Aorta-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for suspicious regions of interest (ROIs). The device uses a deep learning algorithm to identify ROIs and produces boxes around the ROIs. The boxes are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta. Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for ROIs with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. It does not replace the role of the physician or of other diagnosic testing in the standard of care. Aorta-CAD is indicated for adults only.
Device Story
Aorta-CAD is a CADe software device analyzing digital chest radiographs; utilizes deep learning to identify and label regions of interest (ROIs) as 'Aortic calcification' or 'Dilated aorta'. Operates as a concurrent reading aid for physicians within a PACS environment; output includes DICOM Presentation State files with bounding boxes and text labels. If no ROI is detected, the overlay indicates 'No Aorta-CAD ROI(s)'. Physicians toggle the overlay on/off to assist in clinical interpretation. By highlighting suspicious findings, the device aims to improve detection accuracy for chronic aortic conditions, supporting clinical decision-making without replacing physician judgment or standard diagnostic testing.
Clinical Evidence
Retrospective MRMC study (n=244 cases) evaluated reader performance with/without Aorta-CAD. Primary endpoint: AUC of ROC curve. Results showed significant improvement in reader AUC for both categories: Aortic calcification (+0.0984, p<0.001) and Dilated aorta (+0.0885, p<0.001). Standalone performance on 5,000 cases showed high sensitivity (0.910) and specificity (0.896) with an overall AUC of 0.974.
Technological Characteristics
Software-only device; cloud-based processing. Uses supervised deep learning and computer vision algorithms. Input: digital chest X-rays. Output: DICOM Presentation State overlay with bounding boxes and text labels. Integrates with PACS via standard DICOM protocols. No patient-contacting components; no electrical/EMC testing required.
Indications for Use
Indicated for adults undergoing chest radiography to assist physicians in detecting regions of interest suggestive of Aortic Atherosclerosis or Aortic Ectasia.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
September 20, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Imagen Technologies, Inc % Robert Lindsey Chief Science Officer 151 West 26th Street, 10th Floor NEW YORK NY 10001
Re: K213353
Trade/Device Name: Aorta-CAD Regulation Number: 21 CFR 892.2070 Regulation Name: Medical Image Analyzer Regulatory Class: Class II Product Code: MYN Dated: August 15, 2022 Received: August 15, 2022
### Dear Robert Lindsey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K213353
Device Name Aorta-CAD
### Indications for Use (Describe)
Aorta-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for suspicious regions of interest (ROIs). The device uses a deep learning algorithm to identify ROIs and produces boxes around the ROls. The boxes are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta.
Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for ROls with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. It does not replace the role of the physician or of other diagnosic testing in the standard of care. Aorta-CAD is indicated for adults only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) (and 21 CFR 807.92) the 510(k) Summary for Aorta-CAD is provided below.
#### SUBMITTER 1.
| Applicant: | Imagen Technologies, Inc.<br>594 Broadway, #701<br>New York, NY 10012 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact and Primary<br>Correspondent: | Robert Lindsey, Ph.D.<br>Chief Technology Officer<br>Imagen Technologies, Inc.<br>594 Broadway, #701<br>New York, NY 10012<br>917-830-4721<br>rob@imagen.ai |
| Secondary Correspondent: | Alex J. Cadotte, Ph.D.<br>Director, Software & Digital Health<br>MCRA<br>803 7th Street, NW, 3rd Floor<br>Washington, DC 20001<br>202-742-3828<br>acadotte@mcra.com |
| Date Prepared: | September 13, 2022 |
#### 2. DEVICE
| Device Trade Name: | Aorta-CAD |
|-----------------------------------------------|------------------------|
| Device Common Name or<br>Classification Name: | Medical Image Analyzer |
| Regulation: | 21 CFR 892.2070 |
| Regulatory Class: | II |
| Product Code: | MYN |
{4}------------------------------------------------
#### PREDICATE DEVICE 3.
Imagen Technologies' Chest-CAD has been identified as the predicate device for Aorta-CAD.
#### DEVICE DESCRIPTION 4.
Aorta-CAD is computer-assisted detection (CADe) software designed for physicians to increase the accurate detection of findings on chest radiographs that are suggestive of chronic conditions in the aorta. The ROIs are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta. Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for suspicious ROIs with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. Aorta-CAD's output is available for physicians as a concurrent reading aid and does not replace the role of the physician or of other diagnostic testing in the standard of care for the distinct conditions. Aorta-CAD uses modern deep learning and computer vision techniques to analyze chest radiographs.
