← Product Code [MWP](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP) · K182727

# Trident HD Specimen Radiography System (K182727)

_Hologic, Inc. · MWP · Jan 10, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K182727

## Device Facts

- **Applicant:** Hologic, Inc.
- **Product Code:** [MWP](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP.md)
- **Decision Date:** Jan 10, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

A cabinet X-ray system used to provide digital X-ray images of surgical and core biopsy specimens from various anatomical regions in order to allow rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification in the same room as the procedure or nearby improves workflow, thus reducing the time the patient needs to be under examination.

## Device Story

Trident HD Specimen Radiography System; self-contained cabinet X-ray system for imaging surgical/biopsy specimens. Input: specimen placed in shielded cabinet. Operation: X-ray source (20-50 kV, tungsten anode) generates images captured by TFT-based amorphous selenium digital detector (70 μm pixels). Output: 14-bit digital X-ray images displayed on integrated touch-screen monitor. Used in surgical suites, stereotactic biopsy suites, and pathology labs by clinicians/technicians. System provides rapid verification of lesion/calcification presence in excised tissue; allows immediate confirmation of correct margins/tissue removal. Benefits: improved workflow; reduced patient time under examination. Features: manual or auto-exposure modes; 1.5x/2.0x magnification; DICOM connectivity (worklist, print, storage, query/retrieve).

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including functional testing, usability testing, and compliance with electrical safety (IEC 61010) and EMC standards.

## Technological Characteristics

Cabinet X-ray system; tungsten anode X-ray source (20-50 kV); TFT-based amorphous selenium digital detector; 70 μm pixel size; 7.1 lp/mm spatial resolution. Integrated touch-screen display; Windows 10 OS. Connectivity: DICOM (worklist, print, storage, query/retrieve). Safety: door interlock, passcode key, fully shielded cabinet. Standards: UL 61010-1, IEC 61010-1, IEC 61010-2-091, EN 61326-1, ISO 14971.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Trident Specimen Radiography System ([K111508](/device/K111508.md))

## Reference Devices

- Selenia FFDM system ([P010025](/device/P010025.md))
- Affirm Prone Biopsy System ([K153486](/device/K153486.md))
- Selenia Dimensions systems ([P010025](/device/P010025.md)/S013, [P080003](/device/P080003.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 10, 2019

Hologic, Inc. % Meghan Wakeford Regulatory Affairs Specialist 36 Apple Ridge Road DANBURY CT 06810

Re: K182727

Trade/Device Name: Trident HD Specimen Radiography System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MWP Dated: December 11, 2018 Received: December 13, 2018

Dear Ms. Wakeford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hse 2. Nils

for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name

Trident HD Specimen Radiography System

Indications for Use (Describe)

A cabinet X-ray system used to provide digital X-ray images of surgical and core biopsy specimens from various anatomical regions in order to allow rapid verification that the correct tissue has been excised during the biopsy procedure.

Doing the verification in the same room as the procedure or nearby improves workflow, thus reducing the time the patient needs to be under examination.

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Traditional 510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92

| Date Prepared:                                           | December 11, 2018                                                   |
|----------------------------------------------------------|---------------------------------------------------------------------|
| Manufacturer:                                            | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA       |
| Establishment Registration #:                            | 1220984                                                             |
| Contact Person:                                          | Meghan Wakeford<br>Regulatory Affairs Specialist<br>P: 203.702.7819 |
| Identification of the Device:                            |                                                                     |
| Proprietary/Trade Name:                                  | Trident® HD Specimen Radiography System                             |
| Classification Name:                                     | Cabinet X-Ray System                                                |
| Regulatory Number:                                       | 21 CFR 892.1680                                                     |
| Product Code:                                            | MWP                                                                 |
| Device Class:                                            | Class II                                                            |
| Review Panel:                                            | Radiology                                                           |
| Identification of the Legally Marketed Predicate Device: |                                                                     |

| Trade Name:              | Trident® Specimen Radiography System |
|--------------------------|--------------------------------------|
| Classification Name:     | Cabinet X-Ray System                 |
| Regulatory Number:       | 21 CFR 892.1680                      |
| Product Code:            | MWP                                  |
| Device Class:            | Class II                             |
| Review Panel:            | Radiology                            |
| Submitter/510(k) Holder: | Hologic, Inc.                        |
| Clearance:               | K111508 (cleared August 19, 2011)    |

#### Device Description:

The Trident® HD Specimen Radiography System is a self-contained, direct digital imaging system for imaging surgical and biopsy specimens. The Trident® HD system includes the following major components: an image display monitor, touch-screen control display, and an imaging cabinet.

