← Product Code [MWP](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP) · K072557

# FAXITRON SPECIMEN RADIOGRAPHY SYSTEM, MODEL MX-20 (K072557)

_Faxitron X-Ray Corp. · MWP · Oct 30, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K072557

## Device Facts

- **Applicant:** Faxitron X-Ray Corp.
- **Product Code:** [MWP](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP.md)
- **Decision Date:** Oct 30, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Intended Use

The MX-20 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

## Device Story

Cabinet X-ray system for high-detail radiographic imaging of surgically excised medical specimens. Input: harvested tissue specimens. Operation: X-ray source (10-35 kV) with <20 micron focal spot provides up to 5X geometric magnification; integrated digital camera captures images. Output: high-resolution, DICOM-compliant digital X-ray images or radiographic film. Used in clinical settings (OR or nearby) by healthcare professionals to verify correct tissue excision during biopsy. Benefits: faster case completion, reduced patient examination time, potential reduction in patient recalls.

## Clinical Evidence

Bench testing only. Device tested for compliance with 21 CFR 1020.40 (Cabinet X-ray Systems), electrical safety (UL certification), and European Low Voltage and EMC directives.

## Technological Characteristics

Cabinet X-ray system; 10-35 kV energy range; 300 µA tube current; <20 micron focal spot; 0.03" window filtration; 30 x 35 cm X-ray coverage. Digital system: 12 x 12 cm active area (options 5x5 or 5x10 cm); 10 lp/mm spatial resolution. Connectivity: DICOM 3.0 compliant (Store, Print, Modality Worklist), network-ready. Power: 100-240 VAC. Software: Faxitron Specimen Radiography software.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Faxitron Micro 50 Specimen Radiography System ([K852634](/device/K852634.md))
- Faxitron DX-50 Digital Radiography System ([K061361](/device/K061361.md))
- piXarray Digital Specimen Radiography System ([K052433](/device/K052433.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC72557
Page 1 of 4

Image /page/0/Picture/1 description: The image shows the logo for Faxitron X-Ray Corporation. The logo is in black and white and features the company name in a bold, sans-serif font. The word "Faxitron" is larger than "X-Ray" and is placed above the company's name.

007 3002007

# PREMARKET NOTIFICATION 510 (k) Summary As required by 21 CFR 807.92

Device name – as required by 807.92(a)(2):

| Trade Name:                       | Faxitron MX-20 Specimen Radiography System                   |
|-----------------------------------|--------------------------------------------------------------|
| Common/Classification Name:       | Specimen x-ray System/Cabinet X-ray System                   |
| Classification Regulation:        | 21 CFR 892.1680                                              |
| Device Class:                     | Class II                                                     |
| Product Code (Procode):           | MWP                                                          |
| Panel:                            | Radiologic Devices Panel                                     |
| Premarket Notification submitter: |                                                              |
| Company Name:                     | Faxitron X-ray Corporation                                   |
| Company Address:                  | 225 Larkin Drive, Suite 1<br>Wheeling, II 60090-7209         |
| Contact:                          | Douglas C. Wiegman,<br>VP Engineering and Regulatory Affairs |
| Preparation Date:                 | September 6, 2007                                            |

A. LEGALLY MARKETED PREDICATE DEVICES – as required by 21 CFR 807.92(a)(3)

The Faxitron MX-20 Digital Specimen Radiography System is substantially equivalent to the following to the following predicate devices:

 Ο Faxitron Micro 50 Specimen Radiography System-K852634
   

 Ο Faxitron DX-50 Digital Radiography System-K061361

| Faxitron X-ray Corporation | MX-20 510(k) Submission | Page 9 of 30 |
|----------------------------|-------------------------|--------------|
|----------------------------|-------------------------|--------------|

{1}------------------------------------------------

KC72557
Page 2 of 4

- o piXarray Digital Specimen Radiography System-K052433
- B. DEVICE DESCRIPTION as required by 21 CFR 807.92(a)(4)

The Faxitron MX-20 Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens. The exceptionally high magnification capability (up to 5X) from the 0.02 mm focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 30 x 35 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images through the use of an integrated camera and Faxitron Specimen Radiography software.

- C. DEVICE CLAIMS as required by 807.92(a)
The Faxitron MX-20 Specimen Radiography System is a fully shielded Cabinet X-ray System that has been designed to comply with 21 CFR 1020.40. The system allows up to 5 times geometric magnification of excised specimens with minimal qeometric distortion through the use of a focal spot size that is less than 20 microns. The x-ray coverage of the device allows the use of radiographic film sizes up to 30 x 35 cm. The device can also be configured to provide high resolution, DICOM compliant, digital images through the use an integrated digital camera and Faxitron Specimen Radiography software. The Faxitron Software supports the DICOM Store, Print and Modality Worklist services.