For each image within a study, Aorta-CAD generates a DICOM Presentation State file (output overlay). If any ROI is detected by Aorta-CAD in the study, the output overlay for each image includes which radiographic finding(s) were identified and what chronic condition in the aorta is suggested by these findings, such as "Aortic calcification suggestive of Aortic Atherosclerosis." In addition, if ROI(s) are detected in an image, bounding boxes surrounding each detected ROI are included in the output overlay for that image and are labeled with the radiographic findings, such as "Aortic calcification". If no ROI is detected by Aorta-CAD in the study, the output overlay for each image will include the text "No Aorta-CAD ROI(s)" and no bounding boxes will be included. Regardless of whether an ROI is detected, the overlay includes text identifying the X-ray study as analyzed by Aorta-CAD and a customer configurable message containing a link to our instructions for users to access labeling documents. The Aorta-CAD overlay can be toggled on or off by the physician within their Picture Archiving and Communication System (PACS) viewer, allowing for concurrent review of the X-ray study.
#### INTENDED USE/INDICATIONS FOR USE ട.
Aorta-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for suspicious regions of interest (ROIs). The device uses a deep learning algorithm to identify ROIs and produce boxes around the ROIs. The boxes are labeled with one of the following radiographic findings: Aortic calcification or Dilated aorta.
Aorta-CAD is intended for use as a concurrent reading aid for physicians looking for ROIs with radiographic findings suggestive of Aortic Atherosclerosis or Aortic Ectasia. It does not replace the role of the physician or of other diagnostic testing in the standard of care. Aorta-CAD is indicated for adults only.
{5}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE 6.
### Comparison of Indications
The predicate device for Aorta-CAD is Chest-CAD (K210666). Chest-CAD has the following FDA-cleared Indications for Use:
Chest-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies using machine learning techniques to identify, categorize, and highlight suspicious regions of interest (ROI). Any suspicious ROI identified by Chest-CAD is assigned to one of the following categories: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. The device is intended for use as a concurrent reading aid for physicians. Chest-CAD is indicated for adults only.
Chest-CAD and Aorta-CAD both analyze chest radiographs and both detect ROIs in the chest. Both devices identify and categorize ROIs. Chest-CAD and Aorta-CAD are indicated for use as a concurrent reading aid. Both devices are intended as an aid to the physician and not intended to replace the role of the physician or of other diagnostic testing in the standard of care. The differences in Indications for Use do not constitute a new intended use, as both devices are intended to assist physicians by identifying and marking ROIs in chest radiographs.
### Technological Comparisons
Table 1 provides a comparison of the Technological Characteristics of Aorta-CAD to the predicate Chest-CAD.
| | Proposed Device | Predicate |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number | TBD | K210666 |
| Applicant | Imagen Technologies, Inc. | Imagen Technologies, Inc. |
| Device Name | Aorta-CAD | Chest-CAD |
| Classification Regulation | 892.2070 | 892.2070 |
| Product Code | MYN | MYN |
| Image Modality | X-ray | X-ray |
| Study Type | Chest | Chest |
| Clinical Output | Identify and mark regions of interest (ROIs) on chest radiographs and label the box around the ROI as one of the following: Aortic calcification or Dilated aorta. | Identify and mark regions of interest (ROIs) on chest radiographs and label the box around the ROI as one of the following: Cardiac, Mediastinum/Hila, Lungs, Pleura, Bones, Soft Tissues, Hardware, or Other. |
Table 1: Technological Comparison
{6}------------------------------------------------
| | Proposed Device | Predicate |
|------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Intended Users | Physicians | Physicians |
| Intended User Workflow | Device intended for use as a concurrent reading aid for physicians interpreting chest radiographs. | Device intended for use as a concurrent reading aid for physicians interpreting chest radiographs. |
| Patient Population | Adults with Chest Radiographs | Adults with Chest Radiographs |
| Machine Learning Methodology | Supervised Deep Learning | Supervised Deep Learning |
| Platform | Secure cloud-based processing and delivery of chest radiographs | Secure cloud-based processing and delivery of chest radiographs |
| Image Source | Digital X-ray | Digital X-ray |
| Image Viewing | Image displayed on PACS system | Image displayed on PACS system |
| Privacy | HIPAA Compliant | HIPAA Compliant |
As outlined in the table above, Aorta-CAD's technological characteristics are similar to those of Chest-CAD. Aorta-CAD differs from Chest-CAD in that Aorta-CAD simultaneously identifies and categorizes ROIs as one of two categories compared to Chest-CAD which simultaneously identifies and categorizes ROIs as one of eight categories. The fundamental purpose of both devices is to identify ROIs on chest X-rays for further consideration by the physicians, and the differences in technological characteristics do not raise different concerns of safety and effectiveness.