The system is self-contained. Shielding is incorporated within the cabinet chamber system design, eliminating the need for separate shielding. The unit is mounted on casters to allow for easy transportation.

Dedicated specimen radiography systems are intended for use in the following environments:

- . The surgical suite
- The stereotactic biopsy suite
- The pathology lab ●

{4}------------------------------------------------

Specimen radiography units are utilized to confirm removal of the intended tissue, or site marker in surgical and core biopsy specimens from various anatomical regions. By generating a high-resolution X-ray of the specimen, the presence of a lesion or calcification in the extracted sample can be confirmed by the user reviewing the digital image.

#### Indications for Use:

A cabinet X-ray system used to provide digital X-ray images of surgical and core biopsy specimens from various anatomical regions in order to allow rapid verification that the correct tissue has been excised during the biopsy procedure.

Doing the verification in the same room as the procedure or nearby improves workflow, thus reducing the time the patient needs to be under examination.

#### Standards:

- UL 61010-1:2012 Safety Requirements for Electrical Equipment for Measurement, Control, ● and Laboratory Use; Part 1: General Requirements
- IEC 61010-1:2010 Safety Requirements for Electrical Equipment for Measurement, Control, ● and Laboratory Use, Part 1: General Requirements
- IEC 61010-2-091:2012 Safety Requirements for Electrical Equipment for Measurement, ● Control. and Laboratory Use, Part 2-091: Particular Requirements for Cabinet X-ray Systems
- EN 61326-1:2013 Electrical Equipment for Measurement, Control, and Laboratory Use EMC ● Requirements, General Requirements
- EN 55011 and CISPR 11 (Class A) Industrial, Scientific and Medical (ISM) Radio-Frequency ● Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement
- ISO 14971: 2012 – Medical devices – Application of Risk Management to Medical Devices

### FDA Guidance Documents:

- "Applying Human Factors and Usability Engineering to Medical Devices," issued on February 3. ● 2016
- . "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices," issued on September 1, 2016
- . "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued on May 11, 2005
- "Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software," ● issued on January 14, 2005
- "Radio Frequency Wireless Technology in Medical Devices," issued on August 14, 2013 ●
- . "Off-the-Shelf Software Use in Medical Devices," issued on September 9, 1999

### Comparison with Predicate Device:

The Trident® HD System and its predicate device, the Trident® Specimen Radiography System, have the same intended use, general configuration, principles of operation, and operating parameters.