- D. DEVICE TECHNICAL SPECIFICATIONS as required by 807.92(a)(4)
Cabinet Specifications:

- o Energy Range: 10-35 kV
- Tube Current: 300 µA o
- Focal Spot Size: < 20 microns ্
- Window Filtration: 0.03" o
- X-ray Coverage: 30 x 35 cm o
- Power: 100-240 VAC, 50/60 Hz, 150VA o
- Dimensions: ୍

External: 17" W x 18.5" D x 29" H Internal: 14" W x 16" D x 19" H

- o Weight: 150 lbs.
Digital System Specifications:

- o Active image Area: 12 x 12 cm Options: 5 x 5 cm or 5x 10 cm
- Typical Spatial resolution: 10 lp/mm

Faxitron X-ray Corporation

MX-20 510(k) Submission

Page 10 of 30

{2}------------------------------------------------

KC72557
Page 3 of 4

- DICOM 3.0 compliant software includes Store, Print and Modality ে Worklist
- Network ready workstation and monitor included o
- E. INTENDED USE as required by 807.92(a)(5)

Indications for Use: The MX-20 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure.

Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

- F. LEVEL OF CONCERN as required by recent FDA guidance
Faxitron has determined that the submitted device has a "moderate" software Level of Concern and has provided that documented record as part of this submission.

## G. TECHNOLOGICAL CHARACTERISTICS SUMMARY

The MX-20 Specimen Radiography System has the same indications for use as the predicate devices listed above. The technological characteristics of the MX-20 Specimen Radiography System have been compared to the predicate devices sited and is covered in detail in the Substantial Equivalence section of this submission. Differences in the MX-20 Specimen Radiography system as compared to the predicate devices sited are summarized in the table below.

| SYSTEM AND THE PREDICATE DEVICES SITED |                   |                      |                   |                       |
|----------------------------------------|-------------------|----------------------|-------------------|-----------------------|
| Characteristic                         | Faxitron<br>MX-20 | Faxitron<br>Micro 50 | Faxitron<br>DX-50 | Bioptics<br>piXiarray |
| Energy range                           | 10 – 35 kV        | 15 – 50 kV           | 10 – 35 kV        | 5- 45 kV              |
| Tube current                           | 0.3 mA            | 0.35 mA              | 0.1 mA            | 0.5 mA                |
| X-ray                                  | 41.6 cm           | 25 cm                | 14.6 cm           | 19 cm                 |
| Coverage                               |                   |                      |                   |                       |
| Focal Spot Size                        | < 20 microns      | 50 microns           | 50 microns        | 50 microns            |
| Magnification Capability               | Up to 5 times     | Up to 2 times        | Up to 1.5 times   | Up to 1.4 times       |

TABLE OF DIFFERENCES BETWEEN THE MX-20 SPECIMEN RADIOGRAPHY SYSTEM AND THE PREDICATE DEVICES SITED

{3}------------------------------------------------

KC72557
Page 4 of 4

As can be seen from the Table above the energy ranges for the MX-20 Specimen Radiography System and the predicate devices listed are very similar. One advantage that the MX-20 Specimen Radiography System has over the predicate devices is the size of the focal spot and the geometric magnification capability of the device. In addition the MX-20 Specimen Radiography Systems uses the same digital imaging hardware and software as the Faxitron DX-50 Digital Radiography System.

## H. NONCLINICAL PERFORMANCE DATA TESTING AND REVIEW - as required by 807.92(b)(1)

The MX-20 Specimen Radiography System is a Cabinet X-ray System and has been designed and tested to comply with the performance standards set forth in 21 CFR 1020.40 Cabinet X-ray Systems. Testing and performance data pertaining to this standard has been included as part of the submission. This device has also been tested for electrical safety and is Classified by Underwriters Laboratories in the US and Canada. Additional testing to certify compliance to the European Low voltage and EMC directives has been successfully completed and documented in this submission.

- I. Substantial Equivalence Summary
The MX-20 Specimen Radiography System has the same indications for use as the predicate devices sited. The technical characteristics of the MX-20 are very similar to the predicate devices. One major difference that we believe allows the MX-20 Specimen Radiography System to perform better than the predicate devices sited is the size of the focal spot (<20 microns) and the optimized cabinet geometry which allows greater x-ray coverage and higher geometric maqnification of excised specimens with minimal geometric distortion.

#### J. CONCLUSIONS

We conclude that the documentation and testing included in this submission indicates that the MX-20 Specimen Radiography System is safe and effective and substantially equivalent to the predicate devices sited.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 3 0 2007

Mr. Douglas C. Wiegman VP Engineering Faxitron X-ray Corporation 225 Larkin Drive, Suite 1 WHEELING IL 60090-7209

Re: K072557

Trade/Device Name: MX-20 Specimen Radiography System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MWP Dated: September 6, 2007 Received: September 11, 2007

Dear Mr. Wiegman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# STATEMENT OF INDICATIONS FOR USE

510(K) Number (if known): _ K072557

Device Name: MX-20 Specimen Radiography System

Indications for Use: The MX-20 is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure.

Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

Prescription Use (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Faxitron X-ray Corporation

Page 8 of 30 MX-20 510/k) Subn

ductive. Abdominal, and

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K072557](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MWP/K072557)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
