                              | extraoral source x-ray system, which is        | extraoral source x-ray system, which is        |
|                              | intended for dental radiographic examination   | intended for dental radiographic examination   |
|                              | of the teeth, jaw, and oral structures,        | of the teeth, jaw, and oral structures,        |
|                              | specifically for panoramic examinations and    | specifically for panoramic examinations and    |
|                              | implantology and for TMJ studies and           | implantology and for TM3 studies and           |
|                              | cephalometry, and it has the capability, using | cephalometry, and it has the capability, using |
|                              | the CBVT technique, to generate dento-         | the CBVT technique, to generate dento-         |
|                              | maxillo-facial 3D images.                      | maxillo-facial 3D images.                      |
|                              | The device uses cone shaped x-ray beam         | The device uses cone shaped x-ray beam         |
|                              | projected on to a flat panel detector, and     | projected on to a flat panel detector, and     |
|                              | the examined volume image is reconstructed     | the examined volume image is reconstructed     |
|                              | to be viewed in 3D viewing stations.           | to be viewed in 3D viewing stations.           |
|                              | 2D Images are obtained using the               | 2D Images are obtained using the               |
|                              | standard narrow beam technique.                | standard narrow beam technique.                |
|                              | CBCT Cone beam                                 | CBCT Cone beam                                 |
| 3D technology                | Computed Tomography                            | Computed Tomography                            |
|                              | CBCT Computed tomography                       | CBCT Computed tomography                       |
|                              | Panoramic                                      | Panoramic                                      |
| Performance<br>Specification | Cephalometric(optional)                        | Cephalometric(optional)                        |
|                              | - One_shot type                                | -One_shot type                                 |
|                              | - Scan type                                    |                                                |
| Functional<br>Option         | Base                                           | Base                                           |
|                              | α -3D : CT+PANO                                | α -3D : CT+PANO                                |
|                              | Option                                         | Option                                         |
|                              | a-Multi 3D: CT+PANO+One-shot CEPH(option)      | a-Multi 3D: CT+PANO+One-shot CEPH(option)      |
|                              | a -SM3D: CT+PANO+SCAN CEPH(option)             |                                                |
|                              | Computed Tomography(CT) :                      | Computed Tomography(CT) :                      |
|                              | Flat panel X-ray sensor                        | Flat panel X-ray sensor                        |
| Delector                     | Pano : Flat panel X-ray sensor                 | Pano : Flat panel X-ray sensor                 |
| Type                         | Ceph(Optional)                                 | Ceph(Optional)                                 |
|                              | - Flat panel X-ray sensor(One-shot type)       | - Flat panel X-ray sensor                      |
|                              | - CdTe Direct flat panel sensor[Scan type]     |                                                |
| Focal size                   | 0.5mm                                          | 0.5mm                                          |
| Field of<br>View(CT)         | 90x90mm                                        | 90x90mm                                        |
|                              |                                                |                                                |

## Statement of Substantial Equivalence

{3}------------------------------------------------

| X-ray Voltage           | 60~90kVp                            | 60-90kVp                   |
|-------------------------|-------------------------------------|----------------------------|
| X-ray Current           | 4~17mA                              | 4~17mA                     |
| Total Filtration        | 2.6 mm Al equivalent                | 2.6 mm Al equivalent       |
| Magnification           | CT : 1.39                           | CT : 1.39                  |
|                         | Pano : 1.31                         | Pano : 1.31                |
|                         | Ceph[One-shot type] : 1.13          | Ceph[One-shot type] : 1.13 |
|                         | Ceph[Scan type] : 1.11              |                            |
| Scan time               | CT : 14sec                          | CT : 14sec                 |
|                         | Pano : below 14sec                  | Pano : 14sec               |
|                         | Ceph[One-shot type] : 0.3sec~3.0sec | Ceph : 0.3sec~3.0sec       |
|                         | Ceph[Scan type] : below 18sec       |                            |
| Applicable<br>Standards | IEC 60601-1                         | IEC 60601-1                |
|                         | IEC 60601-1-1                       | IEC 60601-1-1              |
|                         | IEC 60601-1-3                       | IEC 60601-1-3              |
|                         | IEC 60601-2-7                       | IEC 60601-2-7              |
|                         | IEC 60601-2-28                      | IEC 60601-2-28             |
|                         | IEC 60601-2-32                      | IEC 60601-2-32             |
|                         | IEC 60601-2-44                      | IEC 60601-2-44             |
|                         | IEC 60601-1-2                       | IEC 60601-1-2              |
| Certificate<br>Product  | CE0120(MDD93/42/EEC)                | CE0120(MDD93/42/EEC)       |

Safety details, for instance the non-clinical performance, in regards to intended use, safety characteristics, PANORAMA sensor (Detector) and CBCT sensor (Detector) and One-shot CEPH sensor(Detector) are equivalent. The only difference is the additional option of Scan type CEPH sensor.

Remaining sensors are the same, the non-clinical considerations thereof are also equivalent, and the report regarding non-clinical & clinical consideration for the added Scan CEPH sensor is provided separately.

It shares the same technological characteristics as the predicate devices. Minor technological differences do not raise any new questions regarding safety or effectiveness of the device. Based on the non-clinical and clinical considerations and the outcome of experts review of image comparisons for both devices, new RAYSCAN α-Expert 3D is substantially equivalent, in terms of safety and effectiveness, to the predicate device RAYSCAN α- Expert 3D[K122981].

{4}------------------------------------------------

510(k) Submission- RAYSCAN a-Expert 3D

K131695
Page 5 of 5

## Safety and Effectiveness Information

Electrical, mechanical, environmental safety and performance testing according to standards IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32 and IEC 60601-2-44 was performed, and EMC testing was conducted in accordance with the standard IEC 60601-1-2.

In addition, non-clinical & Clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed.

All test results were satisfactory.

### Conclusions

Based on a comparison of intended use, indications, constructions, construction materials, principal of Operations, features and technical data, the RAYSCAN α-Expert 3D system are safe and effective to perform its intended use as well as substantially equivalent to the predicate device

{5}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/12 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 1. 2013

Ray Co., Ltd. % Mr. Andrew Paeng Consultant 4747 Hoen Avenue SANTA ROSA CA 95405

Re: K131695

Trade/Device Name: RAYSCAN a-Expert 3D Regulation Number: 21 CFR 872.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: October 1, 2013 Received: October 11, 2013

Dear Mr. Paeng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{6}------------------------------------------------

Page 2 - Mr. Andrew Paeng

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Indications for Use

510(k) Number (if known): K131695

Device Name: RAYSCAN a-Expert 3D

Indications For Use:

RAYSCAN a-Expert 3D, panoramic x-ray imaging system with cephalostat, is an extraoral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dentomaxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations.

2D Image is obtained using the standard narrow beam technique.

Over-The-Counter Use _ Prescription Use _____________________________________________________________________________________________________________________________________________________________ x AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Michael D. O'Hara

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K131695 510(k) ==

Page 1 of

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OAS/K131695](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/OAS/K131695)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