{5}------------------------------------------------

## Substantial Equivalence:

|                                | Trident® Specimen<br>Radiography System                                                                                                                                                                                                                                    | Trident® HD Specimen<br>Radiography System                                                                                                                                                                                                                                  |                                                                                                                                                           |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | Predicate (K111508)                                                                                                                                                                                                                                                        | Proposed                                                                                                                                                                                                                                                                    | Comparison                                                                                                                                                |
| Indications for Use            | A cabinet X-ray system<br>used to provide digital<br>X-ray images of surgical<br>and core biopsy<br>specimens from various<br>anatomical regions in<br>order to allow rapid<br>verification that the<br>correct tissue has been<br>excised during the<br>biopsy procedure. | A cabinet X-ray system<br>used to provide digital X-<br>ray images of surgical<br>and core biopsy<br>specimens from various<br>anatomical regions in<br>order to allow rapid<br>verification that the<br>correct tissue has been<br>excised during the biopsy<br>procedure. | Same                                                                                                                                                      |
|                                | Doing the verification in<br>the same room as the<br>procedure or nearby<br>improves workflow, thus<br>reducing the time the<br>patient needs to be under<br>examination.                                                                                                  | Doing the verification in<br>the same room as the<br>procedure or nearby<br>improves workflow, thus<br>reducing the time the<br>patient needs to be under<br>examination.                                                                                                   |                                                                                                                                                           |
| Level of Concern               | Moderate                                                                                                                                                                                                                                                                   | Moderate                                                                                                                                                                                                                                                                    | Same                                                                                                                                                      |
| Method of Use                  | Cabinet X-ray system<br>used for imaging small<br>to medium biopsy and<br>surgical specimens                                                                                                                                                                               | Cabinet X-ray system<br>used for imaging small to<br>large biopsy and surgical<br>specimens                                                                                                                                                                                 | Expanded specimen<br>size capabilities with<br>larger detector                                                                                            |
| Mechanism of<br>Action         | Sample verification:<br>obtain correct margins,<br>specimen of interest, etc.                                                                                                                                                                                              | Sample verification:<br>obtain correct margins,<br>specimen of interest, etc.                                                                                                                                                                                               | Same                                                                                                                                                      |
| Physical                       |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Construction                   | Shielded cabinet on<br>moveable cart                                                                                                                                                                                                                                       | Fully integrated system<br>with movable base<br>casters                                                                                                                                                                                                                     | Similar; redesigned for<br>ease of use and<br>transportation                                                                                              |
| Size (W x D x H)<br>(in.)      | 26.5 x 37.5 x 68.3                                                                                                                                                                                                                                                         | 24 x 26 x 66.5                                                                                                                                                                                                                                                              | Similar; reduced<br>footprint for ease of use<br>and transportation                                                                                       |
| Weight (lb.)                   | 275<br>325 with UPS                                                                                                                                                                                                                                                        | 278 (with MFD)<br>291 (with HDT)                                                                                                                                                                                                                                            | Similar; minor weight<br>changes due to design<br>change                                                                                                  |
| Digital Image Receptor         |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Detector<br>Technology         | TFT-based direct<br>capture technology                                                                                                                                                                                                                                     | TFT-based direct capture<br>technology                                                                                                                                                                                                                                      | Same                                                                                                                                                      |
| X-ray Absorption<br>Material   | Amorphous selenium                                                                                                                                                                                                                                                         | Amorphous selenium                                                                                                                                                                                                                                                          | Same                                                                                                                                                      |
| Active Imaging<br>Area Size    | 12 cm x 14 cm                                                                                                                                                                                                                                                              | 12 cm x 14 cm (MFD)<br>20 cm x 20 cm (HDT)                                                                                                                                                                                                                                  | Trident® HD is<br>available with two                                                                                                                      |
|                                |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             | detectors for variety in<br>image size and<br>preference                                                                                                  |
| Digital Detector               | Detector used on<br>predicate, K111508, was<br>approved for use with<br>the Selenia FFDM<br>system, in P010025.                                                                                                                                                            | MFD used on the Affirm<br>Prone Biopsy System,<br>cleared under K153486.<br><br>HDT used on the Selenia<br>Dimensions systems,<br>approved in<br>P010025/S013 and<br>P080003.                                                                                               | Both the predicate and<br>the proposed devices<br>use detectors<br>cleared/approved for<br>use with Hologic digital<br>mammography and<br>biopsy systems. |
| Pixel Size                     | 70 μm pixels                                                                                                                                                                                                                                                               | 70 μm pixels                                                                                                                                                                                                                                                                | Same                                                                                                                                                      |
| Limiting Spatial<br>Resolution | 7.1 lp/mm                                                                                                                                                                                                                                                                  | 7.1 lp/mm                                                                                                                                                                                                                                                                   | Same                                                                                                                                                      |
| Output Image                   | 14-bit image data                                                                                                                                                                                                                                                          | 14-bit image data                                                                                                                                                                                                                                                           | Same                                                                                                                                                      |
| Image<br>Enhancement           | Enhanced visualization<br>adjustment for image<br>sharpness                                                                                                                                                                                                                | Enhanced visualization<br>adjustment for image<br>sharpness                                                                                                                                                                                                                 | Same                                                                                                                                                      |