### PERFORMANCE DATA 7.
## Biocompatibility Testing
There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.
# Electrical Safety and Electromagnetic Compatibility (EMC)
The subject device is a software-only device. Therefore, electrical safety and EMC testing was not necessary to establish the substantial equivalence of this device.
# Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software level of concern for Aorta-CAD is Moderate since a malfunction of, or a latent design flaw in, the
{7}------------------------------------------------
software device may lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
# Bench Testing
Imagen conducted a standalone performance assessment on 5,000 chest radiograph cases representative of the intended use population. The results of the standalone testing demonstrated that Aorta-CAD detects ROIs with high sensitivity (0.910; 95% Wilson's Confidence Interval: 0.896. 0.922), high specificity (0.896; 95% Wilson's Confidence Interval: 0.889, 0.902), and high Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve (0.974, 95% Bootstrap Confidence Interval: 0.971, 0.977).
The AUC of the ROC curve was also estimated for each category and Figure 1 shows AUCs remain high across the two categories (further details in Table 2). The AUC of the ROC curve was 0.972 for Aortic calcification suggestive of Aortic Atherosclerosis and was 0.948 for Dilated aorta suggestive of Aortic Ectasia. Sensitivity and specificity were calculated for each category. As shown in Table 3, sensitivity was 0.922 for Aortic calcification suggestive of Aortic Atherosclerosis and sensitivity was 0.830 for Dilated aorta suggestive of Aortic Ectasia. Specificity was 0.897 for Dilated aorta suggestive of Aortic Ectasia and was 0.894 for Aortic calcification suggestive of Aortic Atherosclerosis. The Free-Response ROC (FROC) curve was also estimated for each Aorta-CAD category and Figure 2 shows the box-level sensitivity versus the false positives per image. The FROC curves terminate at the device's box-level sensitivity for each category due to the cascaded nature of the Aorta-CAD predictions.
{8}------------------------------------------------
Image /page/8/Figure/2 description: This image is a plot that shows sensitivity on the y-axis and 1-specificity on the x-axis. There are two curves on the plot, one for 'Aortic calcification suggestive of Aortic Atherosclerosis' and one for 'Dilated aorta suggestive of Aortic Ectasia'. The 'Aortic calcification suggestive of Aortic Atherosclerosis' curve is blue, and the 'Dilated aorta suggestive of Aortic Ectasia' curve is red. The plot also includes a dashed line that represents the line of no discrimination.
#### Figure 1: Standalone Results - Aorta-CAD ROC Curve by Category
Table 2: AUC of the ROC Curve for Aorta-CAD Predictions by Category
| Category | Ground Truth<br>Positive n (%) | AUC | 95% Bootstrap CI |
|-----------------------------------------------------------------|--------------------------------|-------|------------------|
| Aortic calcification<br>suggestive of Aortic<br>Atherosclerosis | 1662 (33.2) | 0.972 | 0.967, 0.976 |
| Dilated aorta suggestive<br>of Aortic Ectasia | 247 (4.9) | 0.948 | 0.939, 0.957 |
Abbreviations: AUC = Area Under the Curve; CI = Confidence Interval; ROC = Receiver Operating Characteristic.
{9}------------------------------------------------
| Category | Sensitivity | Specificity | Positive<br>Predictive Value | Negative<br>Predictive Value |
|-----------------------------------------------------------------|-------------------------|-------------------------|------------------------------|------------------------------|
| | 95%<br>Wilson's CI | 95%<br>Wilson's CI | 95%<br>Wilson's CI | 95%<br>Wilson's CI |
| Aortic calcification<br>suggestive of Aortic<br>Atherosclerosis | 0.922<br>(0.908, 0.934) | 0.894<br>(0.883, 0.904) | 0.812<br>(0.794, 0.829) | 0.958<br>(0.951, 0.965) |
| Dilated aorta<br>suggestive of Aortic<br>Ectasia | 0.830<br>(0.778, 0.872) | 0.897<br>(0.888, 0.906) | 0.296<br>(0.263, 0.331) | 0.990<br>(0.987, 0.993) |
#### Sensitivity, Specificity, Positive Predictive Value, and Negative Predictive for Aorta-Table 3: CAD Predictions by Category
Abbreviations: CI = Confidence Interval.