| Acquisition Workstation        |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Display Monitor                | 2.3 MP color LCD<br>display                                                                                                                                                                                                                                                | 2 MP High luminescence<br>diagnostic monitor                                                                                                                                                                                                                                | Similar                                                                                                                                                   |
| Operating System               | Windows 7                                                                                                                                                                                                                                                                  | Windows 10                                                                                                                                                                                                                                                                  | Similar; both the<br>predicate and proposed<br>devices use Windows<br>OS                                                                                  |
| DICOM Services                 | Worklist, print, storage                                                                                                                                                                                                                                                   | Worklist, print, storage,<br>query retrieve                                                                                                                                                                                                                                 | Similar; increased<br>DICOM capabilities<br>through ability to pull<br>priors                                                                             |
| User Interface                 | Keyboard, mouse                                                                                                                                                                                                                                                            | Integrated touch screen,<br>track pad                                                                                                                                                                                                                                       | Improved for ease of<br>use                                                                                                                               |
| X-ray Source                   |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Energy Range                   | 20-35 kV                                                                                                                                                                                                                                                                   | 20-50 kV                                                                                                                                                                                                                                                                    | Expanded upper range<br>to support larger<br>detector and imaging<br>area                                                                                 |
| Anode Type                     | Tungsten                                                                                                                                                                                                                                                                   | Tungsten                                                                                                                                                                                                                                                                    | Same                                                                                                                                                      |
| Tube Current                   | 1 mA                                                                                                                                                                                                                                                                       | 1 mA                                                                                                                                                                                                                                                                        | Same                                                                                                                                                      |
| Exposure                       | Up to 10 mAs                                                                                                                                                                                                                                                               | Up to 20 mAs                                                                                                                                                                                                                                                                | Increase to support<br>thicker specimen<br>samples                                                                                                        |
| Focal Spot Size                | 50 μm                                                                                                                                                                                                                                                                      | 50 μm                                                                                                                                                                                                                                                                       | Same                                                                                                                                                      |
| Filtration                     | 125 μm beryllium                                                                                                                                                                                                                                                           | 200 μm beryllium                                                                                                                                                                                                                                                            | Larger target angle;<br>larger image area                                                                                                                 |
| Exposure Modes                 |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Manual                         | User selects kV and<br>mAs                                                                                                                                                                                                                                                 | User selects kV and mAs                                                                                                                                                                                                                                                     | Same                                                                                                                                                      |
| Auto                           | System determines<br>optimum kV and mAs                                                                                                                                                                                                                                    | System determines<br>optimum kV and mAs                                                                                                                                                                                                                                     | Same                                                                                                                                                      |
| System Performance             |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Time to Preview                | < 20 seconds                                                                                                                                                                                                                                                               | < 20 seconds                                                                                                                                                                                                                                                                | Same                                                                                                                                                      |
| Cycle Time                     | 30 Seconds                                                                                                                                                                                                                                                                 | < 60 seconds                                                                                                                                                                                                                                                                | Increased cycle time to<br>support thicker<br>specimen samples                                                                                            |
| Cabinet                        |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                             |                                                                                                                                                           |
| Safety Features                | Door interlock, key<br>switch, fully shielded                                                                                                                                                                                                                              | Door interlock, passcode<br>key, fully shielded                                                                                                                                                                                                                             | Similar                                                                                                                                                   |
| Indicators                     | Power, door open, ready,<br>X-ray ON                                                                                                                                                                                                                                       | Power, door open, ready,<br>X-ray ON                                                                                                                                                                                                                                        | Same                                                                                                                                                      |
| Magnification                  | Specimen tray positions<br>at 1.5x and 2.0x, auto-<br>sensed                                                                                                                                                                                                               | Specimen tray positions<br>at 1.5x and 2.0x, auto-<br>sensed                                                                                                                                                                                                                | Same                                                                                                                                                      |

{6}------------------------------------------------

{7}------------------------------------------------

#### Summary of Testing:

The Trident® HD system successfully performed system design control verification and validation tests, which are summarized in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005) based on a moderate level of concern.

The Trident® HD system complies with IEC 61010 standards, as performed by the third-party test house, UL. No clinical studies have been performed. Substantial equivalence has been demonstrated by nonclinical testing. Additional bench testing, including functional testing and usability testing, was also performed on the Trident® HD system. The comparative and other performance testing showed that the overall system demonstrated equivalent performance and equivalent safety and effectiveness as the predicate Trident® Specimen Radiography System (K111508).

#### Conclusion:

Based on the information submitted in this premarket notification, The Trident® HD system is substantially equivalent to the Trident® system (K111508). The design, operation, basic construction and materials used are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K182727](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K182727)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