#### Standalone Results - Free-Response ROC (FROC) Curve by Aorta-CAD Category Figure 2:
Image /page/9/Figure/6 description: The image is a plot comparing the box-level sensitivity versus false positives per image for two different conditions. The blue line represents "Aortic calcification suggestive of Aortic Atherosclerosis", while the red line represents "Dilated aorta suggestive of Aortic Ectasia". The plot shows that the blue line has a higher box-level sensitivity than the red line for lower values of false positives per image. The box-level sensitivity ranges from 0.0 to 1.0, and the false positives per image range from 0.0 to 0.7.
{10}------------------------------------------------
## Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
# Clinical Data
Imagen conducted a fully-crossed multiple reader, multiple case (MRMC) retrospective reader study to determine the impact of Aorta-CAD on reader performance in detecting Aortic calcification suggestive of Aortic Atherosclerosis and Dilated aorta suggestive of Aortic Ectasia in chest radiograph cases. The primary objective of this study was to determine whether the accuracy of readers aided by Aorta-CAD ("Aided") was superior to the accuracy of readers when unaided by Aorta-CAD ("Unaided") per category as determined by the case-level Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve.
Clinical readers each evaluated 244 cases in Aorta-CAD's Indications for Use under both Aided and Unaided conditions. Each case was previously evaluated by a panel of U.S. board-certified radiologists who assigned a ground truth binary label indicating the presence of Aortic calcification suggestive of Aortic Atherosclerosis and Dilated aorta suggestive of Aortic Ectasia. The MRMC study consisted of two independent reading sessions separated by a washout period of at least 28 days in order to avoid memory bias. For each case, each reader was required to provide a binary determination of the presence of an ROI for each category and to provide a confidence score representing their certainty.
The accuracy of readers in the intended use population was superior when aided by Aorta-CAD than when unaided by Aorta-CAD for each category as calculated by the Dorfman, Berbaum, and Metz (DBM) modeling approach. The results of the clinical study are shown in Figure 3 and Figure 4.
{11}------------------------------------------------
### Figure 3: Reader Study Results - Aided and Unaided ROC Curves for Aortic calcification suggestive of Aortic Atherosclerosis
Image /page/11/Figure/3 description: This image is a plot comparing the sensitivity and specificity of two modalities, "Aided" and "Unaided". The x-axis represents "1 - Specificity", while the y-axis represents "Sensitivity". The "Aided" modality is represented by a solid blue line, and the "Unaided" modality is represented by a dotted blue line. The plot shows that the "Aided" modality has higher sensitivity and specificity than the "Unaided" modality.
Reader Study Results - Aided and Unaided ROC Curves for Dilated aorta Figure 4: suggestive of Aortic Ectasia
Image /page/11/Figure/5 description: This image is a plot of sensitivity vs 1-specificity. There are two curves on the plot, one for 'Aided' and one for 'Unaided'. The 'Aided' curve is solid, while the 'Unaided' curve is dotted. The 'Aided' curve is generally higher than the 'Unaided' curve, indicating that the 'Aided' modality has better performance.
{12}------------------------------------------------
In particular, the clinical study results demonstrated improvements when Aided versus Unaided:
- Reader AUC estimates significantly improved for both categories (p-values < 0.001). ●
- Reader AUC improvement for Aortic calcification suggestive of Aortic Atherosclerosis was 0.0984 (95% Confidence Interval: 0.0984, 0.0985) and for Dilated aorta suggestive of Aortic Ectasia was 0.0885 (95% Confidence Interval: 0.0885, 0.0886).
#### CONCLUSIONS 8.
The conclusions drawn from the standalone and clinical studies demonstrate that Aorta-CAD is as safe, as effective, and performs as well as Chest-CAD. The special controls for the Medical Image Analyzer (CADe) 21 CFR 892.2070 regulation are satisfied by demonstrating effectiveness of the device in both the standalone testing and the clinical testing, showing superiority of Aided versus Unaided reads in the clinical testing, and communicating testing results in the labeling. Aorta-CAD's technological characteristics, including but not limited to the intended end-users, imaging modality, output display on X-ray studies, and assistive functionality during chest radiograph interpretation workflows, are similar to those of Chest-CAD. The technological differences identified and discussed in §6 do not raise different concerns of safety and effectiveness. Thus, Aorta-CAD is substantially equivalent to Chest-CAD for the intended use of computer-assisted detection.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.